Manuel d’utilisation Milwaukee 2216-20

20 pages 3.38 mb
Télécharger

Aller à la page of 20

Summary
  • Milwaukee 2216-20 - page 1

    MIL W AUKEE ELECTRIC T OOL CORPORA TION – TEST & MEASUREMEN T PRODUCT REP AIR SER VICE INSTRUCTIONS — CALIBRA TION 2 216- 20 Light Commercial Digital Multimeter Environmental Condition Perform all calibration at an ambient temperature of 23°C ± 2°C / 73.4°F ± 3.6°F a nd relative humidity of 80% - allow the digital multimeter to si t a ...

  • Milwaukee 2216-20 - page 2

    i T ABLE OF CONTENTS Page Introduction 1 Precautions and Safety Information 1 Symbols 1 Safety 2 S pecifications 3 General S pecification 3 V oltage S pecif ication 4 Current S pecifications 5 Resist ance, Continuity , Capacit ance S pecifications 6 Frequency , Rotary Switch, Auto Power Of f S pecification 7 Physical and environment characteristics ...

  • Milwaukee 2216-20 - page 3

    1 Introduction W arning T o avoid shock or injury , do not perform the verification test s or calibration procedu res described in the manual unless y ou are qualified to do so. The information provided in this document is for the use of qu alified personnel only . Caution The 2216-20 cont ain p art s that can be damaged by st atic discharge. Follo ...

  • Milwaukee 2216-20 - page 4

    2 SAFETY Review the following safety precaution s to avoid injury and prevent damag e to this product or any product s connected to it. T o avoid potential hazards, use the product only as spe cified. CAUTION: These st atement s identify conditions or practices that could result in damage to the equipment or other pro perty . W ARNING: These statem ...

  • Milwaukee 2216-20 - page 5

    3 SPECIFICA TIONS All specifications are warranted unless noted typical and apply to the 2216 -20 DMM. S tated accuracies are at 23°C±5°C at tha n 80% relative humidity and without the battery indicator displayed. Gen era l sp ec ifi ca tions n o i t p i r c s e D s c i t s i r e t c a r a h C s t n u o c 0 0 0 6 t n u o c y a l p s i D c e s / ...

  • Milwaukee 2216-20 - page 6

    4 Measurement Characteris tics Accuracy is ±(% reading + number of digit s) at 23°C ± 5°C, less than 80% R.H. T emperature coeffici ent: 0.1 * (S pecifi ed accuracy)/°C , < 18°C , > 28°C (1) DC V olt age Range Resolution Accuracy 600.0mV 0.1mV 6.000V 0.001V 60.00V 0.01V 600.0V 0.1V ± (0.5% reading + 2 digits) Input Impedance: 10M Ω ...

  • Milwaukee 2216-20 - page 7

    5 (3) DC Current Range Resolution Accuracy 6.000A 0.001A 10.00A 0.01A ± (1% reading + 3 digit s) Response Ti me: < 1.5 sec. Maximum measurement time: 5 minutes at 10A with at least 20 minutes rest time. Overload Protection: AC/D C 20A for 10 sec ond. (4) AC Current Range Resolution Accuracy (Sine W av e) 6.000A * 0.001A 10.00A 0.01A ± (1.5% re ...

  • Milwaukee 2216-20 - page 8

    6 (6) Continuity Range Resolution Accuracy 600.0 Ω 0.1 Ω ± (1% reading + 5 digit s) Open Circuit V olt age: Approx. 0.6V Continuity: Built-in buzzer sounds whe n measured resistance i s less than 30 Ω and sounds off when measured resist ance is more than 200 Ω , Between 30 Ω to 200 Ω the buzzer maybe sound or of f either . Continuity I ...

  • Milwaukee 2216-20 - page 9

    7 (1 1) Auto Po wer Off The meter au tomatically turns the power off after powering up and no operation for 20 ± 5 minutes. T o power up, turn the knob back to OFF f irst. (12) Measuring Rate: 3 t i mes / sec. (13) Fuse: 10A/600V Fast Action Fuse (1 4) Batt er y : Alkaline 1.5V Battery * 2 (15) Low Battery V oltage: 2.3V ± 0.1V (16) Backlight The ...

  • Milwaukee 2216-20 - page 10

    8 Physical and Environment al Characteri stics n o i t p i r c s e D s c i t s i r e t c a r a h C Dime nsio ns (H ×W ×D) 170mm × 80mm × 57m m g K 3 . 0 ) y r e t t a b h t i w ( t h g i e W n o i t p i r c s e D s c i t s i r e t c a r a h c l a t n e m n o r i v n E C ° 0 5 + o t 0 1 - g n i t a r e p o e r u t a r e p m e T Non-Operating -4 ...

  • Milwaukee 2216-20 - page 11

    9 Required Equipment Required equipment is listed in T able B. If the recommended models are not available, equip ment with equivalent specifications may be used. Repairs or se rvicing should be per formed only by qualified personn el. T able B. Required Equipment Equipment Required Characteristics Recommended Model Calibrator AC Voltage Range: 0 ? ...

  • Milwaukee 2216-20 - page 12

    10 Basic Maintenance W arning T o avoid shock, remove the test leads an d any input signals before opening the case o r replacing the battery or fuses. Opening the Meter Ca se Caution T o avoid unintentional shock circuit, always place the uncovered Meter assembl y on a protective surface. When the ca se of the Meter is open, circuit connections ar ...

  • Milwaukee 2216-20 - page 13

    11 T esting Fuses T o test the internal fuses of the meter . 1. T urn the rotary selector s witch to the Ω position. 2. T o test FS1, plug a test lead into V Ω Hz input terminal, and touch the probe to the A input terminal. The displ ay should indicate between 0.0 to 0.2 Ω . (Bussmann KTK-10 recommended). If display reads higher than 0.2 Ω ...

  • Milwaukee 2216-20 - page 14

    12 Performance T est s The following performance test s verify the co mplete operability of the Meter an d check the accuracy of each Meter function against the Meter’s specifications. Accuracy specificatio ns are valid for a period of one year after calibration, when measured at an operating temperature of 18°C to 28°C and at A maximum of 80% ...

  • Milwaukee 2216-20 - page 15

    13 T esting the V olt age Function To verify accuracy in the AC and DC voltage ranges, do the following: 1. T urn t he ro t ar y switch to “ V ” position . 2. Connect the Calibrator to the V Ω and CO M inputs on the Meter. 3. Set the Calibrator for the voltage and frequency from step 1 to 12 in Table 1. 4. Compare the reading on the Meter dis ...

  • Milwaukee 2216-20 - page 16

    14 T esting the Resist ance Function T o verify the accuracy of the resist ance function, do the following: 1. Connect the Calibrator to V Ω Hz and CO M on t he Me ter . 2. T u rn t he ro tary swi tch to Ω . 3. Apply the inputs for st ep 1-6 in T able 3. 4. Compare the Meter display readings to the displ ay readings in T able 3. 5. If the displa ...

  • Milwaukee 2216-20 - page 17

    15 T esting the Ampere Function T o verify the accuracy of AC current measurement function s, do the following: 1. Connect the Calibrator to the A and CO M inpu t s on t he Me ter . 2. T u rn the ro ta ry swit ch to A~. 3. Apply the inputs for step s 1-4 in T able 5. 4. For each input, comp are the readings on the Meter display to the reading in T ...

  • Milwaukee 2216-20 - page 18

    16 T able 7 Frequency T est: Step Source Level Waveform Readin g 1 8.0Vp-p 2.00Hz Square Wa ve 1.98Hz to 2.02Hz 2 8.0Vp-p 80.00Hz Square Wa ve 79.90Hz to 80.10Hz 3 8.0Vp-p 800.0Hz Square Wa ve 799.0Hz to 801.0Hz 4 8.0Vp-p 8.000kHz Square Wa ve 7.990kHz to 8.010kHz 5 10.0Vp-p 50.00kHz Square Wave 49.93kHz to 50.07kHz 6 230.0Vp-p 10.00 Hz Sine Wave 9 ...

  • Milwaukee 2216-20 - page 19

    17 Calibration Procedure Recalibrate the meter: It is recommended that the meter may be calibrated once year . Use the following procedure to calibrate the meter . Envir on ment a l C ondi t ion 1. Perform calibration at an ambient temperature and relative humidity (23°C±2°C a nd R.H. 80%). Allow the instrument to sit at this te mperature for at ...

  • Milwaukee 2216-20 - page 20

    18 (E) DCA Calibration (V R7) 1. Rotate the rotary switch to DCA Function. 2. Connect the output of a DCA calibrator to V and COM input terminals of the meter. 3. Output DCA 5.000A from the calibrator. 4. Adjust VR7 until the display shows 5.000A ± 0.001A. 5. Turn off the output of the calibrator an d disconnect the meter from the calib rator. (F) ...

Fabricant Milwaukee Catégorie Home Safety Product

Les documents que nous recevons du fabricant de l'appareilMilwaukee 2216-20 peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Milwaukee
- manuels d’utilisations 2216-20
- fiches produit Milwaukee
- dépliants
- ou étiquettes-énergie Milwaukee 2216-20
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Milwaukee 2216-20.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Milwaukee 2216-20, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Milwaukee 2216-20. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Milwaukee 2216-20.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Milwaukee 2216-20, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Milwaukee 2216-20 par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Milwaukee 2216-20.

Le manuel d’utilisation complet Milwaukee, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Milwaukee 2216-20 - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Milwaukee 2216-20, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lMilwaukee 2216-20 qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Milwaukee 2216-20 - qui devraient nous aider dans les premières étapes lors de l'utilisation Milwaukee 2216-20
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Milwaukee 2216-20
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Milwaukee 2216-20 dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Milwaukee 2216-20?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Milwaukee 2216-20, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Milwaukee 2216-20 il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)