Инструкция обслуживания Omega Engineering HH505

15 страниц 0.3 mb
Скачать

Перейти на страницу of 15

Summary
  • Omega Engineering HH505 - page 1

    OMEGA HH505 Digital Thermometer ...

  • Omega Engineering HH505 - page 2

    OMEGAnet SM On-Line Service http://www.omega.com Internet e-mail info@omega.com Servicing Nor th America: USA: ISO 9001 Certified Canada: One Omega Drive, Box 4047 976 Bergar Stamford, CT 06907-0047 Laval (Quebec) H7L5A1 Tel: (203) 359-1660 Tel: (514) 856-6928 FAX: (203)359-7700 FAX: (514) 856-6886 e-mail: info@omega.com e-mail: info@omega.com omeg ...

  • Omega Engineering HH505 - page 3

    1 The symbol on the instrument indicates that the operator must refer to an explanation in this manual. WARNING To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC. WARNING To avoid damage or burns, do not make temperature measurement in microwave ovens. INTRODUCTION This instrume ...

  • Omega Engineering HH505 - page 4

    2 SPECIFICATIONS ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable. Measurement Range: Pt385(100 W ) -200°C to 800°C, (-328°F to 1472°F) Pt3926(100 W ) -200°C to 630°C, (-328°F to 1166°F) Resolution: 0.1°C or 0.2°F Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F ...

  • Omega Engineering HH505 - page 5

    3 ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 4½ digit liquid crystal display (LCD) with maximum reading of 19999. Overload: "----.-" is display. Battery: Standard 9V battery. Battery Life: 100 hours typical. Auto ...

  • Omega Engineering HH505 - page 6

    OPERATING INSTRUCTIONS Power Switch The key turns the thermometer on or off. In the data SET mode, can not power off, must leave data SET mode then power off. °C/°F Selecting the Temperature Scale (Main display) Reading the main displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set to th ...

  • Omega Engineering HH505 - page 7

    5 5 6 Pt385/Pt3926 T1 Input RTD Probe Select The Pt385/Pt3926 key switch the T1 input selects the Pt385 or Pt3926 RTD probe as input. When the thermometer is turned on, it is set to the probe selected that was in use when the thermometer was last turned off. MIN MAX with Time record Mode Press MIN MAX key to enter the MIN MAX Recording mode, (displ ...

  • Omega Engineering HH505 - page 8

    6 REL Relative mode Pressing REL key to enter the Relative mode, zero the display, and store the displayed Reading as a reference value and annunciator REL is displayed. Present temperature reading displayed on second display. Press REL key again to exit the relative mode. The relative value can also be entered by the user. (See "SET mode" ...

  • Omega Engineering HH505 - page 9

    7 9 SET mode (Relative value set, Time set and Hi/Lo Limits value set) 1. Press SET key to enter Relative valuse SET mode (Press ENTER key can escape relative valuse set mode), REL set mode. = = = =.= is displayed in main display. Relative value is entered via overlay numbers, then press overlay ENTER key, stored the relative value, enter Time set ...

  • Omega Engineering HH505 - page 10

    8 10 11 12 Hi Limit value Recall display mode Press Hi RECALL key to display the Hi set Limit value on second display. Press Hi RECALL key again to exit this mode. Lo Limit value Recall display mode Press Lo RECALL key to display the Hi set Limit value on second display. Press Lo RECALL key again to exit this mode. Hi/Lo Limits mode (only Main disp ...

  • Omega Engineering HH505 - page 11

    9 WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the rmometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of th ...

  • Omega Engineering HH505 - page 12

    10 TEMPERATURE VS RESISTANCE TABLE(ITS90) °C Pt385 Pt3926 -200°C 18.521 W 16.996 W -100°C 60.256 W 59.479 W 0°C 100.000 W 100.000 W 100°C 138.505 W 139.272 W 200°C 175.856 W 177.362 W 300°C 212.052 W 214.275 W 400°C 247.092 W 250.018 W 500°C 280.977 W 284.591 W 600°C 313.708 W 317.994 W 700°C 345.280 W - 800°C 375.700 W - ...

  • Omega Engineering HH505 - page 13

    1 HH 504 / HH 505 CA LIBRA TION PROCEDURE Note: The following calibration procedure s hould perform only by qualified technicians who have access to the items as fol lowing. Equipment: The class of calibrator ha d better 10 times g reater than the measur ed meter. 1. T urn of f the meter then set the adjusted jumper to J1. 2. Insert the 383.00 Ω c ...

  • Omega Engineering HH505 - page 14

    WARRANTY / DISCLAIMER OMEGA ENGINEERING, INC . warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA W arranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA's customers ...

  • Omega Engineering HH505 - page 15

    Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE þ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies þ Wire: Thermocouple, RTD & Thermistor þ Calibrators & Ice Point References þ Recorders, Controllers & Process Monitors þ Infrared Pyrometers PRESSURE ...

Производитель Omega Engineering Категория Thermometer

Документы, которые мы получаем от производителя устройства Omega Engineering HH505 мы можем разделить на несколько групп. Это в частности:
- технические чертежи Omega Engineering
- инструкции обслуживания HH505
- паспорта изделия Omega Engineering
- информационные брошюры
- энергетические этикетки Omega Engineering HH505
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания Omega Engineering HH505.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа Omega Engineering HH505, инструкции обслуживания, короткие инструкции или инструкции пользователя Omega Engineering HH505. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия Omega Engineering HH505.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства Omega Engineering HH505, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании Omega Engineering HH505 пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей Omega Engineering HH505.

Полная инструкция обслуживания Omega Engineering, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией Omega Engineering HH505 - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания Omega Engineering HH505, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся Omega Engineering HH505, которое найдем в данном документе
3. Советы по использованию основных функций устройства Omega Engineering HH505 - которые должны облегчить нам первые шаги во время использования Omega Engineering HH505
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем Omega Engineering HH505
5. FAQ - чаще всего задаваемые вопросы