  • Omron Healthcare HEM-711AC - page 1

    Automa tic Blood Pressur e Monitor with Arm C uff Model HEM-711A C INSTRUCTION MANUAL ESP AÑOL ENGLISH 280239_HEM-7011-E_S.qxd 2/19/08 3:00 PM Page 1 ...

  • Omron Healthcare HEM-711AC - page 2

    Befor e Using the Monitor Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Operating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Risk ...

  • Omron Healthcare HEM-711AC - page 3

    Thank you for purchasing the Omron ® HEM-711A C IntelliSense ® Automatic Blood Pressure Monitor with Arm Cuff. INTRODUCTION 3 Fill in for future referenc e. DA TE PURCHASED: __________________ SERIAL NUMBER: ____________________ Staple your purchase receipt here Y our new digital blood pressure monitor uses the oscillometric method of blood press ...

  • Omron Healthcare HEM-711AC - page 4

    T o assure the corr ect use of the pr oduct basic safety measur es should always be f ollowed inc luding the warnings and cautions listed in this instruction manual. W ARNING Indicates a potentially hazardous situation which, if not av oided, could result in death or serious injury . SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL CAUTION Indicates ...

  • Omron Healthcare HEM-711AC - page 5

    5 OPERA TING THE DEVICE (continued) Read the Special Conditions section (page 27) of the instruction manual if your systolic pressure is known to be more than 220 mmHg. Inflating to a higher pressure than necessary may result in bruising where the cuff is applied. Operate the device only as intended. Do not use the de vice for any other purpose. Di ...

  • Omron Healthcare HEM-711AC - page 6

    CARE AND MAINTENANCE Do not subject the monitor to strong shocks such as dropping the unit on the floor . Do not submerge the de vice or any of the components in water . Store the device and the components in a clean, safe location. Changes or modification not appro ved by Omron Healthcare will void the user w arranty . Do not disassemble or attemp ...

  • Omron Healthcare HEM-711AC - page 7

    Operating Instructions 7 BEFORE T AKING A MEASUREMENT T o ensure a reliable reading follo w these recommendations: 1. A void eating, drinking alcohol, smoking, e xercising, and bathing for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement. 2. Stress raises blood pressure. A void taking measurements d ...

  • Omron Healthcare HEM-711AC - page 8

    8 Air Jack Main Unit: Batter y Compartment ST AR T/STOP Button MEMORY Button Date/Time Setting Button Display AC Adapter Jack Components: Air T ube Air Plug (Connects to Air Jack) Arm Cuf f Storage Case AC Adapter KNOW Y OUR UNIT INSTRUCTION MANUAL ESP ESPAÑOL ENGLISH Automatic Blood Pressure Monitor with Arm Cuff Model HEM-711AC Instruction Manua ...

  • Omron Healthcare HEM-711AC - page 9

    9 UNIT DISPLA Y Diastolic Blood Pressure Systolic Blood Pr essure Memory Symbol Pulse Display Date/Time Display Low Battery Indicator Av erage V alue Symbol Heartbeat Symbol/ Hypertension Indicator Deflation Symbol Irregular Heartbeat Symbol Movement Error S ymbol 280239_HEM-7011-E_S.qxd 2/19/08 3:00 PM Page 9 ...

  • Omron Healthcare HEM-711AC - page 10

    IRREGULAR HEAR TBEA T SYMBOL ( ) When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the measurement values. An irregular heartbeat rhythm is def ined as a rhythm that is more than 25% slower or 25% faster than the a verage rhythm detected while th ...

  • Omron Healthcare HEM-711AC - page 11

    11 The American Heart Association recommends the following guideline for upper limit of normal for home blood pressure readings: AHA Home Guideline for Upper Limit of Normal BP Systolic Blood Pressure 135 mmHg Diastolic Blood Pressure 85 mmHg This is a general guideline because blood pressure depends on age and morbidity . The AHA recommends a lo w ...

  • Omron Healthcare HEM-711AC - page 12

    HEAR TBEA T SYMBOL/HYPER TENSION INDICA TOR ( ) T AKING A MEASUREMENT The Heartbeat Symbol flashes on the display during the measurement. The Heartbeat Symbol flashes at ev ery heartbeat. When the measurement is complete, the Heartbeat Symbol flashes on the display with your blood pressure and pulse rate if the reading is abov e 135 for the Systoli ...

  • Omron Healthcare HEM-711AC - page 13

    13 1. Press the ▼ indicator on the battery cov er and slide the cov er off in the direction of the arrow . 2. Install 4 “ AA” size batteries so the + (positiv e) and - (negativ e) polarities match the polarities of the battery compartment as indicated. 3. Replace the battery cov er . BA T TER Y INST ALLA TION Battery Cover 280239_HEM-7011-E_S ...

  • Omron Healthcare HEM-711AC - page 14

    BA TTER Y REPLACEMENT When the Low Battery Indicator appears on the display screen turn the monitor off and remo ve all the batteries. Replace with four new batteries at the same time. Longlife alkaline batteries are recommended. CAUTION If battery fluid should get in your eyes, immediately rinse with plenty of clean water . Contact a physician imm ...

  • Omron Healthcare HEM-711AC - page 15

    15 NO TES: • Use only the authorized Omron A C Adapter that came with this monitor . • W e recommend you install the batteries ev en if the A C Adapter is used. If no batteries are installed, you may need to reset the date and time if the A C Adapter is disconnected. The measurement results will not be deleted. CAUTION Use only Omron authorized ...

  • Omron Healthcare HEM-711AC - page 16

    TO CONNECT THE AC ADAPTER Insert the A C Adapter Plug into the A C Adapter Jack on the back side of the monitor as illustrated. Plug the A C Adapter into a 120V A C outlet (60 cycles). TO DISCONNECT THE AC ADAPTER Hold and pull the Housing to remov e the AC Adapter from the electrical outlet. Do not remov e by pulling on the cord. Remov e the AC Ad ...

  • Omron Healthcare HEM-711AC - page 17

    17 Set the monitor to the current date and time before taking a measurement for the first time. If the date and time are not set correctly , the measurement values stored in the memory and the a verage values may not be correct. The Blood Pressure Monitor automatically stores up to 60 individual measurement values with the dates and times. The moni ...

  • Omron Healthcare HEM-711AC - page 18

    SET TING THE DA TE AND TIME 2 . SETTING THE YEAR The year can be set between 2006 and 2030. When the display reaches 2030, it will return to 2006. Press the MEMOR Y button to adv ance by increments of one year . Press the Date/T ime Setting button to set the current year . The month flashes on the display . NO TE: Press and hold the MEMOR Y button ...

  • Omron Healthcare HEM-711AC - page 19

    19 4. SETTING THE D A Y Press the MEMOR Y button to adv ance by increments of one day . Press the Date/T ime Setting button to set the current day . The hour flashes on the display . 5. SETTING THE HOUR The time is set using AM or PM. Press the MEMOR Y button to adv ance by increments of one hour . Press the Date/T ime Setting button to set the cur ...

  • Omron Healthcare HEM-711AC - page 20

    RESETTING THE DA TE AND TIME T o adjust the date or time after it has been setup when you f irst use the unit: Start with the monitor in the off position. 1. Press and hold the Date/T ime Setting button. The year flashes on the display . 2. Follo w the instruction from step 2 on page 18. NO TE : If you need to change the Date and T ime, always adju ...

  • Omron Healthcare HEM-711AC - page 21

    21 CAUTION Read all of the information in the instruction manual and any other literature in the box before operating the unit. CAUTION This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on inf ants or persons who cannot express their intentions. APPL YING THE CUFF ON THE LEFT ...

  • Omron Healthcare HEM-711AC - page 22

    3. Apply the cuff to your left upper arm so the blue arro w is centered on the inside of your arm and aligned with your middle finger . The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2" abov e your elbow . 4. Wrap the cuff f irmly in place around your arm using the closure strip. APPL YING THE A ...

  • Omron Healthcare HEM-711AC - page 23

    23 APPL YING THE ARM CUFF APPL YING THE CUFF ON THE RIGHT ARM When taking a measurement using the right arm use this instruction for Step 4 on page 22. Apply the cuff to your right upper arm so the blue arro w and the air tube runs along the inside of your elbow . The bottom of the cuff should be approximately 1/2'' abov e your elbow . NO ...

  • Omron Healthcare HEM-711AC - page 24

    T AKING A MEASUREMENT 24 The monitor automatically determines your ideal inflation lev el. In rare circumstances when a higher inflation may be necessary , the monitor automatically re-inflates the cuff up to 30 mmHg higher than the initial inflation and restarts the measurement. CAUTION Operate the device only as intended. Do not use the de vice f ...

  • Omron Healthcare HEM-711AC - page 25

    2. Press the ST AR T/STOP button. All symbols appear on the display . The cuff starts to inflate automatically . As the cuff inflates, the monitor automatically determines your ideal inflation lev el. This monitor detects the pulse during inflation. Do not mov e your arm and remain still until the entire measurement process is completed. NO TE : T ...

  • Omron Healthcare HEM-711AC - page 26

    26 T AKING A MEASUREMENT 4. When the measurement is complete, the arm cuff completely deflates. Y our blood pressure and pulse rate are displayed. 5. Press the ST AR T/STOP button to turn the monitor off. NO TE : The monitor will automatically turn off after f ive minutes. When the measurement is complete, the Heartbeat Symbol flashes on the displa ...

  • Omron Healthcare HEM-711AC - page 27

    27 T AKING A MEASUREMENT SPECIAL CONDITIONS CAUTION Inflating to a higher pressure than necessary may result in bruising where the cuff is applied. If your systolic pressure is known to be more than 220 mmHg, press and hold the ST AR T/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure. NO TES: • Do n ...

  • Omron Healthcare HEM-711AC - page 28

    28 The monitor is designed to store the blood pressure and the pulse rate in the memory ev ery time a measurement is completed. The monitor automatically stores up to 60 sets of measurement values (blood pressure and pulse rate). When 60 sets of measurement values are stored, the oldest record is deleted to sav e the most recent values. A VERAGING ...

  • Omron Healthcare HEM-711AC - page 29

    USING THE MEMORY FUNCTION 29 2. Press the MEMOR Y button to display the most recent set of measurement values on the screen. The values are displayed from the most recent to the oldest. NO TES: • The date and time are alternately displayed with the measurement values. • Press the MEMOR Y b utton to display the next set of v alues. • Press and ...

  • Omron Healthcare HEM-711AC - page 30

    Care and Maintenanc e 30 USING THE MEMORY FUNCTION TO DELETE ALL V ALUES STORED IN THE MEMOR Y Y ou cannot partially delete values stored in the memory . All v alues will be deleted. 1. First press the MEMOR Y b utton. While holding it do wn, press the ST AR T/STOP button together for more than 2 seconds. All v alues will be deleted. 2. Press the S ...

  • Omron Healthcare HEM-711AC - page 31

    T o keep your digital blood pr essure monitor in the best condition and pr otect the unit from dama ge, follo w the directions listed belo w: Keep the monitor in the stor age case when not in use. Make sure the A C Adapter is placed under the main unit so that is does not damage the display . A void kinking or sharply bending the A C Adapter cord. ...

  • Omron Healthcare HEM-711AC - page 32

    32 CARE AND MAINTENANCE Remo ve the batteries if the unit will not be used for three months or longer . Always replace all the batteries with new ones at the same time. Use the unit consistent with the instruction pr o vided in this manual. CAUTION Changes or modification not appro ved by Omron Healthcare will void the user w arranty . Do not disas ...

  • Omron Healthcare HEM-711AC - page 33

    SYMBOL CA USE C ORRECTION Cuff under -inflated. Cuff not applied correctly . Remov e the arm cuff. Read, “T aking a Measurement”. Reapply the arm cuff and tak e another measurement. Cuff o ver-inflated T ake another measurement and remain still until the measurement is complete. Monitor could not detect pulse wa ve. Air plug is not connected. B ...

  • Omron Healthcare HEM-711AC - page 34

    34 ERROR INDIC A T ORS AND TROUBLESHOOTING TIPS PROBLEM CA USES AND SOLUTIONS No power . No display appears on the unit. Replace all four batteries with new ones. Check the battery installation for proper placement of the battery polarities. Measurement values appear too high or too low . Blood pressure varies constantly . Many factors including st ...

  • Omron Healthcare HEM-711AC - page 35

    FC C ST A TEMENT Note: PO TENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. The product ...

  • Omron Healthcare HEM-711AC - page 36

    36 W ARRANT Y Y our HEM-711AC IntelliSense ® Automatic Blood Pressure Monitor , excluding the arm cuff, is w arranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase, when used in accordance with the instructions provided with the monitor . The arm cuf f is warranted to be free from defects i ...

  • Omron Healthcare HEM-711AC - page 37

    Model: HEM-71 1AC Display: LCD Digital Display Measurement Range: Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min. Accuracy/Calibration: Pressure: ±3mmHg or 2% of reading Pulse: ±5% of reading Inflation: Automatic by electric pump Deflation: Automatic pressure release valv e Rapid Air Release: Automatic exhaust v alve Pressure Detection: Capacitiv ...

  • Omron Healthcare HEM-711AC - page 38

    Manufactured by OMR ON HEAL THCARE Made in China Distributed by: OMRON HEAL THCARE, INC. 1200 Lakeside Dri ve Bannockburn, Illinois 60015 www .omronhealthcare.com Copyright © 2008 Omron Healthcare, Inc. 280239_HEM-7011-E_S.qxd 2/19/08 3:00 PM Page 38 ...

Manufacturer Omron Healthcare Category Blood Pressure Monitor

Documents that we receive from a manufacturer of a Omron Healthcare HEM-711AC can be divided into several groups. They are, among others:
- Omron Healthcare technical drawings
- HEM-711AC manuals
- Omron Healthcare product data sheets
- information booklets
- or energy labels Omron Healthcare HEM-711AC
All of them are important, but the most important information from the point of view of use of the device are in the user manual Omron Healthcare HEM-711AC.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Omron Healthcare HEM-711AC, service manual, brief instructions and user manuals Omron Healthcare HEM-711AC. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Omron Healthcare HEM-711AC.

A complete manual for the device Omron Healthcare HEM-711AC, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Omron Healthcare HEM-711AC by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Omron Healthcare HEM-711AC.

A complete Omron Healthcare manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Omron Healthcare HEM-711AC - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Omron Healthcare HEM-711AC, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Omron Healthcare HEM-711AC, that we can find in the current document
3. Tips how to use the basic functions of the device Omron Healthcare HEM-711AC - which should help us in our first steps of using Omron Healthcare HEM-711AC
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Omron Healthcare HEM-711AC
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Omron Healthcare HEM-711AC in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Omron Healthcare HEM-711AC?

Use the form below

If you did not solve your problem by using a manual Omron Healthcare HEM-711AC, ask a question using the form below. If a user had a similar problem with Omron Healthcare HEM-711AC it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)