Manuale per l’uso Anritsu ME7840A

126 pagine 1.54 mb
Scarica

Vai alla pagina of 126

Summary
  • Anritsu ME7840A - page 1

    MODEL ME7840A POWER AMPLIFIER TEST SYSTEM OPERA TION AND MAINTENANCE MANUAL 490 JARVIS DRIVE - MORGAN HILL, CA 95037-2809 P/N: 10410-00225 REVISION: A PRINTED: JUNE 2000 COPYRIGHT 2000 ANRITSU CO ...

  • Anritsu ME7840A - page 2

    WARRANTY The ANRITSU product(s) listed on the title page is (are) warranted against defects in materials and workmanship for three years from the date of shipment. ANRITSU’s obligation covers repairing or replacing products which prove to be defective during the warranty period. Buyers shall prepay transportation charges for equipment returned to ...

  • Anritsu ME7840A - page 3

    ...

  • Anritsu ME7840A - page 4

    ...

  • Anritsu ME7840A - page 5

    Safety Symbols T o prevent the risk of personal injury or loss related to equipment malfunction, ANRITSU Company uses the following symbols to indicate safety-related information. For your own safety , please read the informa - tion carefully BEFORE operating the equipment. Symbols used in manuals DANGER This indicates a very dangerous procedure th ...

  • Anritsu ME7840A - page 6

    Safety 1 ME7840A OMM Always refer to the operation manual when working near locations at which the alert mark, shown on the left, is attached. If the operation, etc., is performed without heeding the advice in the operation manual, there is a risk of personal injury . In addition, the equipment perfor - mance may be reduced. Moreover , this alert m ...

  • Anritsu ME7840A - page 7

    T able of Contents Chapter 1 General Information SCOPE OF THIS MANUAL ...............................1 - 1 INTRODUCTION ....................................1 - 1 RELA TED MANUALS .................................1 - 1 CONVENTIONS .....................................1 - 2 SERIAL NUMBER ...................................1 - 2 ONLINE MANUALS ....... ...

  • Anritsu ME7840A - page 8

    Installation .....................................2 - 7 What else is on the CD? ...............................2 - 7 Uninstalling the software ..............................2 - 7 SER VICE CENTERS ..................................2 - 8 Chapter 3 Operations, General INTRODUCTION ....................................3 - 1 PREP ARING THE SYSTEM ....... ...

  • Anritsu ME7840A - page 9

    Gain and IMD Power Sweep ........................... 5-18 POWER SWEEP , ONE TONE ............................. 5-22 Single Frequency Power Sweep .......................... 5-22 I M D ........................................... 5-26 HARMONICS ..................................... 5-29 H O T S 2 2 ........................................ 5-32 Chapter ...

  • Anritsu ME7840A - page 10

    ...

  • Anritsu ME7840A - page 11

    Figure 1-1. Model ME7840A Power Amplifier T est System (P A TS) ...

  • Anritsu ME7840A - page 12

    Chapter 1 General Information 1-1 SCOPE OF THIS MANUAL This manual provides operating and maintenance information for the ME7840A Power Amplifier T est System (P A TS, Figure 1-1). The follow - ing topics are discussed: q Equipment Description q Equipment Installation and Connection q Software Description q Software Installation q Calibration q Sys ...

  • Anritsu ME7840A - page 13

    1-4 CONVENTIONS Throughout this manual, the ME7840A Power Amplifier System may be referenced as P A TS or ME7840A; the MS462XC may be referenced as Scorpion or MS462XC; and the MS4782X T est Set may be referenced as T est Set or MS4782X. 1-5 SERIAL NUMBER All ANRITSU instruments are assigned a unique six-digit serial number , such as “940101.” ...

  • Anritsu ME7840A - page 14

    GENERAL INFORMATION HARDWARE DESCRIPTION ME7840A OMM 1-3 AUT POWER METER OPTIONAL SPECTRUM ANAL YZER SYSTEM CONTROLLER (PC) RF1 STEP A TTENUA TOR 70dB, 10dB/step POWER SUPPL Y COMBINER RF2 OPTIONAL MODULA TION SYNTHESIZER -30dB -30dB -30dB OPTIONAL EXTERNAL PRE-AMP N N LIMITER N K K K K K N N N K KK N N DIRECT RECEIVER ACCESS SCORPION MS4782D T est ...

  • Anritsu ME7840A - page 15

    HARDWARE DESCRIPTION GENERAL INFORMATION 1-4 ME 7840A OMM AUT POWER METER OPTIONAL SPECTRUM ANAL YZER SYSTEM CONTROLLER (PC) RF1 a 1 a 2 b 1 b 2 STEP A TTENUA TOR 70dB, 10dB/step POWER SUPPL Y COMBINER RF2 OPTIONAL MODULA TION SYNTHESIZER -30dB -30dB -30dB OPTIONAL EXTERNAL PRE-AMP N N LIMITER N K K K K K N N N K KK N N DIRECT RECEIVER ACCESS SCORP ...

  • Anritsu ME7840A - page 16

    1-9 SYSTEM DESCRIPTION Brief descriptions of the DRA Scorpion (MS462XC) and MS4782X T est Set are given below . DRA Scorpion The Scorpion Direct Receiver Access (DRA) version V ector Network Mea- surement System (VNMS) (Figure 1-2) functions under control of the soft- ware residing in the PC through GPIB commands. The software supports tuning and a ...

  • Anritsu ME7840A - page 17

    MS4782X T est Set The Series MS4782X T est Set contains a W ilkinson type combiner that combines the two RF signals from Ports 1 and 3 (RF1 & RF3) of the Scor - pion. External preamplifiers can optionally be provided at the combiner input to boost the input RF power to the amplifier-under - test (AUT). The combiner has power input rating of 30- ...

  • Anritsu ME7840A - page 18

    q Bandwidth: Should be sufficiently wide to cover the frequency band of the AUT . q Isolation: Should be no less than (P o - 27 dB), where P o is the power output in dBm of the AUT . Thus for Po = 47 dBm (50 watts), a 20 dB isolation is required. For a P o greater than 47 dBm up to 50 dBm (100 watts), two circulators in series can be used. Where S ...

  • Anritsu ME7840A - page 19

    SYSTEM DESCRIPTION GENERAL INFORMATION 1-8 ME 7840A OMM Noise In Noise +28V NO OPERATOR SERVICE- ABLE PARTS INSIDE. REFER SERVICING TO QUALIFIED PERSONNEL. WARNING CAUTION FOR CONTINUED FIRE PROTECTION REPLACE ONLY WITH SPECIFIED TYPE AND RATED FUSE. IEEE 488.2 GPIB Dedicated GPIB Serial Port Ethernet VGA SCSI-2 Ext Analog Output Ext Trigger LPT Li ...

  • Anritsu ME7840A - page 20

    S 22 ) are not desired, then a circulator is not required. Instead, a through line should be connected between connectors C1 and C2 (Figure 1-5). The MS4782A is shipped from the factory with this through line in place. 1-10 K F ACT OR K factor is a parameter which is sometimes used to indicate the stability of an amplifier . K factor is a function ...

  • Anritsu ME7840A - page 21

    The functions of the T est Executive, if supplied by ANRITSU or by the customer , would include: q T est Sequence development and management q Part number and serial number management q Interface with the Database TEST EXECUTIVE GENERAL INFORMATION 1-10 ME 7840A OMM ...

  • Anritsu ME7840A - page 22

    1-14 SOFTW ARE STRUCTURE The software structure is illustrated in Figure 1-6 (below). GENERAL INFORMATION SOFTWARE STRUCTURE ME7840A OMM 1-11 Anrit su Measurement, Calibration & T uning Modules VNA PowerT ools V4.00 GPIB Local/Network Database T est Executive: Automated T esting, T est Plan Development ( optional item ) Te s t S t a t i o n PC ...

  • Anritsu ME7840A - page 23

    1-15 MS7840A OPTIONS The following options are available: 1-16 OPTIONAL ACCESSORIES The accessories described below are available from ANRITSU. q Model ML2430A Power Meter q Model MG3672A Digital Modulation Signal Generator/with MG0314A W -CDMA Modulation Unit q Model MS2602A Spectrum Analyzer q Model MS8607A Digital Mobile Radio T ransmitter T est ...

  • Anritsu ME7840A - page 24

    MS4782D Test Set. The same cable is also used to connect the Test Set to the MS462XC receivers on the rear panel. 1-17 PREVENTIVE MAINTENANCE The ME7840A Power Amplifier T est System does not require any preven - tive maintenance. 1-18 USER SUPPLED TEST SET The ANRITSU MS4782X is the recommended test set for P A TS; however , users supply their own ...

  • Anritsu ME7840A - page 25

    PERFORMANCE SPECIFICATIONS GENERAL INFORMATION 1-14 ME 7840A OMM Characteristic V alue Notes Amplifier Under T est Power Output 100 W maximum With MS4782D T est Set and with23 dB isolation between AUT power output and MS4623C, Port 2 50 W maximum With MS4782A T est Set Bandwidth through T est Set 800 MHz to 2.4 GHz With MS4782D T est Set (Note 1) 8 ...

  • Anritsu ME7840A - page 26

    1-20 RECOMMENDED ITEMS T able 1-2 provides a list of test equipment and other items needed for cal - ibration and performance verification. 1-21 USER SUPPLIED ITEMS The following items required for the operation of the ME7840A Power Amplifier T est System must be supplied by the user . Item Critical Specification Personal Computer (PC) Pentium II o ...

  • Anritsu ME7840A - page 27

    Chapter 2 Installation 2-1 INTRODUCTION This chapter describes the installation, connection and set-up of the equipment that comprises the ME7840A Power Amplifier T est Station. 2-2 EQUIPMENT COMPLEMENT The ME7840A is made up of the following pieces of equipment: q Scorpion â V ector Network Measurement System (VNMS), Direct Receiver Access (DRA) ...

  • Anritsu ME7840A - page 28

    A listing of the non-optional-accessories always supplied with P A TS is listed below . The cables and terminators are shown in Figures 2-3 and 2-2, on page 2-3. Item Part No. Quantity Broadband T ermination 28N50LF 2 Control Cable 803-49 1 RF Cables, T ype N connectors (front) 15NN50-0.25B 3 RF Cables, SMA connectors (rear) 15NN50-0.35B 4 2-4 INST ...

  • Anritsu ME7840A - page 29

    INSTALLATION INSTALLATION - HARDWARE ME7840A OMM 2-3 GPIB CABLE GPIB CONNECTOR DEDICA TED GPIB CONNECTOR (Used during power level calibration) CONTROL CABLE EXT PREAMP RF CABLES (4ea) EXTERNAL I/O CONNECTOR CONTROL CABLE AC POWER CORD AC POWER CORD CONTROL IN CONNECT OR CIRCULA T OR C1 C2 C3 OUTPUT INPUT Figure 2-2. ME7840A Rear Cable Connections E ...

  • Anritsu ME7840A - page 30

    Step 3. Install the four RF cables between the rear of the MS462XC and the T est Set. T orque the cables to 8 inch-pounds (SMA Connectors) or 12 inch-pounds (T ype N connectors). T able 2-1 lists the applicable connectors by designation or function. Step 4. Install the control cable between the External I/O connector on the rear panel of the MS462X ...

  • Anritsu ME7840A - page 31

    INSTALLATION INSTALLATION - HARDWARE ME7840A OMM 2-5 EXTERNAL ANALOG IN CONNECTOR CURRENT SENSOR CABLE P ASSAGE CURRENT PROBE HEAD Figure 2-4. Current Probe Connection ...

  • Anritsu ME7840A - page 32

    INSTALLATION - HARDWARE INSTALLATION 2-6 ME 7840A OMM Figure 2-5. Software Installation W indow ...

  • Anritsu ME7840A - page 33

    2-5 INST ALLA TION - SOFTW ARE ME7840A software is delivered on a CD. The ANRITSU Power T ools li - brary V ersion 4.0 or higher , which is required to be installed on the target system (the PC which is receiving the installation of ME7840A software). Power T ools V ersion 4.0 is also included in the ME7840A Software CD. Installation Run the approp ...

  • Anritsu ME7840A - page 34

    2-6 SERVICE CENTERS ANRITSU Service Centers are listed in T able 2-2. SERVICE CENTERS INSTALLATION 2-8 ME 7840A OMM UNITED ST A TES ANRITSU COMP ANY 685 Jarvis Drive Morgan Hill, CA 95037-2809 T elephone: (408) 776-8300 1-800-ANRITSU F AX: 408-776-1744 FRANCE ANRITSU S.A 9 Avenue du Quebec Zone de Courtaboeuf 91951 Les Ulis Cedex T elephone: 016-09 ...

  • Anritsu ME7840A - page 35

    Chapter 3 Operations, General 3-1 INTRODUCTION The basic operation of the Model ME7840A Power Amplifier T est Station (without external preamplifiers) is described in this and the following two chapters: Calibration Operations and Measurement Operations. “Calibra - tion Operations” describes the software measurement calibration func - tion and ...

  • Anritsu ME7840A - page 36

    Step 6. When ready for measurements, cautiously apply DC power to the AUT . Step 7. Check that the normally factory-installed RF coaxial “jumper” cable is installed on the rear panel of the T est Set: n MS4782X: Between Ext Preamp Input and Output connectors. n MS4782D: Between connectors C1 and C2, unless a circulator is installed. Step 8. If ...

  • Anritsu ME7840A - page 37

    OPERATIONS, GENERAL PREPARING THE SYSTEM ME7840A OMM 3-3 Optional Current Probe Customer Supplied Power Meter for flat test port power calibration Customer Supplied Power Supply for AUT * Connect when directed by procedure for power calibrations * Amplifier Under T est (AUT) T o: Ext. Analog In connector GPIB GPIB MS4782A or D T est Set MS462X3C DR ...

  • Anritsu ME7840A - page 38

    3-3 USING P A TS SOFTW ARE P A TS software requires a computer with GPIB capability running W in - dows 95 or better (98/NT/2000). The software is started by selecting the “Start” button, then “Programs,” “Pats V1.1,” then “Pats” (Figure 3-2). Software Organization The P A TS software has two main and three collateral functions. The ...

  • Anritsu ME7840A - page 39

    3-4 COLLA TERAL FUNCTIONS The P A TS software collateral functions are found on the top menu bar un - der “Files,” “T ools,” and “Help.” (below) File The file menu (below) provides for saving files from the PC to the MS462XC, recalling files from the MS462XC to the PC, and for exiting the system. Click on File to display a drop-down lis ...

  • Anritsu ME7840A - page 40

    q Save Cal Files to PC. Displays a dialog box (similar to that shown for Recall Cal Files from PC , below) showing the cal files stored in the default folder (T emp). This function is useful for saving the front panel setup and calibration data to a file on the PC. The user has the option of overwriting an existing file by selecting an exist - ing ...

  • Anritsu ME7840A - page 41

    T ools The “T ools” menu (below) displays a drop-down list that provides three collateral options described below . The “Run Calibration Files” option is described in Chapter 4. a. Check Communications to VNA: Runs a test to check that the VNA communicates with the PC. Run this op- tion first to ensure that the P A TS setup is correct and f ...

  • Anritsu ME7840A - page 42

    b. Calibration File Options: Provides options for finding and recalling calibration files (below). These files are used to establish test setup for the measurements de - scribed in Chapter 5. If no file is selected in an area (S Parameter , Power Sweep, IMD, etc., the associated mea - surement will assume the P A TS default setting. Clicking on “ ...

  • Anritsu ME7840A - page 43

    Chapter 4 Operations, Calibration 4-1 INTRODUCTION Measurements always include a degree of uncertainty due to imperfec - tions in the measurement system. The measured value is always a combi - nation of the actual value plus the systematic measurement errors. Cali - bration, as it applies to network analysis, characterizes the systematic measuremen ...

  • Anritsu ME7840A - page 44

    Alternatively , Scorpion front panel settings may be set manually and ap - plied during the automated measurements under P A TS software as fol - lows: q On the Measurement W indow click the Pause button q On the Scorpion front panel press the “Clr / Local” button to take manual control of the Scorpion. q Use Scorpion front panel controls to se ...

  • Anritsu ME7840A - page 45

    4-5 GENERAL Calibration operations are described in this and the following paragraphs Step 1. Click on “Start,” “Programs,” “Pats V1.1,” then “Pats” (Figure 3-2, page 3-4) to start the software. The P A TS flash screen (below) briefly appears, then the main screen. The main screen (below) provides access to all software op- tions. S ...

  • Anritsu ME7840A - page 46

    sure that the equipment is properly connected to the bus. If so, the following screen will be displayed: Step 3. Click on “Run Calibration Procedure” to display a list of cali - bration specification files (below). (Refer to paragraph 4-3 for a description of these files.) Running the files is described in succeeding numbered paragraphs. 4-6 PO ...

  • Anritsu ME7840A - page 47

    Step 3. Double-click on “A_Power Level.txt.” . The program starts with a linear power calibration (below). Step 4. Before hitting the “Next” button, connect the power meter to at the point where power enters the device-under-test (AUT), which becomes the power reference plane as shown in the prompt. The use of cables and/or adapters does no ...

  • Anritsu ME7840A - page 48

    NOTE The linear power calibration takes up-to-5 minutes to complete. It calibrates the output power at the refer - ence plane for both internal sources (power out of Port 1 then Port 3) according to the target power across the frequency range set in the calibration speci - fication file. When the test completes, the “Next” but - ton will become ...

  • Anritsu ME7840A - page 49

    Step 6. Before pressing “Next” disconnect the power sensor and con - nect a throughline between T est Ports 1 and 2 on the MS4782X T est Set, as shown below . Step 7. Press “Next” to perform a receiver calibration. When this cali - bration finishes, the program returns to the main screen. OPERATIONS, CALIBRATION POWER LEVEL ME7840A OMM 4-7 ...

  • Anritsu ME7840A - page 50

    4-7 HOT S22 The “B_HotS22.txt” file provides prompts for setting up a Hot S 22 specifi - cation file. Hot S 22 is a return loss measurement of an amplifier ’s output port (Port 2) while stimulus is applied to its input port. The prompts cause the Scorpion to be placed into a state where it is ready to make cali - brated measurements on the us ...

  • Anritsu ME7840A - page 51

    Step 3. Press “Next” for the next prompt (below). The 3653 or 3753LF Calibration Kits contain a disk with calibration data. The data has to be installed from the MS462C front panel. T o do so, place the floppy disk from the kit into the MS462XC disk drive. Press the Cal key then COMPONENTS UTILITIES/IN - ST ALL KIT INFO FROM FLOPPY DISK. softke ...

  • Anritsu ME7840A - page 52

    Step 6. Press “OK” for the next step (below). Connect the Load to Port 2. Step 7. Press “OK” to complete the Hot S 22 calibration. HOT S22 OPERATIONS, CALIBRATION 4-10 ME 7840A OMM ...

  • Anritsu ME7840A - page 53

    4-8 S-P ARAMETERS CAL The “C_S-Parameters Cal.txt” file to provides prompts for setting up S-parameters specification file. The prompts cause the Scorpion to be placed into a state where it is ready to make calibrated measurements on the user ’s AUT . At the completion of the program, the state information is saved to a *.cal file. T o run th ...

  • Anritsu ME7840A - page 54

    Step 3. Press “Next” for the next step (below). The 3653 or 3753LF Calibration Kits contain a disk with calibration data. The data has to be installed from the MS462C front panel. T o do so, place the floppy disk from the kit into the MS462XC disk drive. Press the Cal key then COMPONENTS UTILITIES/IN - ST ALL KIT INFO FROM FLOPPY DISK softkeys. ...

  • Anritsu ME7840A - page 55

    Step 6. Press “OK” for the next step (below). Connect the Short to T est Port 1 and the Open to T est Port 2. Step 7. Press “OK” for the next step (below). Connect the Throughline between T est Ports 1 and 2. OPERATIONS, CALIBRATION S-PARAMETERS CAL ME7840A OMM 4-13 ...

  • Anritsu ME7840A - page 56

    Step 8. Press “OK” to complete the S Parameters calibration (below). S-PARAMETERS CAL OPERATIONS, CALIBRATION 4-14 ME 7840A OMM ...

  • Anritsu ME7840A - page 57

    4-9 CAL FILE SETUP The “D_Cal File Setup.txt” file provides prompts for setting up One-tone power sweep, IMD, T wo-T one Power Sweep, and Harmonic specification file. The prompts cause the Scorpion to be placed into a state where it is ready to make calibrated measurements on the user ’s AUT . At the com - pletion of the program, the state in ...

  • Anritsu ME7840A - page 58

    Step 3. Press “Next” for the next prompt (below). The 3653 or 3753LF Calibration Kits contain a disk with calibration data. The data has to be installed from the MS462C front panel. T o do so, place the floppy disk from the kit into the MS462XC disk drive. Press the Cal key then COMPONENTS UTILITIES/IN - ST ALL KIT INFO FROM FLOPPY DISK. softke ...

  • Anritsu ME7840A - page 59

    Step 5. Press “OK” to continue to the IMD cal file setup prompt (be - low). Follow the prompt and connect a throughline between T est Ports 1 and 2 on the MS4782X T est Set. OPERATIONS, CALIBRATION CAL FILE SETUP ME7840A OMM 4-17 ...

  • Anritsu ME7840A - page 60

    Step 6. Press “Next” to begin the Power Sweep cal file setup (below). CAL FILE SETUP OPERATIONS, CALIBRATION 4-18 ME 7840A OMM ...

  • Anritsu ME7840A - page 61

    Step 7. Press “Next” to perform Harmonic Cal File Setup calibration (below). Step 8. Press “Next” to end the calibration and return to the main screen. OPERATIONS, CALIBRATION CAL FILE SETUP ME7840A OMM 4-19/4-20 ...

  • Anritsu ME7840A - page 62

    Chapter 5 Operations, Measur ement 5-1 INTRODUCTION This chapter describes the five measurements available with the ME7840A software: S-Parameters, Power Sweep, IMD (intermodulation distortion), Harmonics, and Hot S 22 . 5-2 OPERA TION, GENERAL Refer to Chapter 3 for general operation and setup of the ME7840A. 5-3 MEASUREMENT CALIBRA TION Measureme ...

  • Anritsu ME7840A - page 63

    n C5, Hot S 22 ¾ “Hots22_1.cal” Step 5. Select appropriate file “Open” (below). Calibration(s) Not Completed Sometimes users may wish to run a measurement using default settings, that is, without spending the time required to calibrate. The procedure below gives instructions. Step 1. Click on “T ools,” and open “Check Communications ...

  • Anritsu ME7840A - page 64

    n D1, S-Parameters and Hot S 22 ü Points = 201 ü Start Frequency = 800 MHz ü Stop Frequency = 2400 MHz ü Input Powe r=0d B n D2, Power Sweep (2 tone power sweep) ü Main T one = 950 MHz ü Offset = 971 kHz ü Start Power = - 15 dBm ü Stop Powe r=0d B m ü Data Points = 21 n D3, IMD ü Main T one = 900 MHz ü Offset Frequency = 977 kHz ü Input ...

  • Anritsu ME7840A - page 65

    5-5 S-P ARAMETER TESTS: S 21 ,S 11 ,S 22 ,S 12 , ALL This test measures the scattering parameters (S-parameters). The test module comprises six tests: S 11 ,S 12 ,S 21 ,S 22 , and ALL. (The K Factors test is described in paragraph 5-6.) In each case, the measurement screens are the similar . This procedure will present a test screen for the “S 21 ...

  • Anritsu ME7840A - page 66

    n Select Start T est: Observe that the test screen (Figure 5-1) appears. Button Panel Options: q Print: Prints a copy of the screen graphic to a printer . q W rite Data: T wo data files are written to the C:Program FilesPatsDatafiles folder; the file name will have a time stamp. One datafile has a “.csv” file extension and the other has a ? ...

  • Anritsu ME7840A - page 67

    q Change Params: Returns to User Input screen. q Autoscale Checkbox: T urns autoscale on or off. Autoscale automatically scales the mea - surement results. q Reset/Zoom: Disabled for this test. q Max/Min: T ype in maximum and minimum display value. Use Rescale after changing val - ues. q Rescale: Rescales the measurement results. Markers Screen Opt ...

  • Anritsu ME7840A - page 68

    5-6 S-P ARAMETER TEST : K F ACTOR This test describes the K Factors measurement. This measurement is described in Chapter 1, paragraph 1-10. Procedure. Prepare the ME7840A as shown in Chapter 3. Enter the appropriate test values in the displayed User Input screen. The input fields are described below: The inputs required for this test are de - scri ...

  • Anritsu ME7840A - page 69

    n Select Start T est: Observe that the test screen (Figure 5-1) appears. Button Panel Options: q Print: Prints a copy of the screen graphic to a printer . q W rite Data: T wo data files are written to the C:Program FilesPatsDatafiles folder; the file name will have a time stamp. One datafile has a “.csv” file extension and the other has a ? ...

  • Anritsu ME7840A - page 70

    q Change Params: Returns to User Input screen. q Autoscale Checkbox: T urns autoscale on or off. Autoscale automatically scales the mea - surement results. q Reset/Zoom: Disabled for this test. q Max/Min: T ype in maximum and minimum display value. Use Rescale after changing val - ues. q Rescale: Rescales the measurement results. Markers Screen Opt ...

  • Anritsu ME7840A - page 71

    5-7 POWER SWEEP , TWO TONE This measurement module has three variations of possible Measurement T ypes: Power Sweep (single frequency), Multi Frequency Power Sweep, Gain and IMD. Single Frequency Power Sweep Procedure. Prepare the ME7840A as shown in Chapter 3. Enter the appropriate test values in the displayed User Input screen. The input fields a ...

  • Anritsu ME7840A - page 72

    n T est Set Setting: Byte value for setting test set attenuation and source switch (below). OPERATIONS, MEASUREMENT POWER SWEEP, TWO TONE ME7840A OMM 5-11 Internal Source V alue External Source V alue 0d B 8 0d B 0 10 dB 9 10 dB 1 20 dB 10 20 dB 2 30 dB 1 1 30 dB 3 40 dB 12 40 dB 4 50 dB 13 50 dB 5 60 dB 14 60 dB 6 70 dB 15 70 dB 7 ...

  • Anritsu ME7840A - page 73

    Select Start T est and observe that the test screen (Figure 5-3) appears. Button Panel Options: q Print: Prints a copy of the screen graphic to a printer . q W rite Data: Excel data is written to C:Pro - gram FilesPatsDatafiles - the file name will have a time stamp. For example: “S17-16-19-29.csv” was created on the 17th day of the month at ...

  • Anritsu ME7840A - page 74

    q Autoscale Checkbox: T urns autoscale on or off. Autoscale automatically scales the mea - surement results. q Reset/Zoom: Disabled for this test. q Max/Min: T ype in maximum and minimum display value. Use Rescale after changing val - ues. q Rescale: Rescales the measurement results. q Pause/Go: Pause stops data transfer from VNA to PC and places M ...

  • Anritsu ME7840A - page 75

    Multiple Frequency Power Sweep Procedure. n Measurement T ype: Drop down options. Set to “Multi Freq Power Sweep ” for this test. n Main T one (MHz): CW Frequency setting for source 1. The user may set a lower tone and upper tone. The software will sweep three frequencies (the two input plus one between) and the RF input power to produce the di ...

  • Anritsu ME7840A - page 76

    n T est Set Setting: Byte value for setting test set attenuation and source switch (below). OPERATIONS, MEASUREMENT POWER SWEEP, TWO TONE ME7840A OMM 5-15 Internal Source V alue External Source V alue 0d B 8 0d B 0 10 dB 9 10 dB 1 20 dB 10 20 dB 2 30 dB 1 1 30 dB 3 40 dB 12 40 dB 4 50 dB 13 50 dB 5 60 dB 14 60 dB 6 70 dB 15 70 dB 7 ...

  • Anritsu ME7840A - page 77

    Select Start T est . Observe that the test screen (Fig - ure 5-4) appears. Button Panel Options: q Print: Prints a copy of the screen graphic to a printer . q W rite Data: Excel data is written to C:Pro - gram FilesPatsDatafiles - the file name will have a time stamp. For example: “S17-16-19-29.csv” was created on the 17th day of the month a ...

  • Anritsu ME7840A - page 78

    q Pause/Go: Stops data transfer from VNA to PC and places MS4623C in “Local” mode. q Return: Returns to the main P A TS screen. q Change Params: Returns to User Input screen. q Y1, Y2: Provides for rescaling the left or right axis. q Rescale: Rescales the measurement results. q Max/Min: T ype in maximum and minimum display value. Use Rescale af ...

  • Anritsu ME7840A - page 79

    Gain and IMD Power Sweep Procedure: n Measurement T ype: Drop down options. Set to “Gain and IMD ” for this test. n Main T one (MHz): CW Frequency setting for source 1. n Offset (kHz): Offset frequency for Source 2 (Source 2 will be set to Main T one + Offset) n Start Power (dBm): Power will be swept starting at the start power (input RF power ...

  • Anritsu ME7840A - page 80

    n T est Set Setting: Byte value for setting test set attenuation and source switch (below). OPERATIONS, MEASUREMENT POWER SWEEP, TWO TONE ME7840A OMM 5-19 Internal Source V alue External Source V alue 0d B 8 0d B 0 10 dB 9 10 dB 1 20 dB 10 20 dB 2 30 dB 1 1 30 dB 3 40 dB 12 40 dB 4 50 dB 13 50 dB 5 60 dB 14 60 dB 6 70 dB 15 70 dB 7 ...

  • Anritsu ME7840A - page 81

    Select Start T est . Observe that the test screen (Fig - ure 5-5) appears. Button Panel Options: q Print: Prints a copy of the screen graphic to a printer . q W rite Data: Excel data is written to C:Pro - gram FilesPatsDatafiles - the file name will have a time stamp. For example: “S17-16-19-29.csv” was created on the 17th day of the month a ...

  • Anritsu ME7840A - page 82

    q Return: Returns to the main P A TS screen. q Change Params: Returns to User Input screen. q Y1, Y2: Provides for rescaling the left or right axis. q Rescale: Rescales the measurement results. q Max/Min: T ype in maximum and minimum display value. Use Rescale after changing val - ues. Markers Screen Options (left): q Markers slider: Moves marker b ...

  • Anritsu ME7840A - page 83

    5-8 POWER SWEEP , ONE TONE This measurement module has three variations of possible Measurement T ypes: Power Sweep (single frequency), Multi Frequency Power Sweep, Gain and IMD. Single Frequency Power Sweep Procedure. Prepare the ME7840A as shown in Chapter 3. Enter the appropriate test values in the displayed User Input screen. The input fields a ...

  • Anritsu ME7840A - page 84

    n T est Set Setting: Byte value for setting test set attenuation and source switch (below). OPERATIONS, MEASUREMENT POWER SWEEP, ONE TONE ME7840A OMM 5-23 Internal Source V alue External Source V alue 0d B 8 0d B 0 10 dB 9 10 dB 1 20 dB 10 20 dB 2 30 dB 1 1 30 dB 3 40 dB 12 40 dB 4 50 dB 13 50 dB 5 60 dB 14 60 dB 6 70 dB 15 70 dB 7 ...

  • Anritsu ME7840A - page 85

    Select Start T est and observe that the test screen (Figure 5-3) appears. Button Panel Options: q Print: Prints a copy of the screen graphic to a printer . q W rite Data: Excel data is written to C:Pro - gram FilesPatsDatafiles - the file name will have a time stamp. For example: “S17-16-19-29.csv” was created on the 17th day of the month at ...

  • Anritsu ME7840A - page 86

    q Autoscale Checkbox: T urns autoscale on or off. Autoscale automatically scales the mea - surement results. q Max/Min: T ype in maximum and minimum display value. Use Rescale after changing val - ues. q Rescale: Rescales the measurement results. Markers Screen Options (left): q Markers slider: Moves marker between data points. q Gain: Gain is in d ...

  • Anritsu ME7840A - page 87

    5-9 IMD This test measures inter-modulation distortion. Procedure Prepare the ME7840A as shown in Chapter 3. Enter the appropriate test values in the displayed User Input screen. The input fields are described below: n Measurement T ype: Already set to “IM” for IMD – input disabled. n Main T one (MHz): CW Frequency setting for source 1. n Off ...

  • Anritsu ME7840A - page 88

    Select Start T est . Observe that the test screen (Fig - ure 5-7) appears. Button Panel Options: q Print: Prints a copy of the screen graphic to a printer . q W rite Data: Excel data is written to C:Pro - gram FilesPatsDatafiles - the file name will have a time stamp. For example: “S17-16-19-29.csv” was created on the 17th day of the month a ...

  • Anritsu ME7840A - page 89

    q Return: Returns to the main P A TS screen. q Change Params: Returns to the User In - put screen. q Rescale: Rescales the measurement results. q Max/Min: T ype in maximum and minimum display value. Use Rescale after changing val - ues. q Autoscale Checkbox: T urns autoscale on or off. Autoscale automatically scales the mea - surement results. Mark ...

  • Anritsu ME7840A - page 90

    5-10 HARMONICS The harmonics test measures and displays the harmonics in the device under test. Procedure Prepare the ME7840A as shown in Chapter 3. Enter the appropriate test values in the displayed User Input screen. The input fields are described below: n Measurement T ype: Already set to “HA” for Har - monics – input disabled. n Start Fre ...

  • Anritsu ME7840A - page 91

    Select Start T est and observe that the test screen (Figure 5-8) appears. Button Panel Options: q Print: Prints a copy of the screen graphic to a printer . q W rite Data: Excel data is written to C:Pro - gram FilesPatsDatafiles - the file name will have a time stamp. For example: “S17-16-19-29.csv” was created on the 17th day of the month at ...

  • Anritsu ME7840A - page 92

    q Change Params: Returns to User Input screen. q Autoscale Checkbox: T urns autoscale on or off. Autoscale automatically scales the mea - surement results. q Max/Min: T ype in maximum and minimum display value. Use Rescale after changing val - ues. q Rescale: Rescales the measurement results. Markers Screen Options (left): q M1-M2 Checkbox: T urns ...

  • Anritsu ME7840A - page 93

    5-1 1 HOT S22 The output match of a power amplifier under excited conditions is critical information that has a strong bearing on efficiency , output power , stabil - ity , and often the economic viability of the design. One way of dealing with this information—primarily in power devices operating well away from compression or in somewhat matched ...

  • Anritsu ME7840A - page 94

    n T est Set Setting: Byte value for setting test set attenuation and source switch: The P A TS Hot S 22 measurement provides the following additional fea- tures: q Ability to vary the AUT input excitation power level (Source 2 of Scorpion, at Port 1) while observing the Hot S 22 response. Pressing “U” on the PC keyboard increases the input exci ...

  • Anritsu ME7840A - page 95

    Select Start T est and observe that the test screen (Figure 5-9) appears. Button Panel Options: q Print: Prints a copy of the screen graphic to a printer . q W rite Data: Excel data is written to C:Pro - gram FilesPatsDatafiles - the file name will have a time stamp. For example: “S17-16-19-29.csv” was created on the 17th day of the month at ...

  • Anritsu ME7840A - page 96

    q Change Params: Returns to User Input screen. q Autoscale Checkbox: T urns autoscale on or off. Autoscale automatically scales the mea - surement results. q Reset/Zoom: Disabled for this test. q Max/Min: T ype in maximum and minimum display value. Use Rescale after changing val - ues. q Rescale: Rescales the measurement results. Markers Screen Opt ...

  • Anritsu ME7840A - page 97

    Chapter 6 Performance V erification Pr ocedur e 6-1 INTRODUCTION The following are the specific tests that should be used to verify the per - formance of the Power Amplifier T est System (P A TS) q Source Output Level Accuracy T est q Return Loss Measurement Confidence T est q System Dynamic Range T est q Receiver Magnitude Display Linearity T est ...

  • Anritsu ME7840A - page 98

    6-4 SOURCE OUTPUT ACCURACY This test verifies the source output accuracy of the MS462XC. The test procedure uses the Power ALC V erification built-in function of the MS462XC Basic Measurement software. Setup: Step 1. Connect a GPIB cable between the power meter GPIB connec - tor and the MS462XC dedicated GPIB connector , as shown in Figure 6-1. Ste ...

  • Anritsu ME7840A - page 99

    Step 4. Connect power sensor to MS462XC Port 1 and select ST ART VERIFICA TION softkey . Step 5. V erify that Source 1 passes this test. Step 6. Press the SELECT SOURCE soft key to select 2 Step 7. Connect power sensor to MS462XC Port 3 and select ST ART VERIFICA TION soft key . Step 8. V erify that Source 2 passes this test. Step 9. Select the men ...

  • Anritsu ME7840A - page 100

    RETURN RETURN Step 5. Press Display key and select: DISPLA Y MODE SINGLE CHANNEL RETURN Step 6. Press C h4 key and then Display key . Step 7. Select: GRAPH TYPE LOG MAGNITUDE Step 8. Press Ch 1 key and select LOG MAGNITUDE. T est Procedure: Insert the Calibration Component Coefficients disk into the MS462XC floppy disk drive. Step 1. Press Cal key ...

  • Anritsu ME7840A - page 101

    Step 6. Connect SC5237 6 dB Offset T ermination to T est Port 1o f MS4782 T est Set. Record the measured value in T able 6-1 and verify the measured S 11 value is 6 ±0.4 dB. Remove SC5237 T ermination. Step 7. Connect SC5270 20 dB Offset T ermination to T est Port 1 of MS4782 T est Set. Record the measured value in T able 6-1 and verify the measur ...

  • Anritsu ME7840A - page 102

    6-6 SYSTEM DYNAMIC RANGE This test verifies the P A TS system dynamic range. T est Procedure: Step 1. Press Default key then 0 key of MS462XC to reset the system. Step 2. Press Seq key . Step 3. Select: TTL I/O. P ARALLEL OUT SETUP SET P ARALLEL OUT PORT (0-255) Step 4. Change the value from 0 to 8. Step 5. Select: RETURN RETURN Step 6. Press Ch 3 ...

  • Anritsu ME7840A - page 103

    PERFORM CAL: 2 PORT CAL METHOD: ST ANDARD LINE TYPE: COAXIAL SELECT CALIBRA TION TYPE: TRANSMISSION FREQUENCY RESPONSE P A TH: FOR WARD P A TH (S 21 ) USE OF ISOLA TION IN CALIBRA TION: INCLUDE FREQUENCY RANGE: 800 MHz T O 2400 MHz SELECT CALIBRA TION DA T A POINTS: NORMAL DA T A POINTS: 401 POINTS TEST SIGNAL/POR T 1 POWER: 10 dBm Step 10. After t ...

  • Anritsu ME7840A - page 104

    6-7 RECEIVER DISPLA Y LINEARITY This test verifies the magnitude display linearity of the receiver . Setup: Step 1. Setup the equipment as shown above. Step 2. Allow the system to warm up for at least one hour . Step 3. Press Default key then 0 key . Step 4. Press Seq key and select: TTL I/O. P ARALLEL OUT SETUP SET P ARALLEL OUT PORT (0-255) Step ...

  • Anritsu ME7840A - page 105

    Step 9. T oggle C.W . MODE OFF soft key to C.W . MODE ON . Step 10. Change C.W . Frequency to 1 GHz. Step 1 1. Press Config key and select: DA T A POINTS POINTS DRA W IN C.W . – 51 POINT(S) Step 12. Press Av g key and select: SELECT I.F . BANDWIDTH I. F . BW 10 Hz Step 13. Press Power key and select SOURCE 1 SETUP . Step 14. Adjust SOURCE 1 POWER ...

  • Anritsu ME7840A - page 106

    Step 25. Press Power key and select SOURCE 1 SETUP . Step 26. Set PORT 1 A TTN to 10 dB. Record the Power Meter Readout and the Marker 1 Readout to the table on next page. Repeat this step for other PORT 1 A TTN setting listed in the table. Step 27. V erify if the difference of the two measured values are within specifications (T able 6-2). RECEIVE ...

  • Anritsu ME7840A - page 107

    Chapter 7 Pr eamplifier Operations 7-1 INTRODUCTION This chapter describes the use of external preamplifiers to boost signal power at the input of an test amplifier . 7-2 USE OF EXTERNAL PREAMPLIFIERS External preamplifiers can be used to boost signal level at the input of the Amplifier Under T est (AUT). The maximum power per tone at the AUT input ...

  • Anritsu ME7840A - page 108

    an approximate source power (at the Scorpion Port 1, Port 3, or both). For tests involving power sweep, the MS7840A enables a linear power calibration over the desired power range (and the frequency range). For this calibration, the user can specify his target power start and stop val - ues and a nominal power offset between this reference plane an ...

  • Anritsu ME7840A - page 109

    PREAMPLIFIER OPERATIONS USE OF EXTERNAL PREAMPLIFIERS ME7840A OMM 7-3 AUT POWER METER OPTIONAL SPECTRUM ANAL YZER SYSTEM CONTROLLER (PC) B2 RF1 B1 A1 A2 STEP A TTENUA TOR 70dB, 10dB/step POWER SUPPL Y COMBINER RF2 OPTIONAL MODULA TION SYNTHESIZER -30dB -30dB -30dB N N LIMITER N K K K K K N N N K KK N N DIRECT RECEIVER ACCESS SCORPION MS4782D T est ...

  • Anritsu ME7840A - page 110

    USE OF EXTERNAL PREAMPLIFIERS PREAMPLIFIER OPERATIONS 7-4 ME 7840A OMM AUT POWER METER OPTIONAL SPECTRUM ANAL YZER SYSTEM CONTROLLER (PC) B2 RF1 B1 A1 A2 POWER SUPPL Y COMBINER RF2 -30dB -30dB N N N K K K K K N N N K KK N N DIRECT RECEIVER ACCESS SCORPION T est Set (Partially Shown) EXT I/O SOURCE 2 SOURCE 1 STEP A TTENUAT ORS 70dB, 10dB/step RF3 F ...

  • Anritsu ME7840A - page 111

    Appendix A Calibration Specification Files A-1 INTRODUCTION This appendix describes the Calibration Specification Files that are used to specify calibrations that are performed when accessing “T ools,” “Run Calibration Procedures” in the P A TS software. It lists file types, describes how to open the files for edit, provides technical detai ...

  • Anritsu ME7840A - page 112

    TECHNICAL DETAILS CAL SPECIFICATION FILES A-2 ME 7840A OMM I GENERAL 1. General Structure of file: 2. General syntax: a. Comma separates all fields. b. Spaces before or after a keyword or value in each field is acceptable. c. Place comma between last value and comment. d. Place a "//" before any comment. e. MUST have "END" as la ...

  • Anritsu ME7840A - page 113

    CAL SPECIFICATION FILES TECHNICAL DETAILS ME7840A OMM A-3 k. In each calibration specification/section, use the following structure: 3. Header: 4. “T ype” Keywords, Legal V alues: a. pow Power Level Calibration - linear cal - flat power cal - receiver cal b. sp S Parameter Calibration - 12 term - Hot S 22 - Autocal c. imd Instrument Setup Sets ...

  • Anritsu ME7840A - page 114

    A-4 ME 7840A OMM TECHNICAL DETAILS CAL SPECIFICATION FILES II CALIBRA TION SPECIFICA TIONS, TYPES 1. pow (Power Level) Equipment Required: Power meter connected to Dedicated GPIB port Power sensor with appropriate frequency range Throughline Header Section Keywords ncal Number of calibration sequences of this type to run. For type = pow , each sequ ...

  • Anritsu ME7840A - page 115

    CAL SPECIFICATION FILES TECHNICAL DETAILS ME7840A OMM A-5 Detail Section Keywords CAL, 1 First calibration name, level Name of *.cal file to save when calibration sequence completes. f1 Start frequency in MHz for global calibration NOTE: Set a wide frequency range. Miminum value is 10; ensure f1 is less than 2. f2 Stop frequency in MHz. Example Ref ...

  • Anritsu ME7840A - page 116

    TECHNICAL DETAILS CAL SPECIFICATION FILES A-6 ME 7840A OMM 2. sp and Hot S 22 (S Parameters & Hot S 22 ) Header Section Keywords: ncal Number of calibration sequences of this type to run. npts Number of frequency points to use during the calibration sequence. Increasing this number will increase calibration time. Legal values are (3, 15, 51, 10 ...

  • Anritsu ME7840A - page 117

    CAL SPECIFICATION FILES TECHNICAL DETAILS ME7840A OMM A-7 <pow2> For Hot S 22 only . Source 2 power setting. f1 Start frequency in MHz for global calibration NOTE: Set a wide frequency range. Miminum value is 10; ensure f1 is less than 2. f2 Stop frequency in MHz. Example Refer to Figure A-2 for Hot S 22 (page A-12) and Figure A-4 (page A-14) ...

  • Anritsu ME7840A - page 118

    TECHNICAL DETAILS CAL SPECIFICATION FILES A-8 ME 7840A OMM 3. powsweep (Power Sweep - IMD mode or 2-tone mode): Header Section Keywords: ncal Number of calibration sequences of this type to run. npts NOTE: npts is IGNORED. Always uses 3, which refers to number of frequency values allowed (must be 3 in current software). if IF bandwidth: Legal value ...

  • Anritsu ME7840A - page 119

    CAL SPECIFICATION FILES TECHNICAL DETAILS ME7840A OMM A-9 4. harm (Harmonics): Header Section Keywords: ncal Number of calibration sequences of this type to run. if IF bandwidth: Legal values are 10, 30, 100, 300, 1000, 3000, 10000, 30000 avgs Averages: (Optional). Software will set averaging to value (and turn on) during cal. pow Start power level ...

  • Anritsu ME7840A - page 120

    TECHNICAL DETAILS CAL SPECIFICATION FILES A-10 ME 7840A OMM 5. imd (Imodulation Distortion): Header Section Keywords: ncal Number of calibration sequences of this type to run. pow Start power level for sweep: Set on for Source 1. if IF bandwidth: Legal values are 10, 30, 100, 300, 1000, 3000, 10000, 30000 avgs Averages: (Optional). Software will se ...

  • Anritsu ME7840A - page 121

    A-5 POWER LEVE.TXT The Power Level.txt file is shown in Figure A-1 A-6 CAL FILES SETUP .TXT The Cal Files Setup.txt file is shown in Figure A-2. CAL SPECIFICATION FILES POWER LEVE.TXT ME7840A OMM A-11 ndut,1, //Number of DUTs type,pow, //Power Level Calibration (requires Power Meter) //Perform Global Linear, Flat Power and Receiver Calibrations POW ...

  • Anritsu ME7840A - page 122

    CAL FILES SETUP.TXT CAL SPECIFICATION FILES A-12 ME 7840A OMM ndut,1, //Number of DUTs type,powsweep type,harm type,imd POWSWEEP_START ncal,1, //Number of Power Sweep calibrations npts,3, //Number of points (default for all) if,3000, //IF default avgs,20, //averages default pow,-15, //powerlevel start default switch,8, //switch setting CAL,1, name, ...

  • Anritsu ME7840A - page 123

    A-7 HOT S 22 .TXT The HotS 22 .txt file is shown in Figure A-3. CAL SPECIFICATION FILES HOT S 22 .TXT ME7840A OMM A-13 ndut,1, //Number of DUTs type,sp, //S Parameter calibration SP_START ncal,1, //Number of S-Param calibrations npts,201, //Number of points (default for all) con,FN,FN, //Connector Type Options MN,FN,MK,FK,MS,FS if,3000, //IF defaul ...

  • Anritsu ME7840A - page 124

    A-8 S-P ARAMETERS. TXT The S-Parameters.txt file is shown in Figure A-4. S-PARAMETERS. TXT CAL SPECIFICATION FILES A-14 ME 7840A OMM ndut,1, //Number of DUTs type,sp, //S Parameter calibration SP_START ncal,1, //Number of S-Param calibrations npts,201, //Number of points (default for all) con,FN,FN, //Connector Type Options MN,FN,MK,FK,MS,FS if,300 ...

  • Anritsu ME7840A - page 125

    Index ! Cal Files Setup Calibration Specification File . A-12 A ADAPTIVE P STOP ................1 - 9 C CAL FILE SETUP .............. 4-15 - 19 CAL FILES SETUP .TXT ..........A - 1 1 - 1 2 collateral functions ................3 - 4 COLLA TERAL FUNCTIONS .........3 - 5 - 8 Connectors and Ports ...............1 - 7 CONVENTIONS ................ 1-2,6- ...

  • Anritsu ME7840A - page 126

    RETURN LOSS CONFIDENCE TEST ....6 - 3 - 5 S SCOPE OF THIS MANUAL ............1 - 1 SERIAL NUMBER ................1 - 2 SER VICE CENTERS ..............2 - 8 - 9 Single Frequency Power Sweep ...... 5-10,5-22 Software Organization ...............3 - 4 SOFTW ARE STRUCTURE ...........1 - 1 1 SOURCE OUTPUT ACCURACY .........6 - 2 S-P ARAMETER TESTS ..... ...

Fabbricante Anritsu Categoria Stereo Amplifier

I documenti che otteniamo dal fabbricante del dispositivoAnritsu ME7840A possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Anritsu
- manuali per l’uso ME7840A
- schede prodotto Anritsu
- opuscoli
- o etichette energetiche Anritsu ME7840A
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Anritsu ME7840A.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Anritsu ME7840A, manuali di manutenzione, brevi manuali o manuali utente Anritsu ME7840A. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Anritsu ME7840A.

Manuali per l’uso simili

Manuale completo del dispositivo Anritsu ME7840A, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Anritsu ME7840A dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Anritsu ME7840A.

Il manuale per l’uso completo Anritsu, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Anritsu ME7840A - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Anritsu ME7840A, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Anritsu ME7840A che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Anritsu ME7840A - che ci dovrebbe aiutare nei primi passi durante l'uso Anritsu ME7840A
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Anritsu ME7840A
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Anritsu ME7840A in un dato paese, se il problema non può essere risolto da noi stessi.