Manual Fluke 1523

36 pages 3.74 mb
Download

Go to site of 36

Summary
  • Fluke 1523 - page 1

    1523, 1524 Thermometer Readout User’ s Guide Rev . 891001_EN ...

  • Fluke 1523 - page 2

    Fluke Corporation, Har t Scientific Division 799 E. Utah V alley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • T elef ax: +1.801.763.1010 E-mail: suppor t@har tscientic.com www .har tscientific.com Specications subject to change without notice. • Cop yr ight © 2008 • Printed in USA ...

  • Fluke 1523 - page 3

    Limited W arranty & Limitation of Liability Each product from Fluke Corporation, Hart Scientic Division (“Hart”) is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year(s) for the Reference Thermometer . The warranty period begins on the date of the shipment. Parts, p ...

  • Fluke 1523 - page 4

    DA T A, WHETHER ARISING FROM BREACH OF W ARRANTY OR BASED ON CONTRACT , TOR T , RELIANCE OR ANY OTHER THEOR Y . Since some countries or states do not allow limitation of the term of an implied warranty , or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer . ...

  • Fluke 1523 - page 5

    i T able of Contents 1 Befor e Y ou Start ....................................................................... 1 1.1 Introduction ............................................................................................... 1 1.2 Standard Equipment ................................................................................. 1 1.3 Safety I ...

  • Fluke 1523 - page 6

    1 523, 1 524 Ther mometer R eadout ii F igures Figure 1 Clamp-On Ferrite ................................................................................. 6 Figure 2 Input/Output Connections - 1523 ........................................................ 9 Figure 3 Input/Output Connections - 1524 .................................................... ...

  • Fluke 1523 - page 7

    iii T ables T able 1 International Symbols ............................................................................ 4 T able 2 1523 Input/Output Connections ............................................................ 9 T able 3 1524 Input/Output Connections .......................................................... 10 T able 4 1523 Key Functio ...

  • Fluke 1523 - page 8

    ...

  • Fluke 1523 - page 9

    Bef ore Y ou Start Standar d Equipment 1 Bef or e Y ou Start 1 Introduction 1 .1 The Reference Thermometer Readouts (1523, 1524) are designed to be reliable, stable, temperature measur - ing instruments that can be used in the eld or laboratory . They offer accuracy , portability , and speed for near- ly every eld calibration application. The ...

  • Fluke 1523 - page 10

    1 523, 1 524 Ther mometer R eadout Safety Infor mation 2 Safety Inf ormation 1 .3 The Reference Thermometer is designed in accordance with EN 61010-1 {2nd Edition}, and CAN/CSA 22.2 No 61010.1-04. Use this instrument only as specied in this manual, otherwise the protection provided by the instrument may be impaired. A W arning identies condit ...

  • Fluke 1523 - page 11

    Bef ore Y ou Start Safety Infor mation 3 appears damaged or operates abnormally . Protection may be impaired. When in doubt, have the instrument serviced. Always use an isolated R TD or PR T (metal sheath isolated from lead wires).   Make sure the battery door is closed and latched before you operate the reference thermometer .   Do not ...

  • Fluke 1523 - page 12

    1 523, 1 524 Ther mometer R eadout Safety Infor mation 4 T able 1 International Symbols Symbol Description Symbol Descr iption AC (Alternating Current) PE Ground AC-DC Hot Surface (Burn Hazard) Battery Read the User’ s Guide (Important Information) Complies with European Union directives Off DC On Double Insulated Canadian Standards Association E ...

  • Fluke 1523 - page 13

    Bef ore Y ou Start Using Clamp-On F errit es 5 CE Comments 1 .4 EMC Directiv e 1 .4.1 Hart Scientic’ s equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMC Directive, 2004/108/EC ). The Declaration of Conformity for your instrument lists the specic stan- dards to which the unit was tested. The instrume ...

  • Fluke 1523 - page 14

    1 523, 1 524 Ther mometer R eadout Emissions T esting 6 Figure 1 clamp -on ferrite probe cable Clamp-On Ferrite Emissions T esting 1 .6 The instrument fullls the limit requirements for Class B. Low V oltage Dir ective (Saf ety) 1 .7 In order to comply with the European Low V oltage Directive (2006/95/EC), Fluke equipment has been de- signed to m ...

  • Fluke 1523 - page 15

    Bef ore Y ou Start Authorized Ser vice Center s 7 1 .8 A uthorized Ser vice Cent ers Please contact one of the following authorized Service Center to coordinate service on your Fluke product: Fluke Corporation Hart Scientic Division Phone: +1.801.763.1600 Fluke Nederland B.V . Phone: +31-402-675300 Fluke Int'l Corporation - CHINA Phone: +86 ...

  • Fluke 1523 - page 16

    ...

  • Fluke 1523 - page 17

    Quick Start Setup 9 Quic k Star t 2 Set up 2.1 SAVE NEXT HOME ENTER RESET TREND RECALL HOLD STATS ° C ° F SETUP mV RS 232 1 2 V DC 30 V MAX 1 5 2 3 CALIBRATION THERMOMETER READOUT Figure 2 Input/Output Connections - 1523 1523 Input/Output Connections T able 2 No. Name Description 1 Serial Serial interface connector 2 Connector , T1 Sensor Connect ...

  • Fluke 1523 - page 18

    1 523, 1 524 Ther mometer R eadout Setup 10 SAVE NEXT HOME LOG RESET ENTER RECALL HOLD STATS ° C ° F SETUP TREND mV RS 232 1 2 V DC 30 V MAX T 1 T 2 1 5 2 4 CALIBRATION THERMOMETER READOUT Figure 3 Input/Output Connections - 1524 1524 Input/Output Connections T able 3 No Name Description 1 Serial Serial interface connector 2 Connector , T1 Sensor ...

  • Fluke 1523 - page 19

    Quick Start Setup 11 SAVE NEXT HOME LOG RESET ENTER RECALL HOLD STATS ° C ° F SETUP TREND mV RS 232 1 2 V DC 30 V MAX T 1 T 2 1 5 2 4 CALIBRATION THERMOMETER READOUT 1 10 11 4 3 2 5 6 7 8 9 12 Figure 4 Keys ...

  • Fluke 1523 - page 20

    1 523, 1 524 Ther mometer R eadout Setup 12 1523 Key Functions T able 4 No Key Description 1 Power on or off 2 Y ellow Second or Special Function Key 3 T urns the backlight on or off 4 STATS 1st Press: MAX, 2nd Press: MIn, 3rd Press: A VE, 4th Press: STD DEV 5 ° C ° F Units, °C/°F 6 HOLD 1st press - Holds value on screen "-- HOLD --" ...

  • Fluke 1523 - page 21

    Quick Start Setup 13 No Key Description 11 RECALL 1st Press - Enters Recall menu, 2nd Press - Exits Recall Menu 12 NEXT Moves down to next option on screen. 13 + STATS “RESET” - Resets Stats Data 14 + ° C ° F “ Ω mV” - T oggles from °C to Ω or Ω to °C (PRT , thermistor), °C to mV or mV to °C (TC) 15 + HOLD “TREND” - Starts G ...

  • Fluke 1523 - page 22

    1 523, 1 524 Ther mometer R eadout Setup 14 Figure 5 1523 Menu ...

  • Fluke 1523 - page 23

    Quick Start Setup 15 Figure 6 1523 Menu (cont) ...

  • Fluke 1523 - page 24

    1 523, 1 524 Ther mometer R eadout Setup 16 Figure 7 1523 Menu (cont) ...

  • Fluke 1523 - page 25

    Quick Start Setup 17 1524 Key Functions T able 5 No Key Description 1 Power on or off 2 Y ellow Second or Special Function Key 3 T urns the backlight on or off 4 STATS 1st Press: Max, 2nd Press: Min, 3rd Press: Ave, 4th Press: STD DEV 5 ° C ° F Units, °C/°F 6 HOLD 1st press - Holds value on screen "-- HOLD --" across bottom of screen. ...

  • Fluke 1523 - page 26

    1 523, 1 524 Ther mometer R eadout Setup 18 No Key Description 10 ENTER Selects highlighted selection, Saves a new selection. 11 RECALL 1st press - Enters Recall Menu, 2nd press - Exits Recall Menu 12 NEXT Moves down to next option on screen. 13 + STATS “RESET” - Resets Stats Data 14 + ° C ° F “ Ω mV” - T oggles from °C to Ω or Ω t ...

  • Fluke 1523 - page 27

    Quick Start Setup 19 Figure 8 1524 Menu ...

  • Fluke 1523 - page 28

    1 523, 1 524 Ther mometer R eadout Setup 20 Figure 9 1524 Menu (cont) ...

  • Fluke 1523 - page 29

    Quick Start Setup 21 Figure 10 1524 Menu (cont) ...

  • Fluke 1523 - page 30

    1 523, 1 524 Ther mometer R eadout Setup 22 Figure 11 1524 Menu (cont) ...

  • Fluke 1523 - page 31

    Quick Start Setup 23 Figure 12 1524 Menu (cont) ...

  • Fluke 1523 - page 32

    1 523, 1 524 Ther mometer R eadout Specications 24 Figure 13 1524 Menu (cont) Specications 2.2 Specications are based on a one year calibration cycle and apply from 13 °C to 33 °C unless stated otherwise. All specications assume a ve minute warm up period. General Specifications T able 6 Operating T emperatur e † –10 °C to 60 ...

  • Fluke 1523 - page 33

    Quick Start Specications 25 P ow er requir ements 3 AA alkaline batteries 12 V dc universal power supply Size 96 x 200 x 47 mm (3.75 x 7.9 x 1.86 inches) W eight 0.65 kg (1.4 lb) Safety EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 † Environmental conditions for all specifications: 13 °C to 33 °C Millivolt Measurement T able 7 Range Resolut ...

  • Fluke 1523 - page 34

    1 523, 1 524 Ther mometer R eadout Specications 26 Ohms Measurement, Thermistor T able 9 Ohms Range Accur acy ± Ω, 4 W ire 200 Ω to 50 k Ω ± (0.01 % + 0.5 Ω ) 50 k Ω to 500 k Ω ± (0.03 %) T emperature Coefficient ( –10 °C to 13 °C , +33 °C to 60 °C): 0.002 %/°C + 0.1 Ω (0 Ω to 50 k Ω ) 0.06 %/°C + 0.1 Ω (50 k Ω ...

  • Fluke 1523 - page 35

    Quick Start Specications 27 T ype Range Measure Accur acies (ITS-90) L –200 °C to 0 °C 0 °C to 900 °C 0.36 °C 0.23 °C M –20 °C to 0 °C 0 °C to 400 °C 400 °C to 1400 °C 0.26 °C 0.25 °C 0.22 °C N –200 °C to 0 °C 0 °C to 1300 °C 0.72 °C 0.28 °C R –20 °C to 0 °C 0 °C to 500 °C 500 °C to 1750 °C 1.09 °C 0.97 °C 0. ...

  • Fluke 1523 - page 36

    1 523, 1 524 Ther mometer R eadout Specications 28 T emperature, RTD Ranges, T able 11 and Accuracies (ITS-90) Accur acy ± ° C 4 W ire P robe ± 0.011 at –100 °C ± 0.015 at 0 °C ± 0.019 at 100 °C ± 0.023 at 200 °C ± 0.031 at 400 °C ± 0.039 at 600 °C Resolution: 0.001°C (0.001°F) T emperature, Thermistor T able 12 Accur acy ± ° ...

Manufacturer Fluke Category Thermometer

Documents that we receive from a manufacturer of a Fluke 1523 can be divided into several groups. They are, among others:
- Fluke technical drawings
- 1523 manuals
- Fluke product data sheets
- information booklets
- or energy labels Fluke 1523
All of them are important, but the most important information from the point of view of use of the device are in the user manual Fluke 1523.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Fluke 1523, service manual, brief instructions and user manuals Fluke 1523. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Fluke 1523.

Similar manuals

A complete manual for the device Fluke 1523, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Fluke 1523 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Fluke 1523.

A complete Fluke manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Fluke 1523 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Fluke 1523, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Fluke 1523, that we can find in the current document
3. Tips how to use the basic functions of the device Fluke 1523 - which should help us in our first steps of using Fluke 1523
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Fluke 1523
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Fluke 1523 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Fluke 1523?

Use the form below

If you did not solve your problem by using a manual Fluke 1523, ask a question using the form below. If a user had a similar problem with Fluke 1523 it is likely that he will want to share the way to solve it.