Summary
  • ResMed Pixi - page 1

    User Gui de English Pixi TM PEDIATRIC MASK ...

  • ResMed Pixi - page 2

    i Contents/ Inhalt/ Cont enu/ Dotazione/ Innehåll/ Indhold/ Innhold/ Sisältö 1 4 8 12 11 Pixi ™ pedia tric mask ...

  • ResMed Pixi - page 3

    ii Mask components/ Mask enk omponenten/ Composants du masque/ Componenti della mascher a/ Maskk omponenter/ Mask edele/ Mask ens k omponenter/ Maskin osat 13 14 15 1 3 2 11 4 6 7 8 9 10 5 16 18 12 A 17 ...

  • ResMed Pixi - page 4

    iii Mask components/ Mask enk omponenten/ Composants du masque/ Componenti della mascher a/ Maskk omponenter/ Mask edele/ Mask ens k omponenter/ Maskin osat Item/ Artikel/ Composant/ Elemento/ Artikel/ Del/ Artikkel/ Osa Description/ Besc hreibung/ Description/ Descrizione/ Beskrivning/ Beskr ...

  • ResMed Pixi - page 5

    iv Item/ Artikel/ Composant/ Elemento/ Artikel/ Del/ Artikkel/ Osa Description/ Besc hreibung/ Description/ Descrizione/ Beskrivning/ Beskri velse/ Beskrivelse/ Nimitys Par t number/ T eilenummer/ Code produit/ Codice prodotto / Artikelnummer/ Delnummer/ Delenummer/ T uotenumero 5. F ra ...

  • ResMed Pixi - page 6

    v Item/ Artikel/ Composant/ Elemento/ Artikel/ Del/ Artikkel/ Osa Description/ Besc hreibung/ Description/ Descrizione/ Beskrivning/ Beskri velse/ Beskrivelse/ Nimitys Par t number/ T eilenummer/ Code produit/ Codice prodotto / Artikelnummer/ Delnummer/ Delenummer/ T uotenumero 14 . Mid ...

  • ResMed Pixi - page 7

    1 English Pixi ™ pedia tric mask Thank y ou for c hoosing the Pixi P ediatric mask. Please read the entire manual before use. Intended use The Pixi P ediatric mask c hannels airfl ow nonin vasively to a patient from a continuous positive air wa y pressure (CP AP) or bile vel device (‘de vice’). The Pixi P ediatric mask is: • to be used by ...

  • ResMed Pixi - page 8

    2 Explanation: CP AP and bilevel devices are intended to be used with special masks (or connectors) which have vent holes to allow continuous flow of air out of the mask. When the device is turned on and functioning properly , new air from the device flushes the exhaled air out through the mask vent holes. However , when the device is not operati ...

  • ResMed Pixi - page 9

    3 English Set up 1 2 3 4 5 6 ...

  • ResMed Pixi - page 10

    4 W ARNING • Ensure that the nasal cushion does not bloc k the nostrils at any time during fit ting or use. • Do not leave long lengths of air tubing around the top of the c hild’ s bed. It could twist around the c hild’ s head or nec k while sleeping. Fitting The fi t ting illustrations sho w the sequence for obtaining the best fi t and ...

  • ResMed Pixi - page 11

    5 English Fitting 1 2 3 4 5 6 ...

  • ResMed Pixi - page 12

    6 Adjustment Pixi Hints! • T o ac hiev e optimal seal, adjust the headgear straps starting at the bot tom strap and working upw ards. • If the c hild experiences nasal dryness or irritation, using a humidifi er is recommended. Adjustment st eps 1. If necessar y , adjust the straps b y removing the V elcro and gently tightening the low er headg ...

  • ResMed Pixi - page 13

    7 English At taching the pr essure por t The pressure port is used to measure the air pressure at y our mask so that the correct pressure is set. This happens during the sleep st udy at the hospital. The pressure port is also used to deliv er oxygen. This can take place in the home or hospital environment. If necessar y , y our clinician will advis ...

  • ResMed Pixi - page 14

    8 Cushion chec k Cushion chec k steps 1–2. Ensure the cushion is not creased or folded or caught inside the soc kets of the mask frame. 3. The Pixi cushion is designed to sit mid-bridge. Ensure that the mask does not bloc k the nose or mouth at any time. 1 2 3 ...

  • ResMed Pixi - page 15

    9 English Remov al Remov al st eps 1–2. Gently hold the mask against the c hild’ s face. Unhook the quic k release clip from the low er headgear arm by pulling forward and aw ay from the arm. 3–4. Keep holding the mask and remo ve the headgear by pulling up and ov er the head towards the f ace. 1 2 3 4 ...

  • ResMed Pixi - page 16

    10 Emerg ency r emov al In the ev ent of an emergency , remo ve the mask by placing y our finger under the low er headgear strap and firmly pulling it awa y from the c hild – this will activate the quic k release clip. Do not use this remo val method under normal circumstances. Disassembly Disassembly steps 1. Detach the headgear from the mask ...

  • ResMed Pixi - page 17

    11 English Disassembly 1 2 3 4 5 ...

  • ResMed Pixi - page 18

    12 Headgear r eassembly Pixi Hints! • Y ou don ’t need to disassemble y our Pixi headgear to wash it. • The V elcro straps and plastic buc kles ha ve been color coded for ease of assembly . Headgear r eassembly steps 1. Lay the headgear and plastic buc kles out on a fl at surface as shown, ensuring the low er headgear arms are facing up. 2. ...

  • ResMed Pixi - page 19

    13 English 4 ...

  • ResMed Pixi - page 20

    14 Reassembly Reassembly steps 1. Insert eac h side of the cushion into the soc kets of the mask frame. Ensure the cushion is oriented correctly . The arro ws on the cushion should be pointing up tow ards the plug tether . The raised v ented area of the cushion should also be sit ting up against the mask frame. 2. Pull the arms of the cushion throu ...

  • ResMed Pixi - page 21

    15 English Cleaning the mask The mask and headgear should only be gently handw ashed. Daily/Af ter eac h use: • Disassemble the mask components according to the disassembly instructions. • Thoroughly clean the separated mask components (e xcluding headgear), by gently r ubbing in warm (appro ximately 30°C) water using mild soap. • T o optimi ...

  • ResMed Pixi - page 22

    16 T r oubleshooting Problem / possible cause Solution Mask is too noisy V ent is bloc k ed or partially bloc ked. If the vent requires cleaning, use a sof t bristle brush. Elbow or plug is incorrectly inserted. Remo ve the elbow and plug from the mask and then reassemble according to the instructions. Mask leaks around the face Mask is not positio ...

  • ResMed Pixi - page 23

    17 English W ARNING • The tec hnical specifications of the mask are provided for your clinician to c hec k that they are compatible with the CP AP or bilevel device. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be ef fective, optimum therapy may not be ac hieved, and leak, or varia ...

  • ResMed Pixi - page 24

    18 T ec hnical specifications Pressur e–flo w curve The mask contains passive v enting to protect against rebreathing. As a result of manuf acturing v ariations, the vent fl ow rate may v ar y . V ent flow rate (L/min) 0 10 20 30 40 50 60 70 2 4 6 8 10 12 14 16 18 2 0 22 24 26 28 30 Pressure (cm H 2 O ) Flow (L/min) 2 14 9 31 16 43 23 53 30 6 ...

  • ResMed Pixi - page 25

    19 English Notes: • The mask sy stem does not contain PVC, DEHP or phthalates. • This product does not contain natural r ubber latex. • The manuf acturer reserves the right to c hange these specifications without notice. Sto ra ge Ensure that the mask is thoroughly clean and dr y before storing it f or any length of time. Store the mask in a ...

  • ResMed Pixi - page 26

    Manufacturer: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia. Distributed by: ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA. ResMed (UK) Ltd 96 Milton Park Abingdon Oxfordshire OX14 4RY UK. See www .resmed.com for other ResMed locations worldwide. For patent information, see www .resmed.com/ip Pixi is a tr ...

  • ResMed Pixi - page 27

    Global leaders in sleep and respiratory medicine www .resmed.com 61821 5/1 20 1 1-09 Pixi USER EUR ENG 1 2 3 4 5 6 618215 ...

メーカー ResMed カテゴリー Nebulizer

ResMed Pixiのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- Pixiの取扱説明書
- ResMedの製品カード
- パンフレット
- またはResMed Pixiの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、ResMed Pixiの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、ResMed Pixiの取り付け説明書、サービスマニュアル、簡易説明書、またはResMed Pixiのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、ResMed Pixiの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

ResMed Pixiデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがResMed Pixiを使用するのを助ける技術的文書のことです。説明書は通常、全てのResMed Pixiユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

ResMedの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. ResMed Pixiの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははResMed Pixiの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。ResMed Pixiに関してこのドキュメントで見つけることができる全てのヒントの目次
3. ResMed Pixiデバイスの基本機能を使うにあたってのヒント。 ResMed Pixiのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。ResMed Pixiに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるResMed Pixiのメーカー/サービスへの連絡先に関する情報。

ResMed Pixiについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもResMed Pixiの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがResMed Pixiで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)