    EU Quantitative measurement equipment for flare in the anterior chamber Laser Flar e Meter KO W A K O WA FM- 600 ...

  • Kowa fm600 - page 2

    1 War ning War ning Caution Caution INTRODUCTION If the instrument should be operated wrongly, there may incur a risk of causing death or serious injury. If the instrument should be operated wrongly, there may result in a bodily injury (not so serious as to cause death) ✽ 1 or damage to property ✽ 2 . INTRODUCTION Accept our congratulations on ...

  • Kowa fm600 - page 3

    2 War ning War ning If any abnormal smell or sound, or overheating or smoke should be detected, be sure to turn OFF the main power immediately and then unplug the instrument from the power outlet. Continued use of the instrument may cause the instrument to malfunction or cause a fire. Contact your Kowa dealer where you have purchased the instrument ...

  • Kowa fm600 - page 4

    3 Description of Labels and Symbols Location of Cautionary Marking Before replacing fuse, be sure to turn OFF the main power and unplug the instrument. Removing the fuse holder cover without unplugging the instrument may cause electrical shock. Type B applied parts (degree of protection of applied part against electric shock). Alternative Current O ...

  • Kowa fm600 - page 5

    4 • Handle the instrument with care, and do not ap- ply strong shock to it. Also, do not leave any objects on the instru- ment. • Always cover the system when not in use in or- der to protect its components. • Avoid high temperature and humidity, direct sunlight, and dust when shipping, installing, and storing the instrument. Strictly observe ...

  • Kowa fm600 - page 6

    5 1. Instrument should be operated only by qualified personnel. 2. The following precautions must be observed when installing the instrument: (1) Install at a location free of moisture. (2) Install at a location where there is no risk of detrimental effects caused by atmospheric pressure, temperature, humidity, ventilation, sunlight, dust, salt, or ...

  • Kowa fm600 - page 7

    6 Accessories Power cable (1) Fuse (2) Dust cover (1) Chin-rest paper (1) Chin-rest’s pin (2) Printer paper (2) Instruction Manual (1) ...

  • Kowa fm600 - page 8

    7 Table of Contents Introduction ................................................................................................................... ............. 1 Precautions Concerning Use .................................................................................................... 4 Precautions (Safety Precautions and Hazar d Prevention) ...

  • Kowa fm600 - page 9

    8 7. Tips for measurements .................................................................................................... 24 7-1 How to b lock laser beams .................................................................................................... ...................... 24 7-2 W arning beep dur ing BG check ........................... ...

  • Kowa fm600 - page 10

    9 Weak scattered light induced by laser beam entered into the anterior chamber is detected and used for measurements. It is known that intensity of the scattered light is proportional to the protein level contained in aqueous humor of the anterior chamber. However, intensity of the scattered light may vary if there is a difference in the protein co ...

  • Kowa fm600 - page 11

    10 2. Names and functions of components ■ 2-1 Names and functions of components 1 Power switch : OFF : ON 2 Power supply inlet Where a power supply cord is connected. A fuse holder is provided. 3 External device I/O terminal (for a communication cable) Used for connecting a PC to which measurement data outputs are transmitted. 4 External device I ...

  • Kowa fm600 - page 12

    11 8 Photoreceiver lens The lens that mainly receives scattered light. 9 Front camera and Eye fixation lamp window Used for monitoring a patient’s eye. The patient may see the eye fixation lamp inside. 0 Light emitting lens Used for emitting laser beam and alignment light. A Calibrator Used for calibrating the device. This may be used as a laser ...

  • Kowa fm600 - page 13

    12 G Self-locking screw Tightening this screw lock the Main unit. H Measurement switch Pushing this switch proceeds with Frontal view alignment screen ➔ Diagonal align- ment screen ( BG check ) ➔ Start measurement. I Print switch Pushing this switch allows contents in the display to be printed or output (only when light is lit). J Chin-rest asc ...

  • Kowa fm600 - page 14

    13 FLARE: 9. 5 WA: C N: 5 ID:1234567890 L FLARE: 9. 5 WA: C N: 5 ID:1234567890 L List List Stop ■ 2-2 Screen display details Detail of each screen display ■ Frontal view alignment screen ■ ■ Diagonal alignment screen ■ Function switch display area Previous data Flare value Additional information Number of data saved Corneal reflection lum ...

  • Kowa fm600 - page 15

    14 Del Scale BG N:3 S/N CELL FLARE : 9 SIG:20 BG2:10 (9%) BG1:12 B S FLARE W/A NB G (%) 1 1.3 7 5.3 2 1.0 1 6.1 3 1.2 0 5.7 4 0.7 2 4.7 5 0.7 2 4.3 AV. :5.2 SD. :0.8 Cont. Edit Stat End List ■ Measurements graph screen ■ Background Signal 1 Background Signal 2 Flare Signal Flare V alue = SIG - ( BG1 + BG2 ) /2 To save no data T o change vertica ...

  • Kowa fm600 - page 16

    15 ■ 3-1 Installation Place the device on a motorized table (optional accessory). Make sure that POWER Switch shown in Fig. 1 is at OFF position (O). Connect the power cable to the power supply inlet and connect the other end of the cable to a nearby wall outlet. ■ 3-2 Printer paper installation The printer lid opens when the round dimple “Pr ...

  • Kowa fm600 - page 17

    16 ■ 4-4 Frontal view switch Use this switch to c h ange from fine adjustment (Diagonal alignment screen) to coarse adjustment (Frontal view alignment screen) or return to fine adjustment (Diagonal alignment screen) from post-measurement (Measurements graph screen) (Fig. 4). ■ 4-3 Measurement switch Use this switch to change from coarse adjustm ...

  • Kowa fm600 - page 18

    17 Stained lens may affect measurements. It is recommended to calibrate regularly in order to obtain accurate measurements. A message will appear during device start-up if the device had no calibration done for more than a month. Calibrate the device as soon as possible when the message appears. 5. Calibration Follow the calibration procedure descr ...

  • Kowa fm600 - page 19

    18 Close the light shade as shown in Fig. 9 for calibration, BG check, and measurement. 5 Fig. 5 is displayed when calibration is completed. 6 Slide the protective cap back into the original position, and turn Calibrator counterclockwise to MEASUREMENT position (Fig. 6). 5. Calibration ■ 5-2 Light shade Remove the white cap using a paper crip or ...

  • Kowa fm600 - page 20

    19 2 As shown in Fig. 4, move the luminous dot into “ ” mark and adjust the position of Main unit or Measurement unit to reduce the size of the luminous dot as small as possible. After all adjustments are completed, push “Measurement switch”. ■ 6-3 Frontal view alignment Perform the frontal view alignment while monitoring Frontal view ali ...

  • Kowa fm600 - page 21

    20 Points for alignment 1. Measurement window must be between the left and right laser beams. 2. Measurement window must be located below the Spot light. 3. Measurement window and the right laser beam must be within the pupil. Points for operation It is recommended to tilt the joy stick lightly in diagonal directions while monitoring Diagonal view ...

  • Kowa fm600 - page 22

    21 BG level indicator BG check finds out the uniformity of background lights by determining the back- ground lights around Measurement window. Level of the uniformity appears in Measurement window shown in the display. This is what we call “BG level indi- cator”. Details are shown in Figure 9. Level 5 is the highest level that indicates backgro ...

  • Kowa fm600 - page 23

    22 ● Waveform evaluation Flare is compared with |BG1 - BG2 | per measurement. (SIG – BBG ✽ ) - |BG1 - BG2 | > 0 is acceptable and no message is displayed. (SIG – BBG ✽ ) - |BG1 - BG2 | 0 is unacceptable and the message “ Inappropriate Data Form. Delete? ” is displayed. Select “ Del ” to delete or “ Store ” to save. ✽ BBG ...

  • Kowa fm600 - page 24

    23 N BG (%) FLARE W/A 1 9.8 17 20 BC 2 8.5 6 22 3 8.0 9 21 4 9.0 8 19 AV: 20.5 SD: 1.1 Cont Edit Stat End N BG % FLARE W/A 1 9.8 17 20 BC 2 8.5 6 22 3 8.0 9 21 4 9.0 8 19 AV: 20.5 SD: 1.1 Del Set N BG (%) FLARE W/A 1 0.7 7.4 82.9 2 0.9 23.0 77.6 B 3 0.9 55.6 81.8 B AV : 80.8 SD : 2.8 NAME: SEX: M / F ID: Eye: R YY / MM / DD HH:MM KOWA FM-600 Ver. 1 ...

  • Kowa fm600 - page 25

    24 ■ 7-4 Recommended procedures when BG check obtains no good result Reading on the eye. BG check may result in good without using the alignment mark (Fig. 4). Important point is to align the device so that no iris image or reflected glare comes into Measurement window (see ■ 6-4 Diagonal view alignment). ■ 7-3 Recommended procedures when BG ...

  • Kowa fm600 - page 26

    25 ■ 7-5 Recommended procedures when measured waveforms are offset to left or right You may encounter waveform offset to the left as shown in Fig. 6 when attempting a measurement at an excessively lower part of the anterior chamber. You can not obtain a correct flare value in this condition. 1 Select “ON” for “Laser : Position” as describ ...

  • Kowa fm600 - page 27

    26 8. Configuration and settings 1 Press and hold “ Print switch ” for more than 4 seconds to display the menu screen (Fig. 1). 2 Select “Setting” using “ ” or “ ” and select “Ent” to display Setting menu screen. Laser Flare Meter FM-600 Ver. 1.0 Setting Calibration End Ent Ent Setting Meas. Process Illumination Laser Output mod ...

  • Kowa fm600 - page 28

    27 Data output settings Date/Time settings Setting menu Description Selection Date Form Selecting “yy/mm/dd ” , “ mm/dd/yy ” , or “ dd/mm/yy ” . YY / MM / DD Date Displaying the date when settings were made. Values may be changed using . — — — Time(24h) Displaying the time (24h) when settings were made. — — — Sleep time sett ...

  • Kowa fm600 - page 29

    28 9. Miscellaneous ■ 9-1 ID input In order to input an ID number, “ID No. Input” must be “On” (refer to “8. Configuration and settings”). Use function keys, ten keys, or a barcode reader to input an ID number. 9-1-1 Inputting an ID number using function keys 1 Select “CLR” to clear the number shown if any 2 Use “ ” to move wi ...

  • Kowa fm600 - page 30

    29 2. Header Name or data Description Data type :LSS50 HEAD Fixed :LSS50 HEAD Fixed ID ID No. 10digits, numeric TempCorrect Temperature compensation ON/OFF 0/1 CaliCorrect Calibrator compensation ON/OFF 0/1 LD_TempCheck LDTemp check ON/OFF 0/1 LD_AlarmCheck LDAlarm check ON/OFF 0/1 asterisk Calibration execution request OFF/ON 0/1 0 Data type (grap ...

  • Kowa fm600 - page 31

    30 ■ 9-3 Display messages This section explains the massages that appear in the screens. LD Temp Check Please wait Latest Calibration Data: (YY/MM/DD HH:mm) Calibrate Now? Laser Attenuating… If Measure / Calibrate Set the positions of Cali. Tool as follows. Measure : Measurement Calibrate : Calibration ALERT Can't hold both R&L eye dat ...

  • Kowa fm600 - page 32

    31 9. Miscellaneous ■ 9-4 Function switches Details of function switches are described in this section. Function switch Description Ent Set End Yes No Stop Store Del Moves the cursor upwards. Moves the cursor downwards. Moves the cursor to the left or right. Confirms the selected setting or item. Confirms the setting. Terminates the current setti ...

  • Kowa fm600 - page 33

    32 Error1 DARK Error2 PMT sens Error3 PMT Temp. Alarm Error4 LD Temp. Uncontrolled Error5 Printer Please Check Printer Paper Error6 – 11 DARK count exceeded the limit. Sensitivity of the photoreceiver ele- ment dropped below the limit. Temperature of the photoreceiver ele- ment exceeded the limit. Temperature of the laser source be- came uncontro ...

  • Kowa fm600 - page 34

    33 11. Maintenance/Inspection This instrument is a controlled medical device, of which the measurements obtained using this instrument may be affected by daily maintenance and inspection. In order to use this instrument safely and correctly, carefully read the items listed in this section. ■ 11-1 Daily maintenance 1 Perform a calibration of this ...

  • Kowa fm600 - page 35

    34 War ning War ning ■ 11-4 Chin-rest paper installation Place a sheet of Chin-rest paper on Chin-rest and hold the sheet with Chin-rest pins. Use a sheet of Chin-rest paper per patient and replace it with a new sheet for a next patient. When replacing a fuse, be sure to turn OFF the Power switch and unplug instrument from the power outlet. Doing ...

  • Kowa fm600 - page 36

    35 Scanning laser source Photoreceiver element Measurement range Measurement accuracy Measurement area Measurement time Measurement distance Printer Monitor Energy saving function Interface Power supply Dimension/Weight Semiconductor laser, Wave length: 635nm, Intensity: 35 ± 15 µ W A photomultiplier tube 1 - 500 Photon count/ms ± 5%(under the c ...

  • Kowa fm600 - page 37

    36 13. Electromagnetic Compatibility This device is a medical electrical instrument. The medical electrical instrument requires special care concerning electromagnetic compatibility (EMC). This section describes its suitability in terms of electromagnetic compatibility of this instrument. When installing or using this instrument, please read carefu ...

  • Kowa fm600 - page 38

    37 [ Compliance verification and guidance ] Guidance and manufacturer's declaration - electromagnetic emissions The KOWA FM-600 is intended for use in the electromagnetic environment specified below. The customer or the user of the KOWA FM-600 should assure that it is used in such an environment. Emissions test Compliance Electromagnetic envir ...

  • Kowa fm600 - page 39

    38 Guidance and manufacturer's declaration - electromagnetic immunity The KOWA FM-600 is intended for use in the electromagnetic environment specified below. The customer or the user of the KOWA FM - 600 should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guida ...

  • Kowa fm600 - page 40

    39 • The relative spectral output of the instrument • The spectrally-weighted photochemical source radiance, both aphakic L B and aphakic L A L B = 0.0352 mW/(cm 2 • sr) 305nm to 700nm L A = 0.0352 mW/(cm 2 • sr) 305nm to 700nm (informative) Spectrally weighted photochemical radiances L B and L A give a measure of the potential that exists ...

  • Kowa fm600 - page 41

    40 For EU market. ...

  • Kowa fm600 - page 42

    41 ...

  • Kowa fm600 - page 43

    © 2006 Kowa Company, Ltd. All rights reserved. LSS50 V1.0 E 061208 MS Printed in Japan Printed on recycled paper. Immermannstrasse 43B 40210 Duesseldorf F.R. Germany Phone: 49(211) 1793540 Facsimile: 49(211) 161952 Hamamatsu Factory 3-1, Shinmiyakoda 1-chome, Hamamatsu city, Shizuoka Pref., 431-2103 Japan 4-14, Nihonbashi-honcho 3-chome, Chuo-ku, ...

Fabbricante Kowa Categoria Inflatable Bouncy Toy

I documenti che otteniamo dal fabbricante del dispositivoKowa fm600 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Kowa
- manuali per l’uso fm600
- schede prodotto Kowa
- opuscoli
- o etichette energetiche Kowa fm600
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Kowa fm600.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Kowa fm600, manuali di manutenzione, brevi manuali o manuali utente Kowa fm600. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Kowa fm600.

Manuali per l’uso simili

Manuale completo del dispositivo Kowa fm600, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Kowa fm600 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Kowa fm600.

Il manuale per l’uso completo Kowa, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Kowa fm600 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Kowa fm600, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Kowa fm600 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Kowa fm600 - che ci dovrebbe aiutare nei primi passi durante l'uso Kowa fm600
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Kowa fm600
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Kowa fm600 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Kowa fm600?

Utilizza il modulo sottostante

Se non hai risolto il problema con Kowa fm600 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Kowa fm600 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)