page 2 Manual Print History Manual Print History Manual Print History Manual Print History Manual Print History The pr int histor y shown below lists the printing dates of all revisions and addenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subsequent updates . Addenda, which are released bet ...
-
Teledyne HFM-60 - page 3
page 3 T T T T T able of Contents able of Contents able of Contents able of Contents able of Contents 1.0 1.0 1.0 1.0 1.0 GENERAL INFORMA GENERAL INFORMA GENERAL INFORMA GENERAL INFORMA GENERAL INFORMA TION TION TION TION TION ............................................................. ............................................................. ...
-
Teledyne HFM-60 - page 4
page 4 ...
-
Teledyne HFM-60 - page 5
page 5 General Information SECTION 1 The Hastings Mass Flowmeter, HFM-60 is designed to accurately measure mass flow up to 70 slpm with an accuracy of better than 3%. Hastings mass flow instruments do not require any periodic maintenance under normal operating conditions with clean gases. No damage will occur from the use of moderate overpressures ...
-
Teledyne HFM-60 - page 6
page 6 1.2 Specifications Accuracy ......................................................................................................................... <±3% F .S. Repeatability ................................................................................................. ±0.2% of F .S.(typical) ±0.4% of F .S. (max.) Maximum pressure . ...
-
Teledyne HFM-60 - page 7
page 7 1.3 Optional 4-20 mA Current Output An option to the standard 0-5 VDC output is the 4-20 mA current output that is propor tional to flow . The 4 - 20 mA signal is produced from the 0 - 5 VDC output of the flowmeter. The cur rent loop output is useful for remote applications where pickup noise could substantially affect the stability of the v ...
-
Teledyne HFM-60 - page 8
page 8 ...
-
Teledyne HFM-60 - page 9
page 9 Installation SECTION 2 This section contains the necessar y steps to assist in getting a new flowmeter in operation as quickly and easily as possible. Please read the following thoroughly before attempting to install the instru- ment. W W W W W ARNING ARNING ARNING ARNING ARNING The customer should determine if their process gas is compatibl ...
-
Teledyne HFM-60 - page 10
page 10 There is a threaded hole in the bottom of the transducer that can be used to secure it to a mount- ing bracket, if desired. Other holes for special mounting can be added to the end cap as desired. The standard inlet and outlet fittings for the HFM-60 are 1/4" NPT to 1/4" Swagelok (optional VCR and VCO fittings are av ailable from ...
-
Teledyne HFM-60 - page 11
page 11 2.6.4 Range Changes: The range of the flowmeter can be changed in the field if recalibra tion facilities are a vailable. The flowmeter ma y require a different shunt which can be purchased from the factor y . A listing of available shunts (order codes) and the corresponding air equivalent flow ranges are as follows: Order Code Shunt Range O ...
-
Teledyne HFM-60 - page 12
page 12 ...
-
Teledyne HFM-60 - page 13
page 13 Theory of Operation SECTION 3 This section contains an overall functional description of the HFM-60. Detailed schematics and parts lists can be found at the end of the manual in Section 5.0. In this section and other sections throughout this manual, it is assumed that the customer is using the wall transformer power supply supplied b y Hast ...
-
Teledyne HFM-60 - page 14
page 14 3.3 Base The 6061 aluminum alloy base has a 1" square cross-section and is 3.375" long (with end cap). The internal flow channel is 0.812" diameter with in and out 304L stainless steel fittings 1/4" NPT to 1/4" Swagelok fittings. 3. 4 Shunt The flo w rate of interest determines the size of the shunt required. As pre ...
-
Teledyne HFM-60 - page 15
page 15 Aluminum epoxy figure 3.3 (a) 10 sccm - 200 sccm (typical configuration) figure 3.3 (b) annular flow passages 500 sccm - 70 slm (typical configuration) flow flow ...
-
Teledyne HFM-60 - page 16
page 16 3.5 Electronic Circuitr y The HFM-60 employs a thermal flow sensor (capillar y tube descr ibed in section 3.2) to measure the flow which is proportional to the total flow through the instr ument. The sensor develops a differential voltage output signal proportional to flow which is approximately 1 mV full scale magnitude. A differential amp ...
-
Teledyne HFM-60 - page 17
page 17 4.1 Authorized Maintenance With proper care in installation and use, the flowmeter will require little or no maintenance. If maintenance does become necessar y , most of the instr ument can be cleaned or repaired in the field. Some procedures may require recalibration. Do not attempt these procedures unless facilities are av ailable. Entr y ...
-
Teledyne HFM-60 - page 18
page 18 4.3 Adjustments Calibration Procedure (See figure 4.1) 1. Connect pow er to pins 1 and 2 as specified in Section 2.5. Allow the instrument to warm up for 30 minutes with 10% flow . 2. Completely shut off the flow and wait for 2 minutes. For the standard 0-5VDC output, adjust the zero potentiometer located on the lower outlet side of the flo ...
-
Teledyne HFM-60 - page 19
page 19 4.4 End Cap Removal The end cap on the inlet side must be remov ed to gain access to the filter and shunt assembly . First remove power and shut off the supply of gas to the instrument. Disconnect the Swagelok fittings on the inlet and outlet sides of the transducer and remove it from the system plumbing. Remov e the two Allen head screws h ...
-
Teledyne HFM-60 - page 20
page 20 4.5 Printed Circuit Board Replacement In the event that any of the PC boards fail, they are easily removed from the instrument and replaced with a spare. This ease in disassembly and replacement substantially reduces instrument downtime. 1 . Replacement of the 4-20 mA option PC board: Unplug the power cable from the trans ducer. If a digita ...
-
Teledyne HFM-60 - page 21
page 21 W arranty and Repair SECTION 5 5.1 W ar ranty Repair P olic y Hastings Instruments warrants this product for a period of one year from the date of shipment to be free from defects in material and workmanship . This warranty does not apply to defects or failures resulting from unauthorized modification, misuse or mishandling of the product. ...
-
Teledyne HFM-60 - page 22
page 22 ...
-
Teledyne HFM-60 - page 23
page 23 Diagrams and Drawings SECTION 6 ...
-
Teledyne HFM-60 - page 24
page 24 ...
-
Teledyne HFM-60 - page 25
page 25 ...
-
Teledyne HFM-60 - page 26
page 26 ...
-
Teledyne HFM-60 - page 27
page 27 ...
-
Teledyne HFM-60 - page 28
page 28 ...
-
Teledyne HFM-60 - page 29
page 29 ...
-
Teledyne HFM-60 - page 30
page 30 ...