  • Agilent Technologies U2781A - page 1

    Agilent T echnologies Agilent U2781A USB Modular Instrument Chassis User’ s Guide ...

  • Agilent Technologies U2781A - page 2

    2 U2781A User’s Guide Notices © Agilent T echnologies, Inc. 2006 N o p a rt o f t h i s m a n u a l m a y be r e p r o d u c e d i n any form or by any means (including elec- tronic storage and retrieval or translation into a foreign language) without prior agree- ment and written consen t from Agilent T echnologies, Inc. as governed by United S ...

  • Agilent Technologies U2781A - page 3

    U2781A User’s Guide 3 Safety Symbols Regulatory Markings Direct current Caution, risk of danger . The CE mark shows that the product complies with all the relevant European le gal Directives (if accompanied by a year , it signifies when the design was proven). The CSA mark is a registered trademark of the Canadian Sta ndards Association. A CSA ma ...

  • Agilent Technologies U2781A - page 4

    4 U2781A User’s Guide In This Guide… This guide cont ains information of product, features, functions and information on ho w to inst all the A gilent U2781A USB modular instrument chassis. 1 Getting Acquainted In th is ch apter yo u are i ntr oduc ed to a n o ve rvie w of the pr odu ct features, applications a nd specif ications. 2 Installatio ...

  • Agilent Technologies U2781A - page 5

    U2781A User’s Guide 5 DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Generic exampl e Manufactu rer’s Name: Agilent T echnologies M icrowave Pro ducts (M) Sdn. Bhd Manufactu rer’s Address: Bayan Lepas Free Industri al Zone, 11900, Ba yan Lepas, Pen ang, Malays ia Declares under sole responsibilit y that the p ...

  • Agilent Technologies U2781A - page 6

    6 U2781A User’s Guide Product Regulations Perf or m ance Crit eria IEC 6132 6-1:1997+ A1:1998 / E N 61326-1: 1997+A1:19 98 CISPR 11:1 990 / EN 550 11:1991 – Group 1 Clas s A IEC 61000- 4-2:1995+A 1:1998 / EN 61000-4-2:1 995 (ESD 4kV CD, 8kV AD) B IEC 6100 0-4-3:1995 / EN 61000- 4-3:1995 (3V/m, 80% AM) A IEC 6100 0-4-4:1995 / EN 61000- 4-4:1995 ...

  • Agilent Technologies U2781A - page 7

    U2781A User’s Guide 7 Contents 1 Getting Acquainted 9 Introduction 10 Features 11 Applications 11 Checking the Contents 11 Product Outlook 12 General Specifications 13 Electrical Specifications 14 Mechanical Specifications 15 2 Installation 17 L-Mount Kit Installation 18 Modular Chassis Installation 19 System Requirements 20 Hardware Requirements ...

  • Agilent Technologies U2781A - page 8

    8 U2781A User’s Guide Contents Identifying Modules Location 36 Georgraphical Address 36 Modules Identification 36 Simultaneous Synchronization (SSI) 38 4 General Main tenance 41 ...

  • Agilent Technologies U2781A - page 9

    9 Agilent U2781A USB Modu lar Instrument Chassis User’s Guide Agilent T echnologies 1 Getting Acquainted Introduction 10 Features 11 Application s 11 Checking the Contents 11 Product Outlook 12 General Spec ifications 13 Electrical Specifications 14 Mechanical Specificatio ns 15 This chapter gives an int roduction an d overview of the pro duct fe ...

  • Agilent Technologies U2781A - page 10

    10 U2781A User’s Guide 1 Getting Acq uainted Introduction The U2781A USB modular instru ment chassis is a 4U height chassis with six USB module s lots. It is a port able chassis with high performanc e added value. It t argets a wide range of applications in both ind ustrial and scientif ic envi ronments. It helps y ou lo wer y our cost of test an ...

  • Agilent Technologies U2781A - page 11

    Getting Acquainted 1 U2781A User’s Guide 11 Features The ke y features of the U2781A modular instrument chassis are as fo ll ow s: • Simult aneous Synchronization (SSI) • St ar trigger • Internal and ext ernal 10 MHz reference clock • Trigg er in and trigger out signals • SCPI commands • IVI- COM driver com patibility • USBTMC 488.2 ...

  • Agilent Technologies U2781A - page 12

    12 U2781A User’s Guide 1 Getting Acq uainted Product Outlook Front V iew Rear View Module L-Moun t Kit Bumper ON/OFF T rigger In USB Port T rigger O ut Ext 10 MHz AC I nput 270 mm 197 mm switch ...

  • Agilent Technologies U2781A - page 13

    Getting Acquainted 1 U2781A User’s Guide 13 General Specifications REMOTE INTERF ACE • USB 2.0 High Speed • USBTMC Class Device POWER CONSUMPTION • 400 V A maximum • Installati on Category II OPERA TING ENVIRONMENT • Operating temperature from 0 ° C to +55 °C • Relative humidity at 15% to 8 5% RH (non-condensing) • Altitude up to ...

  • Agilent Technologies U2781A - page 14

    14 U2781A User’s Guide 1 Getting Acq uainted Electrical Specifications Power Supply AC Input Input voltage ran ge 100 to 240 V AC Input freq uency range 50 to 60 Hz Power consumption 400 V A maximum Effiency 75% Power Supply DC Output Output r ated voltage 12 VDC Max output rated current 16.7 A Max out put rated powe r 200 W Over voltage protecti ...

  • Agilent Technologies U2781A - page 15

    Getting Acquainted 1 U2781A User’s Guide 15 Mechanical Specifications Physical Layout Number of USB module slots 6 Dimension of e ach module slot 25.40 mm (W) x 17 4.54 mm (D) x 105.00 mm (H) Dimension of chassis 270.00 mm (W) x 2 71.20 mm (D) x 197.00 mm (H) W eight 3.7 kg Power LED ON/OFF type USB Backplane Connector 55 pins Ernet male type C I ...

  • Agilent Technologies U2781A - page 16

    16 U2781A User’s Guide 1 Getting Acq uainted ...

  • Agilent Technologies U2781A - page 17

    17 Agilent U2781A USB Modu lar Instrument Chassis User’s Guide Agilent T echnologies 2 Installation L-Mount Kit Inst allation 18 Modular Chassis Installation 19 System Requirements 20 Hardware Requirements 20 Software Requir ements 21 DAQ Driver Installation 21 Hardware V erification 25 Agilent Measurement Manage r Software Installation 25 This c ...

  • Agilent Technologies U2781A - page 18

    18 U2781A User’s Guide 2 Installation L-Mount Kit Installation In order to slot in the USB module into t he U2781A USB modular instrument chassis, you need to f ix the L - Mount kit on both sides of the U2300A series D A Q. Bel o w are sim pl e ins truct ions to ins tall th e L - M ount kit o nto th e U2300A series D A Q. The instructions in page ...

  • Agilent Technologies U2781A - page 19

    Installation 2 U2781A User’s Guide 19 Modular Chassis Installation Fig ures bel o w sho w the i nstal lat ion of the USB m odu le in to the U2781A modular instrument chassis. Follo w the ar ro ws sho wn below . chassis screws onto the modular Secure both the thumb Step 2 Step 1 Slot in the USB module into the modular chassis. Make sure that the 5 ...

  • Agilent Technologies U2781A - page 20

    20 U2781A User’s Guide 2 Installation System Requirements Prior to inst alling the Agilent Measurement Manager sof tware and the USB D A Q dr iver , make sure y our PC meets the f ollo wing minimum system requ irements. Hardware Requirements Processor 500 MHz P entium III or higher required (1 GHz is rec ommended) Operating system Windows 2000/XP ...

  • Agilent Technologies U2781A - page 21

    Installation 2 U2781A User’s Guide 21 Software Requirements Software requirements A gilent IO Libraries Su it e, T&M T oolkit and Microsoft .NET F ramew ork version 1.0 and 2.0 The Agilent IO Lib raries Suite version 14.2 and abov e is recommended. If possible, you should use the current version of the Agilent IO Libr aries Suite . Alternativ ...

  • Agilent Technologies U2781A - page 22

    22 U2781A User’s Guide 2 Installation Installing the USB DAQ driver 1 U npack the U2781A USB modular instru ment chassis. 2 Insert the Product Refer ence CD- ROM in to the CD- ROM dri ve. 3 Installer will automatically execute the Agilent U2300 Series Installation Menu. Click Hardware Driver to beg in the ins t al lat ion o f USB D A Q driv er . ...

  • Agilent Technologies U2781A - page 23

    Installation 2 U2781A User’s Guide 23 5 The Agilent USB D A Q Driver dialog bo x will appear as sho wn in fig ure 2- 2. Cli ck Next to pr ocee d. Figure 2-2 6 Click Install to begin inst allation. Figure 2-3 ...

  • Agilent Technologies U2781A - page 24

    24 U2781A User’s Guide 2 Installation 7 Click Finish when the inst allation is completed. Figure 2-4 ...

  • Agilent Technologies U2781A - page 25

    Installation 2 U2781A User’s Guide 25 Hardware V erification T o verify that th e hardware plugged in and installed is connected properly , r un the A gilent Connect io n Ex pert to d o the h ard ware ver if ic at i on Agilent Co nnection Exper t is one of t he utility files in the A gilent IO Libraries. The Connection Exper t configures connecte ...

  • Agilent Technologies U2781A - page 26

    26 U2781A User’s Guide 2 Installation ...

  • Agilent Technologies U2781A - page 27

    27 Agilent T echnologies Agilent U2781A USB Modu lar Instrument Chassis User’s Guide 3 Features and Functions Introduction 28 USB Backplane 29 T rigger Bus [0...7] 31 T rigger Out 32 External T rigger In (Star T rigger) 32 System Reference Clock 33 Chassis T emperature Monitoring 34 Fan Speed Monitoring 35 Identifying Modu les Location 36 Georgra ...

  • Agilent Technologies U2781A - page 28

    28 U2781A User’s Guide 3 Features and Functions Introduction The Agilent U2781A USB modular instr ument chassis pro vides six USB modular slots and equipped with 200 W universal A C po wer supply and built- in ov er current protection circuit. A 10MHz system reference clock is supplied t o each modules slots. There are two tem perature sensors an ...

  • Agilent Technologies U2781A - page 29

    Features and Functions 3 U2781A User’s Guide 29 USB Backplane Figure 3- 1 illustrates t he 55- pin assignment of t he backplane connector pin. Figure3-1 55-pin backpla ne connector pin assignment Ta b l e 3 - 1 Pin information of SSI conn ector 11 GND +12V +12 V GND USB_D+ USB_D- GND 10 GND +12V +12 V +12V GND GND GND 9 GND +12V +12V +12V GND USB ...

  • Agilent Technologies U2781A - page 30

    30 U2781A User’s Guide 3 Features and Functions Figure3-2 USB backplane block diagram ...

  • Agilent Technologies U2781A - page 31

    Features and Functions 3 U2781A User’s Guide 31 T rigger Bus [0...7] Trigger Bus [0...7] is an 8- bit di git al bus connect ed from slot 1 to s lot 6 to sy nch ron ize diff er e nt USB modules. This trigger bus enables the USB modules of pa ssing trigger signals t o one another . T o hav e one of the mo dules to control the operation of the ot he ...

  • Agilent Technologies U2781A - page 32

    32 U2781A User’s Guide 3 Features and Functions T rigger Out Trigger Out selects one of t he trig ger bus bit [0...7] f or exter nal trigger . The USB de vice in F igure 3- 3 controls the multiplexer switching for the trigger out selection. The SCPI command belo w is used t o set one of the trigger bus [0…7] as external trigger source: TRIGger: ...

  • Agilent Technologies U2781A - page 33

    Features and Functions 3 U2781A User’s Guide 33 Figure3-4 Block diagram of the 10 Mhz Reference Clock and External T rigger In System Reference Clock The 10 MHz ref erence clock can come from two sour ces; inter nal backplane oscillator and exter nal clock source. The inter nal oscillator on t he USB backplane supplies an independent 10 MHz syste ...

  • Agilent Technologies U2781A - page 34

    34 U2781A User’s Guide 3 Features and Functions Chassis T emperature Monitoring The chassis cont ains a temperature control cir cuitr y. It has two therm istor sen sors to sen se the inne r tem per atu re of the chassis. The temperature control cir cuitr y communicates wit h backplane USB de vice through an I 2 C inter face as illustrated in Figu ...

  • Agilent Technologies U2781A - page 35

    Features and Functions 3 U2781A User’s Guide 35 Fan Speed Monitoring The U2781A USB modular instrument chassis is also integ rated with a fan speed control circuit. It is used to monitor the fan status and speed. The control circuit communicates wi t h backplane USB device via I 2 C interface. Refer to Figure 3- 5. T o quer y t he fan st atus, se ...

  • Agilent Technologies U2781A - page 36

    36 U2781A User’s Guide 3 Features and Functions Identifying Modules Location Georgraphical Address Each slot in the chassis is designed wit h a 3- bit address pin, which is designated as a location identity fo r USB modules. The address for all six slots are as belo w: The USB modules are able to read this 3- bit dat a and kno w which slot the mo ...

  • Agilent Technologies U2781A - page 37

    Features and Functions 3 U2781A User’s Guide 37 Figure3-6 Identifying modul es location In order to identify the location of t he modules, a pre- configuration setting is need ed before the synchronization or triggering ev ent begins. Follo w the steps below: 1 Sen d the fo llo wi ng co mma nd to th e mod ular chas sis to trigger it. This command ...

  • Agilent Technologies U2781A - page 38

    38 U2781A User’s Guide 3 Features and Functions 4 During this identif ication operation, the trigger bus is used. Hence, any triggering activiti es on the backplane w ould be blocked. 5 Prior to any triggering acti vities, you must stop the configuration activity by sending the follo wing command: SYST em:IDentity {0|OFF} Simultaneous Synchroniza ...

  • Agilent Technologies U2781A - page 39

    Features and Functions 3 U2781A User’s Guide 39 Figure3-7 Synchronization between modules in the chassis ...

  • Agilent Technologies U2781A - page 40

    40 U2781A User’s Guide 3 Features and Functions ...

  • Agilent Technologies U2781A - page 41

    41 Agilent T echnologies Agilent U2781A USB Modu lar Instrument Chassis User’s Guide 4 General Maintenance This chapter gives a brief desciption on how to maintain the U2781A USB modular instrument chassis. ...

  • Agilent Technologies U2781A - page 42

    42 U2781A User’s Guide 4 Genera l Maintenance To re move t he di r t or mo is tu r e i n t h e c ha ss is pa ne l , t h e c le an in g s t ep s are as follo ws: 1 Turn o f f th e uni t and rem ov e th e po w er c ord and I /O c able . 2 Shake out any dirt that may ha ve accumulated inside the chassis unit. 3 Wipe the chassis with a dr y clot h. N ...

  • Agilent Technologies U2781A - page 43

    © Agilent T echnologi es, Inc. 2006 Printed in Malaysia November 29, 2006 U2781-900 03 www .agilent.com Agilent T echnologies Contact us T o obtain service, warranty or technical support assistance, contact us at the following phone numbers : United States: (tel) 800 829 4444 (fax) 800 829 4433 Canada: (tel) 877 894 4414 (fax) 800 746 4866 China: ...

Manufacturer Agilent Technologies Category Network Card

Documents that we receive from a manufacturer of a Agilent Technologies U2781A can be divided into several groups. They are, among others:
- Agilent Technologies technical drawings
- U2781A manuals
- Agilent Technologies product data sheets
- information booklets
- or energy labels Agilent Technologies U2781A
All of them are important, but the most important information from the point of view of use of the device are in the user manual Agilent Technologies U2781A.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Agilent Technologies U2781A, service manual, brief instructions and user manuals Agilent Technologies U2781A. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Agilent Technologies U2781A.

A complete manual for the device Agilent Technologies U2781A, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Agilent Technologies U2781A by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Agilent Technologies U2781A.

A complete Agilent Technologies manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Agilent Technologies U2781A - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Agilent Technologies U2781A, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Agilent Technologies U2781A, that we can find in the current document
3. Tips how to use the basic functions of the device Agilent Technologies U2781A - which should help us in our first steps of using Agilent Technologies U2781A
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Agilent Technologies U2781A
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Agilent Technologies U2781A in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Agilent Technologies U2781A?

Use the form below

If you did not solve your problem by using a manual Agilent Technologies U2781A, ask a question using the form below. If a user had a similar problem with Agilent Technologies U2781A it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)