  • Emerson Process Management 396PVP - page 1

    Combina tion pH/ORP Sensor Instruction Manual PN 51-396P/rev .H March 2006 MODEL 396P MODEL 396PVP with mating cable Models 396P and 396PVP ...

  • Emerson Process Management 396PVP - page 2

    CAUTION SENSOR/PROCESS APPLICA TION COMP A TIBILITY The wetted sensor materials may not be comp atible with process composition and operating conditions. Application comp atibility is entirely the responsibility of the user . DANGER HAZARDOUS AREA INST ALLA TION This sensor is not Intrinsically Safe. or Explosion Proof. Installations near flammable ...

  • Emerson Process Management 396PVP - page 3

    i MODEL 396P and 396PVP T ABLE OF CONTENTS MODEL 396P and 396PVP COMBINA TION pH/ORP SENSOR T ABLE OF CONTENTS Section Title Page 1.0 DESCRIPTION AND SPECIFICA TIONS ................................................................. 1 1.1 Features and Applications ...................................................................................... ...

  • Emerson Process Management 396PVP - page 4

    MODEL 396P and 396PVP T ABLE OF CONTENTS LIST OF FIGURES Number Title Page 1-1 Cross Section Diagram of the TUpH Patented Reference T echnology ............................................................................... 2 2-1 Dimensional Drawing .................................................................................................... ...

  • Emerson Process Management 396PVP - page 5

    Model 396PVP is of fered with a watertight sensor-to- cable connector which eliminates re-wiring and cable twisting when replacing sensors. The V ariopol VP multiple pin connector is an integral part of each sen- sor model and uses a mating VP cable. Once the cable is installed and wired to the analyzer , sensors are easily replaced without replaci ...

  • Emerson Process Management 396PVP - page 6

    2 MODEL 396P and 396PVP SECTION 1.0 DESCRIPTIONS AND SPECIFICA TIONS Measurement s and Ranges *pH: 0-14 ORP: -1500 to 1500 mv A vailable pH A CCU G LASS T ypes GPLR hemi or flat glass Wetted Materials T itanium, Polypropylene, EPDM, glass; platinum (ORP only) Process Connection 1 in. MNPT front and rear facing threads T emperature Range 0-100°C (3 ...

  • Emerson Process Management 396PVP - page 7

    3 MODEL 396P and 396PVP SECTION 1.0 DESCRIPTIONS AND SPECIFICA TIONS 1.3 ORDERING INFORMA TION The Model 396P Sensor is housed in a molded reinforced polypropylene body with 1 in. MNPT threads suit able for insertion, submersion or flow through installation. The sensor includes a general purpose pH electrode or a platinum ORP electrode, a p atented ...

  • Emerson Process Management 396PVP - page 8

    4 MODEL 396P and 396PVP SECTION 1.0 DESCRIPTIONS AND SPECIFICA TIONS Optional Slotted Tip is available on all hemi bulb sensors, ordered as option -41 Flat Tip is avail- able with flat glass bulb sen- sors Examples of all sensing tip offerings Shrouded Tip is standard on all hemi bulb sensors V ariopol connector shown with mating variopol cable rec ...

  • Emerson Process Management 396PVP - page 9

    5 MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION SECTION 2.0 INST ALLA TION 2. Do not install the sensor on the horizont al. The sensor must be 10° off the horizont al to ensure accuracy . 3. Do not install the sensor up side down. 4. Air bubbles may become trapped in the sensor end between the glass bulb and the sensor body . This problem is mo ...

  • Emerson Process Management 396PVP - page 10

    6 MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION FIGURE 2-1. Dimensional Drawing DWG . NO. REV . 40396P01 D MILLIMETER INCH FIGURE 2-2. Flow-Through T ee with Adapter (PN 915240-xx*) MILLIMETER INCH xx* Process Connection Threads 03 3/4 inch 04 1 inch 05 1-1/2 inch SENSOR CABLE (OR VP CONNECTOR - NOT SHOWN) SENSOR CABLE (OR VP CONNECTOR - NOT SHO ...

  • Emerson Process Management 396PVP - page 11

    MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION FIGURE 2-4. Model 396P with Insertion Mounting Adapter (PN 23242-02). Not for use with Model 396PVP . Mounting adapter allows for sensor removal without twisting or disconnecting interconnecting cable for ease of maintenance. WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH DWG . NO. REV . 40396P0 ...

  • Emerson Process Management 396PVP - page 12

    MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION FIGURE 2-5. Submersion Inst allations WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH DWG . NO. REV . 40396P04 A DWG . NO. REV . 40396P03 A Handrail Mounting Assembly PN 1 1275-01 8 ...

  • Emerson Process Management 396PVP - page 13

    9 MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION FIGURE 2-6. Low Flow Cell PN 23728-00 MILLIMETER INCH FIGURE 2-7. Jet Spray Cleaner PN 12707-00 SENSOR CABLE (OR VP CONNECTOR - NOT SHOWN) ...

  • Emerson Process Management 396PVP - page 14

    10 Figures in this section provide the guidelines for wiring the 396P-01 sensor to various Analyzer/Transmitter instruments. T o determine which wiring guideline to use, locate the model number of the sensor to be installed. 1. If the cable needs to be extended, use a high qual- ity eleven conductor double shielded instrument cable available from R ...

  • Emerson Process Management 396PVP - page 15

    11 MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P Wiring. The Model 396P has an optional built-in pre- amplifier and is offered with a shielded cable. The cable should be handled carefully and kept dry and free of corrosive chemicals at all times. Extreme care should be used to prevent it from being twisted, dam- aged or scraped by rough, sh ...

  • Emerson Process Management 396PVP - page 16

    12 MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P FIGURE 3-2. Wiring Det ails Model 396P-01-50 for use with and without Junction Box (PN 22719-02) for Model 1 181. WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH DWG . NO. REV . 40396P15 B ...

  • Emerson Process Management 396PVP - page 17

    13 FIGURE 3-3. Wiring Det ails Model 396P-01-54 for use with and without Junction Box (PN 22719-02) for Models 1054A, 1054B, 2054, 2081, and Extension Cable (9200254). MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P DWG . NO. REV . 40396P13 D WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH ...

  • Emerson Process Management 396PVP - page 18

    14 MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P FIGURE 3-4. Wiring Det ails Model 396P-01-55 for use with Junction Box (PN 23550-00) for Models 54e, 3081, 4081, 81, and 5081. WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH DWG . NO. REV . 40396P08 B ...

  • Emerson Process Management 396PVP - page 19

    15 MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P FIGURE 3-5. Wiring Det ails Model 396P-01-55 for use with Models 54e, 81, 3081, 4081, and 5081 FIGURE 3-6. Wiring Model 396P-01-( )--55 Sensor to Model 1055-10-22-32 Analyzer ...

  • Emerson Process Management 396PVP - page 20

    16 Figures 4-1 thru 4-22 provide the guidelines for wiring the sensor to various Analyzer/Transmitter instrument s. T o determine which wiring guideline to use, locate the model number of the sensor to be installed. 1. If the cable needs to be extended, use a high qual- ity eleven conductor double shielded instrument cable available from Rosemount ...

  • Emerson Process Management 396PVP - page 21

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-1. Wire Functions for Mating V ariopol Cable used with Model 396PVP FIGURE 4-2. Connector Pins and Their Functions FIGURE 4-3. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating V ariopol Cable (PN 23645-07) for use with Model 81 FIGURE 4-4. Wiring Det ails for Models 396PVP ...

  • Emerson Process Management 396PVP - page 22

    18 MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-8. Wiring Det ails for Models 396PVP or 396P-02-50 with Mating V ariopol Cable (PN 23645-06) for use with Remote Junction Box (PN 23309-03) to Model 1 181 FIGURE 4-7. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating V ariopol Cable (PN 23645-07) for use with Remote Junct ...

  • Emerson Process Management 396PVP - page 23

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-12. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating V ariopol Cable (PN 23645-07) for use with Remote Junction Box (PN 23555-00) to Models 3081, 4081, & 5081 FIGURE 4-1 1. Wiring Details for Models 396PVP or 396P- 02-54 with Mating V ariopol Cable (PN 23645-06) for use ...

  • Emerson Process Management 396PVP - page 24

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-15. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating V ariopol Cable (PN 23645-07) for use with Model 1055 FIGURE 4-16. W iring Model 396P-(02)-( )--54/55-(61) Sensor to Model 1055-01-10-22-32 Analyzer FIGURE 4-17. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating ...

  • Emerson Process Management 396PVP - page 25

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-18. Wiring Det ails for Models 396PVP or 396P-02-54 with Mating V ariopol Cable (PN 23645-06) for use with Models 1054A/B & 2054 FIGURE 4-19. Wiring Det ails for Models 396PVP or 396P-02-54 with Mating V ariopol Cable (PN 23645-06) for use with Remote Junction Box (PN 23309-04) to ...

  • Emerson Process Management 396PVP - page 26

    22 MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-21. Wiring Det ails Model 396P-02-50 for use with Junction Box (PN 23309-03) and Remote Preamplifier , Extension Cable (PN 9200254). DWG . NO. REV . 40396P25 A FIGURE 4-22. Wiring Det ails Model 396P-02-54 for use with Junction Box (PN 23309-04) and Remote Preamplifier (PN 22698-03 ...

  • Emerson Process Management 396PVP - page 27

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-23. Wiring Model 396P-01 to Model Xmt-P-HT -10 FIGURE 4-24. Wiring Models 396P-02 and 396PVP to Model Xmt-P-HT -10 23 ...

  • Emerson Process Management 396PVP - page 28

    1. While obtaining a process solution sample (it is recommended that the sample is taken close to the sensor), record the pH value that is shown on the analyzer/transmitter display . 2. Measure and record the pH of the process solution sample with another temperature compensated, calibrated pH instrument. For best results, st andardization should b ...

  • Emerson Process Management 396PVP - page 29

    25 MODEL 396P and 396PVP SECTION 6.0 MAINTENANCE The Model 396P and 396PVP Sensors require mini- mum maintenance. The sensor should be kept clean and free of debris and sediment at all times. The fre- quency of cleaning by wiping or brushing with a soft cloth or brush is determined by the nature of the solu- tion being measured. The sensor should b ...

  • Emerson Process Management 396PVP - page 30

    26 MODEL 396P and 396PVP SECTION 6.0 MAINTENANCE 6.3 MODEL 396P and 396PVP ORP 4.3.1 Platinum Electrode Check. The platinum elec- trode may be checked as follows: There are two types of standard solutions which may be used to check the ORP electrode/transmitter system. T ype 1: One type of commonly used ORP st andard solution is the saturated quinh ...

  • Emerson Process Management 396PVP - page 31

    27 MODEL 396P and 396PVP SECTION 7.0 DIAGNOSTICS AND TROUBLESHOOTING SECTION 7.0 DIAGNOSTICS AND TROUBLESHOOTING 7.1 MODEL 54e/81/3081/4081/5081/Xmt DIAGNOSTICS AND TROUBLESHOOTING The Model 54 Analyzer and Models 81 and 3081 T ransmitters automatically search for fault conditions that would cause an error in the measured pH value, as does the Mode ...

  • Emerson Process Management 396PVP - page 32

    28 7.2 TROUBLESHOOTING WITHOUT ADV ANCED DIAGNOSTICS. T able 7-2, below , lists common problems, causes and remedies typically encountered in process measurement. Problem Probable Cause Remedy Meter reads off scale. (Display Defective preamplifier Replace preamplifier (for code 02 reads overrange). sensors). For code 01, replace sensor . T .C. elem ...

  • Emerson Process Management 396PVP - page 33

    29 MODEL 396P and 396PVP SECTION 7.0 TROUBLESHOOTING T ABLE 7-3. Model 396P and 396PVP pH/ORP Replacement Parts and Accessories PN DESCRIPTION QUANTITY 1 1275-01 Sensor Handrail Mounting Assembly 200201 1 Flow Cell, CPVC, 1 inch FNPT 23242-02 Mounting Adapter , Insertion, 1¼-inch MNPT (304 S.S.) X 1” FNPT (PEEK) 23309-03 Junction Box, for remote ...

  • Emerson Process Management 396PVP - page 34

    30 MODEL 396P and 396PVP SECTION 8.0 RETURN OF MA TERIAL SECTION 8.0 RETURN OF MA TERIAL 8.1 GENERAL. T o expedite the repair and return of instruments, prop- er communication between the customer and the fac- tory is important. Before returning a product for repair , call 1-949-757-8500 for a Return Materials Authorization (RMA) number . 8.2 W ARR ...

  • Emerson Process Management 396PVP - page 35

    W ARRANTY Seller warrants that the firmware will execute the programming instructions provided by Seller , and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira- tion of the applicable warranty period. Goods are warranted for twelve (12) months ...

  • Emerson Process Management 396PVP - page 36

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .r ...

Fabricant Emerson Process Management Catégorie Stereo System

Les documents que nous recevons du fabricant de l'appareilEmerson Process Management 396PVP peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Emerson Process Management
- manuels d’utilisations 396PVP
- fiches produit Emerson Process Management
- dépliants
- ou étiquettes-énergie Emerson Process Management 396PVP
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Emerson Process Management 396PVP.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Emerson Process Management 396PVP, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Emerson Process Management 396PVP. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Emerson Process Management 396PVP.

Manuel d’utilisation complet de l’appareil Emerson Process Management 396PVP, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Emerson Process Management 396PVP par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Emerson Process Management 396PVP.

Le manuel d’utilisation complet Emerson Process Management, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Emerson Process Management 396PVP - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Emerson Process Management 396PVP, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lEmerson Process Management 396PVP qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Emerson Process Management 396PVP - qui devraient nous aider dans les premières étapes lors de l'utilisation Emerson Process Management 396PVP
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Emerson Process Management 396PVP
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Emerson Process Management 396PVP dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Emerson Process Management 396PVP?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Emerson Process Management 396PVP, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Emerson Process Management 396PVP il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)