    S8™ A uto 25 Positive Airw A y Pressure device User Guide English ...

  • ResMed S8 Auto 25 - page 2

    English S8™ A uto 25 POSITIVE AI RWA Y PRES SURE DEVICE User Guide English ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA ResMed (UK) Ltd (EU Authorised Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK ResMed Offic ...

  • ResMed S8 Auto 25 - page 3

    Cont ents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Contraindications 1 Adv erse Ef f ects 1 The S8 Auto 25 S ystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Masks 2 Humidifier 2 ResScan Data Card 2 Using the S8 Auto 25 . . . . . . . . . . . . . . . . ...

  • ResMed S8 Auto 25 - page 4

    1 Introduction English Introduction The S8 TM Auto 25 self adjusting sl eep apnoea system is indicated for the treatment of obstructive sleep apn oea (OSA) in pati ents > 30 kg. T he S8 Auto 25 is inte nded for home and hospital use. Contraindications The S8 Auto 25 is not a life support device and may stop operating during power failure or cert ...

  • ResMed S8 Auto 25 - page 5

    2 Masks The following ResMed mask systems are recommended fo r use with this device: For information on using masks, see your mask manual. For the latest avai lable masks, see www.resmed.com. Humidifier If you are experie ncing dryness of the nose, throat or mouth, the H4i heated humidifier is recommend ed for use with the S8 Auto 25. ! WARNING • ...

  • ResMed S8 Auto 25 - page 6

    3 Using the S8 Auto 25 English Using the S8 Auto 25 Setting up your S8 1 Connect the power cord to the socket at the rear of your S8 Auto 25. 2 Plug the other end of the power co rd into the power outlet. 3 Connect one end of the air tubin g firmly onto the air outlet. 4 Connect the assembled mask system to the free end of air tubi ng. For informat ...

  • ResMed S8 Auto 25 - page 7

    4 Attaching a H4i Humidifier The H4i humidifier attaches to the front of the S8 Au to 25 device to prov ide heate d humidification. This device automaticall y detects the pr esence of the H4i and no othe r accessories are required fo r its use. For more information on using your H4i, plea se refer to the H4i user guide. ! WARNING • Make sure that ...

  • ResMed S8 Auto 25 - page 8

    5 Using the S8 Auto 25 English Using the Menus The S8 Auto 25 provides a set of functions whi ch are arranged in menus and submenus. Via the LCD screen , the menus and submenus allow you to view and change the settings for a particular function. To navigate an d make selections: 1 Press or to scroll through items within a level. 2 Press to enter a ...

  • ResMed S8 Auto 25 - page 9

    6 How to Select the Mask Type Scroll to MASK and select . Press or until you see the setting you require. The following table shows the setti ng that should be selected for each mask: How to change the EPR level (CPAP mode) Expiratory Pressure Relief (EPR) can be used to help ease any discomfort you may experience when breathing out duri ng treatme ...

  • ResMed S8 Auto 25 - page 10

    7 Using the S8 Auto 25 English Starting Treatment 1 Make sure the power is on. The product name is displayed briefly on the LCD screen , then the standby (Ramp) screen appears. The key and LCD backl ights also turn on. 2 Fit your mask as d escribed in the mask user instructions. ! WARNING A mask should not be used unl ess your S8 Auto 25 is turned ...

  • ResMed S8 Auto 25 - page 11

    8 The LCD displays a mask-fit star rating from zero to five stars. Three to five stars indicates a good fit or better. Zero to two stars indicates that the mask needs to be adjusted. Reminders on the S8 LCD Your clinician may have set y our S8 Auto 2 5 to remind you about i mportant events, such as when to rep lace your mask, when to i nsert your D ...

  • ResMed S8 Auto 25 - page 12

    9 Using the S8 Auto 25 English Using the Data Card If your clinician need s to review your treatment, they will a sk you to use the Data Card to copy data from your S8 device and return the card to them. Copying Data onto a Data Card 1 Switch on your S8 and wait until you see the standby (Ramp) screen. 2 Hold the Data Card with the a rrow facing up ...

  • ResMed S8 Auto 25 - page 13

    10 Traveling with the S8 Auto 25 International Use Your S8 flow generator has an internal power adapter that ena bles it to operate in other countries. It will operate on power supplies of 100–240V and 50–60Hz. No special adjustment i s necessary, but you wil l require an approved power co rd for that country. Use on an Aircraft Please consult ...

  • ResMed S8 Auto 25 - page 14

    11 Cleaning and Maintenance English Cleaning and Maintenance You should regularly carry out the cleaning and main tenance described in this section. Refer to your mask and humidifier manuals fo r detailed instru ctions regarding the care of those devices. Daily Disconnect the air tubing and hang it in a clean, dry pla ce until next use. Weekly 1 Re ...

  • ResMed S8 Auto 25 - page 15

    12 ! WARNING • Do not wash the air filt er . The air filter is not washable or reusab le. • The air filter cover protec ts the devi ce in the even t of accidental liquid spillage onto the devic e. Ensure the air filter and air filt er cover are fit ted at all times. Servicing This product should be inspected by an authorised ResMed service cent ...

  • ResMed S8 Auto 25 - page 16

    13 T roubleshooting English Troubleshooting If there is a p roblem, try the fol lowing suggestions. If the problem cannot be solved , contact your equipment supplier or Re sMed. Do not attempt to open the device. No display Insufficient air delivered fr om the S8 device Device doe s not star t when y ou br eathe into th e mask Device does not stop ...

  • ResMed S8 Auto 25 - page 17

    14 Displays er ror message: Exxxx Call Service (wher e xxxx defines an err or) Displays er ror message: HIGH LEAK!!! Adjust Mask The follow ing message is display ed on the LCD aft er you try to update set tings or copy data to the D ata Card : Card Error Remove Card The follow ing message is display ed on the LCD after y ou try to update the setti ...

  • ResMed S8 Auto 25 - page 18

    15 T echnical Specifications English Technical Specifications Notes: • The manu fact urer rese r ves the right to c hange t hese speci fications without notice . • Pressure may be displayed in cm H 2 0 or hP a. Operating pressure range 4 to 25 cm H 2 O Maximum single fault steady state pressure 40 cm H 2 O Pressure measurement tolerance ±0.5 c ...

  • ResMed S8 Auto 25 - page 19

    16 Symbols Which Appear On The Device Caution; Follow instructions for use; Drip proof; Type CF equipment; Dangerous voltage; Class II equipment; Start/Stop and mask-fit; Manufactu rer; Environmenta l information WEEE 2002/96/E C is a European Di rective that requires th e proper dispo sal of electri cal and electro nic equipment. T his device shou ...

  • ResMed S8 Auto 25 - page 20

    17 General W arnings and Cautions English Note: At a fixed rate of supplement al oxy gen flow , the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pat tern, mask, point o f application and leak rate. • Do not use the S8 Auto 25 if there are obvious external def ects or unexplained c hanges in performanc ...

  • ResMed S8 Auto 25 - page 21

    18 Guidance and Manufactur er’ s Declaration—Electromagnetic Emissi ons and Immunity Guidance and manufactur er’ s declaration—electr omagnetic emissions The S8 A uto 25 de vice is inten ded for use in the electr omagnetic en vironment specif ied below . The cus tomer or the user of the S8 device should assure that the device is used in suc ...

  • ResMed S8 Auto 25 - page 22

    19 General W arnings and Cautions English Guidance and manufactur er ’ s declaration—electromagnetic immunity The S8 de vice is intend ed for use in the electroma gnetic envi ronment speci fied belo w . The customer or the user of th e S8 device should assure th at the device is used in such an environment. Immunity test IE C 6 0 6 01- 1- 2 t e ...

  • ResMed S8 Auto 25 - page 23

    20 Limited Warranty ResMed warrants that your ResMed product shall be free fro m defects in material and workmanship for the period specified bel ow from the date of pu rchase by the initial co nsumer. Th is wa rranty is not transferable. Note: Some models are not av ailable in all regions. If the product fails under conditions of norm al use, ResM ...

メーカー ResMed カテゴリー Respiratory Product

ResMed S8 Auto 25のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- S8 Auto 25の取扱説明書
- ResMedの製品カード
- パンフレット
- またはResMed S8 Auto 25の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、ResMed S8 Auto 25の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、ResMed S8 Auto 25の取り付け説明書、サービスマニュアル、簡易説明書、またはResMed S8 Auto 25のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、ResMed S8 Auto 25の製品を使用するにあたって最も人気のある説明書を閲覧できます。

ResMed S8 Auto 25デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがResMed S8 Auto 25を使用するのを助ける技術的文書のことです。説明書は通常、全てのResMed S8 Auto 25ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

ResMedの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. ResMed S8 Auto 25の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははResMed S8 Auto 25の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。ResMed S8 Auto 25に関してこのドキュメントで見つけることができる全てのヒントの目次
3. ResMed S8 Auto 25デバイスの基本機能を使うにあたってのヒント。 ResMed S8 Auto 25のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。ResMed S8 Auto 25に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるResMed S8 Auto 25のメーカー/サービスへの連絡先に関する情報。

ResMed S8 Auto 25についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもResMed S8 Auto 25の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがResMed S8 Auto 25で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)