Summary
  • Invacare IRC5PO2AW - page 1

    Dealer: This manual MUST be given to the end user. User: BEFORE using this product, read this manual and save for future reference. Operator’s Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Perfecto 2 ™ Series Oxygen Concentrators without SensO 2 ® , Oxygen Concentrators with SensO 2 ...

  • Invacare IRC5PO2AW - page 2

    WARNING Perfecto 2 ™Series 2 Part No 1150713  WARNING DO  NOT  use  this  product  or  any  av ailable  optional  equipment  without  first  completely  reading  and  understanding  these  instructions  and  any  additional  instructional  material  such  as  owner ’ ...

  • Invacare IRC5PO2AW - page 3

    TABLE OF CONTENTS Part No 1150713 3 Perfecto 2 ™Series ACCESSORIES .................................................. 2 SPECIAL NOTES ............................................... 4 LABEL LOCATION ............................................ 6 SECTION 1—GENERAL GUIDELINES ..................... 7 Radio Frequency Interference .. ............... ...

  • Invacare IRC5PO2AW - page 4

    SPECIAL NOTES Perfecto 2 ™Series 4 Part No 1150713 SPECIAL NOTES Signal  w ords  are  used  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  following  table  for ? ...

  • Invacare IRC5PO2AW - page 5

    SPECIAL NOTES Part No 1150713 5 Perfecto 2 ™Series CAUTION “Caution:  Statutory  law  can  restrict  this  device  to  sal e  by  or  on  order  of  a  physician,  or  any  other  practitioner  licens ed  by  the  law  of  the  governmental  agency  in  which ...

  • Invacare IRC5PO2AW - page 6

    LABEL LOCATION Perfecto 2 ™Series 6 Part No 1150713 LABEL LOCATION Serial Number Lab el is located on the resonator intake assembly Specification Label is located on the back near the bottom Perfecto 2 ™ FOR OPERATING INSTRUCTIONS or ALARMS (See Operator’s Manual) DANGER - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room ...

  • Invacare IRC5PO2AW - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1150713 7 Perfecto 2 ™Series SECTION 1—GENERAL GUIDELINES In  order  to  ensure  the  safe  installation,  assembly  and  operation  of  the  Pe r f e c t o 2  concentrator  these  instructions  MUST  be  followed.  WARNING This  section  conta ...

  • Invacare IRC5PO2AW - page 8

    SECTION 1—GENERAL GUIDELINES Perfecto 2 ™Series 8 Part No 1150713 Operating Information For  optimum  performance,  Inv acare  recommends  that  each  concentrator  be  on  and  running  for  a  minimum  of  30  minutes  at  a  time.  Shorter  periods  of  operation  may ...

  • Invacare IRC5PO2AW - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1150713 9 Perfecto 2 ™Series Radio Frequency Interference This  equipmen t  has  been  tested  and  found  to  comply  with  EMC  limits  specified  by  IEC/EN  60601 ‐ 1 ‐ 2.  These  limits  are  designed  to  provide  a  reasonable ? ...

  • Invacare IRC5PO2AW - page 10

    SECTION 2—FEATURES Perfecto 2 ™Series 10 Part No 1150713 SECTION 2—FEATURES Oxygen Outlet Flowmeter Oxygen Purity Indicator Lights / Fault and Power Indicator Lights Power Switch Circuit Breaker Elapsed Time Meter Cabinet Filter REAR VIEW *HF Outlet Fitting Power Cord *NOTE:  This  outlet  fitting  is  to  be  used  on ...

  • Invacare IRC5PO2AW - page 11

    SECTION 3—HANDLING Part No 1150713 11 Perfecto 2 ™Serie s SECTION 3—HANDLING The  concentrator  should  AL W A YS  be  kept  in  the  upright  position  to  prevent  cabinet  damage  while  being  transported. If  the  concentrator  is  to  be  reshipped  by  common  c ...

  • Invacare IRC5PO2AW - page 12

    SECTION 4—TECHNICAL DESCRIPTION Perfecto 2 ™Series 12 Part No 1150713 SECTION 4—TECHNICAL DESCRIPTION The  Invacare  Per f e c t o 2  concentr ator  is  used  by  patients  with  respiratory  disorders  who  require  supplemental  oxygen.  The  device  is  not  intended  to  susta ...

  • Invacare IRC5PO2AW - page 13

    SECTION 5—TYPICAL PRODUCT PARAMETERS Part No 1150713 13 Perfecto 2 ™Serie s SECTION 5—TYPICAL PRODUCT PARAMETERS Alternating Current Type BF equipment Unit runnin g Unit not runnin g Attention - Consider Accompanying Documents DO NOT smoke Class II, Double Insulated Not Protected Electrical Requirements: 230 VAC ± 10% (253 VAC/20 7 VAC), 50 ...

  • Invacare IRC5PO2AW - page 14

    SECTION 5—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™Series 14 Part No 1150713 Flow Range: 0.5 to 5 L/min (maximum ). For flowrates less than 1 L/min, we recommend th e use of the Invacare Pediatric Flowmeter Accessory (IRCPF16AW). Low Flow Alarm: 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). Average Power Consumption ...

  • Invacare IRC5PO2AW - page 15

    SECTION 5—TYPICAL PRODUCT PARAMETERS Part No 1150713 15 Perfecto 2 ™Serie s No AP/APG Not suitable for use in the presence of a flammable anaesthetic mixture. Standards and Regulatory Listing: IRC5PO2AW/IRC5PAW IEC/EN61000-3-2 IEC/EN61000-3-3, IEC/EN 60601-1 , A1, A2 IEC/EN 60601-1 -2 IRC5PO2AW Models Only ISO8359 MDD 93/42/EEC, Annex I and IX ...

  • Invacare IRC5PO2AW - page 16

    SECTION 5—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™Series 16 Part No 1150713 IRC5PO2AW INDICATORS IRC5PAW INDI CATORS LABEL SYMBOL O 2 PURITY INDICATOR LIGHTS (LED) SYSTEM OKAY O 2 over 85% (± 2%) GREEN Indicator Light O 2 Between 73% (± 3%) to 85% (±2%) YELLOW Indicator light A. YELLOW Solid B. YELLOW Flashing Sensor Failure Call a qualif ied ...

  • Invacare IRC5PO2AW - page 17

    SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 17 Perfecto 2 ™Serie s SECTION 6— OPERATING INSTRUCTIONS Introduction Yo u r  oxygen  concentrator  is  intended  for  individual  use.  It  is  an  electronically  operated  device  that  separates  oxygen  from  room  air .  It  p ...

  • Invacare IRC5PO2AW - page 18

    SECTION 6—OPERATING INSTRUCTIONS Perfecto 2 ™Series 18 Part No 1150713 Yo u r  oxygen  concentrator  will  perform  best  when  operated  under  the  conditions  outlined  in  the  Ty p i c a l  Product  P arameters  on  page 13.  Usage  in  environments  other  than  those ...

  • Invacare IRC5PO2AW - page 19

    SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 19 Perfecto 2 ™Serie s 3. Insert  a  flathead  screwdriver  in  the  plate  groove  on  the  top  edge  of  the  filter  access  door  and  gently  pry  the  filter  access  door  off  (FIGURE  6.2). 4. Pull  up  ...

  • Invacare IRC5PO2AW - page 20

    SECTION 6—OPERATING INSTRUCTIONS Perfecto 2 ™Series 20 Part No 1150713 10. After  assembly ,  ensure  that  oxygen  is  flowing  through  the  cannula. FIGURE 6.3 Humidifier Compartmen t Power Switch NOTE:  For  this  pr ocedure,  refe r  to  FIGURE  6.4. 1. Press  power  switch  to  O ...

  • Invacare IRC5PO2AW - page 21

    SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 21 Perfecto 2 ™Serie s Flowrate NOTE:  For  this  procedure,  refe r  to  FIGURE  6.5. NOTE:  Ta k e  care  not  to  set  the  flow  above  RED  ring.  An  oxygen  flow  gr eater  than  5  L/min  will  decr ease  the ...

  • Invacare IRC5PO2AW - page 22

    SECTION 6—OPERATING INSTRUCTIONS Perfecto 2 ™Series 22 Part No 1150713 SensO 2 Oxygen Purity Indicator - Model IRC5PO2AW This  feature  monitors  the  purity  of  oxygen  generated  by  the  oxygen  concentrator .  If  purity  falls  below  factory  preset  standards,  indicator  light ...

  • Invacare IRC5PO2AW - page 23

    SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 23 Perfecto 2 ™Serie s FIGURE 6.6 Explanation of Oxygen Purity Indicator Lights - Model IRC5PO2AW Explanation of Indicator Lights - Model IRC5PAW NOTE:  For  this  procedure,  refe r  to  FIGURE  6.7. IRC5PAW Model Indicator Light Explanation RED  light  () ‐  T ...

  • Invacare IRC5PO2AW - page 24

    SECTION 7—MAINTENANCE Perfecto 2 ™Series 24 Part No 1150713 SECTION 7—MAINTENANCE  WARNING The  Invacare  concentrators  are  specifically  designed  to  minimize  routine  preventiv e  mainte nance.  Only  qualified  personnel  should  perform  preventiv e  maintenance  on  the  ...

  • Invacare IRC5PO2AW - page 25

    SECTION 7—MAINTENANCE Part No 1150713 25 Perfecto 2 ™Serie s 2. Clean  the  cabinet  filter  with  a  va c u u m  cleaner  or  wa s h  in  w arm  soapy  wa t e r  and  rinse  thoroughly . 3. Dry  the  filter  thoroughly  before  reinstallation. FIGURE 7.1 Cleaning the Cabinet Filter ...

  • Invacare IRC5PO2AW - page 26

    SECTION 7—MAINTENANCE Perfecto 2 ™Series 26 Part No 1150713 PREVENTIVE MAINTENANCE RECORD Model  No.  IRC  ______________  Serial  No.________________ On  Each  Inspection *NOTE:  Refer  to  Preventive  Ma intenance  section  of  Service  Manual. **NOTE:  Model  not  available  within  ...

  • Invacare IRC5PO2AW - page 27

    SECTION 8—TROUBLESHOOTING GUIDE Part No 1150713 27 Perfecto 2 ™Serie s SECTION 8—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: Short beeps, long pause 1. Power cord not plugged in. 1. Insert plug in to outlet. Concentrator not operating, power switch On. 2. No power at outlet. 2 . Inspect house circuit break ...

  • Invacare IRC5PO2AW - page 28

    SECTION 8—TROUBLESHOOTING GUIDE Perfecto 2 ™Series 28 Part No 1150713 YELLOW or RED light Illuminated. Alarm: Continuo us On RED light only. * Only applies to IRC5PO2AW models. 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tubing, cannula or humidifier.* 2. Inspect for kinks or blockages. Correct, clean or replace ite ...

  • Invacare IRC5PO2AW - page 29

    SECTION 9—OPTIONAL ACCESSORIES Part No 1150713 29 Perfecto 2 ™Serie s SECTION 9—OPTIONAL ACCESSORIES The  following  optional  accessories  are  also  av ailable:  • Standard  Adult  Nasal  Cannula,  2.1  m  (7  feet) ‐ M3120 • Standard  Humidifier ‐ M5120 • PreciseRX™  P ed ...

  • Invacare IRC5PO2AW - page 30

    RECYCLING INFORMATION Perfecto 2 ™Series 30 Part No 1150713 RECYCLING INFORMATION This  product  has  been  supplied  from  an  environmentally  aw are  manufacturer  that  complies  with  the  Wa s t e  Electrical  and  Electronic  Equipment  (WEEE)  Directive  2002/96/CE. This  p ...

  • Invacare IRC5PO2AW - page 31

    LIMITED WARRANTY Part No 1150713 31 Perfecto 2 ™Serie s LIMITED WARRANTY United  Kingdom:  Invacare  Ltd,  South  Road,  Bridgend  Industrial  Estate,  UK ‐ Bridge nd  CF31  3PY Te l :  (44)  (0)1656  664  321,  Fax:  (44)  (0)1656  667  532 UK@inv acare.com  Ireland: Inv acare  ...

  • Invacare IRC5PO2AW - page 32

    0120 Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 440-329-6000 800-333-6900 Technical Services 440-329-6593 800-832-4707 EU Representative Invacare International Sarl Route de Cité Ouest 2 1196 Gland Switzerland Tel: +41 22 354 60 10 Fax: +41 22 354 60 11 Part No 1150713 Invacare Corporation 2101 E. Lake M ...

Производитель Invacare Категория Oxygen Equipment

Документы, которые мы получаем от производителя устройства Invacare IRC5PO2AW мы можем разделить на несколько групп. Это в частности:
- технические чертежи Invacare
- инструкции обслуживания IRC5PO2AW
- паспорта изделия Invacare
- информационные брошюры
- энергетические этикетки Invacare IRC5PO2AW
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания Invacare IRC5PO2AW.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа Invacare IRC5PO2AW, инструкции обслуживания, короткие инструкции или инструкции пользователя Invacare IRC5PO2AW. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия Invacare IRC5PO2AW.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства Invacare IRC5PO2AW, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании Invacare IRC5PO2AW пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей Invacare IRC5PO2AW.

Полная инструкция обслуживания Invacare, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией Invacare IRC5PO2AW - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания Invacare IRC5PO2AW, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся Invacare IRC5PO2AW, которое найдем в данном документе
3. Советы по использованию основных функций устройства Invacare IRC5PO2AW - которые должны облегчить нам первые шаги во время использования Invacare IRC5PO2AW
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем Invacare IRC5PO2AW
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра Invacare IRC5PO2AW в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся Invacare IRC5PO2AW?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с Invacare IRC5PO2AW, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с Invacare IRC5PO2AW со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)