    Teledyne API, Model 265A Chemiluminescence Ozone Analy zer Manual Addendum, 03389, Rev. B, DCN 5254 HIGHLIGHTS The purpose of this Highlights Page is to list the changes that were incorporated into this Manual per DCN 5254. Chapter / Page Number Description Title Page  Changed Revision number and date.  Changed address.  Added DCN number t ...

  • Teledyne 265A - page 2

    THIS PAGE IS INTENTIONALLY LEFT BLANK ...

  • Teledyne 265A - page 3

    TELEDYNE INSTRUMENTS Advanc ed Polluti on Instrumentati on A Tel e dyn e T ech nologies C ompany MANUAL ADDENDUM MODEL 265A CHEMILUMINESCENCE OZONE ANALYZER (ADDENDUM TO M200A INSTRUCTION MANUAL #02246) TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-2251 TOLL-FREE: 800-324-5190 FAX: 858- ...

  • Teledyne 265A - page 4

    THIS PAGE IS INTENTIONALLY LEFT BLANK ...

  • Teledyne 265A - page 5

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 i PRINTED DOCUMENTS ARE UNCONTROLLED SAFETY MESSAGES Your safety and the safety of others is very im portant. We have provided m any important safety messages in this manual. Please read these m essages carefully. A safety message alerts you to potential hazards that ...

  • Teledyne 265A - page 6

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 ii PRINTED DOCUMENTS ARE UNCONTROLLED TABLE OF CONTENTS SAFETY MESSAGES ....................................................................................................... I TABLE OF CONTENTS ....................................................................... ...

  • Teledyne 265A - page 7

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 iii PRINTED DOCUMENTS ARE UNCONTROLLED LIST OF FIGURES FIGURE 1-1: REMOVAL OF SHIPPING SCREWS & CHECK FOR CORRECT POWER ............ 3 FIGURE 1-2: M265A REAR PANEL .................................................................................................. ...

  • Teledyne 265A - page 8

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 iv PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK ...

  • Teledyne 265A - page 9

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 1 PRINTED DOCUMENTS ARE UNCONTROLLED 1 GETTING STARTED 1.0 Overview The Model 265A is a close derivative of the Model 200A Chemiluminescence NOx Analyzer. Because of this, the Model 200A Instruction manual continues to be valid as a reference manual covering the deta ...

  • Teledyne 265A - page 10

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 2 PRINTED DOCUMENTS ARE UNCONTROLLED 1.2 Electrical and Pneumatic Connections 1. See Figure 1-2 and Figure 1-3 to locate the rear panel electrical and pneumatic connections. 2. Attach the pump to the “Exhaust Out” port on the instrument rear panel. The exhaust fr ...

  • Teledyne 265A - page 11

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 3 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 1-1: Removal of Shipping Screws & Check for Correct Pow er ...

  • Teledyne 265A - page 12

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 4 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 1-2: M265A Rear Panel ...

  • Teledyne 265A - page 13

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 5 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 1-3: Inlet and Exhaust Venting Recommendations ...

  • Teledyne 265A - page 14

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 Figure 1-4: V/F Board Switch Settings 6 PRINTED DOCUMENTS ARE UNCONTROLLED ...

  • Teledyne 265A - page 15

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 7 PRINTED DOCUMENTS ARE UNCONTROLLED 1.3 Initial Operation 1. Turn on the instrument power. The display should imm ediately light, showing the computer’s mem ory configuration, then the instrument type - M265A. A diagram of the software menu trees can be found in F ...

  • Teledyne 265A - page 16

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 8 PRINTED DOCUMENTS ARE UNCONTROLLED Step 2 - Calibrate the instrument: Zero/Span Calibration Procedure Step Number Action Comment 1. Input zero gas Allow zero gas to enter the sample port on the rear of the instrument. 2. Press CAL The M265A enters the calibrate mod ...

  • Teledyne 265A - page 17

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 9 PRINTED DOCUMENTS ARE UNCONTROLLED Step Number Action Comment 1. Scroll the TEST function menu until the O 3 SLOPE is displayed. The SLOPE value for O 3 should be 1.0 ± 0.3. If the SLOPE value is in the acceptable range the instrument will perform optimally. 2. Sc ...

  • Teledyne 265A - page 18

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 10 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK ...

  • Teledyne 265A - page 19

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 11 PRINTED DOCUMENTS ARE UNCONTROLLED 2 SPECIFICATIONS, WARRANTY 2.1 Specifications Table 2.1: M265A Specifications Ranges In 1 ppb increments from 100 ppb to 20,000 ppb Measurement Mode Single range or AutoRange Measurement Units ppb, µg/m 3 Noise at zero 0.3 ppb p ...

  • Teledyne 265A - page 20

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 12 PRINTED DOCUMENTS ARE UNCONTROLLED 2.2 Warranty ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) (02024D) (DCN 4473) Prior to shipme nt, Teledyne API equipm ent is thoroughly inspected and tested. Should equi pment fail ure occur, Teledyne API assures its custo ...

  • Teledyne 265A - page 21

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 13 PRINTED DOCUMENTS ARE UNCONTROLLED 2.3 Test and Calibration Values Table 2-2: Final Test and Calibration Values TEST Values Observed Value Units Nominal Range Reference Section RANGE ppb 10-20000 NOISE ppb 0.0 – 7.0 SAMP FLW cm³/min 500 ± 50 NO FLOW cm³/min 5 ...

  • Teledyne 265A - page 22

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 14 PRINTED DOCUMENTS ARE UNCONTROLLED Table 2-2: Final Test and Calibration Values (Continued) Parameter Observed Value Units Nominal Range Reference Section O 3 Span Conc ppb 0.5 - 5000 O 3 Slope 1.0 ± 0.3 O 3 Offset mV ± 25 Noise at Zero ppb 0.0 – 1.0 Noise At ...

  • Teledyne 265A - page 23

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 15 PRINTED DOCUMENTS ARE UNCONTROLLED 3 THE M265A OZONE ANALYZER 3.1 Principle of Operation The Teledyne API Model 265A analyzer is designed to measure the concentration of ozone using the chemiluminescence reaction below. The signal com es from the light em itted by ...

  • Teledyne 265A - page 24

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 16 PRINTED DOCUMENTS ARE UNCONTROLLED TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 16 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 3-1: Block Diagram ...

  • Teledyne 265A - page 25

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 17 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 3-2 : Pneumatic Diagram ...

  • Teledyne 265A - page 26

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 18 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK ...

  • Teledyne 265A - page 27

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 19 PRINTED DOCUMENTS ARE UNCONTROLLED 4 SOFTWARE FEATURES The M265A software has two major operating modes. The SAMPLE m ode is the normal mode when the instrument is taking data. The software m enu that covers the SAMPLE mode is shown in Figure 4-1 . Wh en the instr ...

  • Teledyne 265A - page 28

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 20 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 4-1: Sample Menu Tree ...

  • Teledyne 265A - page 29

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 Figure 4-2: Setup Menu Tree 21 PRINTED DOCUMENTS ARE UNCONTROLLED ...

  • Teledyne 265A - page 30

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 22 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK ...

  • Teledyne 265A - page 31

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 23 PRINTED DOCUMENTS ARE UNCONTROLLED 5 MAINTENANCE 5.1 Maintenance Schedule NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. Table 5-1: Preventative Maintenance Schedule Item Maintenance Interval Reference Sec ...

  • Teledyne 265A - page 32

    TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 24 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK ...

Manufacturer Teledyne Category Musical Toy Instrument

Documents that we receive from a manufacturer of a Teledyne 265A can be divided into several groups. They are, among others:
- Teledyne technical drawings
- 265A manuals
- Teledyne product data sheets
- information booklets
- or energy labels Teledyne 265A
All of them are important, but the most important information from the point of view of use of the device are in the user manual Teledyne 265A.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Teledyne 265A, service manual, brief instructions and user manuals Teledyne 265A. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Teledyne 265A.

Similar manuals

A complete manual for the device Teledyne 265A, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Teledyne 265A by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Teledyne 265A.

A complete Teledyne manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Teledyne 265A - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Teledyne 265A, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Teledyne 265A, that we can find in the current document
3. Tips how to use the basic functions of the device Teledyne 265A - which should help us in our first steps of using Teledyne 265A
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Teledyne 265A
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Teledyne 265A in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Teledyne 265A?

Use the form below

If you did not solve your problem by using a manual Teledyne 265A, ask a question using the form below. If a user had a similar problem with Teledyne 265A it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)