Manuale per l’uso Emerson Process Management 396PVP

36 pagine 0.77 mb
Scarica

Vai alla pagina of 36

Summary
  • Emerson Process Management 396PVP - page 1

    Combina tion pH/ORP Sensor Instruction Manual PN 51-396P/rev .H March 2006 MODEL 396P MODEL 396PVP with mating cable Models 396P and 396PVP ...

  • Emerson Process Management 396PVP - page 2

    CAUTION SENSOR/PROCESS APPLICA TION COMP A TIBILITY The wetted sensor materials may not be comp atible with process composition and operating conditions. Application comp atibility is entirely the responsibility of the user . DANGER HAZARDOUS AREA INST ALLA TION This sensor is not Intrinsically Safe. or Explosion Proof. Installations near flammable ...

  • Emerson Process Management 396PVP - page 3

    i MODEL 396P and 396PVP T ABLE OF CONTENTS MODEL 396P and 396PVP COMBINA TION pH/ORP SENSOR T ABLE OF CONTENTS Section Title Page 1.0 DESCRIPTION AND SPECIFICA TIONS ................................................................. 1 1.1 Features and Applications ...................................................................................... ...

  • Emerson Process Management 396PVP - page 4

    MODEL 396P and 396PVP T ABLE OF CONTENTS LIST OF FIGURES Number Title Page 1-1 Cross Section Diagram of the TUpH Patented Reference T echnology ............................................................................... 2 2-1 Dimensional Drawing .................................................................................................... ...

  • Emerson Process Management 396PVP - page 5

    Model 396PVP is of fered with a watertight sensor-to- cable connector which eliminates re-wiring and cable twisting when replacing sensors. The V ariopol VP multiple pin connector is an integral part of each sen- sor model and uses a mating VP cable. Once the cable is installed and wired to the analyzer , sensors are easily replaced without replaci ...

  • Emerson Process Management 396PVP - page 6

    2 MODEL 396P and 396PVP SECTION 1.0 DESCRIPTIONS AND SPECIFICA TIONS Measurement s and Ranges *pH: 0-14 ORP: -1500 to 1500 mv A vailable pH A CCU G LASS T ypes GPLR hemi or flat glass Wetted Materials T itanium, Polypropylene, EPDM, glass; platinum (ORP only) Process Connection 1 in. MNPT front and rear facing threads T emperature Range 0-100°C (3 ...

  • Emerson Process Management 396PVP - page 7

    3 MODEL 396P and 396PVP SECTION 1.0 DESCRIPTIONS AND SPECIFICA TIONS 1.3 ORDERING INFORMA TION The Model 396P Sensor is housed in a molded reinforced polypropylene body with 1 in. MNPT threads suit able for insertion, submersion or flow through installation. The sensor includes a general purpose pH electrode or a platinum ORP electrode, a p atented ...

  • Emerson Process Management 396PVP - page 8

    4 MODEL 396P and 396PVP SECTION 1.0 DESCRIPTIONS AND SPECIFICA TIONS Optional Slotted Tip is available on all hemi bulb sensors, ordered as option -41 Flat Tip is avail- able with flat glass bulb sen- sors Examples of all sensing tip offerings Shrouded Tip is standard on all hemi bulb sensors V ariopol connector shown with mating variopol cable rec ...

  • Emerson Process Management 396PVP - page 9

    5 MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION SECTION 2.0 INST ALLA TION 2. Do not install the sensor on the horizont al. The sensor must be 10° off the horizont al to ensure accuracy . 3. Do not install the sensor up side down. 4. Air bubbles may become trapped in the sensor end between the glass bulb and the sensor body . This problem is mo ...

  • Emerson Process Management 396PVP - page 10

    6 MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION FIGURE 2-1. Dimensional Drawing DWG . NO. REV . 40396P01 D MILLIMETER INCH FIGURE 2-2. Flow-Through T ee with Adapter (PN 915240-xx*) MILLIMETER INCH xx* Process Connection Threads 03 3/4 inch 04 1 inch 05 1-1/2 inch SENSOR CABLE (OR VP CONNECTOR - NOT SHOWN) SENSOR CABLE (OR VP CONNECTOR - NOT SHO ...

  • Emerson Process Management 396PVP - page 11

    MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION FIGURE 2-4. Model 396P with Insertion Mounting Adapter (PN 23242-02). Not for use with Model 396PVP . Mounting adapter allows for sensor removal without twisting or disconnecting interconnecting cable for ease of maintenance. WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH DWG . NO. REV . 40396P0 ...

  • Emerson Process Management 396PVP - page 12

    MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION FIGURE 2-5. Submersion Inst allations WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH DWG . NO. REV . 40396P04 A DWG . NO. REV . 40396P03 A Handrail Mounting Assembly PN 1 1275-01 8 ...

  • Emerson Process Management 396PVP - page 13

    9 MODEL 396P and 396PVP SECTION 2.0 INST ALLA TION FIGURE 2-6. Low Flow Cell PN 23728-00 MILLIMETER INCH FIGURE 2-7. Jet Spray Cleaner PN 12707-00 SENSOR CABLE (OR VP CONNECTOR - NOT SHOWN) ...

  • Emerson Process Management 396PVP - page 14

    10 Figures in this section provide the guidelines for wiring the 396P-01 sensor to various Analyzer/Transmitter instruments. T o determine which wiring guideline to use, locate the model number of the sensor to be installed. 1. If the cable needs to be extended, use a high qual- ity eleven conductor double shielded instrument cable available from R ...

  • Emerson Process Management 396PVP - page 15

    11 MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P Wiring. The Model 396P has an optional built-in pre- amplifier and is offered with a shielded cable. The cable should be handled carefully and kept dry and free of corrosive chemicals at all times. Extreme care should be used to prevent it from being twisted, dam- aged or scraped by rough, sh ...

  • Emerson Process Management 396PVP - page 16

    12 MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P FIGURE 3-2. Wiring Det ails Model 396P-01-50 for use with and without Junction Box (PN 22719-02) for Model 1 181. WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH DWG . NO. REV . 40396P15 B ...

  • Emerson Process Management 396PVP - page 17

    13 FIGURE 3-3. Wiring Det ails Model 396P-01-54 for use with and without Junction Box (PN 22719-02) for Models 1054A, 1054B, 2054, 2081, and Extension Cable (9200254). MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P DWG . NO. REV . 40396P13 D WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH ...

  • Emerson Process Management 396PVP - page 18

    14 MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P FIGURE 3-4. Wiring Det ails Model 396P-01-55 for use with Junction Box (PN 23550-00) for Models 54e, 3081, 4081, 81, and 5081. WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH DWG . NO. REV . 40396P08 B ...

  • Emerson Process Management 396PVP - page 19

    15 MODEL 396P and 396PVP SECTION 3.0 WIRING - MODEL 396P FIGURE 3-5. Wiring Det ails Model 396P-01-55 for use with Models 54e, 81, 3081, 4081, and 5081 FIGURE 3-6. Wiring Model 396P-01-( )--55 Sensor to Model 1055-10-22-32 Analyzer ...

  • Emerson Process Management 396PVP - page 20

    16 Figures 4-1 thru 4-22 provide the guidelines for wiring the sensor to various Analyzer/Transmitter instrument s. T o determine which wiring guideline to use, locate the model number of the sensor to be installed. 1. If the cable needs to be extended, use a high qual- ity eleven conductor double shielded instrument cable available from Rosemount ...

  • Emerson Process Management 396PVP - page 21

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-1. Wire Functions for Mating V ariopol Cable used with Model 396PVP FIGURE 4-2. Connector Pins and Their Functions FIGURE 4-3. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating V ariopol Cable (PN 23645-07) for use with Model 81 FIGURE 4-4. Wiring Det ails for Models 396PVP ...

  • Emerson Process Management 396PVP - page 22

    18 MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-8. Wiring Det ails for Models 396PVP or 396P-02-50 with Mating V ariopol Cable (PN 23645-06) for use with Remote Junction Box (PN 23309-03) to Model 1 181 FIGURE 4-7. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating V ariopol Cable (PN 23645-07) for use with Remote Junct ...

  • Emerson Process Management 396PVP - page 23

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-12. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating V ariopol Cable (PN 23645-07) for use with Remote Junction Box (PN 23555-00) to Models 3081, 4081, & 5081 FIGURE 4-1 1. Wiring Details for Models 396PVP or 396P- 02-54 with Mating V ariopol Cable (PN 23645-06) for use ...

  • Emerson Process Management 396PVP - page 24

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-15. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating V ariopol Cable (PN 23645-07) for use with Model 1055 FIGURE 4-16. W iring Model 396P-(02)-( )--54/55-(61) Sensor to Model 1055-01-10-22-32 Analyzer FIGURE 4-17. Wiring Det ails for Models 396PVP or 396P-02-55 with Mating ...

  • Emerson Process Management 396PVP - page 25

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-18. Wiring Det ails for Models 396PVP or 396P-02-54 with Mating V ariopol Cable (PN 23645-06) for use with Models 1054A/B & 2054 FIGURE 4-19. Wiring Det ails for Models 396PVP or 396P-02-54 with Mating V ariopol Cable (PN 23645-06) for use with Remote Junction Box (PN 23309-04) to ...

  • Emerson Process Management 396PVP - page 26

    22 MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-21. Wiring Det ails Model 396P-02-50 for use with Junction Box (PN 23309-03) and Remote Preamplifier , Extension Cable (PN 9200254). DWG . NO. REV . 40396P25 A FIGURE 4-22. Wiring Det ails Model 396P-02-54 for use with Junction Box (PN 23309-04) and Remote Preamplifier (PN 22698-03 ...

  • Emerson Process Management 396PVP - page 27

    MODEL 396P and 396PVP SECTION 4.0 WIRING - MODEL 396PVP FIGURE 4-23. Wiring Model 396P-01 to Model Xmt-P-HT -10 FIGURE 4-24. Wiring Models 396P-02 and 396PVP to Model Xmt-P-HT -10 23 ...

  • Emerson Process Management 396PVP - page 28

    1. While obtaining a process solution sample (it is recommended that the sample is taken close to the sensor), record the pH value that is shown on the analyzer/transmitter display . 2. Measure and record the pH of the process solution sample with another temperature compensated, calibrated pH instrument. For best results, st andardization should b ...

  • Emerson Process Management 396PVP - page 29

    25 MODEL 396P and 396PVP SECTION 6.0 MAINTENANCE The Model 396P and 396PVP Sensors require mini- mum maintenance. The sensor should be kept clean and free of debris and sediment at all times. The fre- quency of cleaning by wiping or brushing with a soft cloth or brush is determined by the nature of the solu- tion being measured. The sensor should b ...

  • Emerson Process Management 396PVP - page 30

    26 MODEL 396P and 396PVP SECTION 6.0 MAINTENANCE 6.3 MODEL 396P and 396PVP ORP 4.3.1 Platinum Electrode Check. The platinum elec- trode may be checked as follows: There are two types of standard solutions which may be used to check the ORP electrode/transmitter system. T ype 1: One type of commonly used ORP st andard solution is the saturated quinh ...

  • Emerson Process Management 396PVP - page 31

    27 MODEL 396P and 396PVP SECTION 7.0 DIAGNOSTICS AND TROUBLESHOOTING SECTION 7.0 DIAGNOSTICS AND TROUBLESHOOTING 7.1 MODEL 54e/81/3081/4081/5081/Xmt DIAGNOSTICS AND TROUBLESHOOTING The Model 54 Analyzer and Models 81 and 3081 T ransmitters automatically search for fault conditions that would cause an error in the measured pH value, as does the Mode ...

  • Emerson Process Management 396PVP - page 32

    28 7.2 TROUBLESHOOTING WITHOUT ADV ANCED DIAGNOSTICS. T able 7-2, below , lists common problems, causes and remedies typically encountered in process measurement. Problem Probable Cause Remedy Meter reads off scale. (Display Defective preamplifier Replace preamplifier (for code 02 reads overrange). sensors). For code 01, replace sensor . T .C. elem ...

  • Emerson Process Management 396PVP - page 33

    29 MODEL 396P and 396PVP SECTION 7.0 TROUBLESHOOTING T ABLE 7-3. Model 396P and 396PVP pH/ORP Replacement Parts and Accessories PN DESCRIPTION QUANTITY 1 1275-01 Sensor Handrail Mounting Assembly 200201 1 Flow Cell, CPVC, 1 inch FNPT 23242-02 Mounting Adapter , Insertion, 1¼-inch MNPT (304 S.S.) X 1” FNPT (PEEK) 23309-03 Junction Box, for remote ...

  • Emerson Process Management 396PVP - page 34

    30 MODEL 396P and 396PVP SECTION 8.0 RETURN OF MA TERIAL SECTION 8.0 RETURN OF MA TERIAL 8.1 GENERAL. T o expedite the repair and return of instruments, prop- er communication between the customer and the fac- tory is important. Before returning a product for repair , call 1-949-757-8500 for a Return Materials Authorization (RMA) number . 8.2 W ARR ...

  • Emerson Process Management 396PVP - page 35

    W ARRANTY Seller warrants that the firmware will execute the programming instructions provided by Seller , and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira- tion of the applicable warranty period. Goods are warranted for twelve (12) months ...

  • Emerson Process Management 396PVP - page 36

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .r ...

Fabbricante Emerson Process Management Categoria Stereo System

I documenti che otteniamo dal fabbricante del dispositivoEmerson Process Management 396PVP possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Emerson Process Management
- manuali per l’uso 396PVP
- schede prodotto Emerson Process Management
- opuscoli
- o etichette energetiche Emerson Process Management 396PVP
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Emerson Process Management 396PVP.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Emerson Process Management 396PVP, manuali di manutenzione, brevi manuali o manuali utente Emerson Process Management 396PVP. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Emerson Process Management 396PVP.

Manuale completo del dispositivo Emerson Process Management 396PVP, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Emerson Process Management 396PVP dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Emerson Process Management 396PVP.

Il manuale per l’uso completo Emerson Process Management, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Emerson Process Management 396PVP - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Emerson Process Management 396PVP, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Emerson Process Management 396PVP che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Emerson Process Management 396PVP - che ci dovrebbe aiutare nei primi passi durante l'uso Emerson Process Management 396PVP
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Emerson Process Management 396PVP
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Emerson Process Management 396PVP in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Emerson Process Management 396PVP?

Utilizza il modulo sottostante

Se non hai risolto il problema con Emerson Process Management 396PVP con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Emerson Process Management 396PVP è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)