  • Genesis Advanced Technologies 2006 - page 1

    USER’S MANUAL GENES I S 2002 TM GENES I S 2006 TM PLASMA THAWING WATER BATH SYSTEM GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.genesisbps.com 0609 ...

  • Genesis Advanced Technologies 2006 - page 2

    GENESIS 2006 SHOWN 2 ...

  • Genesis Advanced Technologies 2006 - page 3

    To assure proper record keeping, please retain this information for future use. DATE PURC HASED ____________________________ SERIAL # _____________________________________ DISTRIBUT OR NAME __________________________ PHONE NUMBER ______________________________ CITY, STATE __________________________________ The GENESIS THAWING SYSTEMS incorporate wa ...

  • Genesis Advanced Technologies 2006 - page 4

    CONTENTS page 1. UNPACKING ........................................................ 5 2. TO THAW BLOOD PRODUCTS ........................ 5 3. TO REMOVE A THAWED BAG ........................ 9 4. TO MAINTAIN THE DEVICE IN THE READY TO THAW STATUS .............................. 9 5. TO CLEAN THE BATH OR TO ADD WATER .. 10 6. CONTROL PANEL AND INDICAT ...

  • Genesis Advanced Technologies 2006 - page 5

    1. UNPACKING 1.1 The following is supplied with your GENESIS 2002/2006  unit. C o n t r o l U n i t W a t e r B a t h T h a w i n g B a g D i s p e n s e r Plasma Overwrap Thawing Bags (1000/Box) P l a s m a B a g H o l d e r L A B a t h C L E A N TM R e f e r e n c e T h e r m o m e t e r T h a w i n g B a g W e i g h t s U s e r ' s M a n ...

  • Genesis Advanced Technologies 2006 - page 6

    indicator will light. To correct this situation, fill bath with water as explained above. Add 2 ml of LABath CLEAN TM each time the water is changed to control bacterial and algae growth. The bath cleaning agent also helps to prev ent the formation of turbidity and odor. 2.3 Plug the device into a 115 volt AC outlet. For GENESIS 2002  , a dedica ...

  • Genesis Advanced Technologies 2006 - page 7

    c. To change the default SET TEMP : You can adjust this temperature within the range of ambient to + 56 ° C, in one degree increments. To adjust, press the SET TEMP key - then press the numeric key pad to indicate the temperature desired. All keys must be pressed within 5 seconds of each other. The display should now show the new SET TEMP. When th ...

  • Genesis Advanced Technologies 2006 - page 8

    thawing time may vary. The amount of air left in the plasma thawing bag during thawing will also affect thawing time. The laboratory personnel can set the desired thawing time for each unit to be thawed. They do not have to all be set for the same thaw time. f. To change the default TIMER : Press the START/CHANGE key for the selected bag location, ...

  • Genesis Advanced Technologies 2006 - page 9

    3. TO REMOVE A THAWED BAG 3.1 Press the ROCK key to stop the rocking motion. a. Remove the thawed bag (and any weights used) from the Plasma Bag Holder. Discard plasma thawing overwrap bag after each use, as they are designed for single use only. b. If there is another bag being thawed, press the ROCK key to resume the rocking motion for the rem ai ...

  • Genesis Advanced Technologies 2006 - page 10

    5. TO CLEAN THE BATH OR TO ADD WATER S A F E T Y : NEVER ADD WATER OR MOVE THE BATH IF THE POWER SWITCH (ON THE LEFT SIDE PANEL) IS TURNED ON. RECOMMENDATION : FOR MAXIMUM SAFETY, UNPLUG THE UNITS POWER CORD FROM THE ELECTRICAL OUTLET. 5.1 TURN THE POWER SWITCH (ON THE LEFT SIDE PANEL) TO OFF POSITION TO TURN THE UNIT POWER OFF. BEFORE REMOVING THE ...

  • Genesis Advanced Technologies 2006 - page 11

    when filled with water. Do not drop the bath on a hard surface. GENESIS 2006  : Remove all, or part of the water. The bath weighs approximately thirty (30) pounds when filled with water. Carefully remove the water bath from the device. 5.6 DO NOT USE BLEACH OR HARSH CLEANERS. Depending on mineral content of water used, the bath may discolor. Pro ...

  • Genesis Advanced Technologies 2006 - page 12

    6. CONTROL PANEL AND INDICATORS GENESIS 2002 TM CONTROL PANEL 12 ...

  • Genesis Advanced Technologies 2006 - page 13

    GENESIS 2006 TM CONTROL PANEL 13 ...

  • Genesis Advanced Technologies 2006 - page 14

    CONTROL KEYS POWER .......... This key turns the unit on and off when the POWER SWITCH on the left panel is turned ON. The POWER SWITCH on the left panel is a master on/off switch to remove electrical power completely from all circuits in the unit. SET TEMP ..... With this key switch you can adjust water temperature within the range of ambient to + ...

  • Genesis Advanced Technologies 2006 - page 15

    INDICATORS HEATING ....... This red indicator is lit when the heating element is actually drawing current from the circuit and heating the water. When it reaches set temperature, the heating indicator light shuts off. FAULT ............ If the water temperature exceeds 58 ° C, an alarm will sound and the red fault light will be illuminated. This m ...

  • Genesis Advanced Technologies 2006 - page 16

    2) The ADD WATER light is on and there is sufficient water in the bath. • With the unit unplugged and raised out of the water, check that there is no obstruction around the float switch in the shroud unit. The float should move freely up and down the stem approximately half-inch. 3) The FAULT light is on. • Check the water temperature. If the w ...

  • Genesis Advanced Technologies 2006 - page 17

    7) Pump does not circulate the water. • Debris may be in the pump impeller housing. Raise the shroud unit and lock in the up position. Inspect pump for debris and remove with tweezers or similar instrum ent. 8) HEATING light is flashing. • This is normal operation. At approximately one degree below the set temperature, the microprocessor begins ...

  • Genesis Advanced Technologies 2006 - page 18

    8. SPECIAL FEATURES • Thaws multiple plasma units simultaneously • Thaws standard size plasma as well as cryoprecipitate • Individual time setting for each bag location • Forced water circulation with gentle rocking motion • Compact size - saves valuable bench space • Quick heating • Simple to operate - easy to read LED display • Li ...

  • Genesis Advanced Technologies 2006 - page 19

    10. SPECIFICATIONS GENESIS 2002 TM GENESIS 2006 TM External Dim ension 21.6” L x 8.3” Wide x 13.2” High 26.6” L x 9.6” Wide x 13.2” High Interior Bath Dimension 13.0” L x 7.0” Wide x 7.5” Deep 18.0” L x 8.0” Wide x 7.5” Deep Bath Volume 2.5 US gallons 3.7 US gallons Controlled Tem p Range Ambient to 56 o C Ambient to 56 o C ...

  • Genesis Advanced Technologies 2006 - page 20

    11. ORDERING/RETURN INFORMATION ORDERING INFORMATION : To place an order for a part or an accessory, please call : GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.genesisbps.com RETURN INFORMATION : Same as above. 1. Obtain a Return Goods Authorization number. 2. Ensure all water is ...

  • Genesis Advanced Technologies 2006 - page 21

    Hospital/Account Name Department and shipping address Name and telephone number of contact Date GENESIS 2002/2006  was purchased S e r i a l # RGA # (Return Goods Authorization #) 6. Mark outside of box clearly, with your return address, and send to GenesisBPS, LLC . (as above) 21 ...

  • Genesis Advanced Technologies 2006 - page 22

    GENESIS 2002  GENESIS 2006  PLASMA THAWING WATER BATH SYSTEM WARRANTY The GENESIS 2002/2006  , Plasma Thawing Water Bath System is warranted against defects in materials and workmanship for a period of one year from the date of purchase. The warranty applies only to products that have not been tampered with, used for purposes other than th ...

Manufacturer Genesis Advanced Technologies Category Bathroom Aids

Documents that we receive from a manufacturer of a Genesis Advanced Technologies 2006 can be divided into several groups. They are, among others:
- Genesis Advanced Technologies technical drawings
- 2006 manuals
- Genesis Advanced Technologies product data sheets
- information booklets
- or energy labels Genesis Advanced Technologies 2006
All of them are important, but the most important information from the point of view of use of the device are in the user manual Genesis Advanced Technologies 2006.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Genesis Advanced Technologies 2006, service manual, brief instructions and user manuals Genesis Advanced Technologies 2006. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Genesis Advanced Technologies 2006.

A complete manual for the device Genesis Advanced Technologies 2006, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Genesis Advanced Technologies 2006 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Genesis Advanced Technologies 2006.

A complete Genesis Advanced Technologies manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Genesis Advanced Technologies 2006 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Genesis Advanced Technologies 2006, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Genesis Advanced Technologies 2006, that we can find in the current document
3. Tips how to use the basic functions of the device Genesis Advanced Technologies 2006 - which should help us in our first steps of using Genesis Advanced Technologies 2006
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Genesis Advanced Technologies 2006
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Genesis Advanced Technologies 2006 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Genesis Advanced Technologies 2006?

Use the form below

If you did not solve your problem by using a manual Genesis Advanced Technologies 2006, ask a question using the form below. If a user had a similar problem with Genesis Advanced Technologies 2006 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)