  • Teledyne 460L - page 1

    INSTRUCTION MANUAL MODEL 460L Nema Ozone Monitor 9480 CARROLL PARK DRI VE SAN DIEGO, CA 92121-52 01 USA Toll-free Phone: 800-324-5 190 Phone: 858-657-9 800 Fax: 858-657-9 816 Email: api-sales@teledyne.c om Website: http://www.teledyne-api.com/ 05012 REV. D DCN 5164 © 2008 T-API 22 January 2009 PRINT DATE: 19 February 2009 ...

  • Teledyne 460L - page 2

    TELEDYNE INSTRUMENTS HIGHLIGHTS Model 460L – NEMA Ozone Monitor Instruction Manual H-ii 05012 Rev D 22 January 2009 PRINTED DOCUMENTS ARE UNCONTROLLED DCN 5164 THIS PAGE IS INTENT IONALLY LEFT BLANK ...

  • Teledyne 460L - page 3

    TELEDYNE INSTRUMENTS 460L Instruction Manual Safety Messages 05228 Rev B i DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED SAFETY MESSAGES Your safety and the safety of o thers is very im portant. We have provided many important sa fety messages in this manual. Please read these messages carefully. A safety message alerts you to potenti al hazard s th ...

  • Teledyne 460L - page 4

    TELEDYNE INSTRUMENTS Safety Messages 460L Instruction Manual ii 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 THIS PAGE IS INTENTIONALLY LEFT BLANK ...

  • Teledyne 460L - page 5

    TELEDYNE INSTRUMENTS 460L Instruction Manual Table of Contents 05228 Rev B iii DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED TABLE OF CONTENTS 1. INTROD UCTION ................................................................................................................ ........... 1 1.1. Preface ................................................... ...

  • Teledyne 460L - page 6

    TELEDYNE INSTRUMENTS Table of Contents 460L Instruction Manual iv 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 9.1.1. Filter Repl acement P rocedure............................................................................................ 45 9.1.2. Mounting the Partic ulate Filter Extern ally .......................................... ...

  • Teledyne 460L - page 7

    TELEDYNE INSTRUMENTS 460L Instruction Manual Table of Contents 05228 Rev B v DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED List of Figures Figure 3-1 M460L Front Co ver Lay out ............................................................................. 7 Figure 3-2 460L Layout ....................................................................... ...

  • Teledyne 460L - page 8

    TELEDYNE INSTRUMENTS Table of Contents 460L Instruction Manual vi 05228 R ev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 LIST OF APPENDICES Appendix A – Spare Parts List Appendix B – List of Schematics ...

  • Teledyne 460L - page 9

    TELEDYNE INSTRUMENTS 460L Instruction Manual Introduction 05228 Rev B 1 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 1. INTRODUCTION 1.1. Preface Teledyne Instruments is pleased that you have purchased the Mo del 460L NEMA Ozone Monitor. Included is a full one-year warran ty (see Section 2.2) and we at Teledyne I nstruments will be pleased to provi ...

  • Teledyne 460L - page 10

    TELEDYNE INSTRUMENTS Introduction 460L Instruction Manual 2 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 1.3. Using This Manual This manual has the followin g data structures: 1.0 Table of Contents: Outlines the conte nts of the manual i n the order the information is presented. This is a good overview of the topics covered in the manua ...

  • Teledyne 460L - page 11

    TELEDYNE INSTRUMENTS 460L Instruction Manual Introduction 05228 Rev B 3 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED interact with each ot her. A close reading of this section is inv aluable for underst anding the monitor’s operation. 11.0 Troublesh ooting Section: This section includes poin ters an d instructions for diagnosing problems with the ...

  • Teledyne 460L - page 12

    TELEDYNE INSTRUMENTS Introduction 460L Instruction Manual 4 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 USER NOTES: ...

  • Teledyne 460L - page 13

    TELEDYNE INSTRUMENTS 460L In structi on Manua l Spec ification s and Wa rranty 05228 Rev B 5 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 2. SPECIFICATIONS AND WARRANTY 2.1. Specifications Table 2-1 460L Specifications Measurement Princi ple UV Absorption (Beer Lambert Law) Ranges 1ppm to 500 ppm: User selectabl e Measurement Units ppm; ppb Accurac ...

  • Teledyne 460L - page 14

    TELEDYNE INSTRUMENTS Specifications an d Warranty 460L Instruction Manual 6 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 2.2. Warranty WARRANTY POLICY (020 24D) Prior to shipment, T-API equipment is thorou ghly inspected and tested. Should equipment failure occur, T-API assures its cus tomers that prom pt service and support will be ava ...

  • Teledyne 460L - page 15

    TELEDYNE INSTRUMENTS 460L Instruction Manual Getting Started 05228 Rev B 7 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 3. GETTING STARTED 3.1. Unpacking 1. Inspect the received packages for external sh ip ping damage. If damaged, please advise the shipper first, then Tele dyne Instruments. 2. Loosen the 2 setscrews located in the top and bottom le ...

  • Teledyne 460L - page 16

    TELEDYNE INSTRUMENTS Getting Started 460L Instruction Manual 8 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 CAUTION Never disconnect electronic cir cuit boards, wiring harnesses or electronic subassemblies while the unit is under power. 4. Check the connectors of the various internal wiring harnesses an d pneumatic hoses to make sure th ...

  • Teledyne 460L - page 17

    TELEDYNE INSTRUMENTS 460L Instruction Manual Getting Started 05228 Rev B 9 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED A bsorption Tube Reference Scrubber UV Lamp Power Supply UV Lamp Power Connect or UV Lamp Ozone Outlet Manifold Ozone Intlet Manifol d UV Detector Measure / Reference Valve Sensor Module PCA UV Lamp Housing Figure 3-3 46 0L Sensor ...

  • Teledyne 460L - page 18

    TELEDYNE INSTRUMENTS Getting Started 460L Instruction Manual 10 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 Table 3-1 460L Ventilation Cle arances AREA MINIMUM REQUIRED CLEARANCE Sides of the monitor 1 inch Above and below the monitor. 1 inch Figure 3-4 460L Mounting Hole Locations and Dimen sions 3.3. Pneumatic Connections 1. Connect ...

  • Teledyne 460L - page 19

    TELEDYNE INSTRUMENTS 460L Instruction Manual Getting Started 05228 Rev B 11 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED NOTE The ozone delivery pr essure should be at ambient pressure +/- 5 PSIG. All tubing used shou ld be made of ozone r esistant material such as Stainl ess Steel, PTFE (Teflon  ) or FEP. API can supply appropriate tubing for c ...

  • Teledyne 460L - page 20

    TELEDYNE INSTRUMENTS Getting Started 460L Instruction Manual 12 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 3.4. Electrical Connections NOTE It is recommended that if multi- strand wires ar e used for the following electrical connections. To ensure a reli able connection the wires should be:  “Tinned” with solder or;  Termina ...

  • Teledyne 460L - page 21

    TELEDYNE INSTRUMENTS 460L Instruction Manual Getting Started 05228 Rev B 13 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 3.5. Signal I/O Connections All digital and analog signal I/ O connections are made vi a a 16-pin connector on the main board (See Figure 3-6 for the locati on of the connecto r.) This connector can be unplugged from the head er ...

  • Teledyne 460L - page 22

    TELEDYNE INSTRUMENTS Getting Started 460L Instruction Manual 14 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 3.5.2. Digital Status Outputs The 460L has six assigned digital status outputs for indicating error and operational status conditions of the monitor as well as the statu s of its O 3 concentration alarms (see Table 3-2). These ou ...

  • Teledyne 460L - page 23

    TELEDYNE INSTRUMENTS 460L Instruction Manual Getting Started 05228 Rev B 15 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 3.5.3. Control Inputs Two digital control inputs are also available for use on the 460L. The control inputs are used for remote control of the 460L by a de vice such as a PL C. They are label ed ZE RO INPUT and AUX INPUT (see Fig ...

  • Teledyne 460L - page 24

    TELEDYNE INSTRUMENTS Getting Started 460L Instruction Manual 16 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 Below each relay is a 3- pin connector that al lows the relay to be connec ted for either norma lly open or norma lly close d opera tion. Table 3-4 describes how to connect the alarm relays. NOTE The relay contacts are rated t o ...

  • Teledyne 460L - page 25

    TELEDYNE INSTRUMENTS 460L Instruction Manual Getting Started 05228 Rev B 17 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 1 Located just below each rela y connector (see Figure 3-10) 2 N.O. = Normally Open operat ion. N.C. = Norma lly Closed opera tion. 3.6. Initial Startup Perform the fo llowing steps when initia lly starting up the 460L or when br ...

  • Teledyne 460L - page 26

    TELEDYNE INSTRUMENTS Getting Started 460L Instruction Manual 18 05228 Rev B PRINTED DOCUMENTS ARE UN CONT ROLL ED DCN 5164 While the standard factory configuration is for RS- 232, the monitor’s serial port can be configured for either RS-232 or RS-485 (see Section 3.7.1 for the procedure).  Use RS-232 for direct connecti on to a nearby (no mor ...

  • Teledyne 460L - page 27

    TELEDYNE INSTRUMENTS 460L Instruction Manual Getting Started 05228 Rev B 19 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 7. Connect the appropriate type of ca ble to the 16 pin Signal I/O connector inside the analyzer. 8. A cable may be included with you i nstrument. If not, Figure 3-12 below illustrates how to construct wiring for both RS -232 and ...

  • Teledyne 460L - page 28

    TELEDYNE INSTRUMENTS Getting Started 460L Instruction Manual 20 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164  The Serial Port of the devi ce being used to communi cate with the 460L sho uld be configured as follow s: Table 3-6 Serial Port Configuration PARAMETER VALUE Baud Rate 9600 bps Data Bits 8 Stop Bits 1 Parity None 3. Find the ...

  • Teledyne 460L - page 29

    TELEDYNE INSTRUMENTS 460L Instruction Manual Fr equently Asked Questions 05228 Rev B 21 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 4. FREQUENTLY ASKED QUESTIONS Q: What do I do if I smell Ozone and suspect a leak? A: Ensure that all the fi ttings are tight. If the fittings are ti ght and ozone is sti ll detected send the monitor back for re pair ...

  • Teledyne 460L - page 30

    TELEDYNE INSTRUMENTS Frequently Asked Questions 460L Instruction Manual 22 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 USER NOTES: ...

  • Teledyne 460L - page 31

    TELEDYNE INSTRUMENTS 460L Instruction Manual Optio nal Hardware and Soft ware 05228 Rev B 23 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 5. OPTIONAL HARDWARE AND SOFTWARE This section includes descriptions of the hardwa re and software options available for the 460L monitor. For assistance with ordering these op tions please contact th e sales dep ...

  • Teledyne 460L - page 32

    TELEDYNE INSTRUMENTS Optional Hardware and Soft wa re 460L Instruction Manual 24 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 USER NOTES: ...

  • Teledyne 460L - page 33

    TELEDYNE INSTRUMENTS 460L Instruction Manual Operating Instructio ns 05228 Rev B 25 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 6. OPERATING INSTRUCTIONS The 460L has been designed for simple and tr ouble-free operation. The sections below detail the operational features of the 460L. 6.1. Front Panel Display 6.1.1. O 3 readout The current ozon e c ...

  • Teledyne 460L - page 34

    TELEDYNE INSTRUMENTS Operating Instructio ns 460L Instruction Manual 26 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 6.2. Concentration Alarms There are two O 3 concentration alarms, HI and HI-HI . Both alarms are triggered when the measured O 3 concentration equals or rises abov e the set limit for t hat alarm. The set limit f or HI al ...

  • Teledyne 460L - page 35

    TELEDYNE INSTRUMENTS 460L Instruction Manual Serial Communicatio ns 05228 Rev B 27 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 7. SERIAL COMMUNICATIONS The 460L comes equipped with a powe rful digital Seria l Communications Port th at can be used for Data Acquisiti on and for changing t he monitor’s co nfigurat ion. This port can be configured f ...

  • Teledyne 460L - page 36

    TELEDYNE INSTRUMENTS Serial Communicatio ns 460L Instruction Manual 28 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 7.2. Serial Port Command Su mmary Table 7-1 below lists the commands availa ble and a sum mary of their function. Table 7-1 Serial Port Comm and Summ ary COMMAND DESCRIPTION ALMACK Acknowledges and clears any active concen ...

  • Teledyne 460L - page 37

    TELEDYNE INSTRUMENTS 460L Instruction Manual Serial Communicatio ns 05228 Rev B 29 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 7.3. Serial Port Command Reference 7.3.1. ALMACK SYNTAX <address>ALMACK<CR> DESCRIPTION Acknowledges any active con centration alarm an d clears them if possible For more inf ormation see Section 6.2. DATA PARA ...

  • Teledyne 460L - page 38

    TELEDYNE INSTRUMENTS Serial Communicatio ns 460L Instruction Manual 30 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 7.3.2. ALMSTAT SYNTAX <address>ALSTAT<CR> DESCRIPTION Returns the current status o f both of the concentratio n alarms. 0= Inactive 1= Active DATA PARAMETERS SENT None RESPONSE <address>:<HI alarm stat ...

  • Teledyne 460L - page 39

    TELEDYNE INSTRUMENTS 460L Instruction Manual Serial Communicatio ns 05228 Rev B 31 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 7.3.3. CZERO SYNTAX <address>CZERO<CR> DESCRIPTION Performs a zero calibration using gas sourced from the ozone gas inlet. DATA PARAMETERS SENT None RESPONSE <address>:<success_fl ag>#<chec ksum& ...

  • Teledyne 460L - page 40

    TELEDYNE INSTRUMENTS Serial Communicatio ns 460L Instruction Manual 32 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 7.3.4. DACSTEP SYNTAX <address>DACSTEP< CR> DESCRIPTION Puts the monitor i nto Analog Output setup mode. The Analog Output steps from zero to full-scale in 25% increments , pausing for 10 secon ds at each level ...

  • Teledyne 460L - page 41

    TELEDYNE INSTRUMENTS 460L Instruction Manual Serial Communicatio ns 05228 Rev B 33 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 7.3.5. O 3 SYNTAX <address>O3<CR> DESCRIPTION Returns the current ozone conce ntration measured by t he monitor. DATA PARAMETERS SENT None RESPONSE <address>:<o3_conc>#<checksu m><CR> EX ...

  • Teledyne 460L - page 42

    TELEDYNE INSTRUMENTS Serial Communicatio ns 460L Instruction Manual 34 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 7.3.6. SETADDR SYNTAX <address>SETADDR:<new_address><CR > DESCRIPTION Changes the communications a ddress to a ne w value. DATA PARAMETERS SENT new_address is the new address for the mo nitor; Range for n ...

  • Teledyne 460L - page 43

    TELEDYNE INSTRUMENTS 460L Instruction Manual Serial Communicatio ns 05228 Rev B 35 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 7.3.7. TDUMP SYNTAX <address>TDUMP<CR> DESCRIPTION This command is best used if th e response is in tended as input for a database program or data acquisition system It returns a string made up of the current v ...

  • Teledyne 460L - page 44

    TELEDYNE INSTRUMENTS Serial Communicatio ns 460L Instruction Manual 36 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 7.3.8. TLIST SYNTAX <address>TLIST<CR> DESCRIPTION Returns a formatted, easy to read list of the p a rameters by name with the current values fo r each. DATA PARAMETERS SENT None RESPONSE Test Parameter List (S ...

  • Teledyne 460L - page 45

    TELEDYNE INSTRUMENTS 460L Instruction Manual Serial Communicatio ns 05228 Rev B 37 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 7.3.9. VGET SYNTAX <address>VGET:<var_index><CR> DESCRIPTION Returns value of an internal conf iguration variable (VAR. ) DATA PARAMETERS SENT var_index is index number for internal VAR as follows: Table ...

  • Teledyne 460L - page 46

    TELEDYNE INSTRUMENTS Serial Communicatio ns 460L Instruction Manual 38 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 7.3.10. VLIST SYNTAX <address>VLIST<CR> DESCRIPTION Returns a formatted, easy to read list of an in ternal configuration vari ables (VAR’s) and their current values. DATA PARAMETERS SENT None RESPONSE VAR Lis ...

  • Teledyne 460L - page 47

    TELEDYNE INSTRUMENTS 460L Instruction Manual Serial Communicatio ns 05228 Rev B 39 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 7.3.11. VSET SYNTAX <address>VSET:<var_index>, <new_value><CR> DESCRIPTION Sets value of an internal con figuration variabl e (VAR.) DATA PARAMETERS SENT var_index index number for internal VAR (See ...

  • Teledyne 460L - page 48

    TELEDYNE INSTRUMENTS Serial Communicatio ns 460L Instruction Manual 40 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 EXAMPLE2 Command: 1VSET:6,1<CR> - Set CONC_UNITS VAR to ppm Response: 1:OK#261<CR> - VSET Successful 1:FAIL#391< CR> - VSET Failed NOTES ANALOG_RANGE: Setting is in wh atever units selected by the CONC_UN ...

  • Teledyne 460L - page 49

    TELEDYNE INSTRUMENTS 460L Instruction Manual Calibration 05228 Rev B 41 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 8. CALIBRATION 8.1. Zero Point calibration Variations in ambient condi tions, most notably changes in ambient humi dity, and minor changes in the performance of certain elec tronic components over time may cause slight offsets in the ...

  • Teledyne 460L - page 50

    TELEDYNE INSTRUMENTS Calibration 460L Instruction Manual 42 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 Where  <address> is the network address of the M460M being ca librated.  929 is a special password allowing you to calibrate the analyzer. 5. List the current values of the M460 M’s VARS. Type: <address>VLIST<C ...

  • Teledyne 460L - page 51

    TELEDYNE INSTRUMENTS 460L Instruction Manual Calibration 05228 Rev B 43 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 8. Enter the New Slope i nto the M460M’s memory enter i t in. Type: <address>VSET:<VAR Number>, <New Slope Value><CR> EXAMPLE 3VSET:16,1.048<CR> Where:  The address of the M460M is 3  The O3 SLOP ...

  • Teledyne 460L - page 52

    TELEDYNE INSTRUMENTS Calibration 460L Instruction Manual 44 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 5. The monitor will au tomatically generate t he nominal signal lev el by producing a 25% increment outpu t level on the current loop. EXAMPLE OUTPUT LEVEL NOMINAL SIGNAL LEVEL 0% 4mA 25% 8mA 50% 12mA 75% 16mA 100% 20mA 6. The displa ...

  • Teledyne 460L - page 53

    TELEDYNE INSTRUMENTS 460L Instruction Manual Maintenance 05228 Rev B 45 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 9. MAINTENANCE: 9.1. Replacing the Particulate Filter Element The 460L is equipped with a particul ate filter on the ozone inlet. Only Teledyne Instruments filter elements should be used. When th e monitor is first installed, the sam ...

  • Teledyne 460L - page 54

    TELEDYNE INSTRUMENTS Maintenance 460L Instruction Manual 46 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 9.1.2. Mounting the Particulate Filter Externally In applications where the 460L is sampling ambient air at its own lo cation and no inlet gas lin e is being used, the particulate filter can be mounted on the o utside of the instru m ...

  • Teledyne 460L - page 55

    TELEDYNE INSTRUMENTS 460L Instruction Manual Maintenance 05228 Rev B 47 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 9.2. Maintaining the Optional H 2 O Coalescing Filter 9.2.1. Draining the Optional Coalescing Filter The coalescing filter component of the optional sample conditioni ng system may accumulate with water (see Figure 9-3 for filter loc ...

  • Teledyne 460L - page 56

    TELEDYNE INSTRUMENTS Maintenance 460L Instruction Manual 48 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 9.2.2. Replacing the Coalescing Membrane The coalescing filter cont ains a replaceable Teflon ® coalescing membrane. Under normal operating condition s, this membrane shou ld last for a long time, how ever a significant reduction in ...

  • Teledyne 460L - page 57

    TELEDYNE INSTRUMENTS 460L Instruction Manual Maintenance 05228 Rev B 49 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 9.3. UV Lamp Replacement 1. Disconnect power from the monitor and open cover. 2. Loosen, but do not remov e, the two la mp set-screws (see Figure 9-5 below). 3. Unplug lamp from the UV lamp power con nector on Sensor Module P CA. 4. ...

  • Teledyne 460L - page 58

    TELEDYNE INSTRUMENTS Maintenance 460L Instruction Manual 50 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 9.4. Cleaning Exterior Surfaces of the 460L If necessary, the exterior surfaces of the 460L can be cleaned with a damp cloth. Do not attempt to clean any of the other surfaces of the monitor. Do not submerge any part of the monitor i ...

  • Teledyne 460L - page 59

    TELEDYNE INSTRUMENTS 460L Instruction Manual Theory of Operation 05228 Rev B 51 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 10. THEORY OF OPERATION 10.1. Basic O 3 Measurement Principle 10.1.1. (Beer’s Law) The detection of ozone molecules in a gas is base d on the principle that ozone is a very strong absorber of UV light wit h a wavelength of ...

  • Teledyne 460L - page 60

    TELEDYNE INSTRUMENTS Theory of Operation 460L Instruction Manual 52 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 To solve this equation for C , the concentration of the absorbing gas (in this case O 3 ), the application of a little algebra is requ ired to rearrange the equation as follow s:            ? ...

  • Teledyne 460L - page 61

    TELEDYNE INSTRUMENTS 460L Instruction Manual Theory of Operation 05228 Rev B 53 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED beam is only required to make pas s through th e Absorption Tube. Therefore no complex mirror system is needed to lengthen the effective path b y bouncing the beam back a nd forth. Finally, the UV then passes through a si mil ...

  • Teledyne 460L - page 62

    TELEDYNE INSTRUMENTS Theory of Operation 460L Instruction Manual 54 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 The Measurement / Reference Cycle consists o f: Table 10-1 Measurem ent / Reference Cycle CYCLE STATUS DURATION ACTIVITY VALVE STATE (FIG. 10-3) - - Measure/Reference Valv e Opens to the Me asure Path. 0.50 sec. Wait Period. ...

  • Teledyne 460L - page 63

    TELEDYNE INSTRUMENTS 460L Instruction Manual Theory of Operation 05228 Rev B 55 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 10.2. Pneumatic Theory of Operation 10.2.1. Basic Pneumatic Flow And Flow Control O 3 Measurem ent C ell Particulate Filter Flow Meter MEASURE REFERENCE VALVE O 3 IN Exhaust Out Monitor Enclos ure PUMP Flow Meter Valve 1 3 2 ...

  • Teledyne 460L - page 64

    TELEDYNE INSTRUMENTS Theory of Operation 460L Instruction Manual 56 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 10.2.3. Optional Sample Conditioning The source air measured by the 46 0L needs to be as dry as possible. Significant amounts of liquid or vaporous water present in the s ource gas ca n fo ul the measurement cell optics. Also ...

  • Teledyne 460L - page 65

    TELEDYNE INSTRUMENTS 460L Instruction Manual Theory of Operation 05228 Rev B 57 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED To provide a dry purge gas for the outer side of the Nafion ® tube, the 460L returns the dried O 3 free air from the measurement cell and O 3 destruct to the outer tube (see Figure 10-4). When the monitor is first started, t ...

  • Teledyne 460L - page 66

    TELEDYNE INSTRUMENTS Theory of Operation 460L Instruction Manual 58 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 10.3. Electronic Theory of Operation Sensor Module Main Board Status Output Dri ver s Sensor OK Rel ay HI Al ar m Rel ay HI -HI Al arm Rel ay Analog and Digit al I /O Connector Status Output Ser ial I /O Analog Outputs Contro ...

  • Teledyne 460L - page 67

    TELEDYNE INSTRUMENTS 460L Instruction Manual Theory of Operation 05228 Rev B 59 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 10.3.2. O 3 Sensor Module The heart of the 460 Monitor is the O 3 sensor module. This electromechanical assembly locat ed at the left hand side of the enclosure includes a ll of the pneumatic, mechanical and electronic compon ...

  • Teledyne 460L - page 68

    TELEDYNE INSTRUMENTS Theory of Operation 460L Instruction Manual 60 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164  Gas Pressure Sensor: This absolute pressure sensor m easures the gas pressure i n the measurement cell upstream of the Pump. The sample pres sure is used by the CPU to calculate O 3 Concentration (see Form ula 10-3 in Sec ...

  • Teledyne 460L - page 69

    TELEDYNE INSTRUMENTS 460L Instruction Manual Theory of Operation 05228 Rev B 61 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 10.3.3. CPU Board In addition to being responsible for all I/O f unctions and managing the state of the monitors various status indicators and al arms (see below), the 460L’s CPU also calculates the monitor’s offset durin ...

  • Teledyne 460L - page 70

    TELEDYNE INSTRUMENTS Theory of Operation 460L Instruction Manual 62 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 and available at the STATUS COM pin of J2 on the main board (see Figure 3-7). This pin is normally connected to the input ground of the external device. NOTE Most PLC’s have internal provisions for limiting the current t ha ...

  • Teledyne 460L - page 71

    TELEDYNE INSTRUMENTS 460L Instruction Manual Theory of Operation 05228 Rev B 63 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 10.3.5. Power Distribution The 460L operates on 90 to 260 V AC power at eith er 50Hz or 60Hz. As illustrated in Figure 10-6 below, power enters the monitor via a standard 3-conductor power co rd through a hole provided in the ...

  • Teledyne 460L - page 72

    TELEDYNE INSTRUMENTS Theory of Operation 460L Instruction Manual 64 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164  UV LA MP PO WER SUPPL Y: Using +1 2 VDC supplied by the main pow er supply, this assembly generates the 30 kHz AC voltage for the mo nitor’s mercury vapor UV lamp. The output of this power supply is va riable. At startu ...

  • Teledyne 460L - page 73

    TELEDYNE INSTRUMENTS 460L Instruction Manual Troubleshooting 05228 Rev B 65 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 11. TROUBLESHOOTING This chapter gives guidelines for diagnosing sy stem and sensor malfunction s using the five digital Status Outputs provided by the 46 0L. All troubl eshooting should be done after the 460L ha s been turned on ...

  • Teledyne 460L - page 74

    TELEDYNE INSTRUMENTS Troubleshooting 460L Instruction Manual 66 05228 Rev B PRINTED DOCUMENTS ARE UN CONT ROLL ED DCN 5164  The O3 sensor is not responding to CPU reques ts for data. This could represent a variety of electronic problems with the sensor modu l e or internal communication between the CPU and the sensor module. 11.2.2. Invalid Read ...

  • Teledyne 460L - page 75

    TELEDYNE INSTRUMENTS 460L Instruction Manual Troubleshooting 05228 Rev B 67 DCN 5164 PRINTED DOCUMENTS ARE UN CONT ROLL ED Table 11-2 Stat us LED / Output Trouble s hooting Truth Table SENSOR OK INVALID READING LAMP LOW ACTIONS 1 0 0 Normal operation, no action required. 0 0 0  The Lamp intensity has drifted and is too high for proper operation. ...

  • Teledyne 460L - page 76

    TELEDYNE INSTRUMENTS Troubleshooting 460L Instruction Manual 68 05228 Rev B PRINTED DOCUMENTS ARE UN CONT ROLL ED DCN 5164 11.4. Technical Assistance If this addendum and its tr ouble-shooting / repair sections do not solve your problems, techn ical assistance may be obtained from: Teledyne Instrument s Advanced Pollution Instrumentation Di vision ...

  • Teledyne 460L - page 77

    TELEDYNE INSTRUMENTS 460L Instruction Manual A Primer on Electro-Static Disc harge 05228 Rev B 69 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 12. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne Instruments co nsiders th e prevention of damage caused b y the discharge of static electricity to be extremely important part of maki ng sure that your an a ...

  • Teledyne 460L - page 78

    TELEDYNE INSTRUMENTS A Primer on Electro-Static Dischar ge 460L Instruction Manual 70 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 Table 12-1 Static Generation Voltages for Typic al Activities MEANS OF GENERATION 65-90% RH 10-25% RH Walking across nylon carpet 1,500V 35,000V Walking across vinyl til e 250V 12,000V Worker at bench 100V 6 ...

  • Teledyne 460L - page 79

    TELEDYNE INSTRUMENTS 460L Instruction Manual A Primer on Electro-Static Disc harge 05228 Rev B 71 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED  Any time a charged surface (including the human body) discharges to a device. Even simple contact of a fing er to the leads of a sensitive device or assembly can allow enough discharge to cause damage. A ...

  • Teledyne 460L - page 80

    TELEDYNE INSTRUMENTS A Primer on Electro-Static Dischar ge 460L Instruction Manual 72 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 other things, like you a nd your clothing, from discharging through the instrumen t and damaging it. 12.4. Basic Principles of Static Control It is impossible to stop the creati on of instantaneous stat ic e ...

  • Teledyne 460L - page 81

    TELEDYNE INSTRUMENTS 460L Instruction Manual A Primer on Electro-Static Disc harge 05228 Rev B 73 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED While this layer may equalizes any charges th at occur across the whole bag, it does not prevent the build-up of static charges. If la ying on a conductive, gr ounded surface, these bags will allow charges t ...

  • Teledyne 460L - page 82

    TELEDYNE INSTRUMENTS A Primer on Electro-Static Dischar ge 460L Instruction Manual 74 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 This will allow any charges you are carrying t o bleed away through the ground conn ection of the workstation and preven t discharges due to fi eld effects and inductio n from occurring. 2. Pause for a secon ...

  • Teledyne 460L - page 83

    TELEDYNE INSTRUMENTS 460L Instruction Manual A Primer on Electro-Static Disc harge 05228 Rev B 75 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 5. Seal the container. If usi ng a bag, fold the end over and fastening it with a n ti-ESD tape.  Folding the open end over isol ates the component(s) inside from the effects of static fields.  Leaving ...

  • Teledyne 460L - page 84

    TELEDYNE INSTRUMENTS A Primer on Electro-Static Dischar ge 460L Instruction Manual 76 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 12.4.2.5. Packing Components for Return to Teledyne Instruments Customer Service Always pack electronic components and as semblies to be sent to Tel edyne Instruments Cust omer Service in anti-ESD bins, tube ...

  • Teledyne 460L - page 85

    TELEDYNE INSTRUMENTS 460L Instruction Manual A Primer on Electro-Static Disc harge 05228 Rev B 77 DCN 5164 PRINTED DOCUMENTS ARE UN CONTROLLED 4. Seal the container. If usi ng a bag, fold the end over and fastening it with a n ti-ESD tape.  Folding the open end over isol ates the component(s) inside from the effects of static fields.  Leaving ...

  • Teledyne 460L - page 86

    TELEDYNE INSTRUMENTS 460L Instruction Manual Spare Parts List 78 05228 Rev B PRINTED DOCUMENTS ARE UN CONTROLLED DCN 5164 USER NOTES: ...

  • Teledyne 460L - page 87

    M460L Instruction Manual Appendix A - Spare Parts list Appendix A - Spare Parts list • 05128 – Spar e Parts List M460L 05229 Rev B A-1 ...

  • Teledyne 460L - page 88

    Appendix A - Spare Parts list M460L Instruction Manual A-2 05229 Rev B ...

  • Teledyne 460L - page 89

    M460L Spare Parts List Part Number Description 025710100 PCA, UV DETECTOR PREAMP, M454/M460 030500400 OVERLAY, FRONT PANEL, O3 SENSOR, M460L 031680100 PCA, FRONT PANEL, M454 PRODUCTS 034920000 PCA, EXPANSION BD, M554/M460X/03/CAL 036280000 ASSY, WATER DROP OUT FILTER, PFA, M450 036750000 AKIT, ELEMENTS, 47MM, 30 MICRON (10 PCS) 037420000 ASSY, UV L ...

  • Teledyne 460L - page 90

    THIS PAGE IS INTENTIONALLY LEFT BLANK ...

  • Teledyne 460L - page 91

    460L Instruction Manual Append ix B – Electronic Schematics Appendix B – Electronic Schematics Table B-1 M460L List of Electr onic Sc hematics DOCUMENT # DESCRIPTION 03169 PCA, FRONT PANEL DISPLAY 04992 PCA, MAINBOARD 03493 PCA, CPU 05230 Rev A B-1 ...

  • Teledyne 460L - page 92

    Appendix B – Electronic Schema tics 460L Instruction Manual B-2 05230 Rev A ...

Manufacturer Teledyne Category Car Video System

Documents that we receive from a manufacturer of a Teledyne 460L can be divided into several groups. They are, among others:
- Teledyne technical drawings
- 460L manuals
- Teledyne product data sheets
- information booklets
- or energy labels Teledyne 460L
All of them are important, but the most important information from the point of view of use of the device are in the user manual Teledyne 460L.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Teledyne 460L, service manual, brief instructions and user manuals Teledyne 460L. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Teledyne 460L.

Similar manuals

A complete manual for the device Teledyne 460L, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Teledyne 460L by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Teledyne 460L.

A complete Teledyne manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Teledyne 460L - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Teledyne 460L, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Teledyne 460L, that we can find in the current document
3. Tips how to use the basic functions of the device Teledyne 460L - which should help us in our first steps of using Teledyne 460L
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Teledyne 460L
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Teledyne 460L in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Teledyne 460L?

Use the form below

If you did not solve your problem by using a manual Teledyne 460L, ask a question using the form below. If a user had a similar problem with Teledyne 460L it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)