  • Teledyne OT-2 - page 1

    i Teledyne Analytical Instruments Model OT -2 System OPERA TING INSTRUCTIONS FOR Model O T -2 Oxygen T ransmitter System P/N: M71265 ECO: #99-0000 10/21/1999 HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM. PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM. ONLY AUTHORIZED PERSONNEL SHO ...

  • Teledyne OT-2 - page 2

    Teledyne Analytical Instruments ii Model OT -2 System Copyright © 1999 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be elec ...

  • Teledyne OT-2 - page 3

    iii Teledyne Analytical Instruments Model OT -2 System T a ble of Contents 1 Introduction 1.1 Ove r view ........................................................................ 1- 1 1.2 F eatures .......................................................................... 1- 1 1.3 T ypical Applications ............................................... ...

  • Teledyne OT-2 - page 4

    Teledyne Analytical Instruments iv Model OT -2 System Appendix Specif ications ........................................................................ A-1 Recommended Spare Parts List .......................................... A-2 Drawing List ............................................................................ A-2 Mater ial Safety Data Sh ...

  • Teledyne OT-2 - page 5

    1-1 Teledyne Analytical Instruments Model OT-2 System Introduction 1 Introduction 1.1 Overview The Teledyne Analytical Instruments Model OT-2 System is designed to accurately monitor the oxygen content in a wide variety of gases at the ppm level. The Trace Oxygen Transmitter is equipped with two oxygen analysis ranges, 0–10 ppm and 0– 100 ppm ( ...

  • Teledyne OT-2 - page 6

    Teledyne Analytical Instruments 1 Introduction Model OT-2 System 1-2 • Accurate (±2% of full scale at constant temperature on 100 ppm, or 1000 ppm scale; ±1 ppm on 0 – 10 ppm scale at constant temperature) • Insensitive to flow variations • Fast response and recovery • Long life, maintenance-free Micro-Fuel Cell oxygen sensor • Unaffe ...

  • Teledyne OT-2 - page 7

    1-3 Teledyne Analytical Instruments Model OT-2 System Introduction 1 1.4 Operator Interface The OT-2 System is housed in a rugged metal NEMA-4 case with the display visible from the front and the controls accessible from inside the front door. See Figure 1-1. The front door has a clear Lexan window for viewing the sample system components and trans ...

  • Teledyne OT-2 - page 8

    Teledyne Analytical Instruments 1 Introduction Model OT-2 System 1-4 ...

  • Teledyne OT-2 - page 9

    2-1 Teledyne Analytical Instruments Model OT-2 System Operational Theory 2 2. 1 Introduction The OT-2 System is composed of three subsystems: 1 . Micro-Fuel Cell Sensor 2 . Electronic Signal Processing, Display and Control 3 . Sample System The Micro-Fuel Cell is an electrochemical galvanic device that translates the amount of oxygen present in the ...

  • Teledyne OT-2 - page 10

    Teledyne Analytical Instruments 2-2 2 Operational Theory Model OT-2 System 2. 2 Micro-Fuel Cell Sensor 2.2.1 Principles of Operation The oxygen sensor used in the Model OT-2 System is a Micro- fuel Cell designed and manufactured by Analytical Instruments. It is a sealed plastic disposable electrochemical transducer. The key components of the Micro- ...

  • Teledyne OT-2 - page 11

    2-3 Teledyne Analytical Instruments Model OT-2 System Operational Theory 2 Refer to Figure 2-2, Cross Section of a Micro-Fuel Cell , which illustrates the following internal description. Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale) At the top end of the cell is a diffusion membrane of Teflon. Beneath the diffusion membrane lies the ...

  • Teledyne OT-2 - page 12

    Teledyne Analytical Instruments 2-4 2 Operational Theory Model OT-2 System Warning:The sensor used in the Model OT-2 System uses electro- lytes which contain substances that are harmful if touched, swallowed, or inhaled. Avoid contact with ANY fluid or powder in or around the unit. What may appear to be plain water could contain one of these harmfu ...

  • Teledyne OT-2 - page 13

    2-5 Teledyne Analytical Instruments Model OT-2 System Operational Theory 2 2.2.4 The Effect of Pressure The Micro-Fuel Cell responds to the partial pressure of oxygen present inside the cell holder. By design, the total gas pressure inside the cell holder is kept approximately the same as the atmospheric pressure. As the atmospheric pressure change ...

  • Teledyne OT-2 - page 14

    Teledyne Analytical Instruments 2-6 2 Operational Theory Model OT-2 System 2.2.5 Calibration Characteristics Given that the total pressure of the sample gas on the surface of the Micro-Fuel Cell input is constant, a convenient characteris- tic of the cell is that the current produced in an external circuit is directly proportional to the rate at wh ...

  • Teledyne OT-2 - page 15

    2-7 Teledyne Analytical Instruments Model OT-2 System Operational Theory 2 2. 3 Electronics and Signal Processing The OT-2 circuitry consists of three stages: an input stage, a temperature compensation/gain stage and an output stage. The input stage consists of a current to voltage converter producing a voltage signal that is proportional to the ox ...

  • Teledyne OT-2 - page 16

    Teledyne Analytical Instruments 2-8 2 Operational Theory Model OT-2 System 2. 6 Sample System The system consists of a shut off valve, particulate filter, pres- sure regulator, scrubber, calibration selector valve, flow control valve, O 2 transmitter and a sample flow meter and incorporates ¼ inch tube fittings for sample inlet and outlet connecti ...

  • Teledyne OT-2 - page 17

    2-9 Teledyne Analytical Instruments Model OT-2 System Operational Theory 2 Figure 2-5 Piping Layout Figure 2-5 is the Piping Layout of the sampling system. In the standard instrument, calibration gases (zero and span) are con- nected directly to the Span Gas In port by teeing to the port with appropriate valves. The sample gas is connected to the S ...

  • Teledyne OT-2 - page 18

    Teledyne Analytical Instruments 2-10 2 Operational Theory Model OT-2 System ...

  • Teledyne OT-2 - page 19

    Model OT-2 System Operation 3 3-1 Teledyne Analytical Instruments Operation Operation of the Model OT-2 System involves installing the unit, making appropriate sample gas connections, and calibration. 3.1 Installation WARNING: Safe operation of the system requires the user to properly ground the system, install the wiring and power the transmitter. ...

  • Teledyne OT-2 - page 20

    3 Operation Model OT-2 System 3-2 Teledyne Analytical Instruments 3.1.2 Mo unting the Analyzer The OT-2 System is designed for bulkhead mounting. Figure 3-1 is an illustration of the OT-2 System and mounting hole dimensions. There are four mounting holes — one in each of the inside corners of the rigid enclosure. Figure 3-1 Mounting Details ...

  • Teledyne OT-2 - page 21

    Model OT-2 System Operation 3 3-3 Teledyne Analytical Instruments All operator controls are located inside the system enclosure, which is hinged on the left edge and provides easy access to the transmitter, sensor and cell block inside the instrument. The main enclosure door will swing open when the slotted latch at the right center of the door is ...

  • Teledyne OT-2 - page 22

    3 Operation Model OT-2 System 3-4 Teledyne Analytical Instruments To connect gas lines to the Model OT-2 System: 1 . Insert the tube into the tube fitting, and finger-tighten the nut until the tubing cannot be rotated freely, by hand, in the fitting. (This may require an additional 1/8 turn beyond finger-tight.) 2 . Hold the fitting body steady wit ...

  • Teledyne OT-2 - page 23

    Model OT-2 System Operation 3 3-5 Teledyne Analytical Instruments Figure 3-3 Front View of Transmitter System Power Requirements: The Model OT-2 System requires a 12 VDC (nominal), reverse polarity protected power source. Observe good wiring practices. For hazardous area applications, the Model OT-2 System must be pow- ered by a source that is cons ...

  • Teledyne OT-2 - page 24

    3 Operation Model OT-2 System 3-6 Teledyne Analytical Instruments 3 . 2 Installing the Micro-Fuel Cell The Micro-Fuel Cell is not installed in the cell block when the instrument is shipped. Install it before the analyzer is placed in service. When the micro-Fuel Cell needs to be installed or replaced, follow the procedures in chapter 5, Maintenance ...

  • Teledyne OT-2 - page 25

    Model OT-2 System Operation 3 3-7 Teledyne Analytical Instruments The span gas cylinder must be equipped with a pressure regulator with a metallic diaphragm. The regulator should be set to approximately 3-5 psig. Rotate the selector valve (a) to the span position and observe the flow meter (b) which should indicate a flow of about 0.4 scfh. For cal ...

  • Teledyne OT-2 - page 26

    3 Operation Model OT-2 System 3-8 Teledyne Analytical Instruments NOTE: This procedure is more critical on the 0-10 ppm range and less critical on the 0-100 or 0-1000 ppm range. 3.3.2 Setting the Span Pot 1. Switch the system from zero gas to span gas and allow time for stabilization. Figure 3-6: Span Pot Location NOTE: The span gas should have an ...

  • Teledyne OT-2 - page 27

    Model OT-2 System Operation 3 3-9 Teledyne Analytical Instruments 3.3.3 H 2 S Scrubber The life expectancy of the stainless steel H 2 S scrubber, in- cluded in this system, is determined by the amount of H 2 S passing through it. With the maximum concentration of 200 PPM H 2 S and a flow rate of 0.2 scfh the expected scrubber life will be over 6 mo ...

  • Teledyne OT-2 - page 28

    3 Operation Model OT-2 System 3-10 Teledyne Analytical Instruments ...

  • Teledyne OT-2 - page 29

    4-1 Teledyne Analytical Instruments Model OT-2 System Maintenance 4 Maintenance 4. 1 Routine Maintenance Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance includes servicing the filter, the scrubber, installing replacement Micro-Fuel cells and recalibration. For recalibration, see section 3.3 Calibration ...

  • Teledyne OT-2 - page 30

    Teledyne Analytical Instruments 4-2 4 Maintenance 4 Maintenance 4 Maintenance 4 Maintenance 4 Maintenance Model OT-2 System To replace the cell: Figure 4-1 1 . Open the enclosure door. 2 . Undo the velcro straps. 3 . Remove the insulation. 4 . The cell block is located in the center of the unit. Unscrew the cap on the bottom of the cell block. The ...

  • Teledyne OT-2 - page 31

    4-3 Teledyne Analytical Instruments Model OT-2 System Maintenance 4 4 . Place the new cell on the cap with the sensing surface face down and the gold contact rings up. 5 . Position the cell and cap so that the cell is inside the cell block. Press up on the cap and rotate. Tighten the cap screws. 4.2.1 Ce l l Warranty The cell warranty is sensor spe ...

  • Teledyne OT-2 - page 32

    Teledyne Analytical Instruments 4-4 4 Maintenance 4 Maintenance 4 Maintenance 4 Maintenance 4 Maintenance Model OT-2 System ...

  • Teledyne OT-2 - page 33

    A-1 Teledyne Analytical Instruments Model OT-2 System Appendix Appendix Specifications: System Enclosure: NEMA 4 Rated, bulkhead mounted Power Requirements: 9-16 VDC and 16-275mA. Ranges: 0 – 10ppm and 0 – 100ppm Oxygen or 0-100 ppm and 0-1000 ppm O 2 . Accuracy: ± 2% of full scale at constant temperature and pressure (temperature and pressure ...

  • Teledyne OT-2 - page 34

    Teledyne Analytical Instruments A-2 Appendix Appendix Appendix Appendix Appendix Model OT-2 System Recommended Spare Parts List Qty. Part Number Description 1 C71345A & C71345B Cell Block Insulation Set 1 C70976B PC Board — 1-5V Transmitter (0-100/0-1000 ppm) 1 C70976A PC Board — 1-5V Transmitter (0-10/0-100 ppm) 1 B71343 Scrubber Assembly ...

  • Teledyne OT-2 - page 35

    A-3 Teledyne Analytical Instruments Model OT-2 System Appendix Material Safety Data Sheet Material Safety Data Sheet Material Safety Data Sheet Material Safety Data Sheet Material Safety Data Sheet Section II – Physical and Chemical Data Section I – Product Identification Product Name: Micro-Fuel Cells Mini-Micro-Fuel Cells, all classes Super C ...

  • Teledyne OT-2 - page 36

    Teledyne Analytical Instruments A-4 Appendix Appendix Appendix Appendix Appendix Model OT-2 System Section III – Physical Hazards Primary route of entry: Ingestion, eye/skin contact Exposure limits: OSHA PEL: .05 mg./cu.m. (Pb) ACGIH TLV: 2 mg./cu.m. (KOH) Effects of overexposure Ingestion: The electrolyte could be harmful or fatal if swallowed. ...

  • Teledyne OT-2 - page 37

    A-5 Teledyne Analytical Instruments Model OT-2 System Appendix NO TE : The oxygen sensors are sealed, and under normal circumstances, the contents of the sensors do not present a health hazard. The following information is given as a guide in the event that a cell leaks. Protective clothing: Rubber gloves, chemical splash goggles. Cleanup procedure ...

  • Teledyne OT-2 - page 38

    Teledyne Analytical Instruments A-6 Appendix Appendix Appendix Appendix Appendix Model OT-2 System ...

Produzent Teledyne Kategorie Oxygen Equipment

Dokumente, die wir vom Produzenten des Geräts Teledyne OT-2 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Teledyne
- Bedienungsanleitungen OT-2
- Produktkarten Teledyne
- Informationsbroschüren
- oder Energieetiketten Teledyne OT-2
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Teledyne OT-2.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Teledyne OT-2, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Teledyne OT-2. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Teledyne OT-2 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Teledyne OT-2, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Teledyne OT-2 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Teledyne OT-2 verständlich ist.

Eine gänzliche Bedienungsanleitung von Teledyne sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Teledyne OT-2 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Teledyne OT-2 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Teledyne OT-2, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Teledyne OT-2 - die uns die ersten Schritte während der Nutzung von Teledyne OT-2 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Teledyne OT-2 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Teledyne OT-2 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Teledyne OT-2?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Teledyne OT-2 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Teledyne OT-2 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)