    MARATHON FA/FR Series 1-Color Fiber Optic Thermometer 2-Color Fiber Optic Thermometer    Operating Instructions Rev. G 01/2010 53001 ...

  • RayTek MARATHON FA - page 2

    ...

  • RayTek MARATHON FA - page 3

    Contacts   Worldwide  Headqu arters  Santa  Cruz,  CA  USA  Tel:  +1  800  227  –  8074  (USA  and  Canada  only)   +1  831  458  –  3900  Fax:  +1  831  458  –  1239  solutions@raytek.com   European  He adquarters  Berlin,  Germany  ...

  • RayTek MARATHON FA - page 4

    W ARRANTY  The  manufacturer  war rants  this  instrument  to  be  free  from  defects  in  materi al  and  workmanship  under  normal  use  and  service  for  the  period  of  two  years  from  date  of  pur chas e.  This  warranty  extends  only  ...

  • RayTek MARATHON FA - page 5

    T ABLE  OF  C ONTENTS  1  SAFETY  INSTRUCT IONS ................................................................................................................... ........... 1 2  PRODUCT  DESCRIPTI ON .................................................................................................................... ........ ...

  • RayTek MARATHON FA - page 6

    6  OPERATION ...................................................................................................................... ............................. 25 6.1  C ONTROL  P ANEL ....................................................................................................................................... 25 6.2  O PERAT ...

  • RayTek MARATHON FA - page 7

    10.4.2.1  Attaching  the  Fiber  Optic  Cable  to  the  Optical  Head .......................................................... 51 10.4.2.2  Attaching  the  Fiber  Optic  Cable  to  the  Electronics  Housing .............................................. 51 10.4.3  Fiber  Calibra tion ... ...

  • RayTek MARATHON FA - page 8

    ...

  • RayTek MARATHON FA - page 9

    Safety  Instructions Marathon  Series  FA/FR  1 1  Safety  Instructi ons  This  document  contains  important  information,  which  should  be  kept  at  all  times  with  the  instrument  during  its  operation al  life.  Other  users  of  this  instrument ? ...

  • RayTek MARATHON FA - page 10

    Product  Description 2  Marathon  Series  FA/FR 2  Product  Description  The  Marathon  FA/FR  fiber  optic  series  of  instruments  are  high ‐ performance  infrared  thermometers.  Each  has  a  front  end  consisting  of  a  small ,  fixed  focus  optica ...

  • RayTek MARATHON FA - page 11

    Product  Description Marathon  Series  FA/FR  3 Another  benefit  is  that  2 ‐ color  sensors  measu re  closer  to  the  highest  temperature  within  the  measured  spot  (spatial  peak  picki ng)  instead  of  an  average  temperature.  A  2 ‐ color  s ...

  • RayTek MARATHON FA - page 12

    Technical  Data 4  Marathon  Series  FA/FR 3  T echnical  Data  3.1  Measurement  Specifications  3.1.1  FA  Models  Temperature  Range  FA1A  475  to  900°C  (887°F  to  1652°F)  FA1B  800  to  1900°C  (1472°F  to  3452°F)  FA1C  1200  to  3000°C ...

  • RayTek MARATHON FA - page 13

    Technical  Data Marathon  Series  FA/FR  5 System  Accuracy 1  no  signal  attenuation  ±(0.3%  T meas *  +  2°C )  up  to  95%  signal  attenuation  ±(1%  T meas *  +  2°C)  for  FR1A/FR1B  up  to  95%  signal  attenuation  ±(1.3%  T meas *  +  2° ...

  • RayTek MARATHON FA - page 14

    Technical  Data 6  Marathon  Series  FA/FR Warm  up  Period  15  minutes  Weight  Optical  Head  100  g  (3.5  oz)  Electronics  Housing  710  g  (9  oz)  3.3  Electrical  Specifications  Power  Supply  24  VDC  ±20%,  500  mA  (max  100  mV  p ...

  • RayTek MARATHON FA - page 15

    Technical  Data Marathon  Series  FA/FR  7 3.4  Dimensions   Figure  2:  Dimensions  of  Optic al  Head(FA  Models)   Figure  3:  Dimensions  of  Optic al  Head  (FR  Models)   Figure  4:  Dimensions  of  Elec tronics  Housing  Fiber Optic Cable min. bend radi ...

  • RayTek MARATHON FA - page 16

    Technical  Data 8  Marathon  Series  FA/FR  Figure  5:  Adjustable  Mou nting  Bracket  for  Optical  Head  ...

  • RayTek MARATHON FA - page 17

    Technical  Data Marathon  Series  FA/FR  9 3.5  Optical  Specifications  The  sensor  comes  as  a  standard  focus  model  or  one  of  two  close  focus  models,  see  following  overview  for  available  options.  For  one ‐ color  temperature  measure ...

  • RayTek MARATHON FA - page 18

    Technical  Data 10  Marathon  Series  FA/FR 3.5.1.2  Close  Focus   Figure  7:  Close  Focus  Spot  Size  Charts  for  FA  models  Spot ∅ S [mm]  Distance D to Object [mm] D:S = 20:1 at focus point FA1A/FA2A CF2 Spot ∅ S [mm]  Distance D to Object [mm] D:S = 20:1 at focus point FA1A/FA ...

  • RayTek MARATHON FA - page 19

    Technical  Data Marathon  Series  FA/FR  11 3.5.2  FR  Models  3.5.2.1  Standard  Focus   Figure  8:  Standard  Focus  Spot  Size  Charts  for  FR  models  Distance D to Object [mm] D:S = 40:1 at focus point Distance D to Object [mm] D:S = 20:1 at focus point Spot ∅ S [mm]  Spot ? ...

  • RayTek MARATHON FA - page 20

    Technical  Data 12  Marathon  Series  FA/FR 3.5.2.2  Close  Focus   Figure  9:  Close  Focus  Spot  Size  Charts  for  FR  models  3.6  Scope  of  Delivery  The  scope  of  delivery  includes  the  following:  • Marathon  FA/FR  Documentation  an d  ...

  • RayTek MARATHON FA - page 21

    Sensor  Location Marathon  Series  FA/FR  13 4  Sensor  Location  Sensor  location  and  configurat ion  depends  on  the  application.  Be fore  deciding  on  a  location,  you  need  to  be  aware  of  the  ambient  temperature  of  the  location,  the ...

  • RayTek MARATHON FA - page 22

    Sensor  Location 14  Marathon  Series  FA/FR 4.5  Sensor  Placement  (1 ‐ Color  Mode)  Optical  head  placement  for  one ‐ color  temperature  measurements  is  more  critical  than  two ‐ color  measurements.  The  sensor  must  have  a  clear  view  of ...

  • RayTek MARATHON FA - page 23

    Sensor  Location Marathon  Series  FA/FR  15  Figure  11:  Sensor  Placem ent  in  2 ‐ Color  Mod e  4.7  Viewing  Angles  The  optical  head  can  be  placed  at  any  angle  from  the  target  up  to  30°  for  one ‐ colo r  mode,  or  45° ...

  • RayTek MARATHON FA - page 24

    Sensor  Location 16  Marathon  Series  FA/FR  Figure  12:  Acceptable  Sensor  Viewing  Angles   Best 90° to targ et Good 1-Color Mode: 30° to 90° to target 2-Color Mode: 45° to 90° to target Bad 1-Color Mode: 0° to 30° to target 2-Color Mode: 0° to 45° to target A cceptable Angles ...

  • RayTek MARATHON FA - page 25

    Installation Marathon  Series  FA/FR  17 5  Installation  5.1  Mounting  the  Sensor  After  all  prepar ations  are  complete  according  to  section  4  Sensor  Location ,  pag e  13  ff.,  you  can  install  the  sensor.  How  and  where  you  anchor ...

  • RayTek MARATHON FA - page 26

    Installation 18  Marathon  Series  FA/FR 5.2  Aiming  An  effective  aiming  technique  is  to  adjust  the  head  until  the  highest  readin g  is  observed  on  the  internal  display.  When  the  highest  reading  is  reached,  hold  the  unit  in ...

  • RayTek MARATHON FA - page 27

    Installation Marathon  Series  FA/FR  19 5.4  Installing  the  Electronics  Housing  The  distance  between  the  electronics  housing  and  a  computer  (via  RS4 85  cable)  can  be  up  to  1200  m  (4000  feet).  This  allows  ample  distance  from ? ...

  • RayTek MARATHON FA - page 28

    Installation 20  Marathon  Series  FA/FR   Figure  15:  Electronics  Housing  Wiring   Incorrect  wiring  can  damage  the  sensor  and  void  the  warra nty!  Before  applying  power,  make  sure  all  connections  are  correc t  and  secure!   The ? ...

  • RayTek MARATHON FA - page 29

    Installation Marathon  Series  FA/FR  21 5.5  Power  Supply  Connections  from  a  24  VDC  (500  mA  or  higher)  power  supply  attach  to  the  appropriate  terminals  on  the  electronic  enclosure’s  terminal  strip.   Isolated  power  is  require ...

  • RayTek MARATHON FA - page 30

    Installation 22  Marathon  Series  FA/FR 2.  Install  all  electronics  wiring  according  to  the  applicable  sections  5.4 ,  5.7.x ,  and  5.8 !  3.  Plug  the  RS485/RS 232  interface  converter  into  your  computer’s  serial  port,  or  attach  it  ...

  • RayTek MARATHON FA - page 31

    Installation Marathon  Series  FA/FR  23  Figure  19:  Wiring  for  2 ‐ Wire  Sensor  Se tup  5.7.3  Connecting  to  Terminal  Block  If  you  need  to  extend  the  wiring  or  to  have  a  complete  wiring  of  all  inputs/outputs,  use  the  T ...

  • RayTek MARATHON FA - page 32

    Installation 24  Marathon  Series  FA/FR 5.8  Installing  of  Multiple  Sensors  in  a  Network  5.8.1  Wiring  For  an  installation  of  two  or  more  sensors  in  a  network,  each  sensor  cable  is  wired  to  its  own  terminal  block.  The ...

  • RayTek MARATHON FA - page 33

    Operation Marathon  Series  FA/FR  25 6  Operation  Once  you  have  the  optical  head  and  electronics  housing  positioned  and  connected  properly,  the  system  is  ready  for  continuous  operatio n.  The  operation  of  the  sensor  can  be  don ...

  • RayTek MARATHON FA - page 34

    Operation 26  Marathon  Series  FA/FR The  sensor  has  a  remote  locking  feature  that  keeps  the  unit  from  being  accident ally  changed  from  the  control  panel  (locked  by  default  in  multidrop  mode).  This  lockout  mode  denies  access ...

  • RayTek MARATHON FA - page 35

    Operation Marathon  Series  FA/FR  27  Figure  25:  Mode  Selector  Button  Sequence  (FR  Models)  6.2.1  Temperature  Display  The  temperature  display  can  be  set  for  either  °C  or  °F  by  pressing  the  C/F  selector  button  ( c  –  ...

  • RayTek MARATHON FA - page 36

    Operation 28  Marathon  Series  FA/FR  The  slope  is  the  deciding  par ameter  for  measur ements  in  2 ‐ color  mode!  The  emissivity  affects  only  m easurem ents  in  1 ‐ color  mode.  For  information  on  det ermining  an  unknown  slope,  and ? ...

  • RayTek MARATHON FA - page 37

    Operation Marathon  Series  FA/FR  29  Figure  26:  Averaging  Example  To  set  and  activate  Averaging,  do  the  following:  1.  Press  the  Mode  button  unt il  the  AVG  LED  is  lit.  2.  Press  the c  button  to  both  set  and  activat ...

  • RayTek MARATHON FA - page 38

    Operation 30  Marathon  Series  FA/FR Peak Hold with decay timer 000.1-299.9 000.0 000.0 -* 0001-9999 Advanced Peak Hold trigger or threshold 300.0 000.0 0250-3000 -* 0000 Advanced Peak Hold timer or threshold 000.1-299.9 000.0 0250-3000 -* 0000 Advanced Peak Hold with decay timer or threshold 000.1-299.9 000.0 0250-3000 -* 0001-9999 Valle ...

  • RayTek MARATHON FA - page 39

    Operation Marathon  Series  FA/FR  31  Figure  27:  Deadband  Example  6.2.11  Ambient  Background  Temperature  Compensation  (FA  Models)  The  FA  model  is  capable  of  improving  the  accuracy  of  target  temperature  measurements  by  taking  into  ...

  • RayTek MARATHON FA - page 40

    Operation 32  Marathon  Series  FA/FR 6.4  Factory  Defaults  To  globally  reset  the  unit  to  its  factory  default  settings,  press  the  c  and  d  buttons  at  the  same  time  for  approximately  2  seconds.  The  baud  rate  will  not  ...

  • RayTek MARATHON FA - page 41

    Options Marathon  Series  FA/FR  33 7  Options  Options  are  items  that  are  factory  installed  and  must  be  specified  at  time  of  order.  The  following  are  available :  • Fiber  optic  cable  lengt hs:  1,  3,  6,  10  m  (3,  10, ? ...

  • RayTek MARATHON FA - page 42

    Accessories 34  Marathon  Series  FA/FR 8  Accessories  8.1  Overview  A  full  range  of  accessori es  for  various  applications  and  indu strial  environments  are  avai lable.  Accessories  include  items  that  may  be  ordered  at  any  time  and  a ...

  • RayTek MARATHON FA - page 43

    Accessories Marathon  Series  FA/FR  35 8.2  Air  Purge  Collar  The  Air  Purge  Collar  accessory  is  used  to  keep  dust,  moisture,  airborne  particles,  and  vapors  away  from  the  optical  head ʹ s  lens.  It  can  be  installed  before  o ...

  • RayTek MARATHON FA - page 44

    Accessories 36  Marathon  Series  FA/FR 8.4  Fitting  System  Flexible  accessory  select ions  allow  you  to  pick  and  choose  the  acce ssories  you  need.   Figure  33:  Flexible  Fitting  System  Part number Description XXXFORFQP Item 1 XXXFORFAP Item 1 + Item 2 XXXFO ...

  • RayTek MARATHON FA - page 45

    Accessories Marathon  Series  FA/FR  37 8.5  RS232/485  Interface  Converter  The  RS232/48 5  interface  converters  have  built ‐ in  smart  switching  and  have  been  designed  to  be  fast,  allowing  for  use  in  either  2 ‐ wire  or  4 ‐ wire  mo ...

  • RayTek MARATHON FA - page 46

    Accessories 38  Marathon  Series  FA/FR 8.6  Industrial  Power  Supply  The  DIN ‐ rail  mount  industrial  power  supply  transforms  an  input  voltage  of  85  –  26 4  VAC  into  an  output  voltage  of  24  VDC  /  1.25  A.  The  power  su ...

  • RayTek MARATHON FA - page 47

    Programming  Guide Marathon  Series  FA/FR  39 9  Programming  Guide  This  section  explains  the  sensor’s  communication  protocol.  Use  them  when  writing  custom  programs  for  your  application s  or  when  communicating  with  you r  sensor  with  ...

  • RayTek MARATHON FA - page 48

    Programming  Guide 40  Marathon  Series  FA/FR Example:  Send  E=0.90  instead  of  E=0. 9;  send  P=001.2  instead  of  P=1.2   After  transmitting  one  command,  the  host  has  to  wait  for  the  response  from  the  unit  before  sending  anothe ...

  • RayTek MARATHON FA - page 49

    Programming  Guide Marathon  Series  FA/FR  41 • Sensor  Model  Type  • Sensor  Seri al  Number  • Relay  Control  • Laser  st atus  • Setpoints  • Deadband  • Current  Output  Mode  (0 ‐ 20  mA  or  4 ‐ 20  mA)   The  following  items  ...

  • RayTek MARATHON FA - page 50

    Programming  Guide 42  Marathon  Series  FA/FR 9.5  Command  List  In  depending  from  the  spec ific  commands,  the  following  char acters  are  used :  ?  ...  host  (e.g.  PC)  requests  for  a  parameter  value  of  the  unit  !  ...  unit  ...

  • RayTek MARATHON FA - page 51

    Programming  Guide Marathon  Series  FA/FR  43 Description Char Format (2) P (1) B (1) S (1) N (1) Legal Values Factory Default Wide Power Q nnnn.nnn √ √ 0000.000-9999. 999 Narrow Power (FR only) R nnnn.nnn √ √ 0000.000-9999. 999 Slope (FR only) S n.nnn √ √ √ √ 0.850-1.150 1.000 Target temperature FR series: 2-color T nn ...

  • RayTek MARATHON FA - page 52

    Programming  Guide 44  Marathon  Series  FA/FR 9.6  Command  Examples  HOST SENSOR HOST SENSOR WHERE USED (1) Description Query  Answer Set  Notification P B S N Burst string format 001?$ 001!$UTSI 001$=UTSI √ √ Show list of commands 001? √ Measured attenuation 001?B 001!B12 √ √ Baud rate 001!D384 001D=384 √ E ...

  • RayTek MARATHON FA - page 53

    Maintenance Marathon  Series  FA/FR  45 10  Maintenance  Our  sales  representatives  and  customer  service  are  alw ays  at  your  disposal  for  questions  regarding  application  assistance,  calibration,  repair,  and  solutions  to  specific  problems.  Pleas ...

  • RayTek MARATHON FA - page 54

    Maintenance 46  Marathon  Series  FA/FR 10.2  Fail ‐ Safe  Operation  The  Fail ‐ Safe  system  is  designed  to  alert  the  operator  and  provide  a  safe  output  in  case  of  any  system  failure.  Basically ,  it  is  designed  to  shutdown  t ...

  • RayTek MARATHON FA - page 55

    Maintenance Marathon  Series  FA/FR  47 The  relay  is  controlled  by  the  temperature  selected  on  the  display.  If  any  failsaf e  code  appears  on  the  display,  the  relay  changes  to  the  “abnormal”  state.  The  exception  is  the  ...

  • RayTek MARATHON FA - page 56

    Maintenance 48  Marathon  Series  FA/FR 10.3  Cleaning  the  Lens  Keep  the  lens  clean  at  all  times.  Any  foreign  matter  on  the  windo w  will  affect  1 ‐ color  measurement  accuracy  and  may  affect  two ‐ color  ac curacy .  However,  car ...

  • RayTek MARATHON FA - page 57

    Maintenance Marathon  Series  FA/FR  49 10.4  Replacing  the  Fiber  Optic  Cable   FA  fiber  cable  assemblies  are  not  field ʺ replaceab le ʺ without  blackbody  reca libration!  As  such,  spare  FA  fiber  cable  assembli es  are  no t  availabl e! ? ...

  • RayTek MARATHON FA - page 58

    Maintenance 50  Marathon  Series  FA/FR  Figure  36:  Removing  the  Fi ber  optic  Cable  from  the  Optical  Head  10.4.1.2  Removing  the  Fiber  Optic  Cable  from  the  Electronics  Hous ing  Complete  the  following  steps  to  disconnect  the  fib ...

  • RayTek MARATHON FA - page 59

    Maintenance Marathon  Series  FA/FR  51 10.4.2  Mounting  the  Fiber  Optic  Cable  10.4.2.1  Attaching  th e  Fiber  Optic  Cable  to  the  Optical  Head  Complete  the  following  steps  to  attach  the  fiber  optic  cable  to  the  optical  head : ? ...

  • RayTek MARATHON FA - page 60

    Maintenance 52  Marathon  Series  FA/FR  Figure  39:  Attaching  the  Fiber  Optic  Cable  to  the  Electronics  Housing  10.4.3  Fiber  Calibration  Each  replacement  fiber  optic  cable  is  calibrated  at  the  factor y  before  shipping.  The  calibr a ...

  • RayTek MARATHON FA - page 61

    Maintenance Marathon  Series  FA/FR  53  Figure  40:  Dialog  for  the  Calibra tion  of  the  Fiber  Cable  ...

  • RayTek MARATHON FA - page 62

    Appendix 54  Marathon  Series  FA/FR 11  Appendix  11.1  Determination  of  Emissivity  Emissivity  is  a  measure  of  an  object’s  ability  to  abso rb  and  emit  infrared  energ y.  It  can  hav e  a  value  between  0  and  1.0.  For  ex ample ...

  • RayTek MARATHON FA - page 63

    Appendix Marathon  Series  FA/FR  55 E MISSIVITY AT 1 µ M FOR M ETALS Aluminum Iron, cast unoxidized 0.1-0.2 oxidized 0.9 oxidized 0.4 unoxidized 0.35 roughened 0.2-0.8 molten 0.35 polished 0.1-0.2 Magnesium 0.3-0.8 Brass Molybdenum polished 0.35 oxidized 0.5-0.9 Burnished 0.65 unoxidized 0.25-0.35 Chromium 0.4 Monel (Ni-Cu) 0.3 Copper Ni ...

  • RayTek MARATHON FA - page 64

    Appendix 56  Marathon  Series  FA/FR 11.3  Typical  Slopes  The  following  slope  settings  are  approximate  and  will  va ry  depending  on  the  metal  alloy  and  surface  finish,  as  well  as  the  applic ation.  These  are  supplied  here  as  ...

  • RayTek MARATHON FA - page 65

    Appendix Marathon  Series  FA/FR  57 11.4  Signal  Reduction  (FR  Models)  The  following  figure s  show  each  sensor  model’s  typical  percentage  of  allowed  sign al  reduction  at  all  temperatures.  Refer  to  these  graphs  to  estimate  what  ...

  • RayTek MARATHON FA - page 66

    Appendix 58  Marathon  Series  FA/FR 11.5  Attenuation  Influence  on  Accuracy  The  ability  of  the  FR  ratio  instruments  to  accurately  measure  the  temperature  of  targets  smaller  than  the  field ‐ of ‐ view  (FOV)  is  a  key  feature.  A ...

  • RayTek MARATHON FA - page 67

    Appendix Marathon  Series  FA/FR  59 11.6  Traceability  of  Instrument  Calibration   C USTOMER I NFRARED T HE RMOMETERS Computer ized cavity (black body sourc e) 500 to 2727 ° C 400 to 800°C Medium Temp Se condary (Working) Standard 400 to 800°C Fluke Medium Temp Primary Standard 760 to 2700°C High Temp Primar y Sta ...

Производитель RayTek Категория Thermometer

Документы, которые мы получаем от производителя устройства RayTek MARATHON FA мы можем разделить на несколько групп. Это в частности:
- технические чертежи RayTek
- инструкции обслуживания MARATHON FA
- паспорта изделия RayTek
- информационные брошюры
- энергетические этикетки RayTek MARATHON FA
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания RayTek MARATHON FA.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа RayTek MARATHON FA, инструкции обслуживания, короткие инструкции или инструкции пользователя RayTek MARATHON FA. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия RayTek MARATHON FA.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства RayTek MARATHON FA, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании RayTek MARATHON FA пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей RayTek MARATHON FA.

Полная инструкция обслуживания RayTek, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией RayTek MARATHON FA - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания RayTek MARATHON FA, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся RayTek MARATHON FA, которое найдем в данном документе
3. Советы по использованию основных функций устройства RayTek MARATHON FA - которые должны облегчить нам первые шаги во время использования RayTek MARATHON FA
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем RayTek MARATHON FA
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра RayTek MARATHON FA в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся RayTek MARATHON FA?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с RayTek MARATHON FA, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с RayTek MARATHON FA со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)