    ® 744 Documenting Process Calibrator HART  Mode Users Guide PN 691292 December 1998 © 1998 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respec tive companies. ...

  • Fluke 744 - page 2

    ...

  • Fluke 744 - page 3

    i Table of Contents Title Page Introduction ................................................................................................................... .. 1 How to Contact Fluke ..................................................................................................... 2 Connecting to a HART Transmitter ........................... ...

  • Fluke 744 - page 4

    744 HART Mode Users Guide ii Calibrating a Supported HART Transmitter .................................................................... 15 Loop Test.................................................................................................................... 16 Output Trim.................................................................... ...

  • Fluke 744 - page 5

    iii List of Tables Table Title Page 1. Supported vs. Non-Supported Instruments............................................................................ 7 ...

  • Fluke 744 - page 6

    744 HART Mode Users Guide iv ...

  • Fluke 744 - page 7

    v List of Figures Figure Title Page 1. Connecting to a HART Transmitter ........................................................................................ 5 2. 744 HART Mode Menu Tree .................................................................................................. 9 3. Block Diagram of a HART Transmitter .................... ...

  • Fluke 744 - page 8

    744 HART Mode Users Guide vi ...

  • Fluke 744 - page 9

    1 Documenting Process Calibrator HART Mode Introduction w Warning To avoid electric shock, read the Safety Information in the 744 Users Manual before you use the Fluke 744 Documenting Process Calibrator. With analog transmitters, you must make a hardware adjustment during calibration. With HART  (Highway- Addressable Remote Transducer) transmitt ...

  • Fluke 744 - page 10

    744 HART Mode Users Guide 2 How to Contact Fluke To contact Fluke, call: USA and Canada: 1-888-99-FLUKE (1-888-993-5853) Europe: +31 402-678-200 Japan: +81-3-3434-0181 Singapore: +65- * -276-6196 Anywhere in the world: +1-425-356-5500 Or visit us on the World Wide Web: www.fluke.com ...

  • Fluke 744 - page 11

    Documenting Process Calibrator HART Mode Connecting to a HART Transmitter 3 Connecting to a HART Transmitter To connect to and begin to communicate with a HART transmitter, refer to Figure 1 and proceed as follows: Note If you only want a communication connection to a transmitter that is powered in a loop, simply clip the HART alligator clips to th ...

  • Fluke 744 - page 12

    744 HART Mode Users Guide 4 pe06s.bmp The Active Device screen provides the following information for all transmitters, supported or generic: • Poll address (if not 0) • Model number and Tag ID • PV (Primary Variable) • PVAO (digital representation of the Analog Output) • PV LRV (PV Lower Range Value) • PV URV (PV Upper Range Value) • ...

  • Fluke 744 - page 13

    Documenting Process Calibrator HART Mode Connecting to a HART Transmitter 5 RTD Transmitter -PS +PS M -IN +IN S Z 7 89 456 123 0 . V MEAS SOURCE SETUP mA V Hz TC RTD CLEAR (ZERO) ENTER TC mA mA V V RTD RTD SOURCE 300V MAX 30V MAX 30V MAX 30V MAX MEAS CAT DOCUMENTING PROCESS CALIBRATOR 744 pe01c.eps Figure 1. Connecting to a HART Transmitter ...

  • Fluke 744 - page 14

    744 HART Mode Users Guide 6 Supported vs. Generic Transmitters The calibrator communicates with virtually all HART transmitters. In addition, the calibrator is programmed to use device-dependent commands for a selection of HART transmitters and their software versions. These are “supported transmitters.” All other transmitters are “generic.? ...

  • Fluke 744 - page 15

    Documenting Process Calibrator HART Mode Supported vs. Generic Transmitters 7 Table 1. Supported vs. Generic Transmitters (Devices) Menu Operation Supported Transmitter Generic Transmitter Single Point Multidrop Single Point Multidrop Top Level Active Device screen •• • • Setup Basic (read/write, cloning capability) •• • • Sensor (r ...

  • Fluke 744 - page 16

    744 HART Mode Users Guide 8 Communication Operations Figure 2 shows the HART Mode menu tree. Availability of some elements in the menus depends on which transmitter you are using, and whether the transmitter is configured by itself on a current loop or in a multidrop connection. Viewing Process Variables From the Active Device screen press the Proc ...

  • Fluke 744 - page 17

    Documenting Process Calibrator HART Mode Communication Operations 9 • Write protect • Alarm state • HART poll address • HART burst mode • HART burst command • Manufacturer • Model • Device HAR T ID • Software revision • Hardware revision • Number of preambles • Software version • Final assembly number •T a g • Message ...

  • Fluke 744 - page 18

    744 HART Mode Users Guide 10 Sensor From the Active Device screen press the Setup and Sensor softkeys to access the read-only Sensor Setup screen. This is where you can view information about the sensor in the transmitter, including serial number, limits, and span. The limits shown are the absolute limits for the sensor. (The Upper Range Value (URV ...

  • Fluke 744 - page 19

    Documenting Process Calibrator HART Mode Communication Operations 11 HART Output From the Active Device screen press the Setup and HART Output softkeys to access the read/write HART Output screen. Here you can change the Poll Address (0 = single transmitter, any other address = multidrop), and control burst mode. pe12s.bmp HART Information From the ...

  • Fluke 744 - page 20

    744 HART Mode Users Guide 12 Service Operations The Service softkey provides access to Loop Test, Pressure Zero Trim (where applicable), Output Trim, and Sensor Trim operations. For generic transmitters, only Loop Test, Output Trim, and Pressure Zero Trim are available. (See Table 1.) The trim (adjustment) operations are described later in this man ...

  • Fluke 744 - page 21

    Documenting Process Calibrator HART Mode Understanding HART Calibration 13 Understanding HART Calibration An analog transmitter has one stage of electrical conversion from a measured physical parameter to a 4-20 mA current loop output. A HART transmitter has the three stages shown in Figure 3. Depending on how the transmitter is used in your applic ...

  • Fluke 744 - page 22

    744 HART Mode Users Guide 14 HART transmitter Computation Input Physical variable 4-20mA loop Sensor trim Pressure zero trim Loop test Output trim Output PV PV AO 744 Calibrator pe02f.eps Figure 3. Block Diagram of a HART Transmitter ...

  • Fluke 744 - page 23

    Documenting Process Calibrator HART Mode Calibrating a Supported HART Transmitter 15 Calibrating a Supported HART Transmitter An “as found” and “as left” transmitter calibration has an easier, more automated process for HART transmitters than for analog transmitters. The calibration procedure is the same as described in the 744 Users Manual ...

  • Fluke 744 - page 24

    744 HART Mode Users Guide 16 7. Proceed with calibration as usual (described in the 744 Users Manual ). 8. Press the Adjust softkey. This returns you to HART mode, and presents you with the Service menu for the transmitter. 9. Do an Output Trim, a Sensor Trim, or both. These are Service Operations, and are described later in this manual. When you h ...

  • Fluke 744 - page 25

    Documenting Process Calibrator HART Mode Calibrating a Supported HART Transmitter 17 Output Trim An Output Trim adjusts the tranmitter’s Output stage. You can perform an Output Trim on both generic and supported transmitters. To perform an Output Trim: 1. From the Active Device screen Press the Service softkey, followed by Output Trim . pe16s.bmp ...

  • Fluke 744 - page 26

    744 HART Mode Users Guide 18 2. Use the u and d keys to select the Sensor Trim operation, and press e . 3. Follow the instructions on the display. For pressure, there is a message to connect a pressure module to the calibrator, and another to press the c key to zero the pressure module. 4. Press Continue . pe18s.bmp 5. Enter the trim value you want ...

  • Fluke 744 - page 27

    Documenting Process Calibrator HART Mode Cloning a Transmitter 19 Cloning a Transmitter Cloning copies the Basic Setup information from one transmitter to another. You can clone generic and supported transmitters. To use the calibrator to clone a transmitter: 1. Connect the calibrator to the transmitter you want to clone. You only need a communicat ...

  • Fluke 744 - page 28

    744 HART Mode Users Guide 20 ...

  • Fluke 744 - page 29

    21 Index —A— abort, 12 active device screen, 4 analog mode, 12 as found, 15 as left, 15 —B— basic setup, 8 browser screen, 6 —C— calibrating a HART transmitter, 15 cloning, 19 communication test leads, 3 —D— device identification, 10 —E— external series resistor, 3 —G— generic transmitters, 6 —H— HART information, 11 HAR ...

  • Fluke 744 - page 30

    744 HART Mode Users Guide 22 —O— output stage, 13 output stage, 16 output trim, 13, 17 —P— phone numbers for Fluke, 2 pressure module, 18 pressure zero trim, 13 process variables, 8 PV, 4 PVAO, 4, 16 —R— recall page, 19 —S— sensor, 10 sensor trim, 13, 17 serial communication, 3 series resistor, 3 service operations, 12 setup basic, ...

Fabbricante Fluke Categoria Water Dispenser

I documenti che otteniamo dal fabbricante del dispositivoFluke 744 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Fluke
- manuali per l’uso 744
- schede prodotto Fluke
- opuscoli
- o etichette energetiche Fluke 744
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Fluke 744.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Fluke 744, manuali di manutenzione, brevi manuali o manuali utente Fluke 744. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Fluke 744.

Manuali per l’uso simili

Manuale completo del dispositivo Fluke 744, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Fluke 744 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Fluke 744.

Il manuale per l’uso completo Fluke, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Fluke 744 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Fluke 744, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Fluke 744 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Fluke 744 - che ci dovrebbe aiutare nei primi passi durante l'uso Fluke 744
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Fluke 744
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Fluke 744 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Fluke 744?

Utilizza il modulo sottostante

Se non hai risolto il problema con Fluke 744 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Fluke 744 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)