Manual Teledyne 6750

96 pages 10.35 mb
Download

Go to site of 96

Summary
  • Teledyne 6750 - page 1

    Teledyne Analytical Instruments OPER A TIN G INSTRUCTIONS FOR Model 6750 T o t al Organic Carbon Analyzer DANGER To xic g ase s and o r fl ammabl e li quids ma y b e pr es en t i n t h i s mon ito ri ng syst e m. Personal protective e q uipment may be required when servicing this instrument. Hazardous volt ages exist on certain components internall ...

  • Teledyne 6750 - page 2

    Model 6750 Teledyne Analytical Instruments ii Copyright © 2005 Teledyne Instruments/ Analytical Instruments All Right s Reserved . No par t of this m anual may be r epro duced, tra nsmitted, transcribed , stored in a retriev al system, or translated into any o th er lan guage o r computer language in whole or in p art, in any f orm or by an y mean ...

  • Teledyne 6750 - page 3

    Total O r ganic Car bon A nalyzer Teledyne Analytical Instruments iii Speci fic Model Info rma tion Instrument S erial Nu mber: _______________________ Instrument Ra ng e: _______________ Calibrated for: _______________ Background Gas: _______________ Zero Gas: _______________ Span Gas: _______________ ...

  • Teledyne 6750 - page 4

    Model 6750 Teledyne Analytical Instruments iv Safety M e ssages Your safety and the saf ety of others is very important. We have provided ma ny important safety messages in this manual. Please read these me ss ages carefully. A safety message alerts you to potential hazards that could hurt you or others. Eac h saf ety message is associated with a s ...

  • Teledyne 6750 - page 5

    Total O r ganic Car bon A nalyzer Teledyne Analytical Instruments v IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE B EHAVIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES . This manual provide s infor mation designe d to guide you through the installation, cali br ation operation and mai ...

  • Teledyne 6750 - page 6

    Model 6750 Teledyne Analytical Instruments vi T ab le of Contents List of Fig ure s .............................................................................. ix List of Ta bles ............................................................................... xi Introduc ti on ...................................................................... ...

  • Teledyne 6750 - page 7

    Total O r ganic Car bon A nalyzer Teledyne Analytical Instruments vii 3.3 Electrical Connections 23 3.3.1 Power Connection 23 3.3.2 Analog Outputs 24 3.3.3 Alarm Connections 25 3.3.4 RS-232 Serial Connection 25 3.4 Gas/liquid Conne ctions 25 3.4.1 Liquid Connection s 25 3.4.2 Gas Connection s 26 3.5 Checking the System 27 Set up and Oper ation .... ...

  • Teledyne 6750 - page 8

    Model 6750 Teledyne Analytical Instruments viii 5.3.8 UV Power Suppl y 58 5.3.9 Mass Flow Controller (P/N ST18001 A) 59 5.3.10 Metering Val ve (P/ N ST16 050) 60 5.3.11 Sparger (P/N ST20025) 61 5.3.12 Pumps 63 5.3.12.1 Pump Mo tor Replacem e nt 63 5.3.12.2 Pump Head Tubin g Replacement 65 5.3.13 4-20 mA Adjustmen t Procedure 67 5.3.13.1 Setting 100 ...

  • Teledyne 6750 - page 9

    Total O r ganic Car bon A nalyzer Teledyne Analytical Instruments ix List of Figures Figure 1-1 : The Model 6750 TOC Ana lyzer (Front door remo ved) . 3 Fig ure 2-1 : NP OC o r TOC – Di re ct Metho d .................................... 6 Fig ure 2-2 : TOC – Tru e Me tho d ..................................................... 6 Fig ure : 2-3 : ...

  • Teledyne 6750 - page 10

    Model 6750 Teledyne Analytical Instruments x Fig ure 5-6 : DC Powe r Sup pl y Modu le........................................... 5 4 Fig ure 5-7 : Rep laci ng CE Co mpute r ............................................. 55 Fig ure 5-8 : UV Lamp Repl acem en t .............................................. 56 Figure 5-9 : Removing the UV Reactor Asse ...

  • Teledyne 6750 - page 11

    Total O r ganic Car bon A nalyzer Teledyne Analytical Instruments xi List of T ables Tab le 4-1 : Prep ara tion o f Sta nda rds (TOC/TIC ) ............................37 Tab le 5-1 : Troub le shoo ting ...........................................................44 ...

  • Teledyne 6750 - page 12

    Model 6750 Teledyne Analytical Instruments xii DANGER COMBUSTIBLE GAS USA G E WARNI NG This i s a genera l purpo se instrument designed for use in a non-h a zardous area. It is the custome r's responsibili ty to ensure safety especially when combustib le gases are being analyzed since the potential of g as lea ks always exist. The customer sho ...

  • Teledyne 6750 - page 13

    Total O r ganic Car bon A nalyzer Introdu c tion Teledyne Analytical Instruments 1 Introduction Tele dyn e Model 6750 Total Organic Carb on Analyzer provi d es accurate and reliable on-line TOC analysis with Windows™ CE Operation, using t h e UV/Heated Persulfate Oxidation Method. This manual includes all necessary information t o h elp you insta ...

  • Teledyne 6750 - page 14

    Introdu c tion Mo del 675 0 Teledyne Analytical Instruments 2 • RS-232C and RS-485 outputs • Separate electronics and liquid compartments 1.2 Options The foll owing op tions are available for the Model 6750 TOC Analyzer a nd are described in Appendix A .3: • Correlated BOD/COD • Dual NDIR Analyzers • Benchma rk /Auto Validation • Auto-C ...

  • Teledyne 6750 - page 15

    Total O r ganic Car bon A nalyzer Introdu c tion Teledyne Analytical Instruments 3 All other t yp es of usage beyond the intended use stated above, are considered a s m isuse of the analyzer. The supplier assumes no responsibilit y for d amage incurred as the r esult of mi s use of the product. 1.5 Operat or Interface The Model 6750 Anal y zer is h ...

  • Teledyne 6750 - page 16

    Introdu c tion Mo del 675 0 Teledyne Analytical Instruments 4 ...

  • Teledyne 6750 - page 17

    Total O r ganic Car bon A nalyzer Theory of Operation Teledyne Analytical Instruments 5 Theory of Ope ration 2.1 Over view The Model 6750 Anal y zer uses the UV/Heated Persulfate method of analysis and i s w ell-suited for many ap plications involving accurate TOC analysis. The b asic analyzer is confi gured for maximum utility using an advanced Mi ...

  • Teledyne 6750 - page 18

    Theory of Operation Mo del 675 0 Teledyne Analytical Instruments 6 To perform c orr ect TOC analysis, the analys is system must be capable of measuring all constituents of organic carbon pre s ent in the sample: N on-Purgeable Organic Carbon (NP OC) and Purgeable Organic Carbon (POC). Correct TOC analysis must also exclude t h e Total Inorg anic Ca ...

  • Teledyne 6750 - page 19

    Total O r ganic Car bon A nalyzer Theory of Operation Teledyne Analytical Instruments 7 (sparges) the CO 2 converted from the TIC and it i s m easured by an independe n t CO 2 detector as “TIC”. POC is also stripped from the sample i n the sparger as HC, but the HC is undetected by the CO 2 detector and does not interfere with the TIC measureme ...

  • Teledyne 6750 - page 20

    Theory of Operation Mo del 675 0 Teledyne Analytical Instruments 8 Figure: 2-3: TOC Measurement Basics Figure 2-4: UV/Persulfate M ethod ...

  • Teledyne 6750 - page 21

    Total O r ganic Car bon A nalyzer Theory of Operation Teledyne Analytical Instruments 9 2.3 The UV/Heated Persulfate Method of Ana lysis In the UV/Heate d Pe rsulfate Method as us ed by the Model 6750, the sample is initially mixed wit h acid and directed to t h e sparger. For TOC-Direct (NPOC) analysis, the CO 2 c o nverted from the inorganic carb ...

  • Teledyne 6750 - page 22

    Theory of Operation Mo del 675 0 Teledyne Analytical Instruments 10 Figure 2-5: Comparison of TOC Measurement Technologies ...

  • Teledyne 6750 - page 23

    Total O r ganic Car bon A nalyzer Theory of Operation Teledyne Analytical Instruments 11 2.4 Subsystems Principl es of Operation The Model 6750 T OC A nalyzer is comprised of five subsystems: 1. Sample Handling 2. Inorganic Ca rbon R emoval – OR – Inorganic Ca rbon An a lysis/TOC-True Me thod 3. Oxidation 4. NDIR CO 2 G as Det ection 5. Electro ...

  • Teledyne 6750 - page 24

    Theory of Operation Mo del 675 0 Teledyne Analytical Instruments 12 actual measurement of the TIC, a more accurate TOC is achieved since (TOC = TC - TIC). “TOC-Direct” mode is preferred for accuracy when no or little volatiles are pr esent. In this mode, the car bonate-free sample is extracted from the organic carbon reactor and measured direct ...

  • Teledyne 6750 - page 25

    Total O r ganic Car bon A nalyzer Theory of Operation Teledyne Analytical Instruments 13 Figure 2-7: Sample Handli ng TO C-True 2.4.2.2 G AS P HASE S AMPL E H ANDLING Referencing Figure 3-5 and “As Built” drawings in the Appendix, a carrie r g as of (99.8%) pure oxygen or CO 2 and hydrocarbon-free air is required. E rron eous TOC analysis will ...

  • Teledyne 6750 - page 26

    Theory of Operation Mo del 675 0 Teledyne Analytical Instruments 14 TOC analysis in a continuous, on-line TOC analyze r. The CO 2 generated by the oxidation of carbon in the react o r is read by the NDIR as a percentage concentration of the CO 2 in the carrier gas. CO 2 (TOC) CO 2 (TOC) + Carrier Any increase or decrease of the carrier flow rate wi ...

  • Teledyne 6750 - page 27

    Total O r ganic Car bon A nalyzer Theory of Operation Teledyne Analytical Instruments 15 be reduce d to approximately 2.0 to 3.0. If the sample pH is not properl y reduced a s d e scribed in Figure 2-8, increas e the norm a lcy of the acid reagent un til the water exiting the sparger drain is between 2.0 and 3.0. Figure 2-8: Relati onship Between T ...

  • Teledyne 6750 - page 28

    Theory of Operation Mo del 675 0 Teledyne Analytical Instruments 16 The construction of t h e patented UV reactor in the Model 6750 affords superior oxidation by precisely controlling the sample tempe r ature, adding a precise amount of u ltra-pure p ersulfate reagent (P/N ST40000) and precisel y controlling the carrier gas flow rate. Pump (P-1) co ...

  • Teledyne 6750 - page 29

    Total O r ganic Car bon A nalyzer Theory of Operation Teledyne Analytical Instruments 17 2.4.4 NDIR CO 2 Gas Detection The key c o mponent for reliable TOC anal ysis is the preci s e and interfe r ence-free detection of CO 2 . The M odel 6750 NDIR unit require s very little maintenance because of its ability to handle the corrosive gases generated ...

  • Teledyne 6750 - page 30

    Theory of Operation Mo del 675 0 Teledyne Analytical Instruments 18 requires no wall r eflectivity and is completely resistant to corrosive gases. The NDIR unit requi r es no optical chopper motor or mechanical device s but u tilizes a single flashing infrar ed source to optically “chop” the infra r ed (IR) beam, which equally irradiates CO 2 / ...

  • Teledyne 6750 - page 31

    Total O r ganic Car bon A nalyzer Theory of Operation Teledyne Analytical Instruments 19 2.4.5.1 B ENCHMARK /A UTO -V ALIDATION Benchma rk is the European NAMUR* s pecified validation techni qu e, whereby on command a chemical calibration standard is automaticall y introduced to the analyzer and the response is compared to the previ ous an alyzer c ...

  • Teledyne 6750 - page 32

    Theory of Operation Mo del 675 0 Teledyne Analytical Instruments 20 ...

  • Teledyne 6750 - page 33

    Total O r ganic Car bon A nalyzer Installation Teledyne Analytical Instruments 21 Installation This manual contains “AS BUILT” drawings and a parts list to aid in the i ns tallation and operation of the ana lyzer. T h ese drawings should be referred to when orde r ing spare parts, operating or servicing the equipme n t. See the Appendix for the ...

  • Teledyne 6750 - page 34

    Installation Model 6750 Teledyne Analytical Instruments 22 Figure 3-1: Analyzer Dimensions Figure 3-2: Re qu ired Door Clearance ...

  • Teledyne 6750 - page 35

    Total O r ganic Car bon A nalyzer Installation Teledyne Analytical Instruments 23 All operat or controls and serviceable components are mounted on the cont ro l panel, which is hinged and doubles as the door that provides acce ss to those components not normally requiring maintenance. Internal components of the Model 6750 TO C Analyzer a r e shown ...

  • Teledyne 6750 - page 36

    Installation Model 6750 Teledyne Analytical Instruments 24 • Power Connection Universal AC Power Source • Primary Input Power CAUT I ON: POWER IS APPLIED TO THE INSTRUMENT’S CIRCUITRY AS LONG AS THE INSTRUMENT IS CONNECTED TO THE FAC ILI T Y POWER SOURCE. THE FACILITY MUST HAVE AN EXTERNAL POWER SWITCH TO R EMOVE PRIMARY POWER FROM THE ANAL Y ...

  • Teledyne 6750 - page 37

    Total O r ganic Car bon A nalyzer Installation Teledyne Analytical Instruments 25 3.3.3 Alarm Connections The Model 6750 i s equipped with two concentration alarms and one anal y zer malfunction alarm. The three alarm-circuit connectors use spring termi n als for making connections t o internal alarm relay connect s. E ach alarm provides a set of F ...

  • Teledyne 6750 - page 38

    Installation Model 6750 Teledyne Analytical Instruments 26 • Reagent s • Drain (s) Connect t h ese lines as per Figure 3-4 and the AS -BUILT drawings in the Appendi x. M ake sure there are no r estrictions or kinks in the sample lines. Figure 3-5: Gas and Liquid Connections to the Analyzer 3.4.2 Gas Connections The gas connections include: • ...

  • Teledyne 6750 - page 39

    Total O r ganic Car bon A nalyzer Installation Teledyne Analytical Instruments 27 Wi th th e e x c ep ti o n of the v ent and carr ier g as in le t, a ll o th er ga s co n ne ct ion s ar e mad e o n th e s am e b otto m p anel as th e l iq u id co nn ec t ion s an d a re s ho wn in F ig u re 3- 5. Th e N D IR v en t a n d ca r r ier ga s c o nn e c ...

  • Teledyne 6750 - page 40

    Installation Model 6750 Teledyne Analytical Instruments 28 ...

  • Teledyne 6750 - page 41

    Total O r ganic Car bon A nalyzer Setup and Operation Teledyne Analytical Instruments 29 Setup and Operation Once the a n alyzer has been installed, it ca n be configure d to your applicati on. To do this you will: • Set system parameters • Calibrate t h e instrument • Define the analysis range. Then choose auto-ranging or select a fixed rang ...

  • Teledyne 6750 - page 42

    Setup and Operation Model 6750 Teledyne Analytical Instruments 30 Figure4-1: Drain Connections 5. On initial start –up or after UV Reactor or Glass Liquid Separator (GLS) servicing, fill the U-Tube with D. I. water as indicated in Figure 4-2. 6. After all of the above checks, turn analyze r ON by acti v ating the facility power switch. The sys te ...

  • Teledyne 6750 - page 43

    Total O r ganic Car bon A nalyzer Setup and Operation Teledyne Analytical Instruments 31 Figure 4-2: Gas Liquid Separator D. I. Filling After reagents and DI have f illed all tubing lines, it is recommended t o p roceed with IR Calibrat ion followed by cal (liquid cali br ation) of the system. Then the analyzer may be placed in service. 4.2 Menus T ...

  • Teledyne 6750 - page 44

    Setup and Operation Model 6750 Teledyne Analytical Instruments 32 • The curre n t TOC value, • [IR] IR response • [Flow] Carrier Gas Flow rate • [m] Availa b le Memory (kb). 4.2.2 Parameter Menu Although your anal y zer has been optimally configured for your applicati on, on initial startup the setup par ameters should be verified Consult F ...

  • Teledyne 6750 - page 45

    Total O r ganic Car bon A nalyzer Setup and Operation Teledyne Analytical Instruments 33 2. From this menu, select:[Parameters] The foll owing s creen appears: This is the Parameters menu from which the anal y zer may be configured t o sui t the requirements of the operator Verify the following settings are correct or select the desired settings: ( ...

  • Teledyne 6750 - page 46

    Setup and Operation Model 6750 Teledyne Analytical Instruments 34 From this menu ve r ify or change the concentration alarm setpoints using the onscree n scro ll arrows. 4.2.4 Stream Sequencer If your analyzer has been confi gur ed for a multi-stream sequencer (P/N ST200009), from the [Edit] Menu, select [Sequencer]. The following me nu appears: Fr ...

  • Teledyne 6750 - page 47

    Total O r ganic Car bon A nalyzer Setup and Operation Teledyne Analytical Instruments 35 4.2.5 D/DBPR This optional utility allows conve n ient ex ternal pH c alibration and input of al k alinity data consistent with the D/DBP R requireme nts If your analyzer has been confi gur ed for the D/DBP R option, from the foll ow ing [Edit] Menu, select [D/ ...

  • Teledyne 6750 - page 48

    Setup and Operation Model 6750 Teledyne Analytical Instruments 36 The foll owing window is illustrative of th e D/DBPR [Run] screen. 4.3 Detail ed Calibration Procedures 1. Although the i ns trument has been calibrated to your specificati ons a t the factory, it should be r echec k ed after satisfact ory installation and periodically calibrated as ...

  • Teledyne 6750 - page 49

    Total O r ganic Car bon A nalyzer Setup and Operation Teledyne Analytical Instruments 37 Table 4-1: Preparation of S tandards (T OC/TIC) ORGANIC COMPOUNDS 100 mg/Liter TOC (AS CARBON) 1000 mg/Liter TOC (AS CARBON) Ethylene Glycol .233 ml/L H 2 0 2.33 ml/L H 2 0 Methanol .337 ml/L H 2 0 3.37 ml/L H 2 0 Ethanol .242 ml/L H 2 0 2.42 ml/L H 2 0 Acetone ...

  • Teledyne 6750 - page 50

    Setup and Operation Model 6750 Teledyne Analytical Instruments 38 CAUT I ON: SODIUM PERSULFATE IS A STRONG OXIDIZING AGENT . TAKE ALL NECESSARY PRECAUTIONS AS WITH HANDLING ANY C ORROSIVE MATERIAL. ALL CHEMICALS, INCLU DIN G WATER, SHOULD BE REAGENT GRADE OR B ETTER AND FREE OF CARBON-CONTAINING COMPOUNDS. THE CONTAINER USED SHOULD BE F LUSHED WITH ...

  • Teledyne 6750 - page 51

    Total O r ganic Car bon A nalyzer Setup and Operation Teledyne Analytical Instruments 39 Note: On the initial Startup, IR gas calib ration should be performed prior to liquid calibration a. Selec t [CALIBRAT E] from the main menu. The following scree n wi ll appear: b. Select [ TOC]. The foll owing s creen appears: c. Flow “ZERO” (D. I. Wa t er ...

  • Teledyne 6750 - page 52

    Setup and Operation Model 6750 Teledyne Analytical Instruments 40 f. If the calibration i s a cceptable, select [OK ] or [CANCEL] if not acceptable. The analyzer is now calibrated and ready for ana lysis. 4.4 Operation After proper installation and calibration, the instrume n t is ready to be placed “ON-LINE” for conti nuous operation. Note: Fo ...

  • Teledyne 6750 - page 53

    Total O r ganic Car bon A nalyzer Setup and Operation Teledyne Analytical Instruments 41 4.5 Shutdown Prior to shutdown of the analyze r o r performing service on the react or, f lush D. I. water through all sample and reagent lines for at least 30 minute s to clean all tubing. CAUT I ON: FAILURE TO OBSERVE ADEQUATE FLUSHING COULD R ESULT IN HARMFU ...

  • Teledyne 6750 - page 54

    Setup and Operation Model 6750 Teledyne Analytical Instruments 42 To view the d ata produced in the last 24 hours , select [LAS T 24 HOURS ]. The foll owing screen will appear: Using the Scroll Ba r, s elect the data to be viewed. Selecting [HISTORICAL DATA] allows you to view any archived data wit h in the memory limits of the computer. A screen a ...

  • Teledyne 6750 - page 55

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 43 Maintenance 5.1 Compupter-aided T es ting Prior to performi ng d etailed troubleshooting procedures listed below, first perform computer-aided testing of the analyzer, as follows: 1. From the scree n, se lect [TEST]. The follow ing menu appears: 2. Select [S YSTEM DIAG ...

  • Teledyne 6750 - page 56

    Maintenance Model 6750 Teledyne Analytical Instruments 44 From this screen t h e operator may perform individual testing of c o mponents related to functionality and settings. 5.2 T roub l eshooting If you are experiencing trouble with your instrument, refer to Table 5-1: Trouble shoot ing. Table 5-1: Troubleshooting PROBLEMS SOLUTION Analyze r wil ...

  • Teledyne 6750 - page 57

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 45 5.3 Module Service 5.3.1 NDIR (Gas Calibration) 1. From the Menu, select [CALIBRATE]. The foll owing s creen will appear: 2. Select [INF RARED ANA LYZER]. The following Screen will appear: 3. Disconnect t ub ing from reactor output and connect IR cali br ation adapter ...

  • Teledyne 6750 - page 58

    Maintenance Model 6750 Teledyne Analytical Instruments 46 Figure 5-1: Disconnecting Tubing f rom Reactor Figure 5-2: IR Calibration A dap ter P/N ST20026 IR Calib ration Adapter ...

  • Teledyne 6750 - page 59

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 47 Figure 5-3: Installing Cali br ation Adapte r 5. To set IR “Zero”, fl ow 200 c c/minute oxy gen (or CO 2 - free ai r) for 5 minutes. Allow NDIR reading to stabilize. Select [ ZERO]. NDIR reading is reset to 0. 6. To set IR “Span” for the 6 inch (shorter) IR ben ...

  • Teledyne 6750 - page 60

    Maintenance Model 6750 Teledyne Analytical Instruments 48 Figure 5-4: Re mov ing the NDIR Unit P/N ST36000 To Replace t h e NDIR Unit: 1. Turn off power. 2. Remove i n let tube from NDIR Assembly. 3. Remove ve n t tube from NDIR As sembly. 4. Unplug interface cable from NDIR Assembly. 5. Remove NDIR Assembly from ca b inet. ...

  • Teledyne 6750 - page 61

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 49 • Remove insula t ion from IR Cell. • Remove lock-tight from three thumbnuts and remove . • Slide ou t and remove Source Assemb l y . ...

  • Teledyne 6750 - page 62

    Maintenance Model 6750 Teledyne Analytical Instruments 50 • Remove IR Ce ll Assembly Note: a. ‘O’-Rings in each of Detector Assemb l y (l eft) and Sou r ce Assemb l y (r ight). b. Sapphire windows loca t ed under ‘O’-Rings ...

  • Teledyne 6750 - page 63

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 51 • *Remove ‘O’-Ring and Sapph ire window taking care t o avoid scratching S apphire window. *Recommend usin g toothpick. • Clean s apphire windows with a soft, lintless tissue (use DI water, if necessa r y ). To Reassemb le • Install IR Cell. • Install Sourc ...

  • Teledyne 6750 - page 64

    Maintenance Model 6750 Teledyne Analytical Instruments 52 • Tighten three thumbnuts un ti l Source Assembly bottoms out on squeeze nuts. • Recommend : Use a lt ernate tightening sequence • Insert lock -t ight on three thumbnuts to prevent backoff. • Install insula ti on. Replace assembly as before. • Recalibrate NDIR (Section 4 . 3). Note ...

  • Teledyne 6750 - page 65

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 53 5.3.3 Master Interface Board (P/N ST130 42-1). The Master Interface Board h as no field a djustment . If malfunctioni ng, r eplace the module P/N ST13042-1 using the instructi ons below. See Figure 5-5. Figure 5-5: Re mov ing the Master Interfac e Board To Remove the P ...

  • Teledyne 6750 - page 66

    Maintenance Model 6750 Teledyne Analytical Instruments 54 5.3.4 D. C. Power Supply The power supply has no field adj ust ment. If malfunctioning, replace the module (P/N ST15083) using the directions below. See Figure 5-6. Figure 5-6: DC Power Supply Modu le To replace the module: 1. Turn off power. 2. Remove plastic cover. 3. Disconnect c onn ecto ...

  • Teledyne 6750 - page 67

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 55 5.3.5 CE Computer (P/N 13039) If malfunctioni ng, r eplace the Computer M odule using the directions below. See Figure 5 -7. Figure 5-7: Re p lacing CE Computer To Replace t h e CE Computer: 1. Turn off power 2. Disconnect electrical connectors. 3. Remove scre ws holdi ...

  • Teledyne 6750 - page 68

    Maintenance Model 6750 Teledyne Analytical Instruments 56 5.3.6 UV Lamp (P/N ST20009) Normal lamp replacement frequency is suggested a t least every 12 months, however, if a lamp is “burned-out” (weak or no blue glow), replace with UV Lamp (P/N ST20009), per directions in Figure 8. Figure 5-8: UV Lamp Replacement To replace the UV lamp: 1. Turn ...

  • Teledyne 6750 - page 69

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 57 CAUT I ON: THE UV LIGHTED LENGTH MUST BE AT LEAST 12 ½ INCHES FOR PROPER OPERATION. DO NOT USE STANDARD 1 2 INCH LIGHTED LENGTH UV LAMPS. 2. Turn off power all ow 10 m inutes for cool dow n. 3. Disconnect electrical connectors. 4. Loosen fittings as ill us trated. 5. ...

  • Teledyne 6750 - page 70

    Maintenance Model 6750 Teledyne Analytical Instruments 58 well as the sa mple inlet line and submerse all three (3) lines int o D . I. water to completely flush system. CAUT I ON: FAILURE TO DO SO M AY RESULT IN ACID BURNS, CLOGGED REACTOR, ETC. FLUSH FOR 30 MINUTES . 2. Turn off power. 3. Disconnect electrical connectors. 4. Disconnect t ub ing. 5 ...

  • Teledyne 6750 - page 71

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 59 Figure 5-10: Re mov ing the UV Power Supply 5.3.9 Mass Flow Controller (P/N ST1800 1A) If proper flow is not detected by the computer, a n a larm will be acti v ated and displayed. The fault may be in the O 2 /Air supply (c h eck for 20 psi and flow of the supply by di ...

  • Teledyne 6750 - page 72

    Maintenance Model 6750 Teledyne Analytical Instruments 60 Figure 5-11: Mass Flow Controller 5.3.10 Metering Valve (P/N ST16050) If no O 2 /air flow to t h e sparger (P/N ST20025) is obs erved (no bubbles), chec k for tubing and fitting leaks , restrict ions and adjust mete r ing valve (be careful to return to previous indicated bubble rate). If flo ...

  • Teledyne 6750 - page 73

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 61 Figure 5-12: Metering Valve 5.3.11 Sparger (P/N ST20025) The sparger may be removed for cleaning with dete rg ent and subsequent flushing with DI wate r. If the sparger inner glass body is observed to be leaking (cracked glass, etc.), or is clogged be yond clea n ing, ...

  • Teledyne 6750 - page 74

    Maintenance Model 6750 Teledyne Analytical Instruments 62 1. Turn mac h ine OFF. 2. Turn carrier gas (Air/O 2 ) OFF. 3. Remove a nd dr ain tubing and sparger. 4. Remove sparge r fro m spring clips. To clean the sp a rger: 1. Run warm water t hrough tubing to flush out any solid contamination. 2. Reinstall sparger i n r everse order. Figure 5-13: Sp ...

  • Teledyne 6750 - page 75

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 63 5.3.12 Pumps If any of the pum ps ar e not pumping (liquid flow is not observed moving up the tubing), check to see if the motor is turning the pump heads and the motor is not binding. Check for proper power and connections to the pu mp motor. If the motor has failed ( ...

  • Teledyne 6750 - page 76

    Maintenance Model 6750 Teledyne Analytical Instruments 64 3. Remove p ump as illustrated to gain access to pump mounting screws. 4. Remove t wo thumbscrews for pump p anel removal. 5. Disc o nnect Pump Motor Connec t o r. ...

  • Teledyne 6750 - page 77

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 65 6. Remove 4 p ump mounting screws for pump remova l. 7. Rep la ce motor and reassemble in reverse step s equence. Figure 5-14: Re mov ing the Pump Note: Actual mounting panel may appear somewha t different but mounting hardware is the same. 5.3.12.2 P U MP H EAD T UB I ...

  • Teledyne 6750 - page 78

    Maintenance Model 6750 Teledyne Analytical Instruments 66 3. Place tu bing in the right groove and against the first two rollers. 4. Hold tub in g with thumb. 5. Nea r groove, insert smaller prong of loading key be tw een the top of the rotor and t ubi ng. 6. Push key in as f ar as possible. 7. Push do w n and turn key counterclockwise (ccw) co m p ...

  • Teledyne 6750 - page 79

    Total O r ganic Car bon A nalyzer Maintenance Teledyne Analytical Instruments 67 11. Pos ition th e other end bell on top and press the end bells together. 12. Be c aref ul not to pinch the tubing. 13. If end be ll s do not snap tightly together, reload tub i n g. 14. If nece s sary, turn key in slot on rotor shaft to adjust tubing (as i n next ste ...

  • Teledyne 6750 - page 80

    Maintenance Model 6750 Teledyne Analytical Instruments 68 5.3.13.1 S ET TING 100 M V C ONT ROL V OLTAGE 1. Using a DC voltmeter, connected to measure DC voltage, connect the black lead to the (-) side of capacitor C1. Located near the lower left corne r of the circuit board. Connect t h e (+) lead to the top of resistor R19, located slightly t o th ...

  • Teledyne 6750 - page 81

    Total O r ganic Car bon A nalyzer Appendix Teledyne Analytical Instruments 69 Appendix A.1Specifications Measurement Me thods: TC (Total Carbon) – UV/Persulfate Oxidation T OC (Total Organic Carbon) acidification and sparging to eliminate inorganic carbon interfe r ence. Tempe r ature: 25 o C ± 5 o Measurement R anges: 0-10 through 0-10,000 ppm, ...

  • Teledyne 6750 - page 82

    Appendix Model 6750 Teledyne Analytical Instruments 70 Sample requireme nts : Inlet Pressure: Atm ospheric to 3 psig Flow Rate: 20c c/min. (nom) Drai n: Gravi ty drain vented to atmosphere Suspended Solids: 1,000 microns (ma x.) Re agents (TOC Mode): Sodium Persulfat e : 4.0 gal/month Phosphoric Acid, 10% v/v: 4.0 gal/m on th (nom.). Calibration: O ...

  • Teledyne 6750 - page 83

    Total O r ganic Car bon A nalyzer Appendix Teledyne Analytical Instruments 71 A.2 Parts Listing Part # Description Image ST16002 ST16003 ST16004 ST16004 ST16005 ST16006 #7013 Pump Head #7014 Pump Head #7015 Pump Head #7016 Pump Head #7017 Pump Head #7018 Pump Head ST18002-1 2 RPM 110V ST18003-1 2 RPM 220V Pump M otor Assembly ST16007 ST16008 ST1600 ...

  • Teledyne 6750 - page 84

    Appendix Model 6750 Teledyne Analytical Instruments 72 Part # Description Image ST15083 12V DC Powe r Supply ST13069-1 Distribution Boa r d ST13070-1 Interface Board ST13039 CE Computer ST100001-6 6” NDIR Assemb l y ST100001-15 15” NDIR Asse m bly ...

  • Teledyne 6750 - page 85

    Total O r ganic Car bon A nalyzer Appendix Teledyne Analytical Instruments 73 Part # Description Image ST100006-1 NDIR Detector Assemb l y ST100009-6 6” NDIR Cell Assemb l y ST100009-15 15” NDIR Cell Asse m bly ST100008-1 NDIR Sou r c e Assembly ST17007 ST20001 NDIR O-Ring NDIR Sapph ir e Window ST20025 Sparger ...

  • Teledyne 6750 - page 86

    Appendix Model 6750 Teledyne Analytical Instruments 74 Part # Description Image ST16050 Metering Va l ue ST200026 IR Calibration Adapter ST20009 UV Process Lamp ST13066-1 AC Boa r d Assembly ...

  • Teledyne 6750 - page 87

    Total O r ganic Car bon A nalyzer Appendix Teledyne Analytical Instruments 75 Part # Description Image ST200003-1 UV Reactor Assembly ST15000- 110V ST15002- 220V UV Power Supply ST18050 Flow Switch ST20024T Gas L iq uid Separator ...

  • Teledyne 6750 - page 88

    Appendix Model 6750 Teledyne Analytical Instruments 76 A.3 Opti ons A.3.1 Option 200008 BENCH MARK (with sin gle point autocal) This option a u tomatically determines analyzer performance to specifie d limits. It is accomplished by introducing a known chemical standard soluti on to the analyzer and performing a computation and analysis to ve r ify ...

  • Teledyne 6750 - page 89

    Total O r ganic Car bon A nalyzer Appendix Teledyne Analytical Instruments 77 A.3.4 Option 200016 EXTERNAL RANGE CHANGE This option provi d es for local or remote a nalyze r r ange change by contact cl osur e . A.3.5 Option 200017 TRUE DUAL RAN GE OPT ION (not an electronic scaling) A.3.6 Option 200009 MULTI-STREAM SEQUENCER This option allows time ...

  • Teledyne 6750 - page 90

    Appendix Model 6750 Teledyne Analytical Instruments 78 Since COD & BOD analysis may va ry wi th s easonal infl u ences or other particular conditions, correlations for those conditions would be used at those time s . After obtai n ing a representative amount of TOC/COD/BOD data, the operator enters those factors into the analyz er computer. The ...

  • Teledyne 6750 - page 91

    Total O r ganic Car bon A nalyzer Appendix Teledyne Analytical Instruments 79 A.4 AS-BU IL T Dra w ings ...

  • Teledyne 6750 - page 92

    Appendix Model 6750 Teledyne Analytical Instruments 80 ...

  • Teledyne 6750 - page 93

    Total O r ganic Car bon A nalyzer Index Teledyne Analytical Instruments 81 Index accuracy, 12, 39 acid, 6, 9, 11, 14, 19 ACID IN por t , 30 acid preparation, 38 air, 70 air gap, 29 air purifier, 13 air supply pr es sure, 29 alarm, 1, 18, 19 concentration, 25, 69 failsafe, 25 high, 25 latching, 25 low, 25 master fault, 1, 25, 69 setpoint, 34 setting ...

  • Teledyne 6750 - page 94

    Index Model 67 50 Teledyne Analytical Instruments 82 enclosure, 3 end bell, 65 end-to-end method, 36 EPA, 9 exhaust connection, 27 external pH calibration, 35 fail-safe, 18 features, 1, 17 figures listing, ix flow rate, 13, 29, 70 Form C conta ct, 25 fuse, 44 gain. See automatic gain control gas connection, 26 gas/liquid separator, 12, 14 general p ...

  • Teledyne 6750 - page 95

    Total O r ganic Car bon A nalyzer Index Teledyne Analytical Instruments 83 parameters menu, 32 parts listing, 71 persulfate, 15, 16, 19 persulfate line, 30 pH , 6, 11, 14 phosphoric acid, 38 POC, 5, 7 potassium hydrogen phthalate, 36 power connection, 23 power requirement, 70 power supply, 23 preferred method of analysis , 6 pressure swing absorpti ...

  • Teledyne 6750 - page 96

    Index Model 67 50 Teledyne Analytical Instruments 84 UV/Heated Persulfate Oxidation Method, 1, 5 . See valve, 19 view window, 3 VOC, 5, 7 volatile hydrocarbon, 11 volatile organic carbon. See VOC volatile organic loss, 9 , 14 volatile organics, 1 voltmeter, 68 wall mounting, 21 warning sign, iv warranty, i i water vapor int er ference, 18 website a ...

Manufacturer Teledyne Category Carbon Monoxide Alarm

Documents that we receive from a manufacturer of a Teledyne 6750 can be divided into several groups. They are, among others:
- Teledyne technical drawings
- 6750 manuals
- Teledyne product data sheets
- information booklets
- or energy labels Teledyne 6750
All of them are important, but the most important information from the point of view of use of the device are in the user manual Teledyne 6750.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Teledyne 6750, service manual, brief instructions and user manuals Teledyne 6750. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Teledyne 6750.

Similar manuals

A complete manual for the device Teledyne 6750, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Teledyne 6750 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Teledyne 6750.

A complete Teledyne manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Teledyne 6750 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Teledyne 6750, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Teledyne 6750, that we can find in the current document
3. Tips how to use the basic functions of the device Teledyne 6750 - which should help us in our first steps of using Teledyne 6750
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Teledyne 6750
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Teledyne 6750 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Teledyne 6750?

Use the form below

If you did not solve your problem by using a manual Teledyne 6750, ask a question using the form below. If a user had a similar problem with Teledyne 6750 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)