    INSTRUCTION MANUAL HFM-E-200/HFC-E-202 SERIES FLOWMETERS/CONTROLLERS TELEDYNE HASTINGS INSTRUMENTS ISO 9001 CER TIFIED ...

  • Teledyne HFM-E-200 - page 2

    HFM-E-200/HFC-E-2 02 Page 2 of 30 Manual Print History The print history shown below lists the printing da tes of all revisions and addenda created for this manual. The revision level letter increases alphabetic ally as the manual undergo es subsequent updates. Addenda, which are released between revisions, c ont ain important change information th ...

  • Teledyne HFM-E-200 - page 3

    HFM-E-200/HFC-E-2 02 Page 3 of 30 Table of Contents 1. INSTALLAT ION A N D OP ERATION ..................................................................................................... ........................ 4 1.1. F EATURES ........................................................................................................................ ...

  • Teledyne HFM-E-200 - page 4

    HFM-E-200/HFC-E-2 02 Page 4 of 30 1. Inst allation and Operation The Hastings HFM-E- 200 mass Flow-meter and HFC-E -20 2 Flow-controller are designed to a ccur ately measure and control mass flow over the range o f 10 sccm to 30 slm, without corrections or compensations for ga s pressure and temperature with an accuracy of better than ±1% FS. Hast ...

  • Teledyne HFM-E-200 - page 5

    HFM-E-200/HFC-E-2 02 Page 5 of 30 1.2. Specifications Accuracy .............................................................................................................. ±1% full scale (F.S.) Repeatability .............................................. <±0.125% of F.S.or ±0.075% of rd g. +0.05% F.S. (max) Maximum operational pressu re ... ...

  • Teledyne HFM-E-200 - page 6

    HFM-E-200/HFC-E-2 02 Page 6 of 30 1.3. Optional 4-20 mA Current Output An option to the standard 0-5 VDC outp ut is the 4-2 0 mA current output that is proportional to flow. The 4 - 20 mA signal is pr oduc ed from the 0 - 5 VDC outp ut of the Flowmeter. The current loop output is useful for remote applications where pi ckup noise could subst ant ia ...

  • Teledyne HFM-E-200 - page 7

    HFM-E-200/HFC-E-2 02 Page 7 of 30 2. Inst allation and Operation This section contains the neces sary steps to assist in getting a new Flowmeter/Controller into operation as quickly and easily as possible. Please read the following thoroughly before attempting to install the instrument. 2.1. Receiving Inspection Carefully unpack the Hastings HFM-E- ...

  • Teledyne HFM-E-200 - page 8

    HFM-E-200/HFC-E-2 02 Page 8 of 30 suggested that all connec t ions be checked for leaks after installation. This can be done by pressurizing the instrument (do not exceed 500 psig unless the Flowmeter is specifically rated for higher pressures) and applying a diluted soap solution to the flow connections rate d for higher pressures) and applying a ...

  • Teledyne HFM-E-200 - page 9

    HFM-E-200/HFC-E-2 02 Page 9 of 30 Pin 1 is the case ground. It should be connected to the cable shield if available and to the AC ground at the power supply. Pin 3 is the output signal from the flow controller. Th is output will be 0-5VDC, 5VDC being 10 0% o f rated or full flow. Pin A is the command input. This sh ould be a 0-5VDC sign al and must ...

  • Teledyne HFM-E-200 - page 10

    HFM-E-200/HFC-E-2 02 Page 10 of 30 periodically during normal operat ion. Zero adjustment is requi red if there is a change in ambient temperature, or vertical orientat ion of the Flowmet er/controller. 2.6.3. High Pressure Operation When operating at high pressure, the increased dens ity of gas will cause natu ral convection to flow through the se ...

  • Teledyne HFM-E-200 - page 11

    HFM-E-200/HFC-E-2 02 Page 11 of 30 2.7.3. Operation with an external sensor. (Fig. 2.2) In some instances, it migh t be desirable to use an external sen sor to provide process infor mation to the control circuitry in the flow controller. For example, you might want to control the pressure in a vacuum system by adjusting the rate at which the syst e ...

  • Teledyne HFM-E-200 - page 12

    HFM-E-200/HFC-E-2 02 Page 12 of 30 3. Theory of Operation This section contains an o v erall functional descript ion of HFC Flow Controllers. Detailed schematics and parts lists can be found at the end of the manual in Section 6.0. In this sect ion and other sections throughout this manual, when a power supply is mentioned, it is a ssumed that the ...

  • Teledyne HFM-E-200 - page 13

    HFM-E-200/HFC-E-2 02 Page 13 of 30 3.4. Shunt: Measurement of flow rates higher th an the 10 sccm full scale is achiev ed by dividing the flow with a fixed ratio shunting arrangement, as is illustrated in Figu re 3.3. This is accomplished by placing the measuring capillary tube parallel with one or more dimensionally similar channe ls, called a lam ...

  • Teledyne HFM-E-200 - page 14

    HFM-E-200/HFC-E-2 02 Page 14 of 30 3.5. Valve: The control valve is an “automatic metering solenoid” valv e. While mo st solenoids operate in either the fully open or fully closed state, the automatic meteri ng solenoid valve is desi gned to control flow (see Figure 3.6). A spring, connected to the plunger assembly, holds a ma gnetic plunger ti ...

  • Teledyne HFM-E-200 - page 15

    HFM-E-200/HFC-E-2 02 Page 15 of 30 4. Maintenance This section contain s ser vice and calibration inform a tion . Some portions of the instrument are delicate. Use extreme care when servicing the flow controlle r. The potentiometer positions and the e lectrical components referred to in the troubleshooting sect ion can be found in Section 6.0 on th ...

  • Teledyne HFM-E-200 - page 16

    HFM-E-200/HFC-E-2 02 Page 16 of 30 SYMPTOM : Flow controller oscillates. CAUSE : Flow controller not adjusted for the dynamics of the flow system. ACTION : Check upstream and downstream pres sures. The gas supply regulator should not have excessive lockup when flow shuts off. Also ensure that there is not a large drop in pressure between the regula ...

  • Teledyne HFM-E-200 - page 17

    HFM-E-200/HFC-E-2 02 Page 17 of 30 4.3.2. Miscellaneous adjustments Periodically, during normal operation, the ZERO should be checked and adjusted when required. If the instrument is not stopping the flow completely when com m and signal is Zero, the orifice may require turning approximately 1/8 turn clockw is e. 4.4. Inlet Removal: The fitting on ...

  • Teledyne HFM-E-200 - page 18

    HFM-E-200/HFC-E-2 02 Page 18 of 30 Where: Formula 1: Formula 2: u P Q D σ 0028 . 0 = σ d Q D Ρ • ΔΡ = 0014 . 0 D = Diameter in inches Q = Flow rate in standard liters per min ute P = Pu - Pd in PSI Pu = Upstream pressure in PSIA Pd = Downstream pressure in PSI A σ = Specific gravity of gas Choose the orifice form Se ction 5.0 that has the c ...

  • Teledyne HFM-E-200 - page 19

    HFM-E-200/HFC-E-2 02 Page 19 of 30 4.8. REPLACEMENT PARTS The following is a list of the available replacement parts and their factory st ock numbers. The HFM-E- 200 and the HFC-E-202 sh unts and sen sor modules are interch angeable. STOCK NO. DESCRIPTION AIR RANGE 81-270 30L - Shunt 24 - 36 81-269 20L - Shunt 16 – 24 81-268 10L - Shunt 8 – 12 ...

  • Teledyne HFM-E-200 - page 20

    HFM-E-200/HFC-E-2 02 Page 20 of 30 5. Conversion Factor T able Rec # Gas Symbol GCF Derived Densit y Density Synonyms Gamma R Z (g/L) 25°C (g/L) 0°C (Cp/Cv) / 1 atm / 1 atm 25 J/gm*K 1 Acetic Acid C 2 H 4 F 2 0.4155 4 2.700 - 2.947 - Et hanoic Acid 1.2 125.88 2.0301 2 Acetic Acid, Anhy dride C 4 H 6 O 3 0.2580 4 4.173 - 4.555 - Acet icanhydride 1 ...

  • Teledyne HFM-E-200 - page 21

    HFM-E-200/HFC-E-2 02 Page 21 of 30 48 Difluoroethylene C 2 H 2 F 2 0.4492 4 2.617 - 2.857 - Vinylidenefluoride, G-1132A, Difluoroethene 1.2 129.85 2.0457 49 Dimethylamine C 2 H 7 N 0.3705 4 1.843 - 2.011 - N-me thylmethanamine 1.2 184.42 1.4793 50 Dimethyl Ether C 2 H 6 O 0.4088 4 1.883 - 2.055 - Methylether; Methane, Oxybis- 1.2 180.48 1.5211 51 D ...

  • Teledyne HFM-E-200 - page 22

    HFM-E-200/HFC-E-2 02 Page 22 of 30 99 Isopentane C 5 H 12 0.2175 4 2.949 - 3.219 - 2- methylbutane 1.2 115.24 1.8975 100 Isopropyl Alcohol C 3 H 8 O 0.2931 4 2.456 - 2.681 - 2- propanol 1.2 138.356 1.7335 101 Isoxazole C 3 H 3 NO 0.4333 4 2.823 - 3.081 - 1-Oxa- 2-azacyclopentadi ene 1.2 120.39 2.1501 102 Ketene C 2 H 2 O 0.5732 4 1.718 - 1.875 - NA ...

  • Teledyne HFM-E-200 - page 23

    HFM-E-200/HFC-E-2 02 Page 23 of 30 152 R123A C 2 HCl 2 F 3 0.2699 4 6.251 - 6.823 - 1,2-Dichloro-1,2,2- trifluoroethane 1.2 54.37 3.1065 153 R125 C 2 HF 5 0.2826 2 4.906 - 5.355 - Pent afluoroethane 1.2 66.50 2.6844 154 R134 C 2 H 2 F 4 0.2996 4 4.170 - 4.552 - 1,1,2,2-te trafluor oethane 1.2 78.42 2.4595 155 R134A C 2 H 2 F 4 0.3110 2 4.170 - 4.55 ...

  • Teledyne HFM-E-200 - page 24

    HFM-E-200/HFC-E-2 02 Page 24 of 30 6. W ARRANTY 6.1. Warranty Repair Policy Hastings Instruments wa rr ants this product for a period of one year from the date of ship ment to be free from defects in material and workmanship. This warran ty does not apply to defects or failur es resulting from unauthorized modifica tion, misuse or mishandling of th ...

  • Teledyne HFM-E-200 - page 25

    HFM-E-200/HFC-E-2 02 Page 25 of 30 7. Drawings 7.1. Notes: 1. Place jumper (Item 42) on pins 4 and 5 of P2 and P3 as sh own. 2. Place one on bottom two pins (1 and 2) of JP 3. 3. Place one on both pins of J P6. ...

  • Teledyne HFM-E-200 - page 26

    HFM-E-200/HFC-E-2 02 Page 26 of 30 ...

  • Teledyne HFM-E-200 - page 27

    HFM-E-200/HFC-E-2 02 Page 27 of 30 ...

  • Teledyne HFM-E-200 - page 28

    HFM-E-200/HFC-E-2 02 Page 28 of 30 ...

  • Teledyne HFM-E-200 - page 29

    HFM-E-200/HFC-E-2 02 Page 29 of 30 ...

  • Teledyne HFM-E-200 - page 30

    HFM-E-200/HFC-E-2 02 Page 30 of 30 ...

Manufacturer Teledyne Category Marine Sanitation System

Documents that we receive from a manufacturer of a Teledyne HFM-E-200 can be divided into several groups. They are, among others:
- Teledyne technical drawings
- HFM-E-200 manuals
- Teledyne product data sheets
- information booklets
- or energy labels Teledyne HFM-E-200
All of them are important, but the most important information from the point of view of use of the device are in the user manual Teledyne HFM-E-200.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Teledyne HFM-E-200, service manual, brief instructions and user manuals Teledyne HFM-E-200. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Teledyne HFM-E-200.

Similar manuals

A complete manual for the device Teledyne HFM-E-200, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Teledyne HFM-E-200 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Teledyne HFM-E-200.

A complete Teledyne manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Teledyne HFM-E-200 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Teledyne HFM-E-200, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Teledyne HFM-E-200, that we can find in the current document
3. Tips how to use the basic functions of the device Teledyne HFM-E-200 - which should help us in our first steps of using Teledyne HFM-E-200
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Teledyne HFM-E-200
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Teledyne HFM-E-200 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Teledyne HFM-E-200?

Use the form below

If you did not solve your problem by using a manual Teledyne HFM-E-200, ask a question using the form below. If a user had a similar problem with Teledyne HFM-E-200 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)