Bedienungsanleitung Teledyne 3110

52 Seiten 2.42 mb
Download

Zur Seite of 52

Summary
  • Teledyne 3110 - page 1

    Teledyne Analytical Instruments OPER A TIN G INSTRUCTIONS FOR Model 3 1 10 Portable T r ace Oxygen Analyzer DANGER To xic gase s a nd o r fl am mabl e li quids may b e pres ent i n t h i s mon ito r i ng sy ste m . Personal protective e q u ipment may be require d when ser vicin g this inst rument. Hazardou s volt a ges exist on certain components ...

  • Teledyne 3110 - page 2

    Model 3110 ii Teledyne Analytical Instruments All Right s Reserved . No par t of this m anual may be r epro duced, tra nsmitted, transcribed , stored in a retriev al system, or translated into any o th er lan guage o r computer language in whole or in p art, in any f orm or by an y means, whe ther it be e lectronic, m echanical, magn etic, optical, ...

  • Teledyne 3110 - page 3

    Portable Trace Oxygen Analyzer Teledyne Analytical Instruments iii Speci fic Model Info rma t ion Instrument S erial Nu mber: _______________________ ...

  • Teledyne 3110 - page 4

    Model 3110 iv Teledyne Analytical Instruments Safety M e ssages Your safety and the saf e ty of others is ver y i mportant. We have provided ma ny i mport ant safety messages in this manual. Please read these me ss ages carefully. A safety message alerts you to potential hazards that could hurt you or others. Eac h saf e ty message is associated wi ...

  • Teledyne 3110 - page 5

    Portable Trace Oxygen Analyzer Teledyne Analytical Instruments v IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE B E HA VIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES . This manual provide s infor m at ion designed to guide you through the installation, cali br a t ion operation and m ...

  • Teledyne 3110 - page 6

    Model 3110 vi Teledyne Analytical Instruments T ab le of Contents List of Fig ures ............................................................................ viii Introduc ti on ................................................................................ 10 1.1 Introduction 10 1.2 Feat ures 11 1.3 Method of Ana l y si s 13 1.4 Micro-Fuel Cell ...

  • Teledyne 3110 - page 7

    Portable Trace Oxygen Analyzer Teledyne Analytical Instruments vii 3.2.3 DATE Screen 27 3.2.4 TIME Screen 28 3.2.5 STANDARD ALARM Screen 29 3.2.6 ALARM Enable 29 3.2.7 FS ALARM S c ree n 30 3.2.8 HI-LOW ALARM Screen 30 3.2.9 LTCH ALARM Screen 31 3.2.10 RANGE 1 S c ree n 32 3.2.11 RANGE 2 S c ree n 32 3.2.12 Range Sc r een 33 3.2.13 SPAN VALUE S c r ...

  • Teledyne 3110 - page 8

    Model 3110 viii Teledyne Analytical Instruments List of Figures Fig ure 2-1 : Model 31 10 Rear Pane l .............................................. 18 Fig ure 3-1 : Front Pa n e l Ke ys ........................................................ 25 Figure 3-2 : Available Menus a n d Their Sequence ...... ....... ....... .... 26 Figure 4-1 : Battery C ...

  • Teledyne 3110 - page 9

    Portable Trace Oxygen Analyzer Teledyne Analytical Instruments ix DANGER COMBUSTIBLE GAS USA G E WARNI NG This i s an intrinsically sa fe instrument whi c h can be used in hazardous areas. It is th e c ustomer's responsibil ity t o ensure safety e specially when combustib le gas es are being analyzed since the potential of g as lea ks always e ...

  • Teledyne 3110 - page 10

    Introdu c tion Mo del 311 0 10 Teledyne Analytical Instruments Introduction 1.1 Introduction The Teledyne Analytical Instruments (TAI) Model 3110 Portable Trace Oxyg en Anal yz er is a portable, intrinsicall y s afe oxygen analyzer capable of analyzing oxygen levels from 2 parts per m i l lion (ppm) oxygen. The instrument ope r a t es from internal ...

  • Teledyne 3110 - page 11

    Portable Trace Oxygen Analyzer Introdu c tion Teledyne Analytical Instruments 11 WARNING Though the 3110 can be applied to monitor percent oxygen, doing so will lead t o m ore fr equent sensor replacement requirements. Figure 1-1: Model 3110 P ortable Trac e Oxygen Analyzer 1.2 Features This instrume n t is designed to be a versatile analytical ins ...

  • Teledyne 3110 - page 12

    Introdu c tion Mo del 311 0 12 Teledyne Analytical Instruments downloaded using the i ns trument’s optional RS -232 interface. • Display: A 2-line 20 cha r a c ter alphanumeric LCD on the front pa n e l displ ays data and operati on a l information through v arious screens. The contrast is adjustable for various lighting conditions. • Four-Bu ...

  • Teledyne 3110 - page 13

    Portable Trace Oxygen Analyzer Introdu c tion Teledyne Analytical Instruments 13 • Quick Disconnect Fitti n gs: Dual self-sealing quic k d i s connect fitti ngs are installed for making easy sample c onn e ctions . 1.3 Met hod of Analys is The sample oxygen i s m e asured by a unique electrochemical transducer whic h fun c t ions as a fuel cell; ...

  • Teledyne 3110 - page 14

    Introdu c tion Mo del 311 0 14 Teledyne Analytical Instruments If you have a warranty claim, ret urn the c ell in question to the factory for e v a lu ation. If it is determined that failure i s due to faulty workmanship or material, the cell will be replaced at no cost to you. WARNING: EVIDENCE OF DAMAGE D U E TO TAM P ERING OR MISHANDLING WILL RE ...

  • Teledyne 3110 - page 15

    Portable Trace Oxygen Analyzer Introdu c tion Teledyne Analytical Instruments 15 affect m eter a ccuracy, the Model 3110 can be used in any position without interfere n c e. ...

  • Teledyne 3110 - page 16

    Introdu c tion Mo del 311 0 16 Teledyne Analytical Instruments Intenti on a lly left blank. ...

  • Teledyne 3110 - page 17

    Portable Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 17 Installation The Model 3110 T r a c e Oxygen Analyzer is designed to be portable and easy to set up a nd configure. To se tup the analyzer: • Read the Manu a l • Charge the b a tt ery • Install t h e s ample and vent gas lines • Install t h e tra ce sensor and pu ...

  • Teledyne 3110 - page 18

    Installation Model 3110 18 Teledyne Analytical Instruments The unit c an operat e continuously for approximately 4 days on a set of fully cha rg e d batter ies. If more frequent charging is required, the batte r i es ar e approaching the end of their useful life and should be replaced. See Battery Replacement in Section 4.3 of this ma nu a l. A low ...

  • Teledyne 3110 - page 19

    Portable Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 19 and calibration gas. For atmospheric pressure sampling, connect a pump and flow control v a lve down stream from the analyzer and draw (rather than push) the sam p l e through the instrument. IMPORTANT: IF A PRESSURE REGULATOR IS USED, IT MUST HAVE A METALLIC DIAPHRAG M. ...

  • Teledyne 3110 - page 20

    Installation Model 3110 20 Teledyne Analytical Instruments For trace a n a lysis app lications, a flowmeter should be installed in the vent l ine bu t it should not incorporate any control valves or restrictive de v i c es. 2.3 Sensor Installati on The Model 3110's trace s ensor requires that the instrument lines be imme d i at ely purged with ...

  • Teledyne 3110 - page 21

    Portable Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 21 2.4 Calibration Calibration i nvo l ves us ing a span gas to set the span of t h e instrument . T he proper sp an gas concentration depe nds on the range that the instrument will be used on. T h e corr ect concentration should be 80- 90% of the range used. For inst a nc ...

  • Teledyne 3110 - page 22

    Installation Model 3110 22 Teledyne Analytical Instruments 2.4.2 Bleeding the Regulator and Purging the G a s Line When using bottled ga s (gas cyl ind er) as a calibration gas for trace analysis applications, the r e gul ator and sample lines must be bled to remove t r a c es of trapped air. Otherwise air that is trapped in the lines especially be ...

  • Teledyne 3110 - page 23

    Portable Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 23 2.6 External Signa l The standa rd 0-1 VD C output signal repres ents the pe rc entage of the curre n t rang e. For instance, if the range was set for 0 to 10 ppm, then 0.1 V would represent 1 ppm, 0.2V would be 2 ppm; 0.3V would be 3 ppm etc . This output signal, when i ...

  • Teledyne 3110 - page 24

    Operation Model 3110 24 Teledyne Analytical Instruments Operation The micro-fuel c e ll is not installed in the analyzer prior to shipment. It i s be a separate item that is to be installed according the procedure i n se c t ion 2.3 of this manual. Once the cell is installed and purged down with zero gas, the integral shut-off valve s in t he quick ...

  • Teledyne 3110 - page 25

    Portable Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 25 arrows on the left a nd right of the bo ttom line identify these screens as setup screens by changing from UP/DO WN to LEFT/RIGHT. The se t up s creens also blink. • Select a val u e —When multiple options or values exist for a function, pre ss ing ENTER selects the cur ...

  • Teledyne 3110 - page 26

    Operation Model 3110 26 Teledyne Analytical Instruments • Increme n t or de crement a value 3.2 Operation and Setup Screens and Menus The Model 3110 ope r a tion and setup fun ctions are arranged in a set of 15 menus. All but t h e POWER ON screen are access ible via the UP/DOWN keys. F igure 3-2 shows the available menus and the sequence of scre ...

  • Teledyne 3110 - page 27

    Portable Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 27 3.2.1 POWER ON Screen The POWER ON s creen automatically app ears on the di spl a y w hen the unit i s first powered up. The display appears briefly and shows the model num b e r and softwar e version. After a few seconds the di spl a y reverts to t h e HO ME screen. 3.2.2 ...

  • Teledyne 3110 - page 28

    Operation Model 3110 28 Teledyne Analytical Instruments 2. Use the UP/DOWN keys to alter t h e mon th. Then press ENTER. The c ursor w ill move over to the next editable field. 3. Use the UP/DOWN keys to alter t h e day. Then press ENTER. The c ursor w ill move over to the next editable field. 4. Use the UP/DOWN keys to alter t h e ye ar. Then pres ...

  • Teledyne 3110 - page 29

    Portable Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 29 At any time you ca n pr e ss the ESC key to abort the entry and return to the DATE screen. 3.2.5 STANDARD ALARM Screen Note: Alarm relays are not in cluded o n the standard version of the 3110 . During an ala r m a b linking “AL” will appear in the upper right corner of ...

  • Teledyne 3110 - page 30

    Operation Model 3110 30 Teledyne Analytical Instruments To enable or disable the alarm: 1. Use the UP/DOWN keys to navigate to the ALARM E N A BLE screen. The n p re ss ENTER to enter the alarm setup screen. 2. Use the UP/DOWN keys to toggle bet w een ENABLED or DISABLED. Then press ENTER to accept the di spl a yed status. 3. Press ENTER again t o ...

  • Teledyne 3110 - page 31

    Portable Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 31 alarm . A H I alarm activates when the oxygen concentration rises above the setpoi n t . The LO alarm activates when the oxygen concentration falls bel ow th e s etpoint. To set the a larm as e ither a HI or LO alarm: 1. Use the UP/DOWN keys to navigate to the HI-LO ALARM s ...

  • Teledyne 3110 - page 32

    Operation Model 3110 32 Teledyne Analytical Instruments To reset a l atched alarm that has triggered use the ALARM ENABLE screen and temporarily DISABLE the alarm. See Section 3.2.6. This will re s tore the alarm to it previous non-alarm condition. Note: A latched alarm can only b e reset w hen the concen tration falls or rises to a non-alarm value ...

  • Teledyne 3110 - page 33

    Portable Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 33 To define or c hang e range 1: 1. Use the UP/DOWN keys to navigate to the RANGE 2 screen. Then press ENTER t o e nt er the setup screen. 2. Use the UP /DOWN keys to change the initial value. Press ENTER to accept the displayed value and to return to the RANGE 2 scree n. Not ...

  • Teledyne 3110 - page 34

    Operation Model 3110 34 Teledyne Analytical Instruments 3.2.13 SPAN VALUE Scree n The SPA N V ALU E screen displays the oxygen concentration of the span gas used for cali br a t ion. This is not a measured value; it is the known span gas concentration that is input to the analyzer by the operator. To change t he span g as concentration: 1. Use the ...

  • Teledyne 3110 - page 35

    Portable Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 35 1. Use the UP/DOWN keys to navigate to the SPAN screen. Then press ENTER t o st a rt th e span calibration. Once the span calibration has begun the screen changes to reveal a span finish selection. 2. When the FIN ISH selection appears, press ENTER to end the span cali br a ...

  • Teledyne 3110 - page 36

    Operation Model 3110 36 Teledyne Analytical Instruments To us e t hi s fe a tur e: 1. Use the UP/DOWN keys to navigate to the FILTER scree n. Then press ENTER t o e nt er the setup screen. 2. Use the ENT/ON key to c h a nge the filter to the active mode. The arrows will poi n t left and r ight and the filter setting will blink. 3. Use the UP/DOWN k ...

  • Teledyne 3110 - page 37

    Portable Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 37 RESET me nu a gain. The data set can be downloaded using the LOG TRANSMIT screen (see Section 3.2.17). To start t h e dat a logger: 1. Use the UP/DOWN keys to navigate to the LOG RESET & START screen and press ENTER. 2. When the scr e en b egins to blink press ENTER aga ...

  • Teledyne 3110 - page 38

    Operation Model 3110 38 Teledyne Analytical Instruments 1 stop bit no flow control Refer to your c o m puter manu al for details on how to setup the optional RS -232 communications port on your compute r. To download a data l og : 1. Use the UP/DOWN keys to navigate to the LOG TRANSMIT screen. The n p re ss ENTER to enter the setup screen. 2. Press ...

  • Teledyne 3110 - page 39

    Portable Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 39 1. Use the UP/DOWN keys to navigate to the POWER DOWN screen. The n p re ss ENTER. 2. Press ENTER again t o t urn the instrument OF F. In a few seconds the analyzer will turn off. ...

  • Teledyne 3110 - page 40

    Maintenance Model 3110 40 Teledyne Analytical Instruments Maintenance & T r oubleshooting 4.1 Routine Mai ntenan ce Other than re p l a cing the sensor and cell holder O-ring, there are no user-serviceable components within the instrument housing. Routine maintenance consists of wiping down the instrument case, cleaning the screen and checking ...

  • Teledyne 3110 - page 41

    Portable Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 41 batte r i es. Refer to th e Spare Parts Listing in the Appendix for the correct replacement battery. Note: The batteries a re not user serviceable. They can only be repla ced by a qualified technic i an. 4.4 Battery Power Supp l y Service The Model 3110 i s designed to be ...

  • Teledyne 3110 - page 42

    Maintenance Model 3110 42 Teledyne Analytical Instruments Figure 4-1: Battery Charger Port on t h e Mode l 3110 4.5 Cell Replacement Th e ch ar act eri st ic s o f t he mic r o- fu el c ell ar e s im il ar to tho se of a N iC ad ba tte ry in th at b oth pr ov ide an al mo st con s ta nt ou tp ut th rou gh th eir us efu l l if e, an d the n fa ll of ...

  • Teledyne 3110 - page 43

    Portable Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 43 in such a way as to obviate any possibility of incurring damage. Under no circumstances should you d isturb the integrity of the cell package until the cell i s to be actually used . If the cell package is punctured a nd a ir p ermitted to enter, the cell will immediately ...

  • Teledyne 3110 - page 44

    Maintenance Model 3110 44 Teledyne Analytical Instruments cell p e rformance wi ll not be affected. On the following pa g e is a graph showing the effects of CO 2 on cell l ife. WARNING: EVIDENCE OF DAMAGE D U E TO TAM P ERING OR MISHANDLING WILL RENDER THE CELL WARRANTY NULL AND VO ID . 4.7 T e mperature Compensation Th e mic ro-f uel ce ll h as a ...

  • Teledyne 3110 - page 45

    Portable Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 45 Effect of CO 2 on B-2 Cell Life 1. This curve a ss umes continuous exposure. 2. Intermitte n t exposure wil l extend life. In general , t he CO 2 effect is cumulative and the ave r a ge CO 2 concentration should be used to find the pr e dicted cell life. 3. Cells ca n b e ...

  • Teledyne 3110 - page 46

    Maintenance Model 3110 46 Teledyne Analytical Instruments 4.8 Leak T es ting WARNING: IF A LEAK IS SUSPECTED I N THE UNIT, DO NOT ATTE M P T TO TIGHTEN THE QUICK DI SCONNECT FITTINGS . THE FITTINGS ARE POTTED I N EPOXY AND TIGHTENING T H EM WILL BREAK THE SEAL. To chec k for l e aks, TAI recommends one of the following procedures: Procedure I: 1. P ...

  • Teledyne 3110 - page 47

    Portable Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 47 Intenti on a lly left blank. ...

  • Teledyne 3110 - page 48

    Appendix Mo del 311 0 48 Teledyne Analytical Instruments Appendix A.1 Specifications TAI Sales Order Number: Instrument S erial Nu mber: Micr o-fue l Cel l Class: B-2, A-2C, B-2CXL, Insta Trans Ranges of An alysis: 2 user defined ra ng e s* Accuracy: ±2% of scale or ±1 PP M (whichever is greate r) a t constant temperature; ±5% of reading or ±1 ...

  • Teledyne 3110 - page 49

    Portable Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 49 A.2 Spare Parts List QTY P/N DESCRIPTION 1 O-165 Cell Ca p O -Ring (Std.) 2 B27296 Batte ry ( m us t be installed by a qualified technition). 1 A761 Batte ry Ch a rger, Univ ersal 1 CP2487 2-pin 0-1 VDC Output Plug 1 A36289 Calibration Kit S td. 2 F1378 Fuse 200 mA(T) 1 A654 ...

  • Teledyne 3110 - page 50

    Appendix Mo del 311 0 50 Teledyne Analytical Instruments Index A2C cell, 43 AC adapter. See charge ada pter AC charger port, 41 accuracy, 14, 48 alarm, 29 enable/disable, 29 failsafe, 29 HI/LOW, 29 latching, 29, 31 resetting, 32 alarm setpoint, 29 analysis range, 10, 48 arrow, 25 atmospheric pressure application, 19 auto-ranging, 12, 32 B2C cell, 4 ...

  • Teledyne 3110 - page 51

    Portable Trace Oxygen Analyzer I ndex Teledyne Analytical Instruments 51 housing, 10, 14 hydrocarbons, 13 impedance, 14, 23 inaccurate analysis, 19 incrementing a value, 26 instrument position, 15 intrinsically safe, 10, 14, 41 leak testing, 46 LED, 12 LO alarm, 31 LOG INTV screen, 36 LOG RE S E T & START screen, 36 LOG TRANSMIT screen, 37, 38 ...

  • Teledyne 3110 - page 52

    Appendix Mo del 311 0 52 Teledyne Analytical Instruments throttle valve, 18 time for mat , 28 TIME screen, 28 time stamping, 28 trace analysis application, 19 trapped air, 20, 21 tube fittings, 19 UP/DOWN key, 24, 25 user interface, 12 vent line, 19 , 22 , 46 vent line r es t riction , 19 warning sign, iv warranty, i i warranty claim, 14 website ad ...

Produzent Teledyne Kategorie Oxygen Equipment

Dokumente, die wir vom Produzenten des Geräts Teledyne 3110 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Teledyne
- Bedienungsanleitungen 3110
- Produktkarten Teledyne
- Informationsbroschüren
- oder Energieetiketten Teledyne 3110
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Teledyne 3110.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Teledyne 3110, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Teledyne 3110. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Teledyne 3110 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Teledyne 3110, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Teledyne 3110 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Teledyne 3110 verständlich ist.

Eine gänzliche Bedienungsanleitung von Teledyne sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Teledyne 3110 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Teledyne 3110 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Teledyne 3110, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Teledyne 3110 - die uns die ersten Schritte während der Nutzung von Teledyne 3110 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Teledyne 3110 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Teledyne 3110 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Teledyne 3110?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Teledyne 3110 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Teledyne 3110 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)