Manual Teledyne 05386D

41 pages 1.79 mb
Download

Go to site of 41

Summary
  • Teledyne 05386D - page 1

    M200EU-NOy CONVERTER ADDENDUM (for use with M200E Operators Manual, PN 04410 and with the M200EU Addendum, PN 05385) © TELEDYNE ADVANCED POLLUTION INSTRUMENTATION 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 USA Toll-free Phone: 800-324-5190 Phone: 858-657-9800 Fax: 858-657-9816 Email: api-sales@teledyne.com Website: http://www.teledyne-api.co ...

  • Teledyne 05386D - page 2

    ...

  • Teledyne 05386D - page 3

    ...

  • Teledyne 05386D - page 4

    ...

  • Teledyne 05386D - page 5

    05386 Rev. D i Safety Messages Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a safety alert symbol, and are placed throughout this manual and in side the instrument. The symbols with messages ...

  • Teledyne 05386D - page 6

    ii 05386 Rev. D ...

  • Teledyne 05386D - page 7

    05386 Rev. D iii Table of Contents Safety Messag es ................................................................................................................ ...................... i Table of Conten ts.............................................................................................................. ..................... iii List ...

  • Teledyne 05386D - page 8

    iv 05386 Rev. D 8.0 M200EU-NO Y SPARE P ARTS LI ST ...................................................................... 37 List of Figures FIGURE 1 REAR PANEL PNEUMATIC CONNECTIONS ................................................................................ 8 FIGURE 2 REPLACING THE PARTICULATE FILTER .......................................... ...

  • Teledyne 05386D - page 9

    05386 Rev. D 5 1.0 INTRODUCTION The M200EU-NOy converter is designed to s upport a NOx analyzer by converting multiple, unstable compounds grouped under the name NOy . The converter is mounted on the analyzer at the sample inlet to minim ize flow time between sam ple in and converter, thereby optimizing measurement accuracy. This manual addendum is ...

  • Teledyne 05386D - page 10

    Model 200EU-NOy Opti on Manual Addendum 6 05386 Rev. D This page intentionally left blank. ...

  • Teledyne 05386D - page 11

    05386 Rev. D 7 2.0 GETTING STARTED The NOy converter has been designed to operate with the M200EU low level NOx analyzer. 2.1 Unpacking The M200EU with NOy Option comes in three boxes:  M200EU analyzer  Bypass Pump Chassis  External Converter housed in a stainless enclosure, and umbilical cable assembly 1. Verify that there is no apparent ...

  • Teledyne 05386D - page 12

    Model 200EU-NOy Opti on Manual Addendum 8 05386 Rev. D Figure 1 Rear Panel Pneumatic Connections ...

  • Teledyne 05386D - page 13

    Model 200EU-NOy Opti on Manual Addendum 05386 Rev. D 9 2.3 Initial Operation 1. After confirming proper supply voltage, turn on the instrum ent power. If you are unfamiliar with the M200EU with NOy Op tion, we recomm end that you read the overview in Section 3 before proceeding.  The power indicator light and display shoul d immediately light, i ...

  • Teledyne 05386D - page 14

    Model 200EU-NOy Opti on Manual Addendum 10 05386 Rev. D This page intentionally left blank. ...

  • Teledyne 05386D - page 15

    05386 Rev. D 11 3.0 SPECIFICATIONS, WARRANTY 3.1 Specifications Table 2. Specifications for the converter and the Bypass Pump Converter Remotely mounted m olybdenum converter with tem perature controller in a NEMA – 4 enclosure. Converter Temperature 315 o C  7 o C with read-out on front panel of bypass pump chassis. Tube/Electrical Cable Asse ...

  • Teledyne 05386D - page 16

    Model 200EU-NOy Opti on Manual Addendum 12 05386 Rev. D 3.2 Warranty TELEDYNE ADVANCED POLLUTION INSTRUMENTATION, INC. 02024c Prior to shipment, API equipment is thoroughly inspected and tested. Should equipment failure occur, API assures its customers that prompt service and support will be available. COVERAGE After the warranty period and through ...

  • Teledyne 05386D - page 17

    05386 Rev. D 13 4.0 THE M200EU-NO Y CONVERTER 4.1 Principle of Operation The T-API Model M200EU with NOy Option is de signed to measure the concentration of NO, NO2, and other compounds that are too unstable to be measured when taken in through the normal ambient air sam ple inlet system. Please refer to the M200EU m anual supplied with this system ...

  • Teledyne 05386D - page 18

    Model 200EU-NOy Opti on Manual Addendum 14 05386 Rev. D This page intentionally left blank. ...

  • Teledyne 05386D - page 19

    05386 Rev. D 15 5.0 CALIBRATION AND ZERO/SPAN CHECKS Unlike most NOx analyzers, the M200EU with th e NOy option does not have a samp le inlet port on the rear panel of the M200EU. The sample port is located on the External Converter; therefore Zero/Span calibration is different than a norm al NOx instrument. If a fitting is located at the sample po ...

  • Teledyne 05386D - page 20

    Model 200EU-NOy Opti on Manual Addendum 16 05386 Rev. D 5.1 Calibration or Cal Check Procedure The calibration of the instrument can be checked or adjusted using gas introduced through the calibration port “CAL IN” on the Bypass Pump Modul e. The calibration gas is routed to a TEE near the sample inlet port on the remote converter assem bly. Si ...

  • Teledyne 05386D - page 21

    Model 200EU-NOy Opti on Manual Addendum 05386 Rev. D 17 Table 4 Enter Expected Span Gas Concentrations Procedure For NO & NOy Step Action Comment 1. Press CAL-CONC-NOy This key sequence causes the M200EU to prompt for the expected NOy concentration. Enter the NOy span concentration value by pressing the key under each digit until the expected v ...

  • Teledyne 05386D - page 22

    Model 200EU-NOy Opti on Manual Addendum 18 05386 Rev. D This page intentionally left blank. ...

  • Teledyne 05386D - page 23

    05386 Rev. D 19 6.0 MAINTENANCE 6.1 Maintenance Schedule The schedule for preventative maintenance is presented in Table 6 below. Table 6 Preventative Maintenance Calendar Item Maintenance Interval Zero/Span Calibration Annually or after maintenance or repairs Zero/Span Checks Daily Particulate Filter Weekly as needed Ozone Flow Check every year an ...

  • Teledyne 05386D - page 24

    Model 200EU-NOy Opti on Manual Addendum 20 05386 Rev. D 6.2 Replacing the Sample Particulate Filters The particulate filter should be inspected often for signs of plugging or contamination. It is also common for dirt particles to cause instrum ent drift, and affect accuracy. To check and change the filter: 1. Fold down the Bypass Pump Module front ...

  • Teledyne 05386D - page 25

    Model 200EU-NOy Opti on Manual Addendum 05386 Rev. D 21 Figure 2 Replacing the Particulate Filter ...

  • Teledyne 05386D - page 26

    Model 200EU-NOy Opti on Manual Addendum 22 05386 Rev. D 6.3 Checking Analyzer Flow Rate The external sample pump is capable of m aintaining the reaction cell pressure at less than 5.0 In-Hg-A. If a higher pressure is noted, th e pump m ay need servicing. Check the pump and pneumatic system for leaks or rebuild pum p. CAUTION! Never operate the anal ...

  • Teledyne 05386D - page 27

    Model 200EU-NOy Opti on Manual Addendum 05386 Rev. D 23 Figure 3 NOy Converter Assembly NOy Out NO Out Sample In ...

  • Teledyne 05386D - page 28

    Model 200EU-NOy Opti on Manual Addendum 24 05386 Rev. D 6.5 Inspecting Pneumatic Lines Particulate matter in the pneumatic lines will affect both flow rate and response tim e. It is important that the pneumatic system be pe riodically inspected and thoroughly cleaned if necessary. Clean by disassembling and passing me thanol through three tim es. D ...

  • Teledyne 05386D - page 29

    Model 200EU-NOy Opti on Manual Addendum 05386 Rev. D 25 Figure 4 Pneumatics diagram (standard and Z/S option) ...

  • Teledyne 05386D - page 30

    Model 200EU-NOy Opti on Manual Addendum 26 05386 Rev. D Figure 5 Pneumatic diagram (OPT 86 option – Sample Driers) ...

  • Teledyne 05386D - page 31

    Model 200EU-NOy Opti on Manual Addendum 05386 Rev. D 27 6.6 Checking for Leaks If a leak checker is not available, it is possibl e to leak check the instrument using the samp le pump plus a shut-off valve. 1. Turn off instrument power and pump power. 2. Cap the sample inlet port, bypass out port, ozone generator air inlet, and zero/span inlets (if ...

  • Teledyne 05386D - page 32

    Model 200EU-NOy Opti on Manual Addendum 28 05386 Rev. D This page intentionally left blank. ...

  • Teledyne 05386D - page 33

    05386 Rev. D 29 7.0 TROUBLESHOOTING, ADJUSTMENTS CAUTION! The operations outlined in this section must be performed by qualified maintenance personnel only! This section of the manual contains inform ation on diagnosing and repairing instrument performance problems. It provides troubleshooti ng procedures that address problems to the board level. A ...

  • Teledyne 05386D - page 34

    Model 200EU-NOy Opti on Manual Addendum 30 05386 Rev. D 7.1 Operation Verification – Diagnostic Techniques 7.1.1 Pneumatic System ...

  • Teledyne 05386D - page 35

    Model 200EU-NOy Opti on Manual Addendum 05386 Rev. D 31 The pneumatic system is diagram ed in ...

  • Teledyne 05386D - page 36

    Model 200EU-NOy Opti on Manual Addendum 32 05386 Rev. D Figure 4 . 7.1.2 Leak Check Refer to Section 5.6 for the leak check procedure. 7.1.3 Bypass Pump Diagnostic Procedures The bypass pump is capable of maintaining a total bypass flowrate of ~2.2 lpm. If flow checks indicate a fall off in bypass flow, the pump may need servicing. Check the pum p, ...

  • Teledyne 05386D - page 37

    Model 200EU-NOy Opti on Manual Addendum 05386 Rev. D 33 Indications that would require re-programming are:  Failure / replacement of controller.  Replacement of thermocouple elem ent.  Controller’s loss of PID parameters leading to instability of the tem p setpoint. In the event that the control parameters are changed or in the event tha ...

  • Teledyne 05386D - page 38

    Model 200EU-NOy Opti on Manual Addendum 34 05386 Rev. D SECONDARY MENU (The fixed characteristics of the system). PRESS DISPLAY ACTION SEL hold until p-n1 SET TO 0 – REVERSE ACTION SEL hold until p-n2 SET TO 2 = TYPE J THERMOCOUPLE SEL p-dF DIGITAL FILTER SET TO 5 SEL P-SL LOWER LIMIT SET TO -40 (-40ºC) SEL P-SU UPPER LIMIT SET TO 350 (350ºC) S ...

  • Teledyne 05386D - page 39

    Model 200EU-NOy Opti on Manual Addendum 05386 Rev. D 35 Figure 6 Electrical Diagram ...

  • Teledyne 05386D - page 40

    Model 200EU-NOy Opti on Manual Addendum 36 05386 Rev. D This page intentionally left blank. ...

  • Teledyne 05386D - page 41

    05386 Rev. D 37 8.0 M200EU-NO Y SPARE PARTS LIST The following spare parts list supplements th e spare parts list for the M200EU analzyer. Note: Use of replacement parts other than those supplied by API may result in non- compliance with European Standard EN 61010-1. PART NO. DESCRIPTION 000940400 ORIFICE, 4 MIL 80 CC, O3 FLOW Rx CELL 000941000 ORI ...

Manufacturer Teledyne Category Table Top Game

Documents that we receive from a manufacturer of a Teledyne 05386D can be divided into several groups. They are, among others:
- Teledyne technical drawings
- 05386D manuals
- Teledyne product data sheets
- information booklets
- or energy labels Teledyne 05386D
All of them are important, but the most important information from the point of view of use of the device are in the user manual Teledyne 05386D.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Teledyne 05386D, service manual, brief instructions and user manuals Teledyne 05386D. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Teledyne 05386D.

Similar manuals

A complete manual for the device Teledyne 05386D, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Teledyne 05386D by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Teledyne 05386D.

A complete Teledyne manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Teledyne 05386D - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Teledyne 05386D, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Teledyne 05386D, that we can find in the current document
3. Tips how to use the basic functions of the device Teledyne 05386D - which should help us in our first steps of using Teledyne 05386D
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Teledyne 05386D
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Teledyne 05386D in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Teledyne 05386D?

Use the form below

If you did not solve your problem by using a manual Teledyne 05386D, ask a question using the form below. If a user had a similar problem with Teledyne 05386D it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)