    760000 AL TERNA TING PRESSURE / L OW AIR L OSS MA TTRESS SYSTEM USER MANUAL Impor tant: Do not operate the Mattress System without rst reading and understanding this manual! Save this manual for future use. 760000-INS-LAB-RevA08 ...

  • Graham Field 760000-INS-LAB-RevA08 - page 2

    2 760000-INS-LAB-RevA08 Contents INTRODUCTION ..................................................................................... 3 INTENDED USE OF THIS DEVICE ........................................................ 3 IMPORT ANT SAFETY PRECAUTIONS .................................................. 3 DANGER ....................................... ...

  • Graham Field 760000-INS-LAB-RevA08 - page 3

    760000-INS-LAB-RevA08 3 Intr oduCtIon Use this manual for set-up and operation of the Alternating Pr essure / Low Air Loss Mattr ess System. Read all instruc- tions befor e using the mattress system. Save this manual for futur e refer ence. Intended use of thIs devICe The intended use of the Alternating Pr essure / Low Air Loss Mattr ess System is: ...

  • Graham Field 760000-INS-LAB-RevA08 - page 4

    4 760000-INS-LAB-RevA08 d anger T o reduce the risk of electr ocution: DANGER: Always unplug this product immediately after using. DANGER: Do not use this product while bathing. DANGER: Do not place or store this product where it can fall or be pulled into a tub or sink. DANGER: Do not place this product in or drop into water or other liquid. DANGE ...

  • Graham Field 760000-INS-LAB-RevA08 - page 5

    760000-INS-LAB-RevA08 5 W ARNING: Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the openings may be blocked. Keep the air opening free of lint, hair , and other similar particles. W ARNING: Never drop or insert any object into any opening or hose. W ARNING: Do not use this product in the p ...

  • Graham Field 760000-INS-LAB-RevA08 - page 6

    6 760000-INS-LAB-RevA08 For U.S. and Canada only E228589 53DG Medical Equipment: Air Pump with respect to electrical shock, re and mechanical hazards only in accordance with UL60601-1 and CAN/CSA C22.2 No. 601.1 pr oduCt desCrIptIon pump and ma ttress sy stem pump and mattress system CPR valves mattress-pump connectors mattress foot end symbol p ...

  • Graham Field 760000-INS-LAB-RevA08 - page 7

    760000-INS-LAB-RevA08 7 pump: rear main power switch power cord receptacle fuse hanger filter pump: front p anel cycle time adjust button; cycle time indicators at right of button decrease pr essure button; pressur e level indicators at right of button panel lock button; panel lock indicator at right of button therapy adjust button; therapy mode in ...

  • Graham Field 760000-INS-LAB-RevA08 - page 8

    8 760000-INS-LAB-RevA08 incr ease pressur e under the buttocks). Press the therapy adjust button at the left end of the r ow to cycle between the settings. The selected therapy mode setting's indicator will illuminate. F r ont panel four th row Panel lock button: The panel lock pr otects the panel set- tings fr om accidental change. Press to l ...

  • Graham Field 760000-INS-LAB-RevA08 - page 9

    760000-INS-LAB-RevA08 9 Ser vice indicator: Mechanical failure, service requir ed; contact your Graham-eld equipment pr ovider . Inst alla tIon unp aCkIng 1. Befor e unpacking, check for obvious damage to the car- ton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment pr ovider . 2. Remove all loose ...

  • Graham Field 760000-INS-LAB-RevA08 - page 10

    10 760000-INS-LAB-RevA08 s Caution: Place the device in a position wher e the car e- giver and/or doctor can access it easily . 5. Connect the mattr ess-pump air hose quick connector; ensur e that the connector has clicked secur ely into place befor e continuing. s Caution: Ensur e that the air hoses ar e not kinked or tucked under the mattr ess. W ...

  • Graham Field 760000-INS-LAB-RevA08 - page 11

    760000-INS-LAB-RevA08 11 Note: The mattress system will automatically enter Autofirm mode for the quickest inflation at set-up; initial inflation takes approximately twenty minutes. 3. The pump will begin delivering air into the mattr ess. 4. When the selected pr essure level is r eached, the pump will enter alternating mode. The mattr ess can t ...

  • Graham Field 760000-INS-LAB-RevA08 - page 12

    12 760000-INS-LAB-RevA08 transport Cap In case of power failur e or trans- port: Disconnect the quick connec- tor and cap the mattr ess with the transport cap to slow deation. Cpr funCtIon If CPR must be performed on the patient while the mattr ess is in use, to immediately deate the mat- tr ess, either: 1. Pull the CPR valves fr om the mattr ...

  • Graham Field 760000-INS-LAB-RevA08 - page 13

    760000-INS-LAB-RevA08 13 general maIntenanCe 1. Check main power cor d for abrasion or excessive wear . 2. Check mattr ess cover for signs of wear or damage. En- sur e mattress cover and tubes ar e connected correctly . 3. Check airow fr om the air hose connector . The airow should alternate between each connector ever y half- cycle time if i ...

  • Graham Field 760000-INS-LAB-RevA08 - page 14

    14 760000-INS-LAB-RevA08 st orage 1. Lay the mattr ess out at and upside down. 2. Roll fr om the head end towar d the foot end. 3. Str etch the foot-end strap around the r olled mattress to pr event unrolling. 4. Stor e mattress and pump in a dust-fr ee environment with no exposur e to direct sunlight. s Caution: Do not fold, cr ease or stack ma ...

  • Graham Field 760000-INS-LAB-RevA08 - page 15

    760000-INS-LAB-RevA08 15 Mattress Specication Model 8" mattress Dimensions (L x W x H) 78.7" x 35.4" x 8" (200 x 90 x 20.3 cm) Weight 16.5 lb (7.5 kg) Warranty GF Health Pr oducts, Inc. warrants the Lumex AltaDyne Plus 760000 Alternating Pr essure / Low Air Loss Mattr ess Sys- tem for a period of twelve months for defects in ...

  • Graham Field 760000-INS-LAB-RevA08 - page 16

    USA, Corporate Headquar ters: Graham-Field Health Products 2935 Nor theast Parkway Atlanta, Georgia 30360 telephone: 800-347-5678, 770-447-1609 fax: 800-726-0601, 678-291-3232 www .grahamfield.com 760000-INS-LAB-RevA08 © 2008 GF Health Products, Inc. ...

Fabricant Graham Field Catégorie Camping Equipment

Les documents que nous recevons du fabricant de l'appareilGraham Field 760000-INS-LAB-RevA08 peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Graham Field
- manuels d’utilisations 760000-INS-LAB-RevA08
- fiches produit Graham Field
- dépliants
- ou étiquettes-énergie Graham Field 760000-INS-LAB-RevA08
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Graham Field 760000-INS-LAB-RevA08.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Graham Field 760000-INS-LAB-RevA08, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Graham Field 760000-INS-LAB-RevA08. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Graham Field 760000-INS-LAB-RevA08.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Graham Field 760000-INS-LAB-RevA08, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Graham Field 760000-INS-LAB-RevA08 par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Graham Field 760000-INS-LAB-RevA08.

Le manuel d’utilisation complet Graham Field, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Graham Field 760000-INS-LAB-RevA08 - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Graham Field 760000-INS-LAB-RevA08, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lGraham Field 760000-INS-LAB-RevA08 qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Graham Field 760000-INS-LAB-RevA08 - qui devraient nous aider dans les premières étapes lors de l'utilisation Graham Field 760000-INS-LAB-RevA08
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Graham Field 760000-INS-LAB-RevA08
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Graham Field 760000-INS-LAB-RevA08 dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Graham Field 760000-INS-LAB-RevA08?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Graham Field 760000-INS-LAB-RevA08, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Graham Field 760000-INS-LAB-RevA08 il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)