    INSTRUCTION MANUAL 201/203/205/207 SERIES FLOW METERS/CONTROLLERS TELEDYNE HASTINGS INSTRUMENTS ISO 9001 CER TIFIED ...

  • Teledyne 201 - page 2

    141-072011 - 201/203 Series Page 2 of 23 Manual Print History The print history shown below lists the printing da tes of all revisions and addenda create d for this manual. The revision level letter increases alph abetically as the manual undergoes subsequent updates. Addenda, which are released between re visi ons, contain important ch ange inform ...

  • Teledyne 201 - page 3

    141-072011 - 201/203 Series Page 3 of 23 Table of Contents 1. GENERAL INFO RMATIO N ........................................................................................................... ................................. 4 1.1. F EATURES ........................................................................................................... ...

  • Teledyne 201 - page 4

    141-072011 - 201/203 Series Page 4 of 23 1. General Information The Hastings HFM-201/HFM-205/H FC-203/HFC- 207 series Mass flow meters (H FM-201/HFM- 205) and controllers (HFC-203/HFC-207) are desi gned to accurately measure and c ontrol mass flow over the range of 30 slm to 2500 slm, with out corrections or compensations for gas pressure and tempe ...

  • Teledyne 201 - page 5

    141-072011 - 201/203 Series Page 5 of 23 1.2. Specifications HFM-201/205 Accuracy 1 and Linearity ....................................................................................................... ±1% F.S. Repeatability ................................................................................................................ .. ±0.05% ...

  • Teledyne 201 - page 6

    141-072011 - 201/203 Series Page 6 of 23 1.4. Optional 4 - 20 mA Current Output An option to the sta ndard 0 - 5 VDC output is the 4 - 20 mA current outpu t that is proportional to flow. The 4 - 20 mA signal is produced from the 0 - 5 VDC ou tput of the flow meter. The current loop output is useful for r emote applications where pickup noise could ...

  • Teledyne 201 - page 7

    141-072011 - 201/203 Series Page 7 of 23 feature. More information about Hastings In struments’ power supplies can be f ound on the Hastings web site. http://www.teledyne-hi.com/p roducts/power pod-series.htm 1.5.2. Interconnecting Cables Cables are available from Hastings, in various leng ths, to connect from the 15 pin "D" connector o ...

  • Teledyne 201 - page 8

    141-072011 - 201/203 Series Page 8 of 23 2. Inst allation and Operation This section contains the necessary steps to get a new flow meter/controller into operation as quickly and easily as possible. Please read the followi ng th oroughly before attempting to in stall the instrument. 2.1. Receiving Inspection Carefully unpack the Hastings 201/203 se ...

  • Teledyne 201 - page 9

    141-072011 - 201/203 Series Page 9 of 23 2.4. Mechanical Connections The flow meter may be mounted in any position as long as the direction of gas flow through the instrument follows the arrow marked on the botto m of the flow meter ca se label. The preferred orientation is with the inlet and outlet f ittings in a horizontal pla ne (if operating wi ...

  • Teledyne 201 - page 10

    141-072011 - 201/203 Series Page 10 of 23 the ground connection at the power supply. Do not connect the m together at the flow co ntroller as the resulting crosstalk could result in flow instabilities. 24 Volt Connections Refer to the diagram at right when connecting 24 Volt units. Connect the positive lead of the power supply to pin 7 of the DE-9 ...

  • Teledyne 201 - page 11

    141-072011 - 201/203 Series Page 11 of 23 some zero offset.) The 201/203 series is intende d for use in non-condens ing environments only. Condensate or any other liquids which enter the flow meter ma y destroy its electronic componen ts. 2.6.2. Zero Check Turn the power supply on if not already energized. Allow for a 1 hour warm-up. Stop all flow ...

  • Teledyne 201 - page 12

    141-072011 - 201/203 Series Page 12 of 23 2.6.5. Operation with a Hastings power supply. There are two controls for each flow controller co nnected to a Hastings power supply. A switch labeled “OPEN; AUTO; CLOSED” (valve override THPS 400 only) and a potentiometer labeled “COMMAND”. For normal operation, the valve over ride switch will be i ...

  • Teledyne 201 - page 13

    141-072011 - 201/203 Series Page 13 of 23 2.6.7. Operation with an external sensor. (Fig. 2.2) In some instances, it might be desirable to use an e xternal sensor to provide process infor mation to the control circuitry in the flow controller. For ex ample, you might want to control the pressure in a vacuum sys tem by adjusting the rate at wh ich t ...

  • Teledyne 201 - page 14

    141-072011 - 201/203 Series Page 14 of 23 3) Step change the command voltage to 10 0%, and observe the flow through the co ntroller. If the overshoo t is too large, remove the jumper. Reset the comm and vo ltage to 10%, and allow the controller to stabilize. 4) To prevent loss of the unused jumper, place it over one pin only on JP4/JP6. 2.6.9. Oper ...

  • Teledyne 201 - page 15

    141-072011 - 201/203 Series Page 15 of 23 3. Theory of Operation This section contains an overall functional de scription of HFC Flow Controllers. Detailed schematics and parts lists can be found at the end of the manual in Section 6.0. In this section and other sections throughout this manual, when a po wer supply is mentioned, it is assumed that ...

  • Teledyne 201 - page 16

    141-072011 - 201/203 Series Page 16 of 23 output is sent to the terminals on the back and to the decoding circuitry in the display which converts it to a 3-d igit output. The controller circuitry utilizes the Command and the Flow voltages as input signals. The 0 - 5VDC comma nd signal is subtracted from the 0 - 5VDC flow signal creating an error si ...

  • Teledyne 201 - page 17

    141-072011 - 201/203 Series Page 17 of 23 3.5. Valve: A solenoid valve is used as a pilot valv e to contro l a much larger pneumatic diaphragm valve. The pilot valve is an “automatic metering solenoid” va lve. While most solenoids operate in either the fully open or fully closed state, the automatic metering solenoid valve is designed to contro ...

  • Teledyne 201 - page 18

    141-072011 - 201/203 Series Page 18 of 23 4. Maintenance This section contains service and calibration info rmation. Some portions of the instrument are delicate. Use extreme care when servicing the fl ow controller. The poten tiometer positions and the electrical components referred to in the trouble shooting section can be fo und in Section 6.0 o ...

  • Teledyne 201 - page 19

    141-072011 - 201/203 Series Page 19 of 23 restriction is pneumatically close to the downstre am end of the fl ow controller. The differential pressure across the unit must be between 10-50 psig. SYMPTOM : Little or no flow, even with Valve Override switch OPEN. CAUSE : Plugged orifice. ACTION : Verify the presence of a 10-50 psig pressure across th ...

  • Teledyne 201 - page 20

    141-072011 - 201/203 Series Page 20 of 23 stability. If the instrument is not shutting co mpletely off when Valve Override switch is in the CLOSE position, the orifice may requi re approximately 1/8 turn clockwi se. 4.4. End Cap Removal: The end cap on the inlet side must be removed to gain access to the filter or shunt assembly. First shut off the ...

  • Teledyne 201 - page 21

    141-072011 - 201/203 Series Page 21 of 23 D) This formula provides approximate results th at tend to be undersized because it neglects pressure drops internal t o the flow controller, compressible gas effects and temperature effects. Multiply the result by ≈ 1. 5 to get the expected minimum orifice size that can reliably pass the desired flows at ...

  • Teledyne 201 - page 22

    141-072011 - 201/203 Series Page 22 of 23 diaphragm. Examine the o-ring on the valve top for damage if required. Install the valve top, ensuring that the ball bear ing in the side is on the downstream side. Tigh ten down the valve top evenly to insure a proper seal at the diaphragm. ...

  • Teledyne 201 - page 23

    141-072011 - 201/203 Series Page 23 of 23 5. W ARRANTY 5.1. Warranty Repair Policy Hastings Instruments warrants this product for a period of one year from the date of shipment to be free from defects in mate rial and workmanship. This warranty does not apply to defec ts or failures resulting from unauthorized modification, misuse or mishandling of ...

メーカー Teledyne カテゴリー Automobile Parts

Teledyne 201のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 201の取扱説明書
- Teledyneの製品カード
- パンフレット
- またはTeledyne 201の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Teledyne 201の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Teledyne 201の取り付け説明書、サービスマニュアル、簡易説明書、またはTeledyne 201のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Teledyne 201の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Teledyne 201デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがTeledyne 201を使用するのを助ける技術的文書のことです。説明書は通常、全てのTeledyne 201ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Teledyneの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Teledyne 201の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははTeledyne 201の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Teledyne 201に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Teledyne 201デバイスの基本機能を使うにあたってのヒント。 Teledyne 201のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Teledyne 201に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるTeledyne 201のメーカー/サービスへの連絡先に関する情報。

Teledyne 201についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもTeledyne 201の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがTeledyne 201で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)