Manual Emerson PN-51-T56

44 pages 0.69 mb
Download

Go to site of 44

Summary
  • Emerson PN-51-T56 - page 1

    Clar it y II T56 T urbidimet er T56 Instruction Manual P N 51- T56 May 2012 ...

  • Emerson PN-51-T56 - page 2

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 This page l eft blank intentionally ...

  • Emerson PN-51-T56 - page 3

    Clar it y II T56 T urbidimet er ESSENTI AL INSTRUC TIONS -Read this p age before proceeding! Y our instrumen t purcha se from Ro semo unt Analy tical, Inc. is one of the finest availab le for your par ticular applicatio n. These instrume nts hav e been designed, and tested to meet many natio nal and intern ationa l standar ds. Experi enc e indica t ...

  • Emerson PN-51-T56 - page 4

    This produc t genera tes, uses, and can radiate radio frequency ener gy and thus can cause radio communication interfere nce. Improper inst allation , or operati on, may increa se such interfere nce. A s tempor arily permitted by regulation, this unit has not been te sted for c ompliance within the limits o f Cla ss A computing devic es, pursuant t ...

  • Emerson PN-51-T56 - page 5

    C on te nts Section 1: Description and Specifications 1 .1 F e atures and Applica tions ......................................................................................1 1.2 Specifi cations – Ge neral .........................................................................................2 1.3 Ordering In formati on ....................... ...

  • Emerson PN-51-T56 - page 6

    ii T able of C onte nts c ont’d Section 6: Programming - T urbidity 6.1 Programming Me asure ments – In troduc tion .......................................................23 6.2 T urbidit y Measu reme nt Programming .................................................................23 Section 7: Calibration 7.1 Calibration – In troduc tion .... ...

  • Emerson PN-51-T56 - page 7

    Sec tion 1: Des crip tion and S peci ficati ons 1.1 F eatures and Applications C OMPLETE SYSTEM includes single or dual inpu t analyze r , sensor(s), and de bubbler a sse mbly CHOOSE U .S. EP A METHOD 180.1 or ISO METHOD 7027 compli ant sensors RANGE 0-200 NTU RESOLUTION 0.001 NTU FULL FE A TURED ANAL Y ZER with fully scalable an alog outputs and f ...

  • Emerson PN-51-T56 - page 8

    1.2 Specifications - General Case: P olycarbon ate . NEMA 4X/CSA 4, IP66. NOTE: T o ensure a NEMA se al, tighte n all four front p anel scr ews to 6 in-lbs o f tor que. Dimensions: 6.2 x 6.2 x 5.2 in. (157 x 157 x 132mm) Conduit openings: A cce pts (6) PG13.5 or 1/2 in. conduit fittings Display: Lar ge 3.75 x 2.2 in. (95.3 x 55.9mm) high r esolutio ...

  • Emerson PN-51-T56 - page 9

    Ordinar y Locations: Pollution Degree 2: Norm ally only non-conductive pollution occ urs. Occasion ally , howe ver , a tem porar y conductivit y caused b y condensa tion must be expected. Altitude : for use u p to 2000 me ter (6562 f t.) RFI/EMI: – EN-61326 L VD: – EN -61010-1 In pu t: O ne o r two i s ol at ed s en so r in pu ts . Me as ur em ...

  • Emerson PN-51-T56 - page 10

    4 Descriptio n and Spec ificatio ns Instruction Manual Clarity II T56 PN-51- T56 May 2012 T56 T urbidity System Components C ode S ensor T ype 01 EP A Sensor 02 ISO Se nsor 40 EP A Sensor (Sec ond Sensor Option) 41 ISO Se nsor (Second Sens or Option) Code Measuring Chamber 10 Debubble r Flow Chambe r 60 T wo Debubble r Flow Ch ambers Code Sensor Ca ...

  • Emerson PN-51-T56 - page 11

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 Sec tion 2.0 – Installa tion 2.1 Unpacking and Inspec tion Th e Cl ar ity II T ur bi di me t er i s a c om pl e t e s ys te m f o r th e de t er mi n at io n o f tu rb id ity in dr in ki n g wa t e r . Th e sy st e m c on si st s of t he an al yz er , s en so r( s) , ca bl e( s) , a nd f lo w ...

  • Emerson PN-51-T56 - page 12

    6 Installation Instruction Manual Clarity II T56 PN-51- T56 May 2012 Fig. 2-1 T56 Panel Mounting Installation dimensions PA N EL SU PP LIE D B Y OTHERS M A XI M UM THIC KN ESS .375 IN (9.52 mm ) PA NEL MOUNT GA SKET BO T T OM VIEW FRONT VIE W SIDE VIEW PA NEL CUT-OUT 6.20 157 .5 6.17 1 56 .8 5 .18 131 .4 1 .27 32 .2 5.45 138 .4 5.45 138 .4 6.01 152 ...

  • Emerson PN-51-T56 - page 13

    Installation 7 Instruction Manual Clarity II T56 PN-51- T56 May 2012 F ig. 2-2 T56 Pipe and W all Mounting Installation dimensions NOTES : UNLESS OT HERWISE SPE CI FIED L C FRONT PA NEL 2 IN CH P IPE MO UNT BR ACK E T 2X SET U-BOLT S FOR 2 " PIPE IN KIT PN 2382 0-00 1 VERTICA L PIP E MOUN TI NG SHOWN. FOR HORI ZONTA L PIPE , ROTA TE BRACKET A ...

  • Emerson PN-51-T56 - page 14

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 8 Ordering Information 2.3 Installation – Debubbler Assembly See Figu re 2 -3 f or i nstal lat ion. Co nnect the s ampl e li ne to the inl et fit tin g. T he f itt ing ac c e p ts 1/4- in ch OD tub ing . S ee S ecti on 2.6 fo r re c omme nded ins tal l at ion o f t he s am ple port. A tta ch a ...

  • Emerson PN-51-T56 - page 15

    Installation 9 Instruction Manual Clarity II T56 PN-51- T56 May 2012 F ig. 2-3 Debubbler and Flo w Chamber ...

  • Emerson PN-51-T56 - page 16

    10 Instruction Manual Clarity II T56 PN-51- T56 May 2012 2.4 Installation – Sensor Uns cre w t he n ut o n t he s id e o f t he de bub ble r . I nse r t th e s ens or i n th e m outh of th e me as uring ch amb e r . Be s ure the p in on the de bu bb le r li nes up w ith t he h ole in the s ens or . R epl ac e the nut . R emo ve t he pro tectiv e ...

  • Emerson PN-51-T56 - page 17

    Sec tion 3.0 Wiring 3.1 General The T56 is ea sy to wire. It i ncludes r emo vable connectors and slide-out sign al input boar ds. The front pa nel is hinged at the bo ttom. The pane l swings down f or easy acc ess to the wiring loca- tions. 3.1.1. R emo vable c onnec tor s and sign al input boar ds The T 56 u ses remo vable sign al input boards an ...

  • Emerson PN-51-T56 - page 18

    3.2 Preparing Conduit Openings The re are s ix co ndui t op en ing s in al l con figura tions o f T56. (No te th at four of the ope nings will b e fi tted wi th pl ugs u pon ship ment .) C on dui t o pe nin gs a cc ep t 1/ 2-i nch c ond uit fi tti ng s o r PG1 3. 5 c abl e g lan ds. T o k ee p the ca se wa te r - ti ght , b loc k u nu sed op en in ...

  • Emerson PN-51-T56 - page 19

    3.4.4 Se nsor wiring t o signal bo ards Plug the pre-te rminated se nsor cable connector directly into the turbi dit y signal bo ard m ating connector . 3.4.5 Se nsor The sens or cable is pre-wired t o a plug that inse r ts into a rece iving socke t on the signal board . See Figures 3-1. The cable a lso pa sses through a strain re lief fitting. T o ...

  • Emerson PN-51-T56 - page 20

    14 Instruction Manual Clarity II T56 PN-51- T56 May 2012 IMPORT ANT NOTE: When u sing EP A/incandesce nt senso rs (PN 8-0108-0002-EP A): • DO NO T power u p the instrumen t without the se nsor connected • DO NO T disconnect and rec onnect a sensor while an an aly zer i s powe red If this is inc onv enient or cannot be a voided: 1. C yc le powe ...

  • Emerson PN-51-T56 - page 21

    Sec tion 4.0 Dis pl ay and Ope ra tion 4.1 User Int er face Th e T56 ha s a l ar ge di spl a y whi ch sho ws two li ve m ea su re me n t re ado uts in l arg e d igi ts and up t o six ad dit ion al pr oc es s var ia ble s o r d ia g- no sti c pa ram et ers c onc urr en tl y . T he di sp la y i s ba ck -l it and th e f or m at can be c u sto mi z ed ...

  • Emerson PN-51-T56 - page 22

    Th e EN TE R/ ME NU k e y is a ls o u se d t o e nt er s e le cti on s o r e n ab le p ro gr amm in g an d ca li br a ti on s te p s. The EXIT key re turns to the previou s menu leve l. Th e I NF O ke y pr ov ide s d eta ile d i nst ruct io ns and ex pla na ti on s d uri ng pr ogr amm ing an d c ali bra ti ng pr oc edu res . I t a lso pr ov id es t ...

  • Emerson PN-51-T56 - page 23

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 Display and Ope ration 17 4.4 Menu System The T 56 me nu system is simil ar to a comput er . Pressing the ENTER/MENU k ey at an y time o pens the t op -level me nu including Calibrati on, Hold, Programming , Displ ay , Data and HAR T functions. T o find a menu ite m, use the directional N avig a ...

  • Emerson PN-51-T56 - page 24

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 18 This page l eft blank intentionally ...

  • Emerson PN-51-T56 - page 25

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 Programming Basics 19 S e ct i o n 5 . 0 . Pr o g r a m mi ng t h e A n a lyz e r - Ba s i cs 5.1 General T yp ica l p rog ram min g s te ps i ncl ude th e f ol lo wi ng lis ted pr oc edu res . E ach o f t hes e p rog ram min g fu ncti ons ar e e as ily an d q uic kly ac co mpl ish ed us in g t ...

  • Emerson PN-51-T56 - page 26

    5.4 Security Acces s When e ntering the c orrec t acces s code f or the Calibration/Hold security level, the Calibr ation a nd Hold men us are a cce ssible . This allows opera tors or technici ans to pe r form rou tine main te- nanc e. This security leve l does no t allow acc ess to the Program or Di spla y menus. Whe n en te ring the correct acce ...

  • Emerson PN-51-T56 - page 27

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 Programming Basics 21 T o reset t o factor y defaults, re set calibration d ata only or res et an alog outputs only , access the Re set screen b y pressing ENTER/MENU from the m ain screen. 5.7 Programm ing Alarm R elay s Th e T56 24V DC (-0 2 or de r co de ) a nd the A C swi tc hin g p o w e r ...

  • Emerson PN-51-T56 - page 28

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 22 Programming Basics 5. T o en ter setpoin ts for rel ays, en ter the de sired value for the proc ess mea suremen t or te mpera ture at which to activate a n alarm ev ent and pre ss ENTER/MENU . 6. T o set de adband as a me asureme nt value, en ter the chan ge in the proc ess va lue needed afte ...

  • Emerson PN-51-T56 - page 29

    Sec tion 6.0 Programming - T urbi dit y 6.1 Programm ing Measurements – Introduction The T56 au toma tically recogniz es each in stalled mea suremen t board upon first power -up and eac h time the analy ze r is powered. Comple tion of Quick S tar t screens upon first powe r up en able measur eme nts, but addition al steps m ay be required to prog ...

  • Emerson PN-51-T56 - page 30

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 24 This page l eft blank intentionally ...

  • Emerson PN-51-T56 - page 31

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 Calibration 25 Sec tion 7.0 Calibra tion 7.1 Calibration – Intro duc tion Calibration is the pr oces s of adjustin g or standard izing the an alyzer t o a lab t est or a calibrated labora tor y instrumen t, or standard izing to some known re fere nce. C ompletion of Quick S tar t upon first po ...

  • Emerson PN-51-T56 - page 32

    7.2.1 Slope Calibra tion – T urbidit y This section describes ho w to c onduc t a 2-point calibra tion of the turbi dit y sens or against a use r-p repar ed 20NTU standar d. The calibrati on requires two step s. Fir st, immerse the se nsor in f ilte red water h aving ve r y low turbidity and measure the sens or output. Next, increa se the turbidi ...

  • Emerson PN-51-T56 - page 33

    Sec tion 8.0 Mai n te n anc e 8.1 Ov er view Th e 5 6 A n alyz er u sed in th e C la rity II tu rbi dim et er n eed s l itt le ro uti ne ma int e n anc e . C le an the an al yz er c as e a nd fro nt p ane l by wi pin g i t w ith a c le an so ft c lo th da mpe ne d wit h w at er ON L Y . D o no t u se sol ve nt s, lik e a lc oho l, th a t mig ht cau ...

  • Emerson PN-51-T56 - page 34

    NOTE: I f you h ave a dual input ana lyzer , you can re apply power a t this point. The initi al reading from the othe r sensor will be mom en tarily ze ro. Af ter a bout 60 sec onds the reading will reac h its final value . 3. Using a small Phi llips sc rewdrive r , remo ve the t wo screws holdi ng the top fl ange of the se nsor to the body . 4. U ...

  • Emerson PN-51-T56 - page 35

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 Mainte nance 29 8.3 Debubbler and Measuring C hamber 8.3.1 Cle aning the se nsor 1. T urn off the sample su pply to the de bubbler . 2. Remo ve the sens or and put it in a sa fe plac e. The calibration c up is a good place t o store the sens or . 3. Loose n the small drain plug in the ba se plug ...

  • Emerson PN-51-T56 - page 36

    30 Mainte nance Instruction Manual Clarity II T56 PN-51- T56 May 2012 8.4 List of R eplacement P ar ts LOCA TION IN DESCRIPTION P ART FIGURE 8-4 NUMBER — Re placem ent l amp board as sembly , USEP A -compli ant sensor 1-0901-0009-EP A — Re placem ent l amp board as sembly , ISO -compli ant sensor 1-0901-0010-ISO — Re placem ent s ensor , USEP ...

  • Emerson PN-51-T56 - page 37

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 Mainte nance 31 Sec tion 9.0 R e turn o f M a te ri al 9.1 General T o expedite the repa ir and return o f instr umen ts, proper comm unication be t ween the cu stome r and the factor y is impor tant. Bef ore returning a product for re pair , call 1-949-757- 8500 for a R eturn Ma teri als Author ...

  • Emerson PN-51-T56 - page 38

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 32 Append ix Appe ndix This procedur e describes ho w to verify linearity bet ween turbidity and TSS. 1. Collect a sample of the pro ces s liquid-you ma y need 10 L or more if y ou use the Cl arit y II for me asuring turbidity . If you u se a laborat or y turbidimete r , you will need le ss v ol ...

  • Emerson PN-51-T56 - page 39

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 War ranty 33 W ARRANT Y Se lle r warr an ts th at the firm ware will ex ec ut e th e p rogra mmi ng instru c tion s p rovi ded by Se lle r , a nd th at the Good s man ufactu red or Serv- ice s pro vided by Seller will be f ree from defects in mate rials or workmanship un der normal u se and car ...

  • Emerson PN-51-T56 - page 40

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 34 Certific ation ...

  • Emerson PN-51-T56 - page 41

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 Certific ation 35 ...

  • Emerson PN-51-T56 - page 42

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 36 This page l eft blank intentionally ...

  • Emerson PN-51-T56 - page 43

    Instruction Manual Clarity II T56 PN-51- T56 May 2012 37 This page l eft blank intentionally ...

  • Emerson PN-51-T56 - page 44

    Credit Cards f or U.S. Purcha ses Only. The right peo ple , the right answe r s, right now . ON-LINE ORDERING NO W AV AILABLE ON OUR WEB SITE Ro semountAn aly tical.c om Spec ificati ons subject to ch ange without no tic e. 8 Emerson Process Management 2400 Barranca Parkway Ir vine, CA 92606 USA T el: (949) 757-8500 F ax: (949) 474-7250 Ro semoun t ...

Manufacturer Emerson Category Weather Radio

Documents that we receive from a manufacturer of a Emerson PN-51-T56 can be divided into several groups. They are, among others:
- Emerson technical drawings
- PN-51-T56 manuals
- Emerson product data sheets
- information booklets
- or energy labels Emerson PN-51-T56
All of them are important, but the most important information from the point of view of use of the device are in the user manual Emerson PN-51-T56.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Emerson PN-51-T56, service manual, brief instructions and user manuals Emerson PN-51-T56. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Emerson PN-51-T56.

Similar manuals

A complete manual for the device Emerson PN-51-T56, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Emerson PN-51-T56 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Emerson PN-51-T56.

A complete Emerson manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Emerson PN-51-T56 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Emerson PN-51-T56, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Emerson PN-51-T56, that we can find in the current document
3. Tips how to use the basic functions of the device Emerson PN-51-T56 - which should help us in our first steps of using Emerson PN-51-T56
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Emerson PN-51-T56
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Emerson PN-51-T56 in a specific country, if it was not possible to solve the problem on our own.