    Sierra Model 951 Digital Flo-Box Instruction Manual Part Number IM-951 Revision A 09-05 5 Harris Court, Building L Monterey , CA 93940 (831) 373-0200 (800) 866-0200 Fax (831) 373-4402 www .sierrainstruments.com Sierra Instruments b.v . Bijlmansweid 2 • 1934 RE Egmond a/d Hoef The Netherlands • +31 (0) 72 5071400 • Fax +31 (0) 72 5071401 Sierr ...

  • Sierra 951 - page 2

    Page 1 DESCRIPTION The Model 951 Digital Flow Box is a microprocessor-based digital indicator and power supply , capable of interfacing directly to a mass flow meter or a mass flow controller . The integrated +/-15 vdc @250ma power supply provides a well regulated, short circuit and thermal overload protected output. It is packaged in a small 1/8 D ...

  • Sierra 951 - page 3

    INST ALLA TION AND WIRING INST ALLA TION The Model 951 enclosure is designated for panel mounting in a 1/8 DIN cutout. The cutout dimensions are shown below . T o panel mount the 951, perform the following steps: 1. Rotate the four pawl screws ( outside screws in each corner) several turns counter-clockwise to retract the pawls. Make sure the pawls ...

  • Sierra 951 - page 4

    Page 3 CAL T ARE H1 H2 L1 L2 Pawl Screw (1 of 4 outside screws) These 4 screws are used to mount the instrument in a panel. Menu Switch S1 (Single Function Switch) Increment Switch S2 (Single Function Switch) Step/Min-Max Switch S3 (Dual Function Switch) Zero/Exit Switch S4 (Dual Function Switch) Calibrate Switch S5 (Single Function Switch) Annunci ...

  • Sierra 951 - page 5

    Page 4 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | J1 P1 6 +15 vdc 8 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | J2 P2 2 1 5 Digital Gnd RXD TXD SPH1 SPL1 SPH2 SPL2 12 13 14 15    ? ...

  • Sierra 951 - page 6

    Page 5 SWITCH DEFINITIONS Menu Switch (S1) The Menu switch (S1) is used to cycle through the flow controller setpoints, alarm setpoints, alarm hysteresis, calibration values and decimal point locations. The sequence is as follows. The designators, in parenthesis, identify text that will be displayed, momentarily , followed by the value. Setpoint fo ...

  • Sierra 951 - page 7

    Page 6 SWITCH DEFINITIONS (cont) Step and Min-Max Switch (S3) This switch has a dual function. When in the MENU sequence, it steps the digit to be incremented, one position to the right. Using S3 in conjunction with S2 allows rapid updating of all parameter values. When  in the MENU sequence, depressing S3 momentarily displays the maximum ...

  • Sierra 951 - page 8

    Page 7 FUNCTION DEFINITIONS The Mode switch (S1) cycles through the following functions. Flow Set Point (StPt) The Setpoint (0-5 vdc) command signal used to control the flow through a flow controller . The output voltage is 0-5 vdc. The setpoint output is defined as: Setpoint Output = StPt / CAL x 5.000 vdc +/- V offset (Zero value) Example: If the ...

  • Sierra 951 - page 9

    Page 8 FUNCTION DEFINITIONS (cont) Alarm Set Point High 2 (SPH2) SPH2 is independent and functions identically to SPH1. SPH2 has its own open collector outputs and front panel LED (H2). Alarm Set Point Low 2 (SPL2) SPL2 is independent and functions identically to SPL1. SPL2 has its own open collector outputs and front panel LED (L2). Alarm Hysteres ...

  • Sierra 951 - page 10

    Page 9 FUNCTION DEFINITIONS (CONT) Decimal Point (dP) dP allows selection of decimal point locations for display . The locations selectable are x.xxxx, xx.xxx, xxx.xx, xxxx.x and xxxxx. ...

  • Sierra 951 - page 11

    Page 10 OPERA TING INSTRUCTIONS TURN-ON INSTRUCTIONS Reference Figures 3 and 4 for proper input and output signal wiring. 1. Apply power to the instrument. 2. V erify the instrument readout blanks momentarily , then displays the value of the input signal. If the display blinks continuously it is because the Model 951 is not receiving a signal. Chec ...

  • Sierra 951 - page 12

    Page 1 1 PROGRAM INSTRUCTIONS (cont) 2. Depress MODE (S1) switch once. The text  followed by the current value of Alarm Set Point High 1 (SPH1) will be displayed. Use S2 and S3 to set Alarm Set Point High 1 to the desired value. Proceed to Step 3 to set Alarm Setpoint Low 1 (SPL1) or depress S4 to exit MODE sequence. If S4 is depressed, ...

  • Sierra 951 - page 13

    Page 12 PROGRAM INSTRUCTIONS (cont) 7. Depress MODE (S1) switch once. The text  followed by the current value of Hysteresis Low will be displayed. Repeat Step 2 to set Alarm Hysteresis Low to the desired value. The maximum value for HL is 99 counts. Proceed to Step 8 or depress S4 to exit. 8. Depress MODE (S1) switch once. The text  ...

  • Sierra 951 - page 14

    Page 13 Prior to connecting the 951 to the Flowmeter or Controller , verify that the Controller's +/-15 vdc requirement is no more than 40 0 ma. The 951 is capable of supplying up to 500ma for short periods of time ( in 1/2 or 1 hour increments). Consult the factory if the current requirement is > 350ma for prolonged periods of time. Connec ...

  • Sierra 951 - page 15

    Page 14 CALIBRA TION INSTRUCTIONS (cont) 6. Depress and hold CAL (S5) for approximately 3 seconds. The FS CAL LED should illuminate and the readout should display the FS Cal value within +/- 1 digit. 7. The CALIBRA TION sequence is now complete. NOTE: Calibration information (zero and scale factor values) are stored in non-volatile memory and are u ...

  • Sierra 951 - page 16

    Page 15 RS232 SERIAL COMMUNICA TIONS (Special Order Only) The Parity (Even or None) and the Baud Rate (2400 or 9600) are selectable by setting DIP switches SW a and SWb, accessible from the right side of the instrument case. ON OFF SW a SWb Parity (SWa) Baud Rate (SWb) ON = Even ON = 2400 OFF = None OFF = 9600 Baud Rate: Programmable for 2400 or 96 ...

  • Sierra 951 - page 17

    Page 16 RS232 SERIAL COMMUNICA TIONS (Special Order Only)       [Z] [CR] [LF] Zero the displayed reading [F] [CR] [LF] Calibrate Full Scale (Input signal is scaled to display Full Scale) [S] [counts] [CR] [LF] Set Full Scale display value. (counts range = 0 .... 99999) ...

  • Sierra 951 - page 18

    MODEL SR5 PROGRAM MENU DIAGRAM Depress MODE ST ART  Set  Depress Depress Increments Flashing Digit Steps Flashing Digit to Right  Set  Depress MODE Depress MODE OPERA TE     ...

メーカー Sierra カテゴリー Microscope & Magnifier

Sierra 951のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 951の取扱説明書
- Sierraの製品カード
- パンフレット
- またはSierra 951の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Sierra 951の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Sierra 951の取り付け説明書、サービスマニュアル、簡易説明書、またはSierra 951のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Sierra 951の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Sierra 951デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがSierra 951を使用するのを助ける技術的文書のことです。説明書は通常、全てのSierra 951ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Sierraの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Sierra 951の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははSierra 951の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Sierra 951に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Sierra 951デバイスの基本機能を使うにあたってのヒント。 Sierra 951のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Sierra 951に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるSierra 951のメーカー/サービスへの連絡先に関する情報。

Sierra 951についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもSierra 951の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがSierra 951で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)