    Model 5081-P-FF/FI F OUNDA TION ® Fieldbus T wo-Wire pH/ORP T ransmitter Instruction Sheet PN 51A-5081P-FF-FI/rev .E January 201 1 ESSENTIAL INSTRUCTIONS READ THIS P AGE BEFORE PROCEEDING! Y our purchase from Rosemount Analytical, Inc. has resulted in one of the finest instruments available for your particular application. These instruments have b ...

  • Emerson 5081-P-FF/FI - page 2

    MODEL 5081-P-FF/FI SPECIFICA TIONS SPECIFICA TIONS - GENERAL Housing: Cast aluminum with epoxy coating. NEMA 4X (IP65) and NEM A 7B. Neoprene O-ring cover seals. Dimensions: 6.3 in. x 6.9 in. x 6.4 in. (160 mm x 175 mm x 161 mm); diameter 6.1 in (155 mm) Conduit Openings: 3/4 in. FNPT Reference Impedance: T ransmitter accept s high impedance (i.e. ...

  • Emerson 5081-P-FF/FI - page 3

    3 MODEL 5081-P-FF/FI WIRING SENSOR WIRING AND SET -UP Wire sensor per Figure 1. Refer to the sensor instruction manual for more details. If your sensor has an integral preamplifier or you are using a remote preamplifier , you will need to change the pream- plifier location in the program menu. Please see Figure 14. Select “trAnS” for integral p ...

  • Emerson 5081-P-FF/FI - page 4

    4 MODEL 5081-P-FF/FI WIRING FIGURE 2. Exploded View of Model 5081-P-FF/FI FIGURE 3. Wiring Diagram — Sensor with Preamplifier DWG . NO. REV . 45081P30 A ...

  • Emerson 5081-P-FF/FI - page 5

    5 WIRING THROUGH A JUNCTION BOX WITH REMOTE PREAMPLIFIER MODEL 5081-P-FF/FI WIRING FIGURE 4. Wiring diagram — sensor Model 381+ with preamplifier Wire sensor as shown in Figure 5. Refer to the sensor instruction manual for more details. FIGURE 5. Wiring Through a Junction Box with Remote Preamplifier DWG . NO. REV . 45081P28 A DWG . NO. REV . 450 ...

  • Emerson 5081-P-FF/FI - page 6

    MODEL 5081-P-FF/FI INST ALLA TION 6 INST ALLA TION UNP ACKING AND INSPECTION Inspect the shipping container . If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unp ack the container . Be sure all items shown on the packing list are present. If items are missing, notify Rosemount Analyt ...

  • Emerson 5081-P-FF/FI - page 7

    7 FIGURE 8. FM Intrinsically Safe Inst allation Label ...

  • Emerson 5081-P-FF/FI - page 8

    8 FIGURE 7. FMRC Explosion-Proof Inst allation MODEL 5081-P-FF/FI HAZARDOUS AREA INST ALLA TION HAZARDOUS AREA INST ALLA TION ...

  • Emerson 5081-P-FF/FI - page 9

    9 FIGURE 9. FM Intrinsically Safe Inst allation (1 of 2) ...

  • Emerson 5081-P-FF/FI - page 10

    10 FIGURE 9. FM Intrinsically Safe Inst allation (2 of 2) ...

  • Emerson 5081-P-FF/FI - page 11

    11 FIGURE 10. CSA Intrinsically Safe Inst allation Label ...

  • Emerson 5081-P-FF/FI - page 12

    12 FIGURE 1 1. CSA Intrinsically Safe Inst allation (1 of 2) ...

  • Emerson 5081-P-FF/FI - page 13

    FIGURE 1 1. CSA Intrinsically Safe Inst allation (2 of 2) 13 ...

  • Emerson 5081-P-FF/FI - page 14

    FIGURE 12. A TEX Intrinsically Safe Inst allation Label 14 ...

  • Emerson 5081-P-FF/FI - page 15

    FIGURE 13. A TEX Intrinsically Safe Inst allation (1 of 2) 15 ...

  • Emerson 5081-P-FF/FI - page 16

    FIGURE 14. A TEX Intrinsically Safe Inst allation (2 of 2) 16 ...

  • Emerson 5081-P-FF/FI - page 17

    17 W/O AGENCY APPROVAL REVISIO NS NO T PE RMIT TED REV REV REV REV REV REV CERTIFIED BY THIS DOCUMENT IS °C Tamb= 70 FM A 5/6/04 J. FLOCK 5/6/04 J. FLOCK .140 .125 1.30 ±. 0 2 2X.650 ±.0 1 5 2.56 ±. 0 2 2.180 .120 ±. 01 5 Ø .125 R .25 OR SUITABIL ITY FOR DIV. 2. MAY IMPAIR INTRINSIC SAFET Y 5081-P-FI-67 MODEL GRPS. E, F & G GRPS. A,B,C,D, ...

  • Emerson 5081-P-FF/FI - page 18

    18 UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN I NCHES REMOVE B URRS & SHA RP EDGES .020 MAX MACHINED FI LLET RADII .020 MAX FINISH MATERIAL PROJEC T ENGR A P VD SIZE DWG NO. REV TITLE SCALE TYPE CHECKED DRAWN THIS DWG CREATE D IN DATE DESCRIPTION BILL OF M ATERIAL QTY ITEM PART NO. RELEASE DATE ECO NO. REV LTR ECO DESCRIPTION BY DATE REVISION ...

  • Emerson 5081-P-FF/FI - page 19

    B FM ANY FM APPROVED ASSOCIATED APPARATUS ANY FM APPROVED ASSOCIATED APPARATUS ANY FM APPROVED ASSOCIATED APPARATUS ANY FM APPROVED TERMINATOR ANY FM APPROVED INTRINSICALLY SAFE APPARATUS ANY FM APPROVED TERM INA TOR ANY FM APPROVED TERMINA TOR ANY FM APPROVED INTRINSICALLY SAFE APPARATUS ANY FM APPROVED TERMINATOR ANY FM APPROVED TERMINA TOR ANY F ...

  • Emerson 5081-P-FF/FI - page 20

    20 IIC/ A,B,C,D,E,F,G IIB/ C,D,E,F ,G GROUPS 360 17.5 5 5081-P-FI 10 2.52 380 17.5 5 5081-P-FI FISCO PARAMETERS 5081-P-FI TABLE III MODEL NO. 10 5.32 Vmax (Vdc) Li (uH) Ci (nF) Pmax (W) Imax (mA) SUPPLY / SIGNAL TERM INALS TB 1-15, 16 La, Lt OR Lo; OR Lc/Rc (La/Ra OR Lo/Ro) AND Li/Ri (La/Ra OR L o/Ro) GROUPS GAS OUTPUT PA RAMETERS 25 21.69 (mH) La ...

  • Emerson 5081-P-FF/FI - page 21

    21 CLASS I, DIV. 2, GRPS A,B,C & D MODEL SA R LR34186 ROSEMO UNT ANALYT ICAL R WARNING-EXPLOSION H AZARD-DO NOT DISCONNECT WHILE CIRCUIT IS LIVE UNLESS AREA IS KNOWN TO BE NON- HAZARDOUS. WARNING-EXPLOSION HAZARD-SUBSTITU- TION OF COMPONENTS M AY IMPAIR SUITABILITY FOR CLASS I, DIV 2. CLASS I, GRPS B,C & D CLASS II, GRPS E, F & G CLASS ...

  • Emerson 5081-P-FF/FI - page 22

    22 Rosemount Analytical, Uniloc Division 2400 Ba rranca P kwy Irvine, CA 92606 UNLESS OTHERWISE SPECI FIED DIME NSIO NS AR E IN IN CHES REMOVE BUR RS & SHARP E DGES .020 M AX MACHINED FILLET RADII .020 MAX FINI SH MATE RIAL PROJE CT ENGR A PVD SIZE DWG NO. REV TITLE E P Y T E L A C S This document contains information proprietary to Rose mou nt ...

  • Emerson 5081-P-FF/FI - page 23

    23 9241502-01/ A DIMENSIONS ARE IN INCHES REMOVE BURRS & SHA RP EDGES .020 MAX MACHIN ED FILLET RADII .0 20 MAX - ITEM PART NO DESCRIPTION QT Y SCALE SIZE REV DW G NO BILL OF M ATERIAL MATERIAL FINIS H DRAW N CHECKED PROJECT ENGR APVD NOTES: UNLESS OTHERW ISE SPECIFIED REV ECO NO RELEASE DATE LTR ECO REVISIONS DESCRIPTION BY DATE CHK This docum ...

  • Emerson 5081-P-FF/FI - page 24

    24 Rosemount Analytical, Uniloc Division 2400 B arran ca Pkw y Irvin e, C A 92606 UN LESS OTH ERW ISE SPEC IFIE D DIMENSIONS ARE IN INCHES REMOVE BURR S & SHARP EDGES .020 M AX MACHINED FIL LET RADII .020 MAX FINISH MATE RIAL PROJECT ENGR AP VD SIZE DWG NO. REV TITLE SCALE TYPE This document c ontains information proprietary to Rosemount Analyt ...

  • Emerson 5081-P-FF/FI - page 25

    25 MODEL 5081-P-FF/FI INFRARED REMOTE CONTROLLER INFRARED REMOTE CONTROLLER (IRC) - KEY FUNCTIONS The infrared remote controller is used to calibrate and program the transmitter and to display diagnostic mes- sages. See figure below for a description of the function of the keys. Hold the IRC within 6 feet of the transmitter , and not more than 15 d ...

  • Emerson 5081-P-FF/FI - page 26

    26 MODEL 5081-P-FF/FI ST ART -UP - pH ST ART -UP CONFIGURA TION AND CALIBRA TION — pH T o perform a two-point buffer calibration, follow these step s: 1. Place the pH sensor in the first buffer solution. 2. On the remote, press CAL, ENTER, ENTER. 3. Use the arrow buttons to select the correct buffer value. Press ENTER. 4. Rinse the sensor and pla ...

  • Emerson 5081-P-FF/FI - page 27

    27 MODEL 5081-P-FF/FI ST ART -UP - ORP ST ART -UP CONFIGURA TION AND CALIBRA TION — ORP T o perform a one point standardization, follow these step s. 1. Place the ORP sensor in the ORP buffer solution. 2. On the Remote press CAL, ENTER, ENTER 3. Use the arrow buttons to adjust the value. Press ENTER 4. Press RESET MENU Sub-menu PROMPT Diag Messag ...

  • Emerson 5081-P-FF/FI - page 28

    MODEL 5081-P-FF/FI MENU MNEMONICS 5081-P DISPLA Y MANUAL TEXT MEANING DBMJcsBuF CALIbrAtE Calibration menu Tue S td S tandardize sensor cG2! 0! cG3 bF1 / bF2 Buffer 1 / Buf fer 2 TMPQF SLOPE Sensor slope Ipme HoLd Set hold GBVMu F AULt Set fault eQo dPn Set output dampening eJBHopTujd dIAGnoS tic Diagnostic submenu sPGGTu rOFFS t Reference offset J ...

メーカー Emerson カテゴリー Satellite Radio

Emerson 5081-P-FF/FIのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 5081-P-FF/FIの取扱説明書
- Emersonの製品カード
- パンフレット
- またはEmerson 5081-P-FF/FIの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Emerson 5081-P-FF/FIの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Emerson 5081-P-FF/FIの取り付け説明書、サービスマニュアル、簡易説明書、またはEmerson 5081-P-FF/FIのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Emerson 5081-P-FF/FIの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Emerson 5081-P-FF/FIデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがEmerson 5081-P-FF/FIを使用するのを助ける技術的文書のことです。説明書は通常、全てのEmerson 5081-P-FF/FIユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Emersonの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Emerson 5081-P-FF/FIの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははEmerson 5081-P-FF/FIの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Emerson 5081-P-FF/FIに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Emerson 5081-P-FF/FIデバイスの基本機能を使うにあたってのヒント。 Emerson 5081-P-FF/FIのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Emerson 5081-P-FF/FIに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるEmerson 5081-P-FF/FIのメーカー/サービスへの連絡先に関する情報。

Emerson 5081-P-FF/FIについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもEmerson 5081-P-FF/FIの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがEmerson 5081-P-FF/FIで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)