    Dealer: This manual must be given to the end user. User: Before using this concentrator, re ad this manual and save for future reference. Operator's Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Platinum ™ Oxygen Concentrators Standard, with /SensO 2 ® and HomeFill ® Model IRC5L ...

  • Invacare IRC5LXAW - page 2

    WARNING Platinum ™ 2 Part No 113486 6  WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UN DERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF YOU ARE ...

  • Invacare IRC5LXAW - page 3

    TABLE OF CONTENTS Part No 1134866 3 Platinum ™ TABLE OF CONTENTS SPECIAL NOTES ............................................... 5 RECYCLING INFO RMATION ............................. 6 SECTION 1—GENERAL GUIDELINES ..................... 7 Maintenance ......................... ...................................................... ........... 9 Ra ...

  • Invacare IRC5LXAW - page 4

    TABLE OF CONTENTS Platinum ™ 4 Part No 113486 6 Routine Maintenanc e ....................... ................................................. 27 SECTION 7—TROUBLES HOOTING GUIDE ............ 31 LIMITED WARRANTY .................................... 33 ...

  • Invacare IRC5LXAW - page 5

    SPECIAL NOTES Part No 1134866 5 Platinum™ SPECIAL NOTES Signal  words  are  used  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  following  table  for  definit ...

  • Invacare IRC5LXAW - page 6

    RECYCLING INFORMATION Platinum™ 6 Part No 1134866 RECYCLING INFORMATION This  product  has  been  supplied  from  an  environmentally  aw are  manufacturer  that  complies  with  the  Wa s t e  Electrical  and  Electronic  Equipment  (WEEE)  Directive  2002/96/CE. This  pro duct ? ...

  • Invacare IRC5LXAW - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1134866 7 Platinum™ SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GU IDELINES contains important information for the safe operation and use of this product. In  order  to  ensure  the  safe  installation,  assembly  and  operation  of  the  Platinum  co ...

  • Invacare IRC5LXAW - page 8

    SECTION 1—GENERAL GUIDELINES Platinum™ 8 Part No 1134866  WARNINGS The  use  of  oxygen  therapy  requires  that  special  care  be  take n  to  reduce  the  risk  of  fire.  Any  materials  that  will  burn  in  air ,  and  some  that  will  not,  are  ...

  • Invacare IRC5LXAW - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1134866 9 Platinum™ Fill  humidifier  with  wa t e r  to  the  lev el  shown  by  the  manufacturer .  DO  NOT  overfill. Keep  unit  at  least  7.5  cm  (3  in)  aw ay  from  w alls,  draperies,  furniture,  and  the  like. ...

  • Invacare IRC5LXAW - page 10

    SECTION 1—GENERAL GUIDELINES Platinum™ 10 Part No 1134866 DO  NOT  reach  for  prod uct  that  has  fallen  into  wat e r .  Unplug  IMMEDIA TEL Y . This  device  is  to  be  used  only  in  accordance  with  the  prescription  of  a  physician  and  this  Owner ...

  • Invacare IRC5LXAW - page 11

    SECTION 2—FEATURES Part No 1134866 11 Platinum™ SECTION 2—FEATURES Oxygen Purity and Fault Indicator Lights (SensO 2 model)/Fault and Power Indicator Lights (LX model) Circuit Breaker Power Switch Elapsed Time Meter (Hour Meter) Cabinet Filter (one of two shown) Flowmet er Oxygen Out let FRONT VIEW TOP VIEW ACCESSOR IES (NOT SHOWN): HomeFill ...

  • Invacare IRC5LXAW - page 12

    SECTION 3—SHIPPING AND HANDLING Platinum™ 12 Part No 1134866 SECTION 3—SHIPPING AND HANDLING NOTE:  For  this  procedure,  refe r  to  FIGURE  3.1  on  page 13. Before  yo u  install  and  operate  the  concentrator ,  please  refer  to  Platinum  Series  Owner ʹ s  Manual. ...

  • Invacare IRC5LXAW - page 13

    SECTION 3—SHIPPING AND HANDLING Part No 1134866 13 Platinum™ FIGURE 3.1 Unpacking ...

  • Invacare IRC5LXAW - page 14

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 14 Part No 1134866 SECTION 4—TYPICAL PRODUCT PARAMETERS Regulatory Listing: EN/IEC 60601-1 Alternating Current Type BF equipment Unit running Unit not running Attention - Consider Accompanying Documents DO NOT smoke Class II, Double Insulation Electrical Requirements: 230 VAC ± 10% (253 VAC/207 ...

  • Invacare IRC5LXAW - page 15

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1134866 15 Platinum™ *Oxygen Output Concentration Levels: All 5LXO2AWQ/5LXAW/ 5LXO2AW/9LXO2AWQ models * (Stated concentration levels achieved after in itial warm-up period (approximately 30 minutes). IRC5LXAW, IRC5 LXO2AW and IRC5LXO2AWQ 93% minimu m at 1 to 3 L/min. 91% minimu m at 4 L/min. 87% mini ...

  • Invacare IRC5LXAW - page 16

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 16 Part No 1134866 Change in maximum recommended flow when back pressure of 7k Pa is applied: 0.7 L/min. Filters: Cabinet (2), Outl et HEPA and Compressor Inlet. Safety System: Current overload or line surge shutdown. High temperature co mpressor shutdown. High Pressure Alarm w/compressor shutdown. ...

  • Invacare IRC5LXAW - page 17

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1134866 17 Platinum™ Regulatory Listing : IRC5LXAW/IR C5LXO2AW/ IRC5LXO2AWQ/IRC9LXO2AWQ ETL tested as complying with: EN55011 EN61000-3-2 EN61000-3-3 IEC 60601-1, A1, A2 IEC 60601-1-2 IRC5LXO2AW, IRC5LXO2AWQ, IRC9LXO2AWQ Models Only ISO8359 MDD CE marked models: IRC5LXO2AW, IRC5LXO2AWQ, IRC9LXO2AWQ E ...

  • Invacare IRC5LXAW - page 18

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 18 Part No 1134866 SENSO 2 INDICATO RS LX INDICATORS LABEL SYMBOL STATUS INDICATOR LIGHTS (LED) SYSTEM OKAY O 2 over 85% (± 2%) GREEN Indicator Light O 2 Between 73% (± 3%) to 85% (±2%) YELLOW Indicator light A. YELLOW Solid B. YELLOW Flashing Sensor Failure Call a qualif ied technician. SYSTEM ...

  • Invacare IRC5LXAW - page 19

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 19 Platinum™ SECTION 5—OPERATING INSTRUCTIONS Introduction Yo u r  oxygen  concentrator  is  intended  for  individual  use  in  the  home.  It  is  an  electronically  operated  device  that  separates  oxygen  from  room  air . ...

  • Invacare IRC5LXAW - page 20

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 20 Part No 1134866 Recommended Guidelines for Optimum Performance NOTE:  Ensure  that  your  concentrator  is  at  least  7.5  cm  (3  in)  aw ay  from  walls,  drap erie s  or  furniture  to  assur e  sufficient  air  flow .  Av o i ...

  • Invacare IRC5LXAW - page 21

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 21 Platinum™ Connect Humidifier (If So Prescribed) NOTE:  For  this  procedure,  refe r  to  FIGURE  5.1,FIGURE  5.2,  FIGURE  5.3  on  page 22  and  FIGURE  5.4  on  page 23. 1. Fill  humidifier  with  distilled  wa t e r  t ...

  • Invacare IRC5LXAW - page 22

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 22 Part No 1134866 3. Remove  the  filter  access  door  located  on  the  side  of  the  concentrator .  Insert  a  flathead  screwdriver  in  the  plate  groove  on  th e  top  edg e  of  the  filter  access  door  ...

  • Invacare IRC5LXAW - page 23

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 23 Platinum™ 6. Attach  oxygen  tubing  from  the  humidifier  bottle  to  oxygen  outlet  connector  on  the  oxygen  concentrator . 7. Attach  the  cannula/patient  supply  tubing  to  the  humidifier  bottle  outlet. 8. After ...

  • Invacare IRC5LXAW - page 24

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 24 Part No 1134866 FIGURE 5.5 Power Switch Flowrate NOTE:  For  this  procedure,  refe r  to  FIGURE  5.6  on  page 25. 1. T urn  the  flowrate  knob  to  the  setting  prescribed  by  you r  physician  or  therapist.  WARNING DO NOT ...

  • Invacare IRC5LXAW - page 25

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 25 Platinum™ NOTE:  The  use  of  some  accessories  such  as  the  pediatric  flowstand  and  the  HomeFill  compr essor  will  deactivate  the  Low  Flow  Alarm. FIGURE 5.6 Flowrate SensO 2 Oxygen Purity Indicators (LX O 2 Sensor Model ...

  • Invacare IRC5LXAW - page 26

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 26 Part No 1134866 FIGURE 5.7 SensO 2 Oxygen Purity Indicators (LX O 2 Sensor Model) GREEN  light  (O 2 ) ‐ Normal  Operation. YELLOW  light  () ‐ Call  supplier  IMMEDIA TEL Y .  Yo u  may  continue  to  use  the  concentrator  unless  in ...

  • Invacare IRC5LXAW - page 27

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 27 Platinum™ FIGURE 5.8 Indicators (LX Model) Main Power Loss Alarm System The  alarm  system  contains  no  battery ‐ and  is  thereby  maintenance  free.  The  alarm  system  is  powered  by  a  capacitor  that  is  continuousl ...

  • Invacare IRC5LXAW - page 28

    SECTION 6—MAINTENANCE Platinum™ 28 Part No 1134866 SECTION 6—MAINTENANCE NOTE:  The  Invacare  concentrators  are  specifically  designed  to  minimize  rou ti n e  pr eventive  maintenance.  In  places  with  high  dust  or  soot  levels ,  maintenance  may  need  to  be ...

  • Invacare IRC5LXAW - page 29

    SECTION 6—MAINTENANCE Part No 1134866 29 Platinum™ 3. Dry  the  filters  thoroughly  before  reinstallation. FIGURE 6.1 Cleaning the Cabinet Filter Cabinet  WARNING Unplug the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. 1. Clean  the  cabinet  with  a  mild  household  cle ...

  • Invacare IRC5LXAW - page 30

    SECTION 6—MAINTENANCE Platinum™ 30 Part No 1134866 FIGURE 6.2 Preventive Maintenance Record ON EACH INSPECTION MODEL NO. I R C SERIAL NO. PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOURS FOR IR ...

  • Invacare IRC5LXAW - page 31

    SECTION 7—TROUBLESHOOTING GUIDE Part No 1134866 31 Platinum™ SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: Short beeps, long pause 1. Power cord NOT plugged in. 1. Insert plug i nto outlet. Concentrator not operating, power switch On. 2. No power at outlet. 2 . Inspect house circuit breakers or fus ...

  • Invacare IRC5LXAW - page 32

    SECTION 7—TROUBLESHOOTING GUIDE Platinum™ 32 Part No 1134866 YELLOW or RED light illumin ated. Alarm: Continuous On red light only. * Only applies t o SensO 2 models. 1. Low oxygen p urity.* 2. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set at 0.5 L/min. or less.* 4. Unit overheating due to blocked air intake. 5. Insufficien ...

  • Invacare IRC5LXAW - page 33

    LIMITED WARRANTY Part No 1134866 33 Platinum™ LIMITED WARRANTY A vailable  At: Belgium  &  Luxemburg: Inv acare  nv ,  Autobaan  22,  B ‐ 8210  Loppem Te l :  (32)  (0)50  83  10  10,  Fax:  (32)  (0)50  83  10  11 belgium@inv acare.com Danmark:  Inv acare  A/S,  Sdr .  Ri ...

  • Invacare IRC5LXAW - page 34

    LIMITED WARRANTY Platinum™ 34 Part No 1134866 Italia: Inv acare  Mecc  San  s.r .l.,  Vi a  dei  Pini  62,  I ‐ 36016  Thiene  (VI) Te l :  (39)  0445  38  00  59,  Fax:  (39)  0445  38  00  34 italia@inv acare.com  Nederland: Inv acare  BV ,  Celsiusstraat  46,  NL ? ...

  • Invacare IRC5LXAW - page 35

    LIMITED WARRANTY Part No 1134866 35 Platinum™ Switzerland: Mobitec  Rehab  AG,  Benkenstrasse  260,  CH ‐ 4108  Wit t e rs w il Te l . :  (41)  (0)61  487  70  80,  Fax.:  (41)  (0)61  487  70  81 office@mobitec ‐ rehab.ch switzerland@inv acare.com United  Kingdom: Inv acare  Ltd,  S ...

  • Invacare IRC5LXAW - page 36

    Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 440-329-6000 800-333-6900 Technical Serv ices 440-329-6593 800-832-4707 Manufacturer: Invacare Medical Equipment (Suzhou) Co., Ltd. Unit 2B, Export Processing Zone No. 200 Suhong Road SIP (Suzhou Industrial Park) Suzhou, Jiangsu, PRC 215021 Tel: 86-512-62586192 F ...

Fabricante Invacare Categoría Oxygen Equipment

Los documentos del dispositivo Invacare IRC5LXAW que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos Invacare
- manuales de instrucciones IRC5LXAW
- hojas de producto Invacare
- folletos informativos
- o etiquetas energéticas Invacare IRC5LXAW
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones Invacare IRC5LXAW.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje Invacare IRC5LXAW, instrucciones de servicio, instrucciones cortas o instrucciones de usuario Invacare IRC5LXAW. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto Invacare IRC5LXAW.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo Invacare IRC5LXAW, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar Invacare IRC5LXAW. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de Invacare IRC5LXAW.

El manual de instrucciones completo de Invacare debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual Invacare IRC5LXAW - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice Invacare IRC5LXAW, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de Invacare IRC5LXAW que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo Invacare IRC5LXAW - que deben facilitarnos los primeros pasos durante el uso de Invacare IRC5LXAW
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con Invacare IRC5LXAW
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de Invacare IRC5LXAW en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de Invacare IRC5LXAW?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con Invacare IRC5LXAW, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con Invacare IRC5LXAW probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)