Rite Life 394の取扱説明書

28ページ 1.01 mb
ダウンロード

ページに移動 of 28

Summary
  • Rite Life 394 - page 1

    393/394 Sensitometer Operation Manual ...

  • Rite Life 394 - page 2

    ...

  • Rite Life 394 - page 3

    i FCC This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equi pment is operated in a commercial environment. This equipment generates, uses, and can radiate radio freque ...

  • Rite Life 394 - page 4

    ii CE DECLA RATION Manufacturer’s Name X-Rite, Incorporated Manufacturer’s Address: 3100 44 th Street, S.W. Grandville, Michigan 49418 U.S.A. Model Name: Sensitometer Model No.: 393 and 394 Directive(s) Confor mance: EMC 89/336/EEC WA RNING This is a class A product. In a domestic environment this product may cause radio interf erence in which ...

  • Rite Life 394 - page 5

    iii Limited Warranty X-Rite, Incorporated warra nts each unit manufactured to be free of defects in mater ial and workmanship for a period of twelve months. I f the fault has been caused by misuse or abnorm al conditi ons of operations, repairs will be billed at a nominal cost . In this case, an estim ate will be submitted before work is started, i ...

  • Rite Life 394 - page 6

    iv Table Of Contents 1. Getting St arted ...................................... 1 Unpacking and In specti on ..................... 1 Instrument De scripti on ........................... 2 Applying Po w er ...................................... 3 2. A pplication and Procedures ................. 4 Sensitometer Monitoring For Process Contro l ........ ...

  • Rite Life 394 - page 7

    1 1. Getting Started Unpacking and Inspection After rem oving the instrument from the shipping carton, inspect for possible damage. I f any damage has occurred during shipping, im mediately contact the transportation company. Do not proceed with installation until the carrier’s agent has inspected the damage. If damage is not evident, ch eck to m ...

  • Rite Life 394 - page 8

    2 Instrument Description The Process Optimization Sensitom eter is a battery operated, dual color, single or dual (394 only ) sided exposure sensitometer, desi gned for quality control of cine and X-ray processing system s. It produces a repeatable stepped exposure on applicable film . Its ease of operation enables anyone with little instruction to ...

  • Rite Life 394 - page 9

    3 Applying Power The sensitometer is shipped from the factory with the battery installed and power switch in the “Off” (left) position—slide switch to the “On” (right) position. The battery switch for the 393 is located on the front of the instrument. T he battery switch for the 394 is located on the bottom of the instrum ent. Because the ...

  • Rite Life 394 - page 10

    4 2. Application and Procedures Sensitometer Monitoring For Process Control The sensitometer exposes film with a known quantity of light through a 21-step light m odulator . The maxim um light is emitted from Step No. 21. Each successive step emits approxim ately 70.7% of the light emitted from the step adjacent to it (.15 log exposure). The proces ...

  • Rite Life 394 - page 11

    5 Variations in processor conditions are m onitored on this step. • Contrast Index: An indicator of the slope of the straight-line portion of the D-Log E Cur ve is called Contrast Index. Select the step closest to but not larger than 2.20D. Subtract from this step the step closest to but not lower then 0.45D. Contrast Index is used to m onitor pr ...

  • Rite Life 394 - page 12

    6 The film selected to m onitor a given processor should be representative of the film used with that processor. Scheduling Sensitometric Control Every processor in use should have a separate control chart plotted to monitor its behavior . The more fr equent the data points, the better the control feedback. A control film should be run at star t-up ...

  • Rite Life 394 - page 13

    7 5. After processing of film is complete, record the date, time, and pr ocessor identification number on the film in the designated areas. Manual Data Recording Procedure Establishing Normal Speed Index, Contrast Index, and Base+Fog The norma l speed index, cont rast index, and base+fog values are established on a representative film, when the pro ...

  • Rite Life 394 - page 14

    8 NOTE: Monitor subsequent film s on the same steps selected for norma l contr ast index. • Date - Month - Day – Year • Processor Number - Processor identification • Emulsion Number - Film batch identification • Developer Type - Developer vendor identification • Fixer Type - Fixer vendor identification • Film Type - Film vendor identi ...

  • Rite Life 394 - page 15

    9 Automatic Data Recording Procedure—Using the 391 Densitometer When the X-Rite 391 densitometer is used to measure process control films, it will autom atically calculate and store the values for: speed index, contrast index, and base+fog. The unit will also store up to 32 measurem ents of film data for twelve processors. Printing Process Contro ...

  • Rite Life 394 - page 16

    10 Processor Troubleshooting When troubleshooting out of tolerance processor conditions, use Speed Index as the prim ary guide. Speed Index is the most predictable indicator for all film types. Base+Fog is predictable, but is the least sensitive. Contrast Index reacts predictably for a given set of film conditions but m a y vary from film to film. ...

  • Rite Life 394 - page 17

    11 3. Using the Sensitometer Setting Exposure Color Optimum sensitometric control occurs when the proper color light exposes the film. E xpose with the sam e color emitted from the intensifyi ng screen recomm ended by the film m anufacturer for the film being exposed. For example, when using blue em itting intensifying screen—expose in “BLUE” ...

  • Rite Life 394 - page 18

    12 Selecting Single or Dual Exposure (394 only) Depending on the type of film that is to be exposed, the sensitometer can have either single sided or double- sided exposure selected. If the film to be exposed is a single em ulsion type, slide the exposure switch to the “SINGLE” position and expose Emulsion Side Down . Expose double emulsion typ ...

  • Rite Life 394 - page 19

    13 activated. When th e green color is selected, switches A,B,C,D bracketed as green are activated. Exposure Setting A B C D 1 MIN OFF OFF OFF OFF 2 ON OFF OFF OFF 3 OFF ON OFF OFF 4 OFF OFF ON OFF 5 ON ON ON OFF 6 OFF OFF OFF ON 7 MAX OFF OFF ON ON Adjusting Exposure Setting (394) The sensitometer has the capability of seven different exposure tim ...

  • Rite Life 394 - page 20

    14 Exposing Film Expose the selected film as follows: 1. Select exposure color—blue or green. 2. Select single or dual exposure (394 only). 3. Adjust exposure setting if required. 4. Insert the film —emulsion side down if single— with the back edge against the stop, and the film centered in the unit. 5. Press the expose button down and hold u ...

  • Rite Life 394 - page 21

    15 6. Release the expose button and rem ove the film imm ediately. 7. Develop the film in the pr ocessor to be m onitored (refer to Section 2). 8. Record data on film im mediately after development (re fer to Section 2). IMPORTA NT NOTES: • Film m ust be inserted a ll the way to the back of the unit in order to be exposed properly. • The sensit ...

  • Rite Life 394 - page 22

    16 4. General Maintenance Repair Information Your 393/394 Instrum e nts ar e covered by a one-year limited warranty —excluding alkaline battery—and should be referred to the factory for repair within the warranty period. Attempts to m ake repairs within this time fram e may void the warranty. X-Rite provides a factory repair service to their cu ...

  • Rite Life 394 - page 23

    17 Cleaning the Instrument Your instrument requires very little m aintenance to achieve years of reliable operation. However, a few simple cleaning procedures should be performe d from time to tim e. General Cleaning The exterior of the instrum e nt m ay be wiped clean with a cloth dampened in water or a m ild cleaner, whenever required. NOTE: DO N ...

  • Rite Life 394 - page 24

    18 Replacing the Battery A low battery condition prevents exposures from occurring . Always Replace The Battery before ref erring the unit for service. 1. Turn Power Switch “Off” and rem ove battery access cover by sliding outward. 2. Disconnect old 9-Volt battery from the circuit and discard. WARNING: Connecting the battery leads backward may ...

  • Rite Life 394 - page 25

    19 5. Technical Specifications Design Conformance: A.N.S.I. PH2.9-1974* Unit-to-Unit Repeatability : ±.02 Log Exposure Exposure Stability: ±.02 Log Exposure per year Temperature Sensitivity: ±.02 Log Exposure from 59°F (15°C) - 86°F (30°C) Electrical Required: 9-Volt alkaline battery Light Modulation: 21-Step Wedge, 0.15D per step Blue Color ...

  • Rite Life 394 - page 26

    20 ...

  • Rite Life 394 - page 27

    ...

  • Rite Life 394 - page 28

    P/N 393-500 Rev. O-4/28/04 X-Rite, Incorporated - World Headquarters 3100 44th Street S.W . • Grandville, Michig an 49418 • USA www.x-ri te.co m • Te l: 1- 888-826-3044 Fax: 1-888-826-3045 or (616)-534-0686 International Tel: 1-888-826-3039 or (616) 534-7663 Fax: (616) 534-0723 X-Rite GmbH Stollwerckstraße 32 • 51149 Köln • Germany Tel: ...

メーカー Rite Life カテゴリー Stud Sensor

Rite Life 394のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 394の取扱説明書
- Rite Lifeの製品カード
- パンフレット
- またはRite Life 394の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Rite Life 394の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Rite Life 394の取り付け説明書、サービスマニュアル、簡易説明書、またはRite Life 394のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Rite Life 394の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Rite Life 394デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがRite Life 394を使用するのを助ける技術的文書のことです。説明書は通常、全てのRite Life 394ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Rite Lifeの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Rite Life 394の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははRite Life 394の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Rite Life 394に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Rite Life 394デバイスの基本機能を使うにあたってのヒント。 Rite Life 394のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Rite Life 394に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるRite Life 394のメーカー/サービスへの連絡先に関する情報。

Rite Life 394についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもRite Life 394の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがRite Life 394で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)