Инструкция обслуживания Continental Scientific Refrigerators and Freezers

36 страниц 6.11 mb
Скачать

Перейти на страницу of 36

Summary
  • Continental Scientific Refrigerators and Freezers - page 1

    Operation, Installation and Instruction Manual for Reach-In Refrigerators and Freezers Cont inental Scient ific A Division of National Refrigeration & Air Conditioning Products, Inc. 539 Dunksferry Road · Bensalem, PA 19020-5908 215-244-1400 · 1-800-523-7138 · Fax: 215-244-9579 ...

  • Continental Scientific Refrigerators and Freezers - page 2

    Operators Manual Table of Contents Page RECEIVING YOUR NEW M ODEL........................................................................ 1 GENERAL INFORMATION AND IMPORTANT OPERAT ING FACTS ................. 1 UNCRATING YOUR NE W MODEL ...................................................................... 2 INSTALLATION AND LOCAT ION ............. ...

  • Continental Scientific Refrigerators and Freezers - page 3

    TROUBLESHOOTING AND SER VICING GUIDE .............................................. 22 W IRING DIAGRAMS .......................................................................................... 26 ...

  • Continental Scientific Refrigerators and Freezers - page 4

    RECEIVING YOUR NEW M ODEL Congratul ations on your r ecent purchase of Continental Scienti fic superior laboratory /pharmacy equipment products! W hen your shipment ar riv es, please thoroughly ex amine the shipping crate for any punctures, dents, or sig ns of roug h handling. It i s in y our best inter est to partiall y remov e or open the shippin ...

  • Continental Scientific Refrigerators and Freezers - page 5

    and sanitation, all models are li sted under the reexami nation ser vi ce of underw riter's l aboratories. UNCRA TING YOUR NEW MO DEL The shipping container shoul d remain on y our model as pr otection against dents or scratches whil e transporting it to the actual set- up location. Remov e the shipping container only at the last possible mome ...

  • Continental Scientific Refrigerators and Freezers - page 6

    efficiency. Important Note: To assure suf ficient air supply and circulation to the condensing unit, a minimum clearance of 12" above th e grill and 3" on each side and back of the cabinet must be prov ided (see figure 1) . If necessary, special v enting or air supply ducts must be installed by the instal ler for this purpose. Do not at a ...

  • Continental Scientific Refrigerators and Freezers - page 7

    approxi m ately 700 pounds of product ( 35 x 20) and assuming the refriger ator itself weig hs 300 pounds, the total combined wei ght of cabinet and product is approxi m ately 1000 pounds. Therefore, the floor in this ex ample must be capable of supporting up to 1000 pounds. INSTA LLING LEGS A ND LEVELING REA CH-IN M ODELS Your new reach-in model i ...

  • Continental Scientific Refrigerators and Freezers - page 8

    Your roll -in model comes with one cart r amp per door opening . T o install the ramp(s), simply open the cabinet door and positi on the ramp slots onto the screw s located along the cabi net bottom f ront. If the door wiper rubs along the ramp to the extreme where the door w ill not self -close , you will need to adjust the door (see "door ad ...

  • Continental Scientific Refrigerators and Freezers - page 9

    FIGURE 3 CA STERS MUST BE TIGHT TO CA BINE T BOTTOM INSTA LLI NG CONDENSA TE EVAPORA TOR No floor drains or plumbing connecti ons are req uired since al l models use an automatic condensate w ater ev aporating system. All models are supplied wi th an electric condensate evaporator pan and mounting bracket packed in the accessor y carton. To install ...

  • Continental Scientific Refrigerators and Freezers - page 10

    FIGURE 4 CONDENSA TE EVA PORA TOR INS TA LLATION REMO VA L OF DOORS A ND DOOR A DJUSTM ENT During installation, it may become necessary to remove the cabi net doors to facilitate passage through narrow doorways or hallw ays. To remove a door, pry off all hing e covers using a shar p tool or knife (see fig ure 5). Sw ing the door to the open-door po ...

  • Continental Scientific Refrigerators and Freezers - page 11

    FIGURE 5 DOOR A DJUSTME NT A ND REMOV A L REMO VA L OF GRILL To remove the grill, loosen all four (4) grill mounting screw s located on the backside of the grill one or two turns. Simply lift grill up off of its mo unting screw s and out. To replace the grill, line up the gr ill m ounting screws w ith the keyhole slots located on the cabinet body, ...

  • Continental Scientific Refrigerators and Freezers - page 12

    cleaning may result in a streaking or staining of the metal. Complete cleaning procedur es and precautions are l isted in the "per iodic cleani ng procedure" under the mai ntenance section. STA RT-UP PROCEDURE ELECTRICA L CONNECTI ONS To insure proper oper ation, your new model must be connected to an i ndivi dual circuit that can suppl y ...

  • Continental Scientific Refrigerators and Freezers - page 13

    the condensing unit is 208- 230 volts i n which the wi ring includes a neutral and a mechanical ground. This w iring should be connected to the appropr iate power source by a q ualified electrician and must conform to all local electr ical codes. SPECIA L VOLTA GE CONNECTIONS W hen models are ordered from the factory w ith special, optional v oltag ...

  • Continental Scientific Refrigerators and Freezers - page 14

    condensing unit, and electrical hook-up should be per f ormed by q ualified refriger ation personnel of a competent refrigeration company only. OPERA TION All cabinets must be g iven sufficient time to reach normal operating temper ature before placing any product inside. Refriger ators are desig ned to maintain an ideal cabinet temperature of 3.3` ...

  • Continental Scientific Refrigerators and Freezers - page 15

    and turn the needle to the desired temperature usi ng y our finger s. Replace the lens cover and recheck the calibrati on. DIGITAL THERMOMETER For a digital thermometer, note how many degrees the digital display needs to be increased or decreased. Pr ess twice directly on the myl ar display towards the rig ht center using y our finger. After pressi ...

  • Continental Scientific Refrigerators and Freezers - page 16

    FIGURE 6 THERMOMETER CA LIBRATION REFRIGERA TOR SYSTEM A ND ADJUSTMENT All self-contai ned refriger ators are desi gned and factory set to maintain an aver age cabinet temperature of 4`c. The temperatur e control is accessible f rom the top of the electrica l console bo x located on the cabinet to p behind the fron t grill (see figure 6). If an adj ...

  • Continental Scientific Refrigerators and Freezers - page 17

    compressor is not oper ating dur ing an off-cycle. Do not set the thermostat too cold where the cabine t tem perature will f all below 2`c becau se the evapo rator will becom e b locked by ice since th e comp ressor of f -cycle will be considerably shortened. This will result in loss of produc t stored within the cabinet and require servi ce to def ...

  • Continental Scientific Refrigerators and Freezers - page 18

    cabinet temperature. On l ow- temperature freez ers, place a screw driv er into the right ther mostat adjustment screw ( do not turn the left di fferent ial adjustment screw) and turn clockwise for a colder cabinet temperatur e or counterclockwise for a warmer cabinet temperature. Fur ther adjustments out of the factory desig n temperature rang e m ...

  • Continental Scientific Refrigerators and Freezers - page 19

    FIGURE 8 DEFROST TIMER SETTING EVA PORATOR A SSEMBLY All refrig erators and freezers have an easi ly accessible, easily serviceabl e, performance rated, f orced-air evaporator assembly w hich utilizes a plasticiz ed fin coil for ex tended life (see figur e 9). LOW-PROFILE, CEILIN G-MOUNT EVAPORATOR ASSEMBLY The low-pr of ile evapor ator system is c ...

  • Continental Scientific Refrigerators and Freezers - page 20

    FIGURE 9 LOW-PROFILE,CEILING- MOUNT EVA PORA TOR A SSEMBLY INTERIOR REA CH-IN A CCESSORIES The standard interior accessory packag e that is supplied from the f actory wi th y o ur reach-in r efrigerator or freezer consi sts of standard pil aster str ips w ith pilaster clips (four (4)clips per shelf), three ( 3 ) epoxy coated shelves per section, an ...

  • Continental Scientific Refrigerators and Freezers - page 21

    the shelf on the clips as show n in figure 10. The standard pil aster and clip ar e shown i n figure 10, and the optional heavy -duty pilaster and clip are shown in figure 11. Important Note: W hen loading shel ves with product, al low space between row s of product for proper air ci rculation, and do not load pr oduct to block back wall (leave at ...

  • Continental Scientific Refrigerators and Freezers - page 22

    FIGURE 11 OPTIONA L HEA VY-DUTY PILA STER FIGURE 12 OPTIONA L A NGLE PA N SLIDES SA FETY PRECA UTIONS The following saf ety precautions shoul d be followed when operati ng any appliances: ¬ Alway s disconnect the power cor d before attempting to w ork on or clean - 19 - ...

  • Continental Scientific Refrigerators and Freezers - page 23

    any eq uipment. ¬ Disconnect the power cord w hen the appliance will be idled for a long period of time. ¬ Do not attempt to serv ice this unit yourself as remov ing any cover s may cause exposure to dang erous voltage. ¬ Alway s route the power cor d so that it is not l ikely to be walked on or pinched by other appliances. Nev er use extension ...

  • Continental Scientific Refrigerators and Freezers - page 24

    contact with stai nless steel. 3. Tincture of iodine, or iron shoul d not come in contact with stainless steel. These solutions, whic h cause stainless steel to di scolor, should be ri nsed off immediately if contact occurs. 4. Gritty, hard abrasives will mar the f inish of stainless steel an d alumin um and are not recommended. SLIDING A ND HINGED ...

  • Continental Scientific Refrigerators and Freezers - page 25

    this cleaning process. Af ter cleaning , restore electrical ser vic e to your model . PA RTS AND SERVICE Al w a ys provide the cabi net model and serial number (l ocated on the data plate on the inside ri ght w all of the cabinet) w henever contacting the factory or your dealer regardi ng questi ons or when ordering parts. M O D E L # SERIA L # ___ ...

  • Continental Scientific Refrigerators and Freezers - page 26

    PROBLEM PROBABLE CA USE CORRECTION Condensing unit will not start - no hum. 1.LINE DISCONNECTED, SW ITCH OPEN. 2.FUSE REMOVED OR BLOW N. 3.OVERLOAD PROTECTOR BLOW N. 4.CONTROL "OFF" DUE TO COLD LOCATION. 5.CONTROL STUCK IN OPEN POSITION. 6.W I RING IMPROPER OR LOOSE. 1.CLOSE START OR DISCONNECT SW ITCH. 2.REPLACE FUSE. 3.DETERM INE REASON ...

  • Continental Scientific Refrigerators and Freezers - page 27

    PROBLEM PROBABLE CA USE CORRECTION Condensing uni t runs, but short cycl es on: 1.OVERLOAD PROTECTOR. 2.THERM OSTAT. 3.HIGH PRESSURE CUT-OUT DUE TO: (a) INSUFFICIENT AIR SUPPLY. (b) OVERCHARGE. (c) AIR IN SYSTEM . 4.LOW PRESSURE CUT-OUT DUE TO: (a) VALVE LEAK. (b) UNDERCHARGE. (c) RESTRICTION IN EXPANSION DEVICE. 1.SEE PROBLEM # 3. 2.DIFFERENTIAL M ...

  • Continental Scientific Refrigerators and Freezers - page 28

    PROBLEM PROBABLE CA USE CORRECTION Product zone temperature too hig h. 1.CONTROL SETTING TOO HIGH. 2.INADEQUATE AIR CIRCULATION. 3.DIRTY CONDENSER 1.ADJUST T-ST AT 2.REARRANGE PRODUCT LOAD TO IMPROVE AIR CIRCULATION. 3.CLEAN CONDENSER COIL Suction line frosted or sweating . 1.OVERCHARGE OF REFRIGERANT. 2.EVAPORATOR FAN NOT RUNNING. 3.EXPANSION VALV ...

  • Continental Scientific Refrigerators and Freezers - page 29

    WIRING DI A G RA M S - 26 - ...

  • Continental Scientific Refrigerators and Freezers - page 30

    - 27 - ...

  • Continental Scientific Refrigerators and Freezers - page 31

    - 28 - ...

  • Continental Scientific Refrigerators and Freezers - page 32

    - 29 - ...

  • Continental Scientific Refrigerators and Freezers - page 33

    - 30 - ...

  • Continental Scientific Refrigerators and Freezers - page 34

    - 31 - ...

  • Continental Scientific Refrigerators and Freezers - page 35

    - 32 - ...

  • Continental Scientific Refrigerators and Freezers - page 36

    - 33 - ...

Производитель Continental Scientific Категория Refrigerator

Документы, которые мы получаем от производителя устройства Continental Scientific Refrigerators and Freezers мы можем разделить на несколько групп. Это в частности:
- технические чертежи Continental Scientific
- инструкции обслуживания Refrigerators and Freezers
- паспорта изделия Continental Scientific
- информационные брошюры
- энергетические этикетки Continental Scientific Refrigerators and Freezers
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания Continental Scientific Refrigerators and Freezers.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа Continental Scientific Refrigerators and Freezers, инструкции обслуживания, короткие инструкции или инструкции пользователя Continental Scientific Refrigerators and Freezers. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия Continental Scientific Refrigerators and Freezers.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства Continental Scientific Refrigerators and Freezers, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании Continental Scientific Refrigerators and Freezers пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей Continental Scientific Refrigerators and Freezers.

Полная инструкция обслуживания Continental Scientific, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией Continental Scientific Refrigerators and Freezers - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания Continental Scientific Refrigerators and Freezers, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся Continental Scientific Refrigerators and Freezers, которое найдем в данном документе
3. Советы по использованию основных функций устройства Continental Scientific Refrigerators and Freezers - которые должны облегчить нам первые шаги во время использования Continental Scientific Refrigerators and Freezers
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем Continental Scientific Refrigerators and Freezers
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра Continental Scientific Refrigerators and Freezers в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся Continental Scientific Refrigerators and Freezers?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с Continental Scientific Refrigerators and Freezers, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с Continental Scientific Refrigerators and Freezers со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)