Electrolux FCW 300 - page 1
820 4 1 70 04 1 2 9 111 0 GB F E USER USER s MANU s MANU AL AL FCW 200 FCW 300 V ACCINE/ ICEP A CK CHEST FREEZER ...
-
Electrolux FCW 300 - page 2
2 USER MANUAL V ACCINE/ ICEP A CK CHEST FREEZER * FCW 200 ** FCW 300 ...
-
Electrolux FCW 300 - page 3
3 INDEX 1 . GENERAL INFORMA TION ..................................................................................................page 4 2. INST ALLA TION ................................................................................................................. ...page 5-8 3. HOW T O USE ..................................................... ...
-
Electrolux FCW 300 - page 4
4 1 . GENERAL INFORMA TION PRODUCT DESCRIPTION The appliance is a vaccine and icepack chest freezer designed for storage and freezing of vaccine and icepacks for medical use mainly at high ambient temperatures. The freezer runs by compressor 230V/50Hz electrical power with refrigerant R1 34a and insulated with polyurethane f oam with cyclopentane. ...
-
Electrolux FCW 300 - page 5
5 2. INST ALL A TION Check the manufacturer's plate at the back of the appliance and make sure that the appliance has correct voltage for local supply , i.e. 230 volt (fig. 1 ). If you have received equipments of wrong voltage ratings, notify concerned authorities and keep the equipments repacked. The appliance is to be installed in a well ven ...
-
Electrolux FCW 300 - page 6
6 2. INST ALL A TION 2. EXAMINE the compressor compar tment (fig. 3). Check carefully for any damage, dislocation or looseness etc. of the components, specially at locations A, B and C as below and rectify if necessar y . L OCA TION A (f ig. 4) Check that all 4 mounting brackets of the compressor are in place and secured. L OCA TION B (f ig 5) Chec ...
-
Electrolux FCW 300 - page 7
7 2. INST ALL A TION L OCA TION C (fig. 6) Check that the outgoing electrical connections from the terminal strip are tight. 3. Refit side cover of the compressor compar tment into position refixing the screws holding the side cover (fig. 7) 4. INST ALL firmly the appliance on a level floor , preferably on wooden blocks or on a wooden platform to p ...
-
Electrolux FCW 300 - page 8
8 2. INST ALL A TION 6. Connect the leads of the power supply cord to a suitable 3-pin plug as shown below (fig. 9) Lead of cord Pin of plug Green/yellow Ear th Blue Neutral Brown Phase NOW THE APPLIANCE IS READ Y TO ST ART 1. Install and prepare the appliance as detailed under `Installation' in the previous chapter . 2.TEMPERA TURE CONTROL Th ...
-
Electrolux FCW 300 - page 9
9 3. HOW TO USE 3. CONTROL P ANEL The green light is illuminated as long as the freezer is connected. The yellow light is illuminated when the Super butt on is depressed. The red light is illuminated when the t emper ature in the freezer is higher than the thermostat setting. The red alarm light can be due to: Inser tion of a large amount of ...
-
Electrolux FCW 300 - page 10
10 3. HOW TO USE 7. If the cabinet temperature does not reach the required low level of 2-3 days, the freez er-thermost at setting can be increased to a higher setting b y tur ning its knob clockwise. 8. Af ter st oring the vaccines, keep the lid properly closed and locked (fig. 1 3). DEFROSTING & CLEANING The moisture in the air , which enters ...
-
Electrolux FCW 300 - page 11
11 4. DEFROSTIN G & CLEANING 3. Open the drainage plug at the bottom, inside the cabinet. Keep a suitable container under the drain hose to collect the defrost water (fig. 1 4). 4. Keep the lid open and allow the frost to melt completely . Never use any heat source other than hot wat er to speed up defrosting. Never use any sharp-edged instrume ...
-
Electrolux FCW 300 - page 12
D AIL Y 1. T ake t emperature readings from the ther mometers and note down the t emper ature and the time of reading. Keep the t emperature records systematically . It is suggested that minimum 2 readings should be taken (in the morning and af ternoon) preferably at the same hours of each day . If the temperature is obser ved to be below or above ...
-
Electrolux FCW 300 - page 13
13 IMPORT ANT : When an appliance is found to be not working at all or not working proper ly , see that the vaccine temperature is within the recommended limits. Do not open the lid unless ver y essential. Obser ve temperature from time to time and if you feel that temperature may e xceed the higher limit before the appliance is repaired - transfer ...
-
Electrolux FCW 300 - page 14
C) ABNORMAL NOISE IN THE APPLIANCE In case of any abnormal noise coming out from the appliance, tr y to locate the sour ce of the noise. Generally it may come from the compressor compar tment. In such cases, 1. T ake out the plug of the appliance from sock et and open the plastic grid of the compressor compar tment (see page 5, actions 1 ) 2. Exami ...
-
Electrolux FCW 300 - page 15
FCW 200 V ACCINE L OADING V accine storage capacit y 1 44 litre Vaccine st orage capacit y 264 litre FCW 200 ICEP ACK L OADING FCW 300 ICEP ACK L OADING Maximum icepack freezing with vaccine 7,2 k g/2 4hour Maximum icepack freezing with vaccine 7,2 k g/2 4hour Maximum icepack freezing without vaccine 22,8 k g/2 4hour Maximum icepack freezing withou ...
-
Electrolux FCW 300 - page 16
Printed by X erox Hungar y Ltd. 2003. 02. 1 2<None>. ...