Manual Acu-Rite 516353-21

20 pages 0.54 mb
Download

Go to site of 20

Summary
  • Acu-Rite 516353-21 - page 1

    REFERENCE MANUAL SENC 50 Acu-Rite Companies Inc. ...

  • Acu-Rite 516353-21 - page 2

    ...

  • Acu-Rite 516353-21 - page 3

    Page Introduction / Supplied items ............................................... 2 Key Points / Tolerances ......................................................... 3 Linear Encoder Dimensions .................................................. 4 Spar Dimensions .................................................................... 5 Mounting Requir ...

  • Acu-Rite 516353-21 - page 4

    The SENC 50 Precision Glass Scale Linear Encoder provides the accuracy and reliability of an Acu-Rite Companies Inc. measuring system in a very small cross-section with digital output (analog output available). Features and options include: • Resolutions of 0.5, 1.0 or 5µm. • Accuracy grades of ± 3 or ± 5 µm / 1000mm. • Position Trac or 5 ...

  • Acu-Rite 516353-21 - page 5

    Use this information to plan your Linear Encoder installation. • Understand your mounting requirements. • Follow kit instructions when using an Acu-Rite Companies Inc. bracket kit. • Mount encoders close to ways to insure system accuracy. • Mount with lip seals down and away from the work area. • Brackets should be short and rigid. • Su ...

  • Acu-Rite 516353-21 - page 6

    Acu-Rite Companies Inc. 4 Linear Encoder Dimensions SENC 50 Measuring length + 4.65 [118] Measuring length + 4.09 [103.9] .280 [7.11] Typ. End cap Scale assembly Scale case Mounting hole .335 [8.51] .140 [3.55] .463 [11.76] .175 [4.57] Typ. 1.85 [47.00] 2.775 [70.50] Reading head assembly for M4 Hex nut Strain relief and cable 2.0 [50] Approx. min ...

  • Acu-Rite 516353-21 - page 7

    680803-13 13 17.650 3.825 3 5.000 680803-16 16 20.650 2.825 4 5.000 680803-18 18 22.650 1.325 3 10.000 680803-15 15 19.650 2.325 4 5.000 680803-14 14 18.650 1.825 4 5.000 SENC 50 Sp ar Dimensions 5 Acu-Rite Companies Inc. Backup spar Linear Encoder L A X No. B Part Number Measuring Length Places 680803-01 1 5.650 .825 2 4.000 680803-09 9 13.650 1.8 ...

  • Acu-Rite 516353-21 - page 8

    Mounting Requirement s SENC 50 Mounting options can be adapted to machine mounting surfaces using spacers, standoffs, and leveling screws. • Measuring length and mechanical configuration of your equipment determines your options. • Length related mounting options shown. • When a backup spar is used, mount the backup spar following the same pr ...

  • Acu-Rite 516353-21 - page 9

    A variety of mounting conditions can be accommodated. • Mechanical configuration of the equipment determines brackets and adapters required to install the linear encoder. • Three typical conditions are shown for reference. • Flush or offset mounting surfaces using a spar. • Reading head mounting bracket required (bottom mounting bracket exa ...

  • Acu-Rite 516353-21 - page 10

    Encoder Installation Procedure SENC 50 These steps apply to all mounting conditions. Although this may not pictorially represent your application, your installation procedure should follow these steps. • Acu-Rite Companies Inc. Acu-Rite Companies Inc. Acu-Rite Companies Inc. Acu-Rite Companies Inc. Acu-Rite Companies Inc. B B B B B racket Kit ins ...

  • Acu-Rite 516353-21 - page 11

    • Slide shipping brackets from reading head and twist to remove from the scale case. • Save the shipping brackets with the Reference Manual. • Drill / tap the first end mounting hole / Attach the scale case. • Align to within .005” TIR. and drill / tap second end hole. • Attach scale case & align to within .005” TIR. of the axis t ...

  • Acu-Rite 516353-21 - page 12

    • Check function of the linear encoder. These steps will confirm proper operation of your installation. Counting tests check channel A and B output function and Reference Mark tests check Reference Mark operation. • Configure the readout for proper counting resolution. • Move axis and compare the display to the movement. • Repeat test for m ...

  • Acu-Rite 516353-21 - page 13

    11 Acu-Rite Companies Inc. SENC 50 Electrical Shielding Connect a ground wire from the terminal on the back of the readout to the machine or earth ground. Attach a ground wire from the machine to a solid earth ground. With the encoder attached to the machine and the cable connected to the readout, check shielding by measuring resistance between con ...

  • Acu-Rite 516353-21 - page 14

    Acu-Rite Companies Inc. 12 T rouble Shooting SENC 50 If you experience difficulties with your installation, there are methods to analyze the difficulty. Operating difficulties can be caused by either the linear encoders or the readout. Checking the readout Checking the readout Checking the readout Checking the readout Checking the readout Follow th ...

  • Acu-Rite 516353-21 - page 15

    13 Acu-Rite Companies Inc. SENC 50 Mechanical Specifications Mechanical Specifications Digital Analog Resolution 0.5µm 1µm 5µm Grating pitch 20µm Scale medium Light transmission reflective off of chrome coated glass Accuracy (@ 20° C) / 1000mm ± 3µm, ± 5µm Max. slew speed (M/s) 1 Connecting cable Length = 5, 13, and 19 ft. Length = 5, 13, ...

  • Acu-Rite 516353-21 - page 16

    Digital Differential Digital Differential Digital Differential Digital Differential Digital Differential Analog Differential Analog Differential Analog Differential Analog Differential Analog Differential Pin 1 Pin 1 Pin 1 Pin 1 Pin 1 Pin 2 Pin 2 Pin 2 Pin 2 Pin 2 Pin 3 Pin 3 Pin 3 Pin 3 Pin 3 Pin 4 Pin 4 Pin 4 Pin 4 Pin 4 Pin 5 Pin 5 Pin 5 Pin 5 P ...

  • Acu-Rite 516353-21 - page 17

    15 Acu-Rite Companies Inc. SENC 50 Electrical Specifications Parameter Analog Output Signals Incremental signals Signal levels Reference Mark signals Signal level Power Supply Square-wave voltage signals. Channels A and B, in 90° quadrature relationship TTL-level Similar phasing, but differential sinusoidal current or 1 volt peak to peak output 7- ...

  • Acu-Rite 516353-21 - page 18

    Acu-Rite Companies Inc. 16 Acu-Rite Companies Inc. W arranty SENC 50 ACU-RITE Companies, Inc. is proud to offer the 3-Year Hassle-Free Warranty for all readout systems, and precision glass scales. This warranty will cover all of the ACU-RITE repair and replacement costs for any readout or precision glass scale returned during the three (3) year war ...

  • Acu-Rite 516353-21 - page 19

    ...

  • Acu-Rite 516353-21 - page 20

    Acu-Rite Companies Inc. Acu-Rite Companies Inc. Acu-Rite Companies Inc. Acu-Rite Companies Inc. Acu-Rite Companies Inc. IS AN IS AN IS AN IS AN IS AN ISO 9001 ISO 9001 ISO 9001 ISO 9001 ISO 9001 CERTIFIED CERTIFIED CERTIFIED CERTIFIED CERTIFIED MANUF MANUF MANUF MANUF MANUF ACTURER ACTURER ACTURER ACTURER ACTURER 516353-21 V e01 8/2009 One Precisio ...

Manufacturer Acu-Rite Category Automobile Accessories

Documents that we receive from a manufacturer of a Acu-Rite 516353-21 can be divided into several groups. They are, among others:
- Acu-Rite technical drawings
- 516353-21 manuals
- Acu-Rite product data sheets
- information booklets
- or energy labels Acu-Rite 516353-21
All of them are important, but the most important information from the point of view of use of the device are in the user manual Acu-Rite 516353-21.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Acu-Rite 516353-21, service manual, brief instructions and user manuals Acu-Rite 516353-21. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Acu-Rite 516353-21.

Similar manuals

A complete manual for the device Acu-Rite 516353-21, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Acu-Rite 516353-21 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Acu-Rite 516353-21.

A complete Acu-Rite manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Acu-Rite 516353-21 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Acu-Rite 516353-21, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Acu-Rite 516353-21, that we can find in the current document
3. Tips how to use the basic functions of the device Acu-Rite 516353-21 - which should help us in our first steps of using Acu-Rite 516353-21
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Acu-Rite 516353-21
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Acu-Rite 516353-21 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Acu-Rite 516353-21?

Use the form below

If you did not solve your problem by using a manual Acu-Rite 516353-21, ask a question using the form below. If a user had a similar problem with Acu-Rite 516353-21 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (1)

Николай

Не работает монитор