  • Invacare IRC10LX - page 1

    Dealer: This manual MUST be given to the end user. User: Before using this prod uct, read this manual and save for futu re reference. Operator’s Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Platinum ™ Series XL, 5, 10 Oxygen Concentrators Standard, with SensO 2 ® and HomeFill ® In ...

  • Invacare IRC10LX - page 2

    WARNING Platinum ™ Series 2 P art No 1118353  WARNING DO  NOT  use  this  product  or  any  av ailable  optional  equipment  without  first  completely  reading  and  understanding  these  instructions  and  any  additional  instructional  material  such  as  owner ’ ...

  • Invacare IRC10LX - page 3

    Part No 1118353 3 Platinum ™ Series TABLE OF CONTENTS TABLE OF CONTENTS ACCESSORIES .................................................. 2 REGISTER YOUR PRODUCT .............................. 4 SPECIAL NOTES ............................................... 5 LABEL LOCATION ............................................ 7 SECTION 1—GENERAL GUIDELINES ...

  • Invacare IRC10LX - page 4

    TABLE OF CONTENTS Platinum ™ Series 4 Part No 1118353 TABLE OF CONTENTS SECTION 6—MAINTENANCE .......................... 23 Routine Maintenanc e ....................... ................................................. 23 Cleaning the Cabinet Filter ..................... ....................... ............. 23 Cleaning the Cabinet ............ ...

  • Invacare IRC10LX - page 5

    SPECIAL NOTES Part No 1118353 5 Platinum ™ Series SPECIAL NOTES Signal  w ords  are  used  in  this  man ual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  following  table  for ? ...

  • Invacare IRC10LX - page 6

    SPECIAL NOTES Platinum ™ Series 6 Part No 1118353 CAUTION “Caution:  Federal  law  restricts  this  device  to  sale  or  rental  by  or  on  order  of  a  physician,  or  any  other  practitioner  licensed  by  the  law  of  the  State  in  which  he/she ...

  • Invacare IRC10LX - page 7

    LABEL LOCATION Part No 1118353 7 Platinum ™ Series LABEL LOCATION HomeFill II Compatib le I/O Operating Instructions (See Operator's Manual) Alarms (See Operator's Manual) W ARNING - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and awa y from areas where oxygen is being delivere ...

  • Invacare IRC10LX - page 8

    SECTION 1—GENERAL GUIDELINES Platinum ™ Series 8 Part No 1118353 SECTION 1—GENERAL GUIDELINES In  order  to  ensure  the  safe  installation,  assembly  and  operation  of  the  Platinum  concentrator  these  instructions  MUST  be  followed.  WARNING This  section  contains  i ...

  • Invacare IRC10LX - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1118353 9 Platinum ™ Series Operating Information For  optimum  performance,  Inv acare  recommends  that  each  concentrator  be  on  and  running  for  a  minimum  of  30  minutes  at  a  time.  Shorter  periods  of  operation  may ...

  • Invacare IRC10LX - page 10

    SECTION 1—GENERAL GUIDELINES Platinum ™ Series 10 Part No 1118353 Other  devices  may  experience  interference  from  ev en  the  low  levels  of  electromagnetic  emissions  permitted  by  the  above  standards.  To  determine  if  the  emissions  from  the  Platinum ? ...

  • Invacare IRC10LX - page 11

    SECTION 2—FEATURES Part No 1118353 11 Platinum ™ Series SECTION 2—FEATURES Oxygen Outlet Flowmeter Oxygen Purity Indicator Lights (SensO 2 model only )/ Fault and Power Indicator Lights (LX model only) Power Switch Circuit Breaker Elapsed Time Meter Cabinet Filter (one of two shown) REAR VIEW ACCESSORIES (NOT SHOWN): Remote Humidifier/Flo wme ...

  • Invacare IRC10LX - page 12

    SECTION 3—HANDLING Platinum ™ Series 12 Part No 1118353 SECTION 3—HANDLING The  concentrator  should  AL W A YS  be  kept  in  the  upright  position  to  prevent  cabinet  damage  while  being  transported. If  the  concentrator  is  to  be  reshipped  by  common  car ...

  • Invacare IRC10LX - page 13

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1118353 13 Platinum ™ Series SECTION 4—TYPICAL PRODUCT PARAMETERS *NOTE:  Stated  concentrati on  levels  achieved  after  initial  warm ‐ up  period  (appr oximately  30  minutes). Electrical Requirements: 120 VAC ± 10% (132 VAC/10 8 VAC), 60 Hz Rated Current In ...

  • Invacare IRC10LX - page 14

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum ™ Series 14 Part No 1118353 Flow Range: Platinum 5 and XL - 0.5 to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory. Platinum 10 - 2 L/min to 10 L/min Low Flow Alarm: Platinum 5 - 0 L/min to 0.5 L/min, Platinum 10 - 0 L/min to 1 L/m ...

  • Invacare IRC10LX - page 15

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1118353 15 Platinum ™ Series Operating Ambient Temperature: 50°F - 95°F (10°C - 35°C) at 20-60% relative humidity Exhaust: Platinum 5 and XL: Less than Ambient + 35°F (+19°C). Platinum 10: Less than Ambient + 45°F (+25°C) Oxygen Output Temperature: Platinum 5 and XL: Less than Ambient + 4°F ...

  • Invacare IRC10LX - page 16

    SECTION 5—OPERATING INSTRUCTIONS Platinum ™ Series 16 Part No 1118353 SECTION 5— OPERATING INSTRUCTIONS Introduction Yo u r  oxygen  concentrator  is  intended  for  individual  use  indoors.  It  is  an  electronically  operated  device  that  separates  oxygen  from  room  air . ? ...

  • Invacare IRC10LX - page 17

    SECTION 5—OPERATING INSTRUCTIONS Part No 1118353 17 Platinum ™ Series Set-Up Plug in Power Cord Plug  in  pow er  cord  to  an  electrical  outlet.  POLARIZED PLUG WARNING As  a  safety  feature,  this  appliance  may  hav e  a  polarized  plug  (one  blade  is  wider  than ? ...

  • Invacare IRC10LX - page 18

    SECTION 5—OPERATING INSTRUCTIONS Platinum ™ Series 18 Part No 1118353 FIGURE 5.2 Humidifier Compartmen t 3. Remove  the  filter  access  door  located  on  the  side  of  the  concentrator .  Insert  a  flathead  screwdriv er  in  the  plate  groove  on  the  top  edge  of ...

  • Invacare IRC10LX - page 19

    SECTION 5—OPERATING INSTRUCTIONS Part No 1118353 19 Platinum ™ Series 7. Attach  oxygen  tubing  from  the  humidifier  bottle  to  oxygen  outlet  connector  on  the  oxygen  concentrator  (FIGURE  5.4). 8. Attach  the  cannula/patient  supply  tubing  to  the  humidifier  ...

  • Invacare IRC10LX - page 20

    SECTION 5—OPERATING INSTRUCTIONS Platinum ™ Series 20 Part No 1118353 Flowrate NOTE:  For  this  procedure,  refe r  to  FIGURE  5.6  on  page 20. 1. T urn  the  flowrate  knob  to  the  setting  prescribed  by  yo ur  physician  or  therapist. CAUTION Never  se t  the  flo ...

  • Invacare IRC10LX - page 21

    SECTION 5—OPERATING INSTRUCTIONS Part No 1118353 21 Platinum ™ Series Front Panel Indicators - Units with SensO 2 NOTE:  For  this  procedure,  refe r  to  FIGURE  5.7. The  SensO 2  feature  monitors  the  purity  of  oxygen  generated  by  the  oxygen  concentrator .  If  purity ? ...

  • Invacare IRC10LX - page 22

    SECTION 5—OPERATING INSTRUCTIONS Platinum ™ Series 22 Part No 1118353 Front Panel Indicators - Units without SensO 2 NOTE:  For  this  procedure,  refe r  to  FIGURE  5.8. FIGURE 5.8 Front Panel Indicators - Units without Se nsO2 Initial Startup of the Concentrator NOTE:  Concentrator  may  be  used  during ? ...

  • Invacare IRC10LX - page 23

    SECTION 6—MAINTENANCE Part No 1118353 23 Platinum ™ Series SECTION 6—MAINTENANCE  WARNING The  Invacare  concentrators  are  specifically  designed  to  minimize  routine  preventiv e  mainte nance.  Only  qualified  personnel  should  perform  preventiv e  maintenance  on  the  c ...

  • Invacare IRC10LX - page 24

    SECTION 6—MAINTENANCE Platinum ™ Series 24 Part No 1118353 2. Clean  the  cabinet  filters  with  a  vac u um  cleaner  or  wa s h  in  w arm  soapy  wa t e r  and  rinse  thoroughly . 3. Dry  the  filters  thoroughly  before  reinstallation. FIGURE 6.1 Cleaning the Cabinet Filter C ...

  • Invacare IRC10LX - page 25

    SECTION 6—MAINTENANCE Part No 1118353 25 Platinum ™ Series FIGURE 6.2 Preventive Maintenance R ecord ON EACH INSPECTION MODEL NO. I R C SERIAL NO. PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOU ...

  • Invacare IRC10LX - page 26

    SECTION 7—TROUBLESHOOTING GUIDE Platinum ™ Series 26 Part No 1118353 SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: SHORT BEEPS, LONG PAUSE Concentrator NOT operating, power switch ON. Beep.... ..Beep... ... Main Power Loss: 1. Power cord NOT plugged in. 1. Insert plug into outlet. 2. No power at outlet. 2. Inspect ...

  • Invacare IRC10LX - page 27

    SECTION 7—TROUBLESHOOTING GUIDE Part No 1118353 27 Platinum ™ Series YELLOW or RED LIGHT Illuminated. Alarm: CONTINUOUS On red light only. * Only applies to 5LXO 2 and 10LXO 2 SensO 2 models. 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tub- ing, cannula or humidifier.* 2. Inspect for kinks or block- ages. Correct, c ...

  • Invacare IRC10LX - page 28

    NOTES Platinum ™ Series 28 Part No 1118353 NOTES ...

  • Invacare IRC10LX - page 29

    NOTES Part No 1118353 29 Platinum ™ Series NOTES ...

  • Invacare IRC10LX - page 30

    LIMITED WARRANTY Platinum ™ Series 30 Part No 1118353 LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURE D AFTER JULY 4, 1975. This warranty is extended only to th e original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer. This wa ...

  • Invacare IRC10LX - page 31

    LIMITED WARRANTY Part No 1118353 31 Platinum ™ Series For warranty service, please contact In vacare Corporation's Service Department at the toll free number on the back cover. Upon receiving notice of an alleged de fect in a prod uct, Invacar e Corporatio n will issue a serialized return authorization. It sh all then be the responsibility o ...

  • Invacare IRC10LX - page 32

    Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707 Part No 1118353 Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted. © 2009 Invacare Corporation Rev J - 4/09 This  pr ...

Manufacturer Invacare Category Oxygen Equipment

Documents that we receive from a manufacturer of a Invacare IRC10LX can be divided into several groups. They are, among others:
- Invacare technical drawings
- IRC10LX manuals
- Invacare product data sheets
- information booklets
- or energy labels Invacare IRC10LX
All of them are important, but the most important information from the point of view of use of the device are in the user manual Invacare IRC10LX.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Invacare IRC10LX, service manual, brief instructions and user manuals Invacare IRC10LX. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Invacare IRC10LX.

Similar manuals

A complete manual for the device Invacare IRC10LX, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Invacare IRC10LX by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Invacare IRC10LX.

A complete Invacare manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Invacare IRC10LX - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Invacare IRC10LX, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Invacare IRC10LX, that we can find in the current document
3. Tips how to use the basic functions of the device Invacare IRC10LX - which should help us in our first steps of using Invacare IRC10LX
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Invacare IRC10LX
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Invacare IRC10LX in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Invacare IRC10LX?

Use the form below

If you did not solve your problem by using a manual Invacare IRC10LX, ask a question using the form below. If a user had a similar problem with Invacare IRC10LX it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)