    Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL TECHNICAL INFORMATION When comparing two products features and specification using sales information it is often difficult to determine which pr oduct will meet the needs of your specific application. This is mostly because each supplier will define specifications using alternate methods ...

  • Fluke 3458/HFL - page 2

    DC Volt age ±( ppm of se tt ing + ppm of r ange) , r el ati ve t o cal i brat i on sta nd ar ds use d . Cali brat i on I nt e rv al 365 day , ±1°C of l ast TC AL R ange 1281 H P3458/HF L 8508A 120. 00000m V - 4 + 3 - 199. 999999m V 6 + 0. 5 - 2. 7 + 0.5 1. 20000000V - 3 + 0. 3 - 1. 99999999V 3 + 0.2 - 2. 7 + 0. 2 12. 000000V - 3 + 0. 05 - 19. 99 ...

  • Fluke 3458/HFL - page 3

    DC C urrent ±(ppm of setting + ppm of range), relative to calib ration standards used. Calibration Interval 365 day , ±1°C of last TCAL Range 1281 HP3458/HFL 8508A 120.000nA - 30 + 400 - 1.200000µA - 20 + 40 - 12.000000µA - 20 + 10 - 120.00000µA - 20 + 8 - 199.99999µA 25 + 2 - 6.5 + 2.0 1.2000000mA - 20 + 5 - 1.9999999mA 25 + 2 - 6.5 + 2.0 1 ...

  • Fluke 3458/HFL - page 4

    Resistance ±(ppm of setting + ppm of range), relative to calibration standa rds used. Calibration Interv al 365 day, ±1°C of las t TC AL Range 1281 HP3458/HFL 8508A 1.99999999 Ohm - - 10 + 2.0 12.00000 Ohm - 10 + 5 - 19.9999999 Ohm 12 + 1.0 - 7.0 + 0.7 120.00000 Ohm - 8 + 5 - 199.999999 Ohm 8 + 0.3 7.0 + 0.25 1.2000000 kOhm - 7 .5 + 0.5 - 1.9999 ...

  • Fluke 3458/HFL - page 5

    A C Current ±(ppm of setting + ppm o f range), rela tive to cal ibrati on standards use d. Calibration Interval 365 day, ±1°C of last TCAL Range Frequency 1281 HP3458/HFL 8508A 120.0000µA 10Hz - 20Hz - 4000 + 200 - 20Hz - 45Hz - 1500 + 200 - 45Hz - 100Hz - 600 + 200 - 100Hz - 5kHz - 600 + 200 - 199.9999µA 1Hz - 10Hz - - 250 + 100 10Hz - 5kHz 2 ...

  • Fluke 3458/HFL - page 6

    A C Voltage ± (ppm of setting + ppm of range), relative to cali bration standards used. Cali bration Interval 365 day, ±1°C of last TCA L Range Frequ ency 1281 HP3458/HFL 8508A 12.00000 mV 1Hz - 40Hz - 300 + 300 - 40Hz - 1kHz - 200 + 110 - 1kHz - 20 kHz - 300 + 110 - 20kHz - 50k Hz - 1000 + 110 - 50kHz - 100 kHz - 5000 + 110 - 100kHz - 300kHz - ...

  • Fluke 3458/HFL - page 7

    A C Voltage ± (ppm of setting + ppm of range), relative to cali bration standards used. Calibration Int erval 365 day, ±1°C of last T CAL Range Frequency 1281 HP3458/HFL 8508A 19.99999V 1Hz - 10Hz - - 100 + 60 10Hz - 40Hz - - 100 + 10 40Hz - 100Hz 80 + 10 - 80 + 10 100Hz - 2kHz 60 + 10 - 60 + 10 2kHz - 10kHz 80 + 10 - 80 + 10 10kHz - 30kHz 200 + ...

  • Fluke 3458/HFL - page 8

    Fluke 8508A 3458/HFL 8508A vs 3458/HFL Functional & Versatility Parameters Comments Full Scale (dc & ac Voltage) 200mV, 2V, 20V, 200V, 1050Vrms 120mV, 1.2V, 12V, 120V, 1050Vdc (ac = 700Vrms) 8508A maintains >10Gohm input impedance to 20Vdc. This compares with 12V on 3458/HFL. The 8508A 1050Vrms measurement is often used to calibrate cali ...

  • Fluke 3458/HFL - page 9

    8508A vs 3458/HFL Functional & Versatility Parameters Comments High Resistance Measurement (compliance voltage) Test Voltage or Full Scale stimulus = 200V Test Voltage or Full Scale stimulus @ 1.2G Ω = 5V Reduces errors when performing high resistance measurement (feature available 8508A - 20M Ω to 20G Ω range only) Tru-Ohms or Offset Com ...

  • Fluke 3458/HFL - page 10

    ACV & DCV 4 wire input. Selectable 2 and 4-wire measurement. Unavailable 4-wire sense measurement support improved accuracy when calibrating calibrators with 4 wire sense output. Remote switched GUARD Terminal IEEE controlled Manual Switch IEEE control of GUARD terminal helps fully automate calibration without need for manual switching. Read ra ...

  • Fluke 3458/HFL - page 11

    8508A vs 3458/HFL Ease of Use Parameters Comments Rear input switched remotely using IEEE 488 commands Yes No The 3458/HFL must be switched manually using a front panel button. Instrument simultaneously displays Frequency as well as the AC voltage & current. Yes Selectable Requires further commands or key presses to determine frequency. Two cha ...

  • Fluke 3458/HFL - page 12

    8508A Options and Accessories: Model Description 8508A 8.5 digit Reference Multimeter 8508A/01 8.5 digit Reference Multimeter with Rear Input Ratio measurement Accessories 8508A-SPRT Standard Platinum Resistanc e Thermometer 8508A-PRT Platinum Resistance Thermometer 8508A-LE AD Comprehensive Measuremen t Lead Kit Y8508 Rack Mount Kit Y8508S Rack Mo ...

  • Fluke 3458/HFL - page 13

    Notes: ...

Produzent Fluke Kategorie Wireless Office Headset

Dokumente, die wir vom Produzenten des Geräts Fluke 3458/HFL erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Fluke
- Bedienungsanleitungen 3458/HFL
- Produktkarten Fluke
- Informationsbroschüren
- oder Energieetiketten Fluke 3458/HFL
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Fluke 3458/HFL.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Fluke 3458/HFL, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Fluke 3458/HFL. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Fluke 3458/HFL ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Fluke 3458/HFL, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Fluke 3458/HFL hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Fluke 3458/HFL verständlich ist.

Eine gänzliche Bedienungsanleitung von Fluke sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Fluke 3458/HFL bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Fluke 3458/HFL befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Fluke 3458/HFL, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Fluke 3458/HFL - die uns die ersten Schritte während der Nutzung von Fluke 3458/HFL erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Fluke 3458/HFL hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Fluke 3458/HFL im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Fluke 3458/HFL?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Fluke 3458/HFL nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Fluke 3458/HFL hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)