    S8 S eries II S8 A utoS et Spirit™ II S8 Au t o Sc o r e ™ I I POSITIVE AIRWA Y PRESSURE D EVICES User Guide English ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA ResMed (UK) Ltd (EU Authorized Representative) 96 Milton Park A ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 2

    Cont ents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Contraindications 1 A dv erse Ef f ects 1 P arts of the S8 Series II system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Masks 2 Humidifiers 2 R esScan Data Card 2 Using the S8 S eries II . . . . . . . . . . . . . . ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 3

    1 Introduction Introduction S8 AutoSet Spirit II The S8 AutoSet Spirit II self-adjusting sleep apnea syste m is indicated for the treatment of obstructive sleep apnoea (OSA) in adult patients. The S8 A utoSet Spirit II self- adjusting sleep apno ea system has two treatment modes: Auto Set and fixed-pressure CPAP. The S8 AutoSet Spirit II system is ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 4

    2 Parts of the S8 Series II system Your S8 Series II device comprises: • S8 Series II de vice • P o w er cord • Carr y bag • 2 m air t ubing . Optional components includ e: • Integrated humidifi er • F ront cap • 3 m air t ubing • DC-1 2 conv erter • Re sMed T ubing W rap. Masks A list of masks (and their settings) compat ible wit ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 5

    3 Using the S8 Series II Using the S8 Series II Setting Up 1 Connect the power cord to the socket at the rear of the S8 Series II device. 2 Plug the other end of the power cord into the power outlet. 3 Connect one en d of the air tub ing firmly onto the ai r outlet. 4 Connect the assembled mask system to the free end of air tubing. For information ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 6

    4 How to Use the Control Panel The control p anel of the S8 Series II include s an LCD screen which displays the menus and treatment screens as well as a ke ypad for navigating through the menus and delivering treatment. The S8 Series II keypad has the following keys: The keypad and LCD are equippe d with backlights that come on when the device is ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 7

    5 Using the S8 Series II 4 Press or until you see the setting that you require. 5 Press to ap ply. Using the Mask -fitting Feature You can use the S8 Series II mask-fitting featur e to help you fit your mask properly. This feature delivers air pressure fo r a three-mi nute period, prior to starting treatment, during which you can check and adjust y ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 8

    6 4 To start treatment, press . 5 Fit your mask as described in the mask user instructions. ! WARNING A mask shou ld not be used u nless the S8 Series II devi ce is powered and operating prope rly. Stopping Treatment To stop treatment at any time, re move your mask and pr ess . If your clinician has enabled SmartStart, si mply remove your mask and ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 9

    7 Changing the Settings on your S8 II Device The Detailed Menu Your clinician may have enabled a detailed menu on your S8 Series II device. The figure below shows the detailed menu in each mode. T o enter th e detailed menu from the sta ndby (RAMP or SETTLING) screen, press . Note: The AutoSet menu is avai lable on the S8 AutoSet Spi rit II only . ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 10

    8 Messages on the S8 Series II LCD Your clinician may have set your S8 Series II device to remind you about important events, such as when to replace your mask , when to insert your Data Card (if your device is Data Card enabled) and so on. The reminder message is displayed on the LCD and is visible if the device is not delivering therapy. The back ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 11

    9 Cleaning and Maintenance 3 Remove the Data Card by gripping the end of the Data Card an d pulling it out. 4 Store the Data Card in its protective folder wh en not in use. 5 Return the card in its protective folder to your clinician using a postal envelope. Updating Settings on your S8 device If your clinician has provided a Data Card with new dev ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 12

    10 ! WARNING Beware of electric shock. Do not immerse the device or power cord in water. Always unplug the de vice before cleaning and ma ke sure it is dry before reconnecting. ! CAUTION • Do not hang the air tubing in direct sunlight as it may harden over time and eventually crac k. • Do not use bleac h, c hlorine, alcohol, or aromati c-based ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 13

    11 Frequently Asked Questions ! CAUTION • Do not at tempt to open the S8 Se ries II case. There are no user serviceable par ts inside. • Inspection and re pair should only be performe d by an authorised ag ent. Under no circumstances sh ould you attempt to serv ice or repair the S8 Se ries II yourself . Frequently Asked Questions How will it fe ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 14

    12 Please consult the medical services dep artment of your carrier if you intend to use your S8 Series II on an aircraft. The S8 Seri es II complies with US FCC Part 15, Class B requirements. Note: Do not use your S8 Series II while the aircraft is ta king off or landing. Can I use my S 8 Series II if main s (AC) power is not ava ilable? A DC-12 co ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 15

    13 T roubleshooting Excessiv e le ak. Adjust posi tion of mask and headgear . Replace caps if missi ng from ports on mask. Air t ubing may not be c onnected properly . Connect firmly at both ends. Air t ubing may be kinked or punct ured. Straighten or replace. Displays er ror messag e: SYSTEM ERRxxxx Call Service! or Exxxx (wher e xxxx defines an e ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 16

    14 Technical Specifications . The manufacturer reserves the right to ch ange the se specification s without notice. The follo wing message is display e d on th e LCD af ter y ou tr y t o update set tings or copy data t o the Data Card : Settings Invalid Remove Card The identification details on the Data Card do not m atc h the identifica tion detai ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 17

    15 General W arnings and Cautions Symbols which appear on the device At tention, consult accompany ing documents; Follow instructions fo r use; Drip proof; T ype CF equipment; Dangerous voltage; Class II equipment; Start/S top; Manufacturer; Environmental infor mation WEEE 20 02/96/EC is a European Directive that requires the proper disposal of ele ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 18

    16 • If o xy gen is used with this de vi ce, the o xygen flow must be t urned of f when the device is not operating . If the o xy gen has been lef t on, turn of f the device, then w ait 30 minutes before t urning on the device again. Explanation: When the P AP device is not in operation and the o xygen flo w is lef t on, oxy gen deliv ered into t ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 19

    17 Limited W arranty Limited Warranty ResMed warrants that your ResMed product shall be free from defe cts in material and workmanship for the period sp ecified below from the date of purchas e by the initi al consume r. This warranty is not transferable. Not e : Some models are not avai lable in all regions. If the product fails under conditions o ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 20

    18 ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 21

    19 Limited W arranty ...

  • ResMed Positive AirwAy Pressure Device S8 Series II - page 22

    20 ...

Производитель ResMed Категория Respiratory Product

Документы, которые мы получаем от производителя устройства ResMed Positive AirwAy Pressure Device S8 Series II мы можем разделить на несколько групп. Это в частности:
- технические чертежи ResMed
- инструкции обслуживания Positive AirwAy Pressure Device S8 Series II
- паспорта изделия ResMed
- информационные брошюры
- энергетические этикетки ResMed Positive AirwAy Pressure Device S8 Series II
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания ResMed Positive AirwAy Pressure Device S8 Series II.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа ResMed Positive AirwAy Pressure Device S8 Series II, инструкции обслуживания, короткие инструкции или инструкции пользователя ResMed Positive AirwAy Pressure Device S8 Series II. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия ResMed Positive AirwAy Pressure Device S8 Series II.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства ResMed Positive AirwAy Pressure Device S8 Series II, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании ResMed Positive AirwAy Pressure Device S8 Series II пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей ResMed Positive AirwAy Pressure Device S8 Series II.

Полная инструкция обслуживания ResMed, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией ResMed Positive AirwAy Pressure Device S8 Series II - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания ResMed Positive AirwAy Pressure Device S8 Series II, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся ResMed Positive AirwAy Pressure Device S8 Series II, которое найдем в данном документе
3. Советы по использованию основных функций устройства ResMed Positive AirwAy Pressure Device S8 Series II - которые должны облегчить нам первые шаги во время использования ResMed Positive AirwAy Pressure Device S8 Series II
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем ResMed Positive AirwAy Pressure Device S8 Series II
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра ResMed Positive AirwAy Pressure Device S8 Series II в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся ResMed Positive AirwAy Pressure Device S8 Series II?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с ResMed Positive AirwAy Pressure Device S8 Series II, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с ResMed Positive AirwAy Pressure Device S8 Series II со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)