    Sierra Monitor Corporation 1991 Tarob Court, Milpitas, CA 95035 (408) 262-6611 Applicability & Effectivity MODEL 5000 S E N T R Y GAS MONITORING SYSTEM Version 6 Sensors Service APPLICABILITY & EFFECTIVITY Effective for all Sentry systems manufactured after September 1, 1995. Instruction Manual Part Number T12001-A1 ...

  • Sierra Monitor Corporation 5000 - page 2

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 69 7.5 COMBUSTIBLE GAS SENSOR MODULE (5100-02) 7.5.1 DESCRIPTION The Combustible Gas M odule includes the sensor and electronic assembl y installed in an explosion proof housing. The sensor screws into one hub of the enclosure and plugs into the bottom electronics ca ...

  • Sierra Monitor Corporation 5000 - page 3

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 70 3. Using a Sierra Monitor M odel 1200-26 Gas Calibrator with a Si erra Monitor M odel 5358- 00 calibration adapter, or M odel 5360 Gas Delivery Fitting. Apply the calibratio n gas at a flow rate of 100 cc/min. until the signal out voltage stops changing (approx. 2 ...

  • Sierra Monitor Corporation 5000 - page 4

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 71 7.5.5 COMBUSTIBLE GAS SCALING FACTORS For combustible gas monitoring, a calibration standard of Me thane or Propane my be used in conjunct ion with scaling factors to cause Sentry concentra tion display and alarm function in % LEL scale of another gas as follows: ...

  • Sierra Monitor Corporation 5000 - page 5

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 72 7.6 OXYGEN MODULE (5100-03) 7.6.1 DESCRIPTION The Oxygen Module includes the electronic assembly i nstalled in an explosion proof housi ng and the electrochemical sensor connected to one hub of the enclosure. Cabling from the controller connects to a three pin spr ...

  • Sierra Monitor Corporation 5000 - page 6

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 73 Unscrew the old sensor from the conduit hub, screw in the new sensor and connect the wiring harness to the transmitter electronics. Repl ace the transm itter into the enclosure, make signal adjustm ents as described above, and replace the enclosure cover. After th ...

  • Sierra Monitor Corporation 5000 - page 7

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 74 7.7 CARBON MONOXIDE MODULE (5100-04) 7.7.1 DESCRIPTION The Carbon Monoxide Module includes t he sensor and electronic assembly installed i n an explosion proof housing. The sensor screws into one hub of the enclosure and plugs into the bottom electronics card via ...

  • Sierra Monitor Corporation 5000 - page 8

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 75 If sensor replacement is necessa ry remove the electronics from the housing and unplug the old sensor from the bottom board, rem ove it from the enclosure hub and reverse the procedure to install the new sensor. After the sensor is installed: • Allow one hour of ...

  • Sierra Monitor Corporation 5000 - page 9

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 76 7.8 HYDROGEN SULFIDE MODULE (5100-05) 7.8.1 DESCRIPTION Model 5100-05 Hydrogen Sulfide Sensor Module includes a sensor assembly and elect ronic assembly i nstalled in an explosion proof housi ng. The sensor assembly includes a reuseable housing and disposable elec ...

  • Sierra Monitor Corporation 5000 - page 10

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 77 1. Confirm that sy stem power has been removed. 2. Remove the transmitter electronics board from the ma in housing and unplug the sensor harness from the transmitter electronics. 3. Unscrew the sensor housing from the bottom of the enclosure 4. Hold the senso r as ...

  • Sierra Monitor Corporation 5000 - page 11

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 78 7.9 TOXIC GAS SENSOR MODULE 7.9.1 DESCRIPTION Toxic Gas Sensor Modules include t he following models and default ranges: 5100-06 Chlorine -10 PPM 5100-07 Hydrogen - 2000 PPM 5100-12 Nitrogen Dioxide - 20 PPM 5100-13 Carbon Monoxide, High R ange - 1,000 PPM 5100-10 ...

  • Sierra Monitor Corporation 5000 - page 12

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 79 The SENSOR FAILURE message for toxic gas sensors does not indicate a specific fail ure of the sensor but indicates that the sensor is not correctl y connected to the electronics. If this message appears check that the sens or harness is correctly installed to the ...

  • Sierra Monitor Corporation 5000 - page 13

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 80 7. Apply span gas. Adjust the gain potenti ometer until TP10 = correct value as described above. To increase voltage at TP10 turn the gain potentiome ter counter clockwise. 8. Remove system power and re-install the sensor and transmitter in th e enclosure. Restore ...

  • Sierra Monitor Corporation 5000 - page 14

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 81 7.10 AMMONIA SENSOR MODULE (5100-25) 7.10.1 DESCRIPTION The Amm onia Sensor Module includes a sensor assembly and electronic assembly installed in an explosion proof housing. The sensor assembly includes a rechargeable electrochemical sensor. The assembly screws i ...

  • Sierra Monitor Corporation 5000 - page 15

    SMC sierra monitor corporation Sentry Instruction Manual - Version 6 SERVICE Page: 82 The SENSOR FAILURE message for am monia sensors does not indicate a specific failure of the sensor but indicates that the sensor is not co rrectly connected to the electronics. If this message a ppears check that the sensor harness is correctly installed to the co ...

メーカー Sierra Monitor Corporation カテゴリー Carbon Monoxide Alarm

Sierra Monitor Corporation 5000のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 5000の取扱説明書
- Sierra Monitor Corporationの製品カード
- パンフレット
- またはSierra Monitor Corporation 5000の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Sierra Monitor Corporation 5000の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Sierra Monitor Corporation 5000の取り付け説明書、サービスマニュアル、簡易説明書、またはSierra Monitor Corporation 5000のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Sierra Monitor Corporation 5000の製品を使用するにあたって最も人気のある説明書を閲覧できます。

Sierra Monitor Corporation 5000デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがSierra Monitor Corporation 5000を使用するのを助ける技術的文書のことです。説明書は通常、全てのSierra Monitor Corporation 5000ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Sierra Monitor Corporationの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Sierra Monitor Corporation 5000の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははSierra Monitor Corporation 5000の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Sierra Monitor Corporation 5000に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Sierra Monitor Corporation 5000デバイスの基本機能を使うにあたってのヒント。 Sierra Monitor Corporation 5000のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Sierra Monitor Corporation 5000に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるSierra Monitor Corporation 5000のメーカー/サービスへの連絡先に関する情報。

Sierra Monitor Corporation 5000についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもSierra Monitor Corporation 5000の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがSierra Monitor Corporation 5000で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)