    Agilent Specifying Calibration Standards for the Agilent 8510 Network Analyzer Product Note 8510-5B ...

  • Agilent Technologies 8510-58 - page 2

    2 3 3 3 4 5 5 7 7 8 9 9 9 11 12 12 12 14 14 15 16 16 17 17 17 18 18 18 18 19 19 19 19 19 20 20 20 21 21 22 22 22 23 23 26 29 Introduction Measurement errors Measurement calibration Calibration kit Standard definition Class assignment Modification procedure Select standards Define standards Standard number Standard type Open circuit capacitance: C 0 ...

  • Agilent Technologies 8510-58 - page 3

    3 Introduction This pr oduct note co vers measurement calibration requirements f or the Agilent 8510B/C network analyzer . All of t he capabilities described in this note also apply to the A gilent 8510A with the follo wing exceptions: response & isolation calibra- tion; short cir cuit induct ance; class assignments for forward/rev erse isolati ...

  • Agilent Technologies 8510-58 - page 4

    4 The arra y coeff icients are computed by measuring a set of “kno wn” devices connected at a fixed point and sol ving as the vector difference between the modeled and measured response. The full 2-por t er r or model sho wn in Figure 1 is an exam ple of only one of the measurement calibra- tions a vailable with the 8510. The measurement calibr ...

  • Agilent Technologies 8510-58 - page 5

    5 Standard definition Standard definition is t he process of mathematical- ly modeling the electrical charact eristics (delay , attenuation and impedance) of each calibration standard. These electrical characteristics can be mathematically deriv ed from the physical dimen- sions and material of each calibration standards or from its actual measured ...

  • Agilent Technologies 8510-58 - page 6

    6 T able 1. Standard definitions table T able 2. Standard class assignments ...

  • Agilent Technologies 8510-58 - page 7

    7 Modification procedure Calibration kit modif ication pro vides the capability to adapt t o measurement calibrations in other con- nector types or to generate more precise error models fr om existing kits. Pro vided the appropri- ate standards are available, cal kit modif ication can be used to establish a reference plane in t he same transmission ...

  • Agilent Technologies 8510-58 - page 8

    8 Define standards A glossar y of standard definition parameters used with the Agilent 8510 is included in this section. Each parameter is described and appropriate con- versions are listed f or im plement ation with t he 8510. T o illustrate, a calibration kit f or WR-62 rec- tangular wav eguide (operating frequency range 12.4 to 18 GHz) will be d ...

  • Agilent Technologies 8510-58 - page 9

    9 Each standard is described using the Standard Definition Table in accordance wit h t he 1- or 2- port model. The St andard Definition t able for a wav eguide calibration kit is sho wn in Table 1. Each standard type (shor t, open, load, thru, and arbi- trar y impedance) ma y be def ined by the parame- ters as specif ied below . • Standard number ...

  • Agilent Technologies 8510-58 - page 10

    10 It is not possible t o remov e fringing capacit ance, but the result ant phase shif t can be modeled as a function of frequency using C 0 t hr ough C 3 (C 0 +C l  f + C 2  f 2 + C 3  f 3 ,with units of F(Hz), C 0 (fF), C 1 (10 -27 F/Hz), C 2 (10 -36 F/Hz 2 ) and C 3 (10 -45 F/Hz 3 ), which are the coeff icients for a cubic polynomial th ...

  • Agilent Technologies 8510-58 - page 11

    11 Note In some cases (when t he phase response is linear with respect t o frequency) the response of an open can be modeled as an equivalent “increment al” length. This method will serve as a f irst order appro xima- tion onl y , but can be useful when data or st an- dards for t he abov e modeling techniques are not available. F or the wa vegu ...

  • Agilent Technologies 8510-58 - page 12

    The inductance as a function of frequency can be modeled b y specifying t he coef ficients of a third- order polynomial (L 0 + L 1  f + L 2  f 2 + L 3  f 3 ), with units of L 0 (nH), L 1 (10 -24 H/Hz), L 2 (10 -33 H/Hz 2 ) and L 3 (10 -42 H/Hz 3 ). F or the wa veguide exam ple, the inductance of t he offset short circuits is negligible. L ...

  • Agilent Technologies 8510-58 - page 13

    13 The conv ention f or definition of offset dela y in wav eguide requires entr y of t he delay assuming no dispersion. F or wa veguide transmission line, the Agilent 8510 calculat es t he ef f ects of dispersion as a function of frequency as f ollows: f co = lo wer cutof f frequency f = measurement frequency Note T o assure accurat e def inition o ...

  • Agilent Technologies 8510-58 - page 14

    14 µ r = relativ e permeability constant of t he medium (equal to 1.0 in air) ε r = relativ e permittivity constant of t he medium (equal to 1.000649 in air) D = inside diameter of outer conductor d = outside diameter of inner conductor The 8510 requires that the charact eristic imped- ance of wav eguide transmission line is assigned t o be equal ...

  • Agilent Technologies 8510-58 - page 15

    15 Therefore, for the WR -62 wav eguide standard defi- nition table, offset loss of zero ohm/sec is entered for all f our standards. Lower/minimum frequency Lo wer frequency defines the minimum frequency at which the st andard is to be used f or the purposes of calibration. Note When defining coaxial of fset st andar ds, it ma y be necessar y t o u ...

  • Agilent Technologies 8510-58 - page 16

    16 Upper/maximum frequency This specifies the maximum frequency at which the standard is valid. In broadband applications, a set of banded standards ma y be necessar y to pr o- vide constant response. For example, coaxial offset standards (i.e., 1 / 4 λ offset short) are generally spec- ified o ver bandwidths of an octav e or less. Bandwidth speci ...

  • Agilent Technologies 8510-58 - page 17

    17 Note Mathematical operations on measurements (and display ed data) after calibration are not corrected for dispersion. Enter W A VEGUIDE int o t he st andar d definition table for all f our standards. Standard labels Labels are entered through the title menu and may contain up to 10 characters. Standard Labels are entered to facilitate menu driv ...

  • Agilent Technologies 8510-58 - page 18

    18 S 11 A,B,C and S 22 A,B,C S 11 A, B,C and S 22 A,B,C cor respond to the S 11 and S 22 ref lection calibrations for por t 1 and port 2 respectivel y. These three classes are used by the Agilent 8510 t o solve for t he systematic errors; directivity , source mat ch, and ref lection tracking. The three classes used b y the 7-mm cal kit are labeled ...

  • Agilent Technologies 8510-58 - page 19

    19 TRL Thru TRL Thru corresponds t o t he measurement of t he S-parameters of a zero-length or shor t t hru connec- tion betw een port 1 and por t 2. The Thr u, Ref lect and Line classes are used ex clusively for t he t hree steps of t he TRL 2-PORT calibration. T ypically , a “delay/thru” with zer o (or the smallest) Of fset Delay is specified ...

  • Agilent Technologies 8510-58 - page 20

    20 Each adapter is specified as a sing le delay/thru standard and up to sev en standards numbers can be specified into t he adapt er class. Standard Class labels Standard Class labels are entered t o facilit ate menu-driven calibration. A label can be any user- selected term which best describes the de vice or class of devices t hat t he operator s ...

  • Agilent Technologies 8510-58 - page 21

    21 Again, cal kit labels should be chosen t o best describe the calibration devices. The “B.1” default suff ix cor responds t o the kit’s mechanical revision (B) and mathematical revision (1). Note T o prev ent confusion, if any st andard definitions in a calibration kit are modified but a new kit label is not entered, the default label will ...

  • Agilent Technologies 8510-58 - page 22

    22 User modified cal kits and Agilent 8510 specifications As noted previously , the resultant accuracy of the 8510 when used with an y calibration kit is depend- ent on how well its standards are def ined and is verified through measurement of a device wit h traceable frequency response. The published Measurement Specifications f or the 8510 N etwo ...

  • Agilent Technologies 8510-58 - page 23

    23 Appendix A Calibration kit entry procedure Calibration kit specif ications can be ent ered into the Agilent 8510 using the 8510 disk drive, a disk drive connected to the system bus, by front panel entr y , or t hr ough program contr ol by an external controller . Disk procedure This is an im por tant feature since the 8510 can internally store o ...

  • Agilent Technologies 8510-58 - page 24

    24 Front panel procedure: (P -band waveguide example) 1. Prior to modifying or generating a cal kit, store one or both of the cal kits in t he 8510’s non- volatile memor y t o a disk. 2. Select CAL menu, MORE. 3. Prepare to modify cal kit 2: press MODIFY 2. 4. T o define a standard: press DEFINE ST AND ARD. 5. Enable st andar d no. 1 to be modifi ...

  • Agilent Technologies 8510-58 - page 25

    25 4. Change the class label for S 11 A: L ABEL CL ASS, S 11 A, ERASE TITLE. 5. Enter t he label of PSHORT 1 b y using the knob, the SELECT sof t key and the SP A CE soft ke y. 6. Com plete the label entry procedure: TITLE DONE, LABEL DONE. F ollow a similar procedure to enter t he remaining standard classes and labels as shown in the table belo w: ...

  • Agilent Technologies 8510-58 - page 26

    26 Appendix B Dimensional considerations in coaxial connectors This appendix describes dimensional considera- tions and required conv entions used in deter min- ing the physical offset lengt h of calibration standards in sexed coaxial connector families. Precise measurement of the ph ysical offset length is required to deter mine the OFFSET DELAY o ...

  • Agilent Technologies 8510-58 - page 27

    27 Female type-N Male type-N 7 mm Coaxial connector T ype-N coaxial connector interface The location of the “calibration plane” in T ype-N standards is the outer conductor mating surfaces as shown below . Note: 1.0mm, 1.85mm and 2.4mm connectors not shown, but similar to 3.5mm calibration planes. ...

  • Agilent Technologies 8510-58 - page 28

    28 ...

  • Agilent Technologies 8510-58 - page 29

    29 Appendix C Cal coefficients model Offset devices like of fset shor ts and offset opens can be modeled by t he follo wing signal f low g raph : Figure 1 Signal flow graph model of offset devices The offset portion of t he open or shor t, is modeled as a per f ectly unif orm lossy air dielectric transmis- sion line. The expected coefficient of ref ...

  • Agilent Technologies 8510-58 - page 30

    30 Their first or der appro ximations, R is small and G=0, are: Equation 3 Since Equation 4 F or coaxial devices ...

  • Agilent Technologies 8510-58 - page 31

    31 then: Equation 5 Equation 6 If the Of fset delay=0, then t he coef ficient of ref lection, Γ = Γ L . ...

  • Agilent Technologies 8510-58 - page 32

    Agilent T echnologies’ T est and Measurement Support, Services, and Assistance Agilent T echnologies aims to maximize t he value y ou receive, while minimizing your risk and problems. W e striv e to ensure that y ou get t he test and measurement capabilities you paid f or and obt ain t he support you need. Our extensive suppor t resources and ser ...

Fabricant Agilent Technologies Catégorie Switch

Les documents que nous recevons du fabricant de l'appareilAgilent Technologies 8510-58 peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Agilent Technologies
- manuels d’utilisations 8510-58
- fiches produit Agilent Technologies
- dépliants
- ou étiquettes-énergie Agilent Technologies 8510-58
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Agilent Technologies 8510-58.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Agilent Technologies 8510-58, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Agilent Technologies 8510-58. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Agilent Technologies 8510-58.

Manuel d’utilisation complet de l’appareil Agilent Technologies 8510-58, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Agilent Technologies 8510-58 par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Agilent Technologies 8510-58.

Le manuel d’utilisation complet Agilent Technologies, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Agilent Technologies 8510-58 - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Agilent Technologies 8510-58, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lAgilent Technologies 8510-58 qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Agilent Technologies 8510-58 - qui devraient nous aider dans les premières étapes lors de l'utilisation Agilent Technologies 8510-58
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Agilent Technologies 8510-58
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Agilent Technologies 8510-58 dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Agilent Technologies 8510-58?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Agilent Technologies 8510-58, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Agilent Technologies 8510-58 il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)