    1521 Handheld Thermometer Readout User’ s Guide Rev. 571203 ENG Hart Scientific ...

  • Fluke 1521 - page 2

    Limited W arranty & Limitation of Liability Each product from Fluke's Hart Scientific Di vision ("Hart") is warranted to be free from defects in mate - rial and workmanship under normal use and service. The warranty period is one year for the Handheld Thermometer Readout. The warranty period begins on the date of the shipment. P ...

  • Fluke 1521 - page 3

    T able of Contents 1 Bef ore Y ou Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 W arnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.2 Cautio ...

  • Fluke 1521 - page 4

    6.6 INFO-CON Connector . . . . . . . . . . . . . . . . . . . . . . . 23 7 Display Functions . . . . . . . . . . . . . . . . . . . . . . . . . 25 7.1 Blank Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 7.2 Min/Max Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 7.3 Hold Mode . . . . . . . . . . . . . . . . . . . . . ...

  • Fluke 1521 - page 5

    11 T roubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 47 11.1 An Error Message Is Displayed . . . . . . . . . . . . . . . . . . . 47 11.2 CE Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 11.2.1 EMC Directiv e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 11.2.1.1 Immunity T esting . . . ...

  • Fluke 1521 - page 6

    iv Figures and T ables T able 1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1 Figure 1 Front of the 1521 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Figure 2 T op and Side V ie w . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Figure 3 INFO-CON Connector . . . . . . . . . . . . . . . . . . ...

  • Fluke 1521 - page 7

    1 Bef ore Y ou Start 1.1 Symbols Used T able 1 lists the International Electri cal Symbols. Some or all of these symbols may be used on the instrument or in this manu al. Symbol Description AC ( Alternating Current) AC- D C Batter y CE Complies with European Uni on Directiv es DC Double Insulated Electric Shock Fuse PE Ground Hot Surf ace (Bur n Ha ...

  • Fluke 1521 - page 8

    Symbol Description Canadian Standards Associati on O VERV OL T AGE (Installat ion) CA TEGORY II , P ollution Degree 2 per IEC1010-1 r e - f ers to the le vel of Impulse Wit hstand V oltage protection provided. Equipment of O VERV OL T AGE CA TEGOR Y II is energy-consuming equipment to be suppl ied from the fix ed install ation. Examples include hou ...

  • Fluke 1521 - page 9

    Carefully place probes on a heat/cold resistant surface or rack until they reach room temperature. • The A C adapter can present safety concerns if misused or damaged. T o av oid the risk of electric shock or fire, do not use the AC adapter outdoors or in a dusty , dirty , or wet environment. If the cord, case, or plug of the adapter is damaged i ...

  • Fluke 1521 - page 10

    • Probes are fragile devices which can be damaged by mechanical shock, ov erheating, and absorption of moisture or fluids in the wires or hub . Damage may not be visibly apparent but nevertheless can cause drift, in - stability , and loss of accuracy . Observe the following precautions: • DO NO T allo w probes to be dropped, struck, bent, or st ...

  • Fluke 1521 - page 11

    Chao Y ang District Beijing 100004, PRC CHIN A Phone: +86-10-6-512-3436 T elefax: +86-10-6-512-3437 E-mail: xingye.han@fluke.com.cn Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Of fice Service Center 60 Alexandra T errace #03-16 The Comtech (Lobby D) 118502 SINGAPORE Phone: +65 6799-5588 T elefax: +65 6799-5588 E-mail: antng@singa.fluke.com ...

  • Fluke 1521 - page 12

    2 Intr oduction The 1521 Handheld Thermometer Readout is a lo w-cost, high-accuracy digital thermometer readout designed to be used wit h PR Ts and thermistors. The unique combination of features makes thi s instrument suitable for a wide v ari - ety of applications in industry . Featur es and capabilities of the 1521 include the following: • Mea ...

  • Fluke 1521 - page 13

    3 Specifications and En vir onmental Conditions 3.1 Specifications Resistance Range 0 Ω to 500 k Ω Resistance Accuracy, PRT, one year † 0 Ω to 25 Ω : 0.002 Ω 25 Ω to 400 Ω : 0.008% (80 ppm) of reading Resistance Accuracy, thermi stor, one year † 0 Ω t o2k Ω :0 . 4 Ω 2k Ω to 200 k Ω : 0.02% (200 ppm) of reading 200 k Ω to 500 k Ω : ...

  • Fluke 1521 - page 14

    Digital Fil ter Exponential with adj ustable time const ant (1 to 60 seconds) Remote Communications RS-232 serial Display LCD, 6-digit x 7-segment with 16-charact er alphanumeric Operating Temperatur e Range 0 to 40°C (32 to 104°F) absolut e 15 to 35°C (59 to 95°F) f ull accuracy Power 12 VDC (AC adapter included) Rechargeable NiMH batter ies ( ...

  • Fluke 1521 - page 15

    4 Quic k Start This section briefly explains the basics of setting up and operating your 1521 Thermometer . 4.1 Unpac king Carefully unpack the 1521 and accessories and inspect them to make sure all components are present and in satisf actory condition. V erify that the following items are present: • 1521 Thermometer • A C Adapter (clamp-on fer ...

  • Fluke 1521 - page 16

    top of the instrument with the ridged side down. It will f it snugly and l ock into place when it is fully inserted. The INFO-CON probe connector includes a memory de vice that stores the unique characteristics of the probe, allo wing the 1521 to measure tempera - ture accurately . If your probe was purchased from Hart for use with the 1521 it shou ...

  • Fluke 1521 - page 17

    5 P ar ts and Controls The functions of the v arious features of the 1521 are described belo w . 5.1 Fr ont Features The front of the 1521 features the LCD display and control b uttons. 13 5 Parts and Controls Front Features ...

  • Fluke 1521 - page 18

    Display -The display sho ws the current temperature (or resist ance) measure - ment on the large numeri c upper portion of the display . It can also show a vari - ety of information on the small er alphanumeric lo wer portion such as minimum, maximum, hold temperature s, delta(x), and other instrumen t settings. Po wer and Backlight button -This b ...

  • Fluke 1521 - page 19

    5.2 T op and Side Features The top and side of the 1521 feature the probe connector , DC power input, se - rial port, and infrared windo w . Probe Connector - At the t op of the thermometer is the opening where the probe connector is inserted. The probe must be connected using an INFO-CON probe connector to measure temperature (see Section 6.6). DC ...

  • Fluke 1521 - page 20

    5.3 Bac k Features See Figure 4 on page 19. Stand -The stand at the back of the 1521 can be flipped do w n to prop up the in - strument for better viewing. Battery Compartment - Behind the stand is the compartment that contains the battery pack. The battery pack can be accessed if necessary by opening the bat - tery cove r with a small Philips scre ...

  • Fluke 1521 - page 21

    5.5 Accessories The 1521's accessories and their features are described here. AC A d a p t e r -The A C adapter recharges the internal battery pack and can also be used to supply po wer to operate the 1521 whi le the battery i s being char ged (see Section 6.2). Serial Cable -The serial cable can be used to connect a computer or a printer to t ...

  • Fluke 1521 - page 22

    Ordering Infor mation Model Type Range Size Cal Uncertainty and Repeatability Typical Drift (1 year) 5626 Pt-100 Ohm, 4-wire –200 to 660°C 1/4” x 12” or 15” (6.35 x 305 or 381 mm) ±0.015°C ±0.03°C 5628 Pt-25 Ohm, 4-wire –200 to 660°C 1/4” x 12” or 15” (6.35 x 305 or 381 mm) ±0.015°C ±0.03°C 5614 Pt-100 Ohm, 4-wir e –200 ...

  • Fluke 1521 - page 23

    6 General Operation This section e xplains the details of t he operation of the 1521 with its compo - nents and accessories. 6.1 Battery The 1521 has a b uilt-in nickel-metal-hydride battery pack that can po wer the instrument for about 18 hours before needing to be recharged. The battery will discharge much more quickl y if the display backlight i ...

  • Fluke 1521 - page 24

    on. The power control circuit i nside the instrument manages battery char ging and will stop charging the battery automatically wh en it’ s fully charged. It nor - mally tak es about four hours to full y charge t he battery . The instrument can be operated while the battery is bein g charged. CA UTION: DO NOT leave the A C adapter plugged in for ...

  • Fluke 1521 - page 25

    adapter supp lied with the instru ment. The DC power source p lugs into th e DC power i nput on the right side of the 1521. W ARNING: The AC adapter has cir cuits with high voltages inside that could pr esent danger of electric shock or f ir e if exposed. If the AC adapter is damaged in any way or becomes hot, discontinue its use immediately , disc ...

  • Fluke 1521 - page 26

    smaller type to the right of the measurement. The possibl e units are de grees Celsius (C), resistance in oh ms ( Ω ), degrees F ahrenheit (F), K elvin (K), and degrees R ankine (R). The unit s can be easily changed by placing the di splay in the units setting mode using th e MODE button (see Section 7.5). The lo wer part of the display has variou ...

  • Fluke 1521 - page 27

    T emperature is generally sensed at the tip of the probe. T o get an accurate tem - perature measurement the probe sheath should be well immersed, with ade - quate depth and fit, into the mediu m to be measured. CA UTION : Pr obes ar e fra gile devices that can be easily damaged by me - chanical shock, o verheating, and absorption of moisture or fl ...

  • Fluke 1521 - page 28

    ognize that accurac y may be significantly de graded using a two-wire connec - tion because of wire resistance. 1521 Handheld Thermometer Readout User’s Guide 24 GND P1 C2 P2 C1 J1 J2 Four-wire Connection GND P1 C2 P2 C1 J1 J2 Three-wire Connection GND P1 C2 P2 C1 J1 J2 T wo-wire Connection Shor t Shor t Figure 6 Probe Wir ing Diagrams ...

  • Fluke 1521 - page 29

    7 Displa y Functions The 1521 operates in an y of se ve ral modes which determine what information is visible in the l ower al phanumeric portion of the display . The MODE button changes the mode. This b utton can be pressed repeatedly unt il a desired mode is set. The various modes are listed below in the order they appear . (See Figure 7). Blank ...

  • Fluke 1521 - page 30

    Each of these operating mod es is described in the sections that follow . 7.1 Blank Mode This mode is identified by “Blank” that temporarily app ears on the display . 1521 Handheld Thermometer Readout User’s Guide 26 Blank Mode Set Units Min/Max Mode Set Rate Hold Mode Set Resolution Cal Mode button button button button button button button ( ...

  • Fluke 1521 - page 31

    In this mode nothing appears on the alphanumeric porti on of the display . Use this mode if you are only interested in the latest measurements and want to sim - plify the appearance of the display . The message “Prb cal expired”, “M tr cal e xpired”, “Recharge Needed”, or “Probe is locked” may be displayed. See Sect ion 11.1. 7.2 Mi ...

  • Fluke 1521 - page 32

    7.5 Units This mode is identified by “Un its: C/ Ω /F/K/R” that a ppears on the display . This mode allo ws you to select the unit of measurement: C for degrees Celsius, Ω for resistance in ohms, F for de grees F ahrenheit, K for K elvin, and R for Rankine. The Up and Down b u ttons are used to select one of the displayed units. The underline ...

  • Fluke 1521 - page 33

    7.8 CAL MODE The calibration menu contains fu nctions for setting pro be and instrument cali - bration parameters. If a probe parameter is changed, the probe parameter is up - loaded into the INF O-CON prob e connector wh en the CAL MODE is exited or after scrolling through the entire menu. Always check to insure that the param - eters hav e been w ...

  • Fluke 1521 - page 34

    Beta [CVD only] - Set the beta probe coefficient b0 [Therm only] - Set the b0 probe coefficient b1 [Therm only] - Set the b1 probe coeff icient b2 [Therm only] - Set the b2 probe coeff icient b3 [Therm only] - Set the b3 probe coefficient Probe W ir es - Set the number of wires of the probe PCal - Set the probe calibration date PDue - Set the probe ...

  • Fluke 1521 - page 35

    ensuring that it is not lost or forgot ten. If necessary the passcode can be re - stored by contacting the manufacturer . T o enter the passcode use the Up and Down b uttons to set each digit of the passcode and the ENT but to n to m ove to the next digit on the r ight. Press ENT tw ice when all digits are entered to move to the next function. If t ...

  • Fluke 1521 - page 36

    ITS-90 PRTs calibrated to ITS-90 R(.01), a, b, c, d, a4, and b4 IEC751 PRTs (RTDs) conforming t o IEC-751 none CVD PRTs characterized by the Cal lendar-Van Dusen equation R0, Alpha, Delta, and Beta YSI400 Thermistors matchi ng the YSI-400 series none THERM Thermistors character ized by the thermistor polynomial b0, b1, b2, and b3 Each of the con ve ...

  • Fluke 1521 - page 37

    A PR T was calibrated to ITS-90 and its calibration certificate states v alues for coeff icients R(273.16K), a4, b4, a8, and b8. Set the 1521 coef fici ents to the certificate v alues as follo ws. 1521 Coefficients Certificat e Value R(.01) R(273.16K) aa 8 bb 8 c0 . 0 d0 . 0 a4 a4 b4 b4 Example 2: A PR T was calibrated to ITS-90 and its calibration ...

  • Fluke 1521 - page 38

    higher), and has moderate accuracy (0.05°C to 1°C). This type uses the Callendar -V an Dusen equation to calcu late temperatur e from resistance: rt C Rt t t R ([ ] ) °= +− − ⎛ ⎝ ⎜ ⎞ ⎠ ⎟ ⎡ ⎣ ⎢ ⎤ ⎦ ⎥ ⎧ ⎨ ⎩ ⎫ ⎬ ⎭ ≥ 0 0 1 1 100 100 10 αδ 1 100 100 1 100 1 100 3 +− − ⎛ ⎝ ⎜ ⎞ ⎠ ⎟ −− ⎛ ...

  • Fluke 1521 - page 39

    Example 2: A thermistor has coef fi cients for the equation ln(R ) as a function of T gi v en as a, b, and c. Set the 1521 coeff icients to the certif icate v alues as follo ws: 1521 Coefficients Certificat e Value b0 a b1 b b2 0.0 b3 c 7.8.14 Pr obe Wires This function sets the number of connecting wires in the probe. The probe wires setting is st ...

  • Fluke 1521 - page 40

    Note : Probe parameters are uploaded to the INFO-CON connector at thi s point. 7.8.17 Filter The digital filter helps to smooth variations in the measurements and improve resolution. A negati ve ef fect of filtering is that it tends to slo w the response to changes in temperature. Y ou can increase the filter time constant to further im - prov e ac ...

  • Fluke 1521 - page 41

    8 Comm unications Interface Remote communications allows an e x ternal de vice, such as a computer , to com - municate with the 1521 to obtain measurement data and cont rol its operati on. Communication is accomplished with v arious commands issued to the 1521 through the RS-232 serial port . 8.1 RS-232 Connection The three-conductor jack for the s ...

  • Fluke 1521 - page 42

    ON (all carriage returns are followed b y a linefeed (ASCII decimal 10)), and the duple x to HALF disabli ng echo. The serial port can be used to transmit measurements to a computer or printer . It can also be used to change set tings of the 1521 from a comput er or interface with software. A full list of commands follows in Section 8.2. Commands s ...

  • Fluke 1521 - page 43

    Description Command Example Response Format Response Example Range Set resolution di gits re[solution]=n res=1 1 to 3 Read time ti[me] time ti: hh:mm:ss ti: 16:23:45 Set time Note: You must set hours, min - utes, and seconds for t he time to be set correctly. ti[me]=hh:mm: s s ti=08:15:0 0 00:00:00 to 23:59:59 Read identificat ion informati on *i[d ...

  • Fluke 1521 - page 44

    Description Command Example Response Format Response Example Range Set probe type p[robe]=IT[S- 90]/ IE[C751]/ C[VD]/ Y[SI400]/ T[HR] pr ITS-90, IEC751, CVD, YSI400, THERM Read ITS-90 R(.01) or CVD R0 r[0] r0 r0: 9999.999 r0: 100.035 Set ITS-90 R(.01) or CVD R0 r[ 0]=n r0=100.106 1 to 900 Read ITS-90 coeffici ent a/b/c/ d/a4/b4 a {a/b/c/d/ a4/b4}: ...

  • Fluke 1521 - page 45

    Description Command Example Response Format Response Example Range n Numeric data supplied by user 9 Numeric data retur ned to user x Character data r eturned to user 41 8 Communications Interface Communication Command List ...

  • Fluke 1521 - page 46

    9 Calibration The 1521 should be calibrated a t re gular interv als to ensure that it conti nues to measure with proper accuracy . Calibration should only be done by qualified, authorized personnel. 9.1 Required Equipment The follo wing items are required to test and cali brate the 1521: • 0 Ω four-wire resistor (short) • 25 Ω four-wire resis ...

  • Fluke 1521 - page 47

    4. V erify the accuracy at 0 Ω , 4k Ω ,1 0k Ω ,4 0k Ω , 100 k Ω , and 500 k Ω . 5. Set the meter calibration date to the present date (see Section 7.8.18). 6. Set the meter due date to one year from the present date (see Section 7.8.19). 1521 Handheld Thermometer Readout User’s Guide 44 ...

  • Fluke 1521 - page 48

    10 Maintenance • This instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product dev elopment. Therefore, with proper care the instrument should require very little maintenance. A void operating the instrument in oily , wet, dirty , or dusty environments. • If the ...

  • Fluke 1521 - page 49

    11 T r oubleshooting In case you run into dif ficulty while operat ing the 1521, this sect ion provides some suggestions that may help you solve the probl em. Belo w are sev eral situa - tions that may arise follo wed by possible causes of the problem and suggested actions you might take . 11.1 An Error Message Is Display ed The follo wing table li ...

  • Fluke 1521 - page 50

    Problem Solution While attempti ng to measure re - sistance the display shows an i n - correct value. Poor or incorrect connecti on of the probe. A common mista ke is to connect the wires of t he probe to the wrong t erminals. Check the wi r - ing carefully ( see Figure 6). Open, shorted, or damaged sensor or lead wires. Check the resi s - tance ac ...

  • Fluke 1521 - page 51

    The instrument was designed specif ically as a test and measuring device. Com - pliance to the EMC di recti ve i s through IEC 61326-1 Electrical equipment for measur ement, contr ol and labor atory use – EMC r equire ments (1998). As noted in the IEC 61326-1, the instrument can ha ve varyi ng configurat ions. The instrument w as tested in a typi ...

Manufacturer Fluke Category Thermometer

Documents that we receive from a manufacturer of a Fluke 1521 can be divided into several groups. They are, among others:
- Fluke technical drawings
- 1521 manuals
- Fluke product data sheets
- information booklets
- or energy labels Fluke 1521
All of them are important, but the most important information from the point of view of use of the device are in the user manual Fluke 1521.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Fluke 1521, service manual, brief instructions and user manuals Fluke 1521. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Fluke 1521.

Similar manuals

A complete manual for the device Fluke 1521, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Fluke 1521 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Fluke 1521.

A complete Fluke manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Fluke 1521 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Fluke 1521, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Fluke 1521, that we can find in the current document
3. Tips how to use the basic functions of the device Fluke 1521 - which should help us in our first steps of using Fluke 1521
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Fluke 1521
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Fluke 1521 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Fluke 1521?

Use the form below

If you did not solve your problem by using a manual Fluke 1521, ask a question using the form below. If a user had a similar problem with Fluke 1521 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)