Omega OS137 - page 1
OS137 SERIES Miniature Non-Contact Infrared T emperature Sensor/T ransmitter MADE IN omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info U ser ’ s Guide Shop online at CA UTION! – This product is not intended for medical use or use on humans ...
-
Omega OS137 - page 2
Servicing Nor th America: U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047 ISO 9001 Certified Stamford, CT 06907-0047 USA Toll Free: 1-800-826-6342 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec), Canada H7L 5A1 Toll-Free: 1-800-826-6342 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: ...
-
Omega OS137 - page 3
i OS137 Series Miniature Non-Contact Infrared T emperature Sensor/T ransmitter T able of Contents Section Page Caution & Safety Information ...................................................................... iii Safety W arnings and IEC Symbols .............................................................. iii Section 1 Introduction ........ ...
-
Omega OS137 - page 4
OS137 Series Miniature Non-Contact Infrared Temperature Sensor/Transmitter ii T able of Figures Figure Description Page 2-1 General W iring Diagram ................................................................ 2 2-2 Alarm Output W iring Diagram .................................................... 2 3-1 Optical Field of V iew .................... ...
-
Omega OS137 - page 5
OS137 Series Miniature Non-Contact Infrared Temperature Sensor/Transmitter iii CAUTION & SAFETY INFORMA TION If the equipment is used in a manner not specified in this manual, the protection provided by the equipment may be impair ed. The Installation category is one (1). There is no user r eplaceable fuse in this product. The output terminals ...
-
Omega OS137 - page 6
1 SECTION 1 - INTRODUCTION The OS137 Series is a precision, miniatur e infrared transmitter . It measures temperature via non-contact, and pr ovides an analog output proportional to the measured temperatur e. The OS137 Series is offer ed in three temperatur e ranges: 0 to 100ºC (32 to 212ºF), -18 to 260ºC (0 to 500ºF) and -18 to 538ºC (0 to 10 ...
-
Omega OS137 - page 7
2 2 - ** means: -MA : 4 to 20 mA current output -V1 : 0 to 5 Vdc output -V2 : 0 to 10 Vdc output -K : K type thermocouple output -MVC : 10 mV/ºC output -MVF : 10 mV/ºF output The following table lists the optional accessories: 2.2 – Electrical Connection The shielded cable provides the power and output connections. Fig 2-1 shows the wiring diag ...
-
Omega OS137 - page 8
3 Operation 3 SECTION 3 – OPERA TION 3.1 – Measuring T emperature Before starting to measur e temperature, make sur e the following check list is met: The power and output connections are made (Fig 2-1). The target is lar ger than the optical field of view of the transmitter (Fig 3-1). Use the Laser Sighting accessory (optional), to ...
-
Omega OS137 - page 9
Operation 3 4 3.2 – Ambient T emperature The transmitter can operate in an ambient temperature of 0 to 70°C (32 to 158°F) without any water cool jacket. It can operate from 0 to 200°C (32 to 392°F) with the water cool jacket accessory , OS137-WC (Fig 3-6 & 3-7). It can operate up to 1 10°C (230°F) with air cooling. There is a warm up pe ...
-
Omega OS137 - page 10
Figure 3-4. Air Purge Collar, OS137-AP Figure 3-5. Stainless Steel Housing Figure 3-6. Water/Air Cool Jacket, OS137-WC Operation 3 5 LIGHT DIAMOND KNURL 2.5 (0.1) WIDE DIMENSIONS mm (in) 1-20 UNEF 127 (5.0) 1.83 m (6 ft) 42.4 (1.67) 25.4 (1.0) DIMENSIONS mm (in) 128.6 (5.062) 7.9 (0.312) 41.2 (1.625) 1-20 UNEF 9.5 (3/8) 1-20 UNEF BUIL T -IN AIR PUR ...
-
Omega OS137 - page 11
3.4 Alarm Setting The unit provides 0-100% alarm setpoint adjustment. Her e is an example of an alarm setting: • An OS137-3-MA (4/20 mA output), the alarm is to be set at 204°C (400°F) temperature. • Connect the alarm output as shown in Fig. 2-2. • Remove the End Cap to get to the Alarm Switch and the Alarm Adjust (Fig. 3-2). • Set the Sl ...
-
Omega OS137 - page 12
SECTION 4 - LASER SIGHT ACCESSOR Y 4.1 – W arnings and Caution Y ou may receive harmful laser radiation exposure if you do not adhere to the warnings listed below: • USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER THAN THOSE SPECIFIED IN THIS GUIDE MA Y RESUL T IN HAZARDOUS RADIA TION EXPOSURE. • DO NOT LOOK A T THE LASER BE ...
-
Omega OS137 - page 13
8 Figure 4-1. Laser Sighting Accessory, OS137-LS Figure 4-2. Laser Warning Label LASER W ARNING LABEL DC POWER JACK LED LASER POWER INDICA TOR LASER BEAM 22.2 (0.875) 50.8 (2.0) 38.1 (1.5) POWER CABLE ON/OFF SWITCH BA TTERY P ACK POWER SUPPL Y OS137 Dimensions mm (in) OUTPUT <1 mW, WAVELENGTH 630-670 nm, CLASS II (2) LASER PRODUCT. COMPLIES WITH ...
-
Omega OS137 - page 14
9 SECTION 5 – SPECIFICA TIONS 5.1 - General T emperature Range: OS137-1 0 to 100ºC (32 to 212ºF) OS137-2 -18 to 260ºC (0 to 500ºF) OS137-3 -18 to 538°C (0 to 1000°F) Accuracy: @22ºC (72ºF) ambient 1.5% of Rdg or 2.0ºC (3.5ºF) whichever is greater Emissivity of 0.95 or greater Repeatability: 1% of Rdg or 1.0ºC (2.0ºF) whichever is grea ...
-
Omega OS137 - page 15
10 5.2 - Laser Sight Accessory Laser W avelength (Color): 630 - 670 nm (red) Operating Distance: Up to 9.1 m (30 ft) Max. Laser Power Output: Less than 1 mW @ 22ºC (72°F) ambient European Classification: Class 2, EN60825-1/1 1.2001 FDA Classification: Class II Laser Product. Complies with 21 CFR 1040.10 Laser Beam Diameter: Less than 5 mm Beam Di ...
-
Omega OS137 - page 16
Emissivity T able 6 11 Material Emissivity ( ε ) Aluminum – pure highly polished plate . . . . . . . . . . . . . . . . . . . . . . . . 0.04 to 0.06 Aluminum – heavily oxidized . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.20 to 0.31 Aluminum – commercial sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...
-
Omega OS137 - page 17
Emissivity Table 6 12 Material Emissivity ( ε ) Asbestos Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.96 Asphalt, tar , pitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.95 to 1.00 Brick – red and rough . . . . . . . . . . . . . . . . . . . . . . . . . . ...
-
Omega OS137 - page 18
OS137 Series Miniature Non-Contact Infrared Temperature Sensor/Transmitter 6 13 NOTES: ...
-
Omega OS137 - page 19
W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from dat e o f p urch ase . OMEGA ’ s W ARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA ’ s ...
-
Omega OS137 - page 20
M4015/0108 Where Do I Find Ever ything I Need for Pr ocess Measurement and Contr ol? OM EGA…Of Course! Shop online at omega.com SM TEMPERA TURE Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies Wire: Thermocouple, RTD & Thermistor Calibrators & Ice Point References Recorders, Co ...
¿Tienes una pregunta acerca de Omega OS137?
Utiliza el formulario que se encuentra abajo
Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con Omega OS137, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con Omega OS137 probablemente compartirá una forma de solucionarlo.