  • ResMed C-Series - page 1

    C-S eries T ango™ POSITI VE AIRWAY PRESSURE DEVICE User Guide English ...

  • ResMed C-Series - page 2

    Manufact urer: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 21 53 Australia Distribut ed by: ResMed Corp 900 1 Spectrum Center Boule vard San Diego CA 921 23 USA ResMed (UK) Ltd 96 Milton Park Abingdon Oxf ordshire OX1 4 4R Y UK See www .resmed.com for other ResMed locations w orldwide. For patent inf ormation, see www .resmed.com/ip C-Se ...

  • ResMed C-Series - page 3

    1 English C-S eries T ango ™ User Guide Intended Use The ResMed Tango CPAP system is indicated for the treatment of obstructive sleep apnoea (OSA) in patients weighing mo re than 30kg (66 lb). It is intended for h ome and hospital use. Contraindications Positive airway pressure therapy ma y be co ntraindicated in so me patients with the following ...

  • ResMed C-Series - page 4

    2 Parts of the Tango Your Tango comprises: • ResMed C-Series T ango flo w generator (shown below) • P o wer cord • 2 m (6’6”) air tubing. Masks The ResMed mask systems recommended for use with the Tango are: For information on using m asks, see your mask manual. Humidifiers For information on using the ResMe d C-Series Heated Humidifier o ...

  • ResMed C-Series - page 5

    3 English The Tango Interface Setting up the Tango ! CAUTION • Be careful to place the device where it cannot be bumped, and where no one will trip over the power cord. • If you put the device on the floor , ma ke sur e the area is free from dust and clear of bedding, cl othes or other objects that could bloc k the air inlet. • Make sure the ...

  • ResMed C-Series - page 6

    4 • Only Re sMed air tubing should b e us ed with the device. A dif ferent type of air tubing may alter the pressure you actually receive and reduce the effec ti veness of your treatment. • Bloc king th e hose and/or air inlet o f the device while in op eration could lead to overheatin g of the device. 1 . Connect the pow er cord to the rear of ...

  • ResMed C-Series - page 7

    5 English Cleaning and Maintenance ! WARNING Beware of electric shock. Do not immerse the flow generator or power cord in water. Always unplug the flow generator before cleaning and make sure it is dry before reconnecting . ! CAUTION • Do not hang the air t ubing in direct sunlight as it m ay harden over time and eventually crac k. • Do not use ...

  • ResMed C-Series - page 8

    6 Changing the Air Filter ! CAUTION Do not wash th e air filt er. 1 . With the power cord and air t ubing disconnected, turn the T ango upside do wn so the air inlet is facing y ou. The air filter is inside the air inlet. 2. Pull the old air filter out of the air inlet. 3. Put the new air filter into the air inlet in the direction shown b y the arr ...

  • ResMed C-Series - page 9

    7 English Technical Specifications The manufacturer reserves th e right to ch ange these specification s without notice. Operating pr essure r ange 4 to 20 cm H 2 O Dimensions (L x W x H) 21 5 mm x 189 mm x 1 1 4 mm (8.5” x 7 .4” x 4.5”) We i g h t 1 .1 k g (2.4 lb) Pow e r S u pp ly Input range 1 0 0–240V , 50/60Hz, 20V A/1 0W (t ypical po ...

  • ResMed C-Series - page 10

    8 Symbols which appear on the product At tention, consult accompanying documents T ype CF equipment Class II equipment Drip proof European Authorised Representative Canadian Standards Association Manufact urer Envi ronmental information WEEE 20 02/96/EC is a European D irective that requires the proper dispos al of elec trical and electronic equipm ...

  • ResMed C-Series - page 11

    9 English ! General Warnings and Cautions A warnin g alerts you to possible inj ury. • Read the entire guide bef ore using the T ango. • The advice in this guide should not supersede instr uctions given by the prescribing phy sician. • The T an go should only be used with masks (and connectors*) recommended by ResMed, or b y a phy sician or r ...

  • ResMed C-Series - page 12

    10 A caution explains special measures for safe and effective use of the device. • At low pressures, the flo w through the exhalation ports of your mask ma y not clear all exh aled gas from the tub ing. Some rebreathing ma y occur . • The airflo w for breathing produced b y this device can be as muc h as 6°C (1 1°F) higher than the temperatur ...

  • ResMed C-Series - page 13

    11 English Guidance and Manufacture r’s Declaration - Electromagnetic Emissions and Immunity Guidance and manufact urer’ s declar ation—electromagnetic emissions The C-Series T ango system is intended f or use in the electromagnetic en vironment specified below . The customer or the user of the C-Series T ango system should assure that the de ...

  • ResMed C-Series - page 14

    12 Guidance and manufact urer’ s declar ation—electromagnetic immunity The C-Series T ango system is intended for use in the electr omagnetic environment sp ec fied below . The customer or the user of the C-Series T ango system should assure that the device is used in suc h an environment . Immunity tes t I E C 6 0 6 01- 1- 2 tes t l e vel Comp ...

  • ResMed C-Series - page 15

    13 English Guidance and manufactur er’ s declaration—electromagnetic immunity The C-Series T ango system is intended for use in t he electromagnetic en vironment specified belo w . The customer or the user of the C-Series T ango system should assure that the device is used in suc h an environmen t. Immunity tes t IEC6060 1 -1 -2 test lev el Com ...

  • ResMed C-Series - page 16

    14 Recommended sepa ration distances between portable and mobile RF communication s equipment and the C-Series T ango syst em. The C-Series T ango system is intended for use in an en vironment in whic h radiat ed RF disturbances are controlled. T he customer or the user of the C-Series T ango system can help pre vent electromagn etic interf erence ...

  • ResMed C-Series - page 17

    15 English Limited Warranty ResMed warrants that your ResM ed product sh all be free from defects in materi al and workmanship for the period specified below fro m the date of purchase by the initial consumer. This warranty is n ot transferable. Note: Some models are not available in all regions. If the product fails under conditions of normal use, ...

  • ResMed C-Series - page 18

    Global leaders in sleep and respiratory medicine www.resmed.com 34840/2 2012-04 C-Series Tango User ROW ENG 34840 ...

メーカー ResMed カテゴリー Sleep Apnea Machine

ResMed C-Seriesのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- C-Seriesの取扱説明書
- ResMedの製品カード
- パンフレット
- またはResMed C-Seriesの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、ResMed C-Seriesの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、ResMed C-Seriesの取り付け説明書、サービスマニュアル、簡易説明書、またはResMed C-Seriesのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、ResMed C-Seriesの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

ResMed C-Seriesデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがResMed C-Seriesを使用するのを助ける技術的文書のことです。説明書は通常、全てのResMed C-Seriesユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

ResMedの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. ResMed C-Seriesの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははResMed C-Seriesの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。ResMed C-Seriesに関してこのドキュメントで見つけることができる全てのヒントの目次
3. ResMed C-Seriesデバイスの基本機能を使うにあたってのヒント。 ResMed C-Seriesのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。ResMed C-Seriesに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるResMed C-Seriesのメーカー/サービスへの連絡先に関する情報。

ResMed C-Seriesについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもResMed C-Seriesの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがResMed C-Seriesで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)