    Dealer: This manual must be given to the end user. User: Before using this concentrator, re ad this manual and save for future reference. Operator's Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Platinum ™ Oxygen Concentrators Standard, with /SensO 2 ® and HomeFill ® Model IRC5L ...

  • Invacare IRC5LXAW - page 2

    WARNING Platinum ™ 2 Part No 113486 6  WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UN DERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF YOU ARE ...

  • Invacare IRC5LXAW - page 3

    TABLE OF CONTENTS Part No 1134866 3 Platinum ™ TABLE OF CONTENTS SPECIAL NOTES ............................................... 5 RECYCLING INFO RMATION ............................. 6 SECTION 1—GENERAL GUIDELINES ..................... 7 Maintenance ......................... ...................................................... ........... 9 Ra ...

  • Invacare IRC5LXAW - page 4

    TABLE OF CONTENTS Platinum ™ 4 Part No 113486 6 Routine Maintenanc e ....................... ................................................. 27 SECTION 7—TROUBLES HOOTING GUIDE ............ 31 LIMITED WARRANTY .................................... 33 ...

  • Invacare IRC5LXAW - page 5

    SPECIAL NOTES Part No 1134866 5 Platinum™ SPECIAL NOTES Signal  words  are  used  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  following  table  for  definit ...

  • Invacare IRC5LXAW - page 6

    RECYCLING INFORMATION Platinum™ 6 Part No 1134866 RECYCLING INFORMATION This  product  has  been  supplied  from  an  environmentally  aw are  manufacturer  that  complies  with  the  Wa s t e  Electrical  and  Electronic  Equipment  (WEEE)  Directive  2002/96/CE. This  pro duct ? ...

  • Invacare IRC5LXAW - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1134866 7 Platinum™ SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GU IDELINES contains important information for the safe operation and use of this product. In  order  to  ensure  the  safe  installation,  assembly  and  operation  of  the  Platinum  co ...

  • Invacare IRC5LXAW - page 8

    SECTION 1—GENERAL GUIDELINES Platinum™ 8 Part No 1134866  WARNINGS The  use  of  oxygen  therapy  requires  that  special  care  be  take n  to  reduce  the  risk  of  fire.  Any  materials  that  will  burn  in  air ,  and  some  that  will  not,  are  ...

  • Invacare IRC5LXAW - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1134866 9 Platinum™ Fill  humidifier  with  wa t e r  to  the  lev el  shown  by  the  manufacturer .  DO  NOT  overfill. Keep  unit  at  least  7.5  cm  (3  in)  aw ay  from  w alls,  draperies,  furniture,  and  the  like. ...

  • Invacare IRC5LXAW - page 10

    SECTION 1—GENERAL GUIDELINES Platinum™ 10 Part No 1134866 DO  NOT  reach  for  prod uct  that  has  fallen  into  wat e r .  Unplug  IMMEDIA TEL Y . This  device  is  to  be  used  only  in  accordance  with  the  prescription  of  a  physician  and  this  Owner ...

  • Invacare IRC5LXAW - page 11

    SECTION 2—FEATURES Part No 1134866 11 Platinum™ SECTION 2—FEATURES Oxygen Purity and Fault Indicator Lights (SensO 2 model)/Fault and Power Indicator Lights (LX model) Circuit Breaker Power Switch Elapsed Time Meter (Hour Meter) Cabinet Filter (one of two shown) Flowmet er Oxygen Out let FRONT VIEW TOP VIEW ACCESSOR IES (NOT SHOWN): HomeFill ...

  • Invacare IRC5LXAW - page 12

    SECTION 3—SHIPPING AND HANDLING Platinum™ 12 Part No 1134866 SECTION 3—SHIPPING AND HANDLING NOTE:  For  this  procedure,  refe r  to  FIGURE  3.1  on  page 13. Before  yo u  install  and  operate  the  concentrator ,  please  refer  to  Platinum  Series  Owner ʹ s  Manual. ...

  • Invacare IRC5LXAW - page 13

    SECTION 3—SHIPPING AND HANDLING Part No 1134866 13 Platinum™ FIGURE 3.1 Unpacking ...

  • Invacare IRC5LXAW - page 14

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 14 Part No 1134866 SECTION 4—TYPICAL PRODUCT PARAMETERS Regulatory Listing: EN/IEC 60601-1 Alternating Current Type BF equipment Unit running Unit not running Attention - Consider Accompanying Documents DO NOT smoke Class II, Double Insulation Electrical Requirements: 230 VAC ± 10% (253 VAC/207 ...

  • Invacare IRC5LXAW - page 15

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1134866 15 Platinum™ *Oxygen Output Concentration Levels: All 5LXO2AWQ/5LXAW/ 5LXO2AW/9LXO2AWQ models * (Stated concentration levels achieved after in itial warm-up period (approximately 30 minutes). IRC5LXAW, IRC5 LXO2AW and IRC5LXO2AWQ 93% minimu m at 1 to 3 L/min. 91% minimu m at 4 L/min. 87% mini ...

  • Invacare IRC5LXAW - page 16

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 16 Part No 1134866 Change in maximum recommended flow when back pressure of 7k Pa is applied: 0.7 L/min. Filters: Cabinet (2), Outl et HEPA and Compressor Inlet. Safety System: Current overload or line surge shutdown. High temperature co mpressor shutdown. High Pressure Alarm w/compressor shutdown. ...

  • Invacare IRC5LXAW - page 17

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1134866 17 Platinum™ Regulatory Listing : IRC5LXAW/IR C5LXO2AW/ IRC5LXO2AWQ/IRC9LXO2AWQ ETL tested as complying with: EN55011 EN61000-3-2 EN61000-3-3 IEC 60601-1, A1, A2 IEC 60601-1-2 IRC5LXO2AW, IRC5LXO2AWQ, IRC9LXO2AWQ Models Only ISO8359 MDD CE marked models: IRC5LXO2AW, IRC5LXO2AWQ, IRC9LXO2AWQ E ...

  • Invacare IRC5LXAW - page 18

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 18 Part No 1134866 SENSO 2 INDICATO RS LX INDICATORS LABEL SYMBOL STATUS INDICATOR LIGHTS (LED) SYSTEM OKAY O 2 over 85% (± 2%) GREEN Indicator Light O 2 Between 73% (± 3%) to 85% (±2%) YELLOW Indicator light A. YELLOW Solid B. YELLOW Flashing Sensor Failure Call a qualif ied technician. SYSTEM ...

  • Invacare IRC5LXAW - page 19

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 19 Platinum™ SECTION 5—OPERATING INSTRUCTIONS Introduction Yo u r  oxygen  concentrator  is  intended  for  individual  use  in  the  home.  It  is  an  electronically  operated  device  that  separates  oxygen  from  room  air . ...

  • Invacare IRC5LXAW - page 20

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 20 Part No 1134866 Recommended Guidelines for Optimum Performance NOTE:  Ensure  that  your  concentrator  is  at  least  7.5  cm  (3  in)  aw ay  from  walls,  drap erie s  or  furniture  to  assur e  sufficient  air  flow .  Av o i ...

  • Invacare IRC5LXAW - page 21

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 21 Platinum™ Connect Humidifier (If So Prescribed) NOTE:  For  this  procedure,  refe r  to  FIGURE  5.1,FIGURE  5.2,  FIGURE  5.3  on  page 22  and  FIGURE  5.4  on  page 23. 1. Fill  humidifier  with  distilled  wa t e r  t ...

  • Invacare IRC5LXAW - page 22

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 22 Part No 1134866 3. Remove  the  filter  access  door  located  on  the  side  of  the  concentrator .  Insert  a  flathead  screwdriver  in  the  plate  groove  on  th e  top  edg e  of  the  filter  access  door  ...

  • Invacare IRC5LXAW - page 23

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 23 Platinum™ 6. Attach  oxygen  tubing  from  the  humidifier  bottle  to  oxygen  outlet  connector  on  the  oxygen  concentrator . 7. Attach  the  cannula/patient  supply  tubing  to  the  humidifier  bottle  outlet. 8. After ...

  • Invacare IRC5LXAW - page 24

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 24 Part No 1134866 FIGURE 5.5 Power Switch Flowrate NOTE:  For  this  procedure,  refe r  to  FIGURE  5.6  on  page 25. 1. T urn  the  flowrate  knob  to  the  setting  prescribed  by  you r  physician  or  therapist.  WARNING DO NOT ...

  • Invacare IRC5LXAW - page 25

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 25 Platinum™ NOTE:  The  use  of  some  accessories  such  as  the  pediatric  flowstand  and  the  HomeFill  compr essor  will  deactivate  the  Low  Flow  Alarm. FIGURE 5.6 Flowrate SensO 2 Oxygen Purity Indicators (LX O 2 Sensor Model ...

  • Invacare IRC5LXAW - page 26

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 26 Part No 1134866 FIGURE 5.7 SensO 2 Oxygen Purity Indicators (LX O 2 Sensor Model) GREEN  light  (O 2 ) ‐ Normal  Operation. YELLOW  light  () ‐ Call  supplier  IMMEDIA TEL Y .  Yo u  may  continue  to  use  the  concentrator  unless  in ...

  • Invacare IRC5LXAW - page 27

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 27 Platinum™ FIGURE 5.8 Indicators (LX Model) Main Power Loss Alarm System The  alarm  system  contains  no  battery ‐ and  is  thereby  maintenance  free.  The  alarm  system  is  powered  by  a  capacitor  that  is  continuousl ...

  • Invacare IRC5LXAW - page 28

    SECTION 6—MAINTENANCE Platinum™ 28 Part No 1134866 SECTION 6—MAINTENANCE NOTE:  The  Invacare  concentrators  are  specifically  designed  to  minimize  rou ti n e  pr eventive  maintenance.  In  places  with  high  dust  or  soot  levels ,  maintenance  may  need  to  be ...

  • Invacare IRC5LXAW - page 29

    SECTION 6—MAINTENANCE Part No 1134866 29 Platinum™ 3. Dry  the  filters  thoroughly  before  reinstallation. FIGURE 6.1 Cleaning the Cabinet Filter Cabinet  WARNING Unplug the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. 1. Clean  the  cabinet  with  a  mild  household  cle ...

  • Invacare IRC5LXAW - page 30

    SECTION 6—MAINTENANCE Platinum™ 30 Part No 1134866 FIGURE 6.2 Preventive Maintenance Record ON EACH INSPECTION MODEL NO. I R C SERIAL NO. PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOURS FOR IR ...

  • Invacare IRC5LXAW - page 31

    SECTION 7—TROUBLESHOOTING GUIDE Part No 1134866 31 Platinum™ SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: Short beeps, long pause 1. Power cord NOT plugged in. 1. Insert plug i nto outlet. Concentrator not operating, power switch On. 2. No power at outlet. 2 . Inspect house circuit breakers or fus ...

  • Invacare IRC5LXAW - page 32

    SECTION 7—TROUBLESHOOTING GUIDE Platinum™ 32 Part No 1134866 YELLOW or RED light illumin ated. Alarm: Continuous On red light only. * Only applies t o SensO 2 models. 1. Low oxygen p urity.* 2. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set at 0.5 L/min. or less.* 4. Unit overheating due to blocked air intake. 5. Insufficien ...

  • Invacare IRC5LXAW - page 33

    LIMITED WARRANTY Part No 1134866 33 Platinum™ LIMITED WARRANTY A vailable  At: Belgium  &  Luxemburg: Inv acare  nv ,  Autobaan  22,  B ‐ 8210  Loppem Te l :  (32)  (0)50  83  10  10,  Fax:  (32)  (0)50  83  10  11 belgium@inv acare.com Danmark:  Inv acare  A/S,  Sdr .  Ri ...

  • Invacare IRC5LXAW - page 34

    LIMITED WARRANTY Platinum™ 34 Part No 1134866 Italia: Inv acare  Mecc  San  s.r .l.,  Vi a  dei  Pini  62,  I ‐ 36016  Thiene  (VI) Te l :  (39)  0445  38  00  59,  Fax:  (39)  0445  38  00  34 italia@inv acare.com  Nederland: Inv acare  BV ,  Celsiusstraat  46,  NL ? ...

  • Invacare IRC5LXAW - page 35

    LIMITED WARRANTY Part No 1134866 35 Platinum™ Switzerland: Mobitec  Rehab  AG,  Benkenstrasse  260,  CH ‐ 4108  Wit t e rs w il Te l . :  (41)  (0)61  487  70  80,  Fax.:  (41)  (0)61  487  70  81 office@mobitec ‐ rehab.ch switzerland@inv acare.com United  Kingdom: Inv acare  Ltd,  S ...

  • Invacare IRC5LXAW - page 36

    Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 440-329-6000 800-333-6900 Technical Serv ices 440-329-6593 800-832-4707 Manufacturer: Invacare Medical Equipment (Suzhou) Co., Ltd. Unit 2B, Export Processing Zone No. 200 Suhong Road SIP (Suzhou Industrial Park) Suzhou, Jiangsu, PRC 215021 Tel: 86-512-62586192 F ...

Производитель Invacare Категория Oxygen Equipment

Документы, которые мы получаем от производителя устройства Invacare IRC5LXAW мы можем разделить на несколько групп. Это в частности:
- технические чертежи Invacare
- инструкции обслуживания IRC5LXAW
- паспорта изделия Invacare
- информационные брошюры
- энергетические этикетки Invacare IRC5LXAW
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания Invacare IRC5LXAW.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа Invacare IRC5LXAW, инструкции обслуживания, короткие инструкции или инструкции пользователя Invacare IRC5LXAW. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия Invacare IRC5LXAW.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства Invacare IRC5LXAW, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании Invacare IRC5LXAW пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей Invacare IRC5LXAW.

Полная инструкция обслуживания Invacare, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией Invacare IRC5LXAW - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания Invacare IRC5LXAW, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся Invacare IRC5LXAW, которое найдем в данном документе
3. Советы по использованию основных функций устройства Invacare IRC5LXAW - которые должны облегчить нам первые шаги во время использования Invacare IRC5LXAW
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем Invacare IRC5LXAW
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра Invacare IRC5LXAW в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся Invacare IRC5LXAW?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с Invacare IRC5LXAW, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с Invacare IRC5LXAW со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)