Manual Teledyne 200

24 pages 0.61 mb
Download

Go to site of 24

Summary
  • Teledyne 200 - page 1

    INSTRUCTION MANUAL 200/202 SERIES FLOWMETERS/CONTROLLERS TELEDYNE HASTINGS INSTRUMENTS ISO 9001 CER TIFIED ...

  • Teledyne 200 - page 2

    140-072011 200/202 Series Page 2 of 24 Manual Print History The print history shown below lists the printing da tes of all revisions and addenda created for this manual. The revision level letter in creases alphabetically as the ma nual undergoes subsequent updates. Addenda, which are released between revisions, cont ain important change informatio ...

  • Teledyne 200 - page 3

    140-072011 200/202 Series Page 3 of 24 Table of Contents 1. GENERAL INFO RMATIO N ....................................................................................................................................4 1.1. F EATURES ....................................................................................................................... ...

  • Teledyne 200 - page 4

    140-072011 200/202 Series Page 4 of 24 1. General Information The Hastings HFM-200 mass flow meter and HFC-202 fl ow controller are designed to accurately measure and control mass flow over the range of 10 sccm to 30 s lm, without corrections or compensations for gas pressure and temperature with an accuracy of better than ±1% FS. Hastings mass fl ...

  • Teledyne 200 - page 5

    140-072011 200/202 Series Page 5 of 24 1.2. Specifications HFM-200 * Accuracy 1 and Linearity ................................................................................................................ ........ ±1% F.S. Repeatability .............................................................................................................. ...

  • Teledyne 200 - page 6

    140-072011 200/202 Series Page 6 of 24 1.4. Optional 4-20 mA Current Output An option to the standard 0 - 5 VDC output is the 4 - 20 mA current output that is pr oportional to flow. The 4 - 20 mA signal is prod uced from the 0 - 5 VDC output of the flow m eter. The current loop output is useful for remote applications where pickup noise could subst ...

  • Teledyne 200 - page 7

    140-072011 200/202 Series Page 7 of 24 1.5.2. Intercon necting Cables Cables are available from Hastings, in various length s , to connect from the 15 pin "D" connector on the back of the Power Pod directly to any of the 200 seri es and 300 seri es flow inst ruments (including digital versions). More information about the available cables ...

  • Teledyne 200 - page 8

    140-072011 200/202 Series Page 8 of 24 2. Inst allation and Operation This section contains the steps necessary to install a new flow me ter/controller into operation as quickly and easily as possible. Please read the following th oroughly before attempting to install the instrument. 2.1. Receiving Inspection Carefully unpack the Hastings HFM-200/H ...

  • Teledyne 200 - page 9

    140-072011 200/202 Series Page 9 of 24 2.4. Mechanical Connections 2.4.1. Standard Configuratio n The flow meter may be mounted in any position as long as the direction of gas flow through the instrument follows the arrow marke d on the bottom of the flow meter case label. The preferred orientation is with the inlet and outlet fittings in a h orizo ...

  • Teledyne 200 - page 10

    140-072011 200/202 Series Page 10 of 24 EXAMPLE: Suppose a 4” LFE is selected. The length of 4” diameter tubing or pi pe, upstream of the sensor inlet tap, will be a minimum of 20 “. The minimum length of 4” tubing or pipe downstream from the sensor outlet tap will be 4”. The length of additional 4” inlet tubing required for the 4” LF ...

  • Teledyne 200 - page 11

    140-072011 200/202 Series Page 11 of 24 24 Volt Connections Refer to the diagram to the right when connecting 24 Volt units. Connect the positive lead of the power supply to pin 7 of the DE-9 connector and negative lead to pin 4. The supply input is diode protecte d such that reversing the input polarity will not damage the instrument. The power su ...

  • Teledyne 200 - page 12

    140-072011 200/202 Series Page 12 of 24 2.6.2. Zero Check Turn the power supply on if not already ene rgized. Allow for a 1 h our warm-up. Stop all fl ow through the instrument and wait 2 minutes. Caution: Do not assume that all metering valves com pletely shut off the flow. Even a slight leakage will cause an indicatio n on the meter and an appare ...

  • Teledyne 200 - page 13

    140-072011 200/202 Series Page 13 of 24 The “COMMAND” pot adjusts the Analog command signal sent to the flo w controller. The setting for each controller connected to the po wer supply can be obse rved. (Depending on how the p ower supply was set up, the display could indicate in flow units or percent of full scale). Hastings Power Su pply Feat ...

  • Teledyne 200 - page 14

    140-072011 200/202 Series Page 14 of 24 2.7.3. Operation with an external sensor (Fig. 2.2) In some instances, it might be desirable to use an external sensor to provide process info rmation to the control circuitry in the flow controller. Fo r example, you might want to control the pressure in a vacuum system by adjusting the rate at which the sys ...

  • Teledyne 200 - page 15

    140-072011 200/202 Series Page 15 of 24 controller circuit establishes contro l at the new command point. This jumper does not affect the system gain and will not dampen out oscillations. To adjust the response, you need a means of pro ducing a step change in the command voltage from 10% of full scale to 100% of full scale. Follow the steps outline ...

  • Teledyne 200 - page 16

    140-072011 200/202 Series Page 16 of 24 3. Theory of Operation This section contains a functional description of Ha stings flow cont rollers. Detailed schematics and parts lists can be obtained by contacting Hastings us ing the contact information found at the end of this document. In this section and other sections througho ut this manual, when a ...

  • Teledyne 200 - page 17

    140-072011 200/202 Series Page 17 of 24 relatively constant over wide ranges of temperat ure and pre ssure, the flow meter ma y be calibrated directly in mass units for those gases. Changes in gas composition usually only require application of a simple multiplier to the air calibration to account for the difference in heat capacity and thus the fl ...

  • Teledyne 200 - page 18

    140-072011 200/202 Series Page 18 of 24 3.5. Valve The control valve is an “automatic m etering solenoid” valve. While most solenoids operate in either the fully open or fully clos ed sta te, the automatic meteri ng solenoid valve is desi gned to control flow (see Figure 3.5). A spring, connected to the plunger a ssembly, holds a magnetic plung ...

  • Teledyne 200 - page 19

    140-072011 200/202 Series Page 19 of 24 4. Maintenance This section contains service and calibration informatio n. Some portions of the instrument are delicate. Use extreme care when servicing the flow controller. 4.1. Authorized Maintenance With proper care in installation and use, the flow controller will require little or no maintenance. If main ...

  • Teledyne 200 - page 20

    140-072011 200/202 Series Page 20 of 24 CAUSE : Plugged orifice. ACTION : Verify the presence o f a 10-50 psig pressure across the instrument. If present, shut off gas supply and power supply. Remove orif ice per Sec tion 4.9. Examine orifice . If plugged, clean or replace as applicable. Reassemble valve. SYMPTOM : Flow meter reads other than 0.00 ...

  • Teledyne 200 - page 21

    140-072011 200/202 Series Page 21 of 24 4.3.2. Miscellaneous adjust ments Periodically, during normal ope ration, the ZERO should be chec ked and adjusted when required. If the instrument is not shutting completely off when the Valve Override CLOSE function is active, or a command of zero flow has been given, then the or ifice may require approxima ...

  • Teledyne 200 - page 22

    140-072011 200/202 Series Page 22 of 24 B) If P u >2P d , use formula 1; otherwise use formula 2. C) Use a consistent set of units for pressure, flow, and density (i.e all lengths, masses, times in the same units, cm, ft, kg, sec etc.) 1 liter/minute = 1.667 x 10 -5 m 3 /sec, 1 gm/lite r = 1 kg/m 3 , 1 psia = 6895 kg/m*sec 2 , 1 Pa = 1 kg/m*se c ...

  • Teledyne 200 - page 23

    140-072011 200/202 Series Page 23 of 24 Lubricate the O-rings slightly with a silicone based grease, and the threads with anti-galling compound. Push the orifice into its hole and screw it in until it is flush with the instrument base. Apply pressure to the inlet side of the instrument. Enable the Valve Override CLOSE function or unplug the instrum ...

  • Teledyne 200 - page 24

    140-072011 200/202 Series Page 24 of 24 5. W arranty and Rep air 5.1. Warranty Repair Policy Hastings Instruments warrants this produc t for a period of one ye ar from the date of shipment to be free from defects in material and work manship. This warra nty does not apply to defects or failures resulting from unauthorized modification, misuse o r m ...

Manufacturer Teledyne Category Automobile Parts

Documents that we receive from a manufacturer of a Teledyne 200 can be divided into several groups. They are, among others:
- Teledyne technical drawings
- 200 manuals
- Teledyne product data sheets
- information booklets
- or energy labels Teledyne 200
All of them are important, but the most important information from the point of view of use of the device are in the user manual Teledyne 200.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Teledyne 200, service manual, brief instructions and user manuals Teledyne 200. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Teledyne 200.

Similar manuals

A complete manual for the device Teledyne 200, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Teledyne 200 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Teledyne 200.

A complete Teledyne manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Teledyne 200 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Teledyne 200, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Teledyne 200, that we can find in the current document
3. Tips how to use the basic functions of the device Teledyne 200 - which should help us in our first steps of using Teledyne 200
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Teledyne 200
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Teledyne 200 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Teledyne 200?

Use the form below

If you did not solve your problem by using a manual Teledyne 200, ask a question using the form below. If a user had a similar problem with Teledyne 200 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)