    MODEL 452 PROCESS OZONE SENSOR ©Teledyne Instruments Advanced Pollution Instrumentation Division (T-API) 9480 Carroll Park Drive San Diego, CA 92121-5201 TOLL-FREE: 800-324-5190 TEL: 858-657-9800 FAX: 858-657-9816 E-MAIL: api-sales@teledyne.com W E B S I T E : www. t e l e d y n e - a p i .c om ...

  • Teledyne 452 - page 2

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E i SAFETY MESSAGES NOTE The M452 contains no user serviceable parts. CAUTION There is risk of dangerous UV exposure when the cover is removed from the sensor. Take necessary precautions to avoid exposure. CAUTION High voltages exist inside the sensor. Please use caution when sensor co ...

  • Teledyne 452 - page 3

    ...

  • Teledyne 452 - page 4

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E iii T ABLE OF CONTENTS SAFETY MESSAGES ............................................................................ I TABLE OF CO NTENTS .................................................................... III LIST OF FIGURES .......................................................... ...

  • Teledyne 452 - page 5

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E iv 5 SENSOR AND SYSTEM TROUBLESHOOTING ...................... 5-1 5.1 S TATUS LED’ S ............................................................................. 5-1 5.2 S TATUS O UTPUTS ......................................................................... 5-2 5.3 S ENSOR O.K. ...

  • Teledyne 452 - page 6

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E v LIST OF FIGURES F IGURE 2-1: M452A H IGH P URITY O ZONE S ENSOR .............................. 2-1 F IGURE 2-2: P RESSURE D ROP VS . F LOW (H IGH P URITY V ERSION ) ........ 2-3 F IGURE 4-1: E LECTRICAL C ONNECTIONS .............................................. 4-2 F IGURE 4-2: D ...

  • Teledyne 452 - page 7

    ...

  • Teledyne 452 - page 8

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E vii W ARRANTY POLICY ADVANCED POLLUTION INSTRUMENTATION DIVISION 02024c Prior to shipme nt, Teledyne API equipm ent is thoroughly i nspected and tested. Should equipm ent failure occur, Teledyne API assures i ts customers that prompt service and support will be available. COVERAGE Af ...

  • Teledyne 452 - page 9

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E viii THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE REMEDIES SET FORTH IN THIS SECTION ARE THE E ...

  • Teledyne 452 - page 10

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 1-1 1 Product Descriptions 1.1 M452 Process Ozone Sensor The Teledyne API M452 is a m icroprocessor-based sensor for measuri ng the concentration of gaseous ozone in processes such as sem iconductor wafer fabrication, water treat ment, and ozone research. The M452 can be used as a fu ...

  • Teledyne 452 - page 11

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 1-2 INTENTIONALLY BLANK ...

  • Teledyne 452 - page 12

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 2-1 2 S pecifications Note: All specificat ions contained herein are subject to change wit hout notice. Please contact Teledyne API to obtain the curre nt specifications. 2.1 Mechanical Specifications 10 Digital out #3 12 Digital out common (emitters) 15 Ground 14 RS485−B 13 RS485? ...

  • Teledyne 452 - page 13

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 2-2 2.2 Performance Specifications Accuracy: Repeatability:  1% of Full Scale. 1% of Full Scale Response Time: 2 sec. To 95% Zero Drift: 1% Full Scale/m onth (non –cumulat ive) 2.3 Operating Limits Measurement R ange: 5, 10, 15, 20% w/w 25, 50, 100, 200, 400 g/Nm3 Proof pressure ...

  • Teledyne 452 - page 14

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 2-3 2.6 Pressure Drop Figure 2-2 below shows the approxim ate pressure drop from the i nlet fitting to the outlet fittin g as a function of volumetric flow rate. Pressure Drop vs. Flow y = 0.0008x 2 + 0.0048x 0.000 0.200 0.400 0.600 0.800 1.000 0 5 10 15 20 25 Flow (LPM)  P (psi) ...

  • Teledyne 452 - page 15

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 2-4 2.7 Options The M452 is availabl e with the following opti ons: 2.7.1 AC Power Adapter This option consists of a Universal Input (100-240 VAC, 50-60Hz) Power Adapter with a standard IEC320 modular AC receptacle. A North American standard 3 prong power cord i s also provided. 2.7. ...

  • Teledyne 452 - page 16

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 3-1 3 Theory of Operation The detection of ozone m olecules is based on absorption of 254 nm UV light due to an intern al el ectronic resonance of the O 3 molecule. The Model 452 uses a mercury lamp constructed so t hat a large majori ty of the light emitted is at the 254 nm wavele n ...

  • Teledyne 452 - page 17

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 3-2 The intensity of li ght is converted into a volta ge by the detector/pream p module. The volt age is converted into a num ber by a voltage-to-frequency (V/F) converter capable of 80,000 count resolution. The digitized signal, along with the other variable s, is used by the CPU to ...

  • Teledyne 452 - page 18

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 4-1 4 Inst allation 4.1 Unpacking Upon receiving the M452 pleas e verify that no appa rent shipping dam age has occurred. If damage has occurred please advise shipper fi rst, then Teledyne API. 4.2 Mechanical Installation Mount the M452 to a stabl e platform usi ng four #8-32UNC scre ...

  • Teledyne 452 - page 19

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 4-2 Analog Out - Analog Out + Power Common +15 Volt Supply +15 Volt Supply Digital In #1 Digital In #2 Digital Out # 1 RS485-A Power Commo n RS485-B Digital Out Common Digital Out #4 Digital Out #3 Digital Out #2 Power Suppl y Common +15V -+ Ozone Concentr ation Meter (0-5 VDC) or Da ...

  • Teledyne 452 - page 20

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 4-3 4.3.2 Analog Output The analog output is a 0-5 volt signal representing the ozone concentrati on measured by the sensor. The output is scaled to the concentration range that the sensor has been set to meas ure. Check the serial number label on the M452 to determ ine the concentra ...

  • Teledyne 452 - page 21

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 4-4 4.3.4 Status Outputs The M452 has four digital st atus outputs for indicati ng error status and when operational parame ters have moved out of norm al lim its. These outputs are in the form of opt o-isolated open-collector t ransistors. They can be used to drive status LED’s on ...

  • Teledyne 452 - page 22

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 4-5 4.3.5 RS232/485 Interface The M452 features a bi-directional digita l serial interface that can be used for sensor control and data acquisit i on. Please contact Tel edyne API for documentation on the use of the RS232/485 interface. ...

  • Teledyne 452 - page 23

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 4-6 4.4 Gas Connections Gas connections to the M452 are made using ¼” com pression tube fittings (M452) or ¼” VCR  face seal fittings (M452A.) The ¼” com pression fittings can be used with ¼” O.D. Stainless Steel or Teflo n  tubing. The M452 is not sensitive to flow ...

  • Teledyne 452 - page 24

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 4-7 4.5 Start-Up and Calibration Procedure 1. Verify that the proper electrical connections have been made (See Section 4.3) and apply power to the M 452. Allow the M452 to warm up for at least 15 m inutes. 2. Purge the M452 with zero gas (usually oxygen) at a mi nimum flow rate of 0 ...

  • Teledyne 452 - page 25

    ...

  • Teledyne 452 - page 26

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 5-1 5 Sensor and System T roubleshooting This chapter gives guideli nes for diagnosing system and sensor malfuncti ons using the four digital Status Outputs provided by t he M452. All troubleshooting shoul d be done after the M452 has been turned on and allowed to warm up for at leas ...

  • Teledyne 452 - page 27

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 5-2 5.2 Status Outputs Table 5-1 below describes the function of the st atus outputs. More detail s as to the meaning of t he status outputs are described in the foll owing sections. Table 5-1: Status Outputs Output # Name On State Off state 1 Sensor O.K. Normal State Reference or Me ...

  • Teledyne 452 - page 28

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 5-3 5.4 Invalid Reading The normal stat e for the Invalid Reading output i s OFF. If this output turns on, this indicates that the M452 is still operational, but a system fault or calibration fault exi sts that may make the current ozone readi ng invalid. The Invalid Reading output i ...

  • Teledyne 452 - page 29

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 5-4 5.6 Cell Dirty The normal stat e for the Cell Dirty output is OFF. If this out put turns on, then the ratio of the measure detector to the reference detector (at zero) is < 0.5. This value is calculated wh en the zero calibration is performed. When this output is on, it i ndic ...

  • Teledyne 452 - page 30

    Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev . E 5-5 INTENTIONALLY BLANK ...

Manufacturer Teledyne Category Stud Sensor

Documents that we receive from a manufacturer of a Teledyne 452 can be divided into several groups. They are, among others:
- Teledyne technical drawings
- 452 manuals
- Teledyne product data sheets
- information booklets
- or energy labels Teledyne 452
All of them are important, but the most important information from the point of view of use of the device are in the user manual Teledyne 452.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Teledyne 452, service manual, brief instructions and user manuals Teledyne 452. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Teledyne 452.

Similar manuals

A complete manual for the device Teledyne 452, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Teledyne 452 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Teledyne 452.

A complete Teledyne manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Teledyne 452 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Teledyne 452, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Teledyne 452, that we can find in the current document
3. Tips how to use the basic functions of the device Teledyne 452 - which should help us in our first steps of using Teledyne 452
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Teledyne 452