    Agilent Specifying Calibration Standards for the Agilent 8510 Network Analyzer Product Note 8510-5B ...

  • Agilent Technologies 8510-58 - page 2

    2 3 3 3 4 5 5 7 7 8 9 9 9 11 12 12 12 14 14 15 16 16 17 17 17 18 18 18 18 19 19 19 19 19 20 20 20 21 21 22 22 22 23 23 26 29 Introduction Measurement errors Measurement calibration Calibration kit Standard definition Class assignment Modification procedure Select standards Define standards Standard number Standard type Open circuit capacitance: C 0 ...

  • Agilent Technologies 8510-58 - page 3

    3 Introduction This pr oduct note co vers measurement calibration requirements f or the Agilent 8510B/C network analyzer . All of t he capabilities described in this note also apply to the A gilent 8510A with the follo wing exceptions: response & isolation calibra- tion; short cir cuit induct ance; class assignments for forward/rev erse isolati ...

  • Agilent Technologies 8510-58 - page 4

    4 The arra y coeff icients are computed by measuring a set of “kno wn” devices connected at a fixed point and sol ving as the vector difference between the modeled and measured response. The full 2-por t er r or model sho wn in Figure 1 is an exam ple of only one of the measurement calibra- tions a vailable with the 8510. The measurement calibr ...

  • Agilent Technologies 8510-58 - page 5

    5 Standard definition Standard definition is t he process of mathematical- ly modeling the electrical charact eristics (delay , attenuation and impedance) of each calibration standard. These electrical characteristics can be mathematically deriv ed from the physical dimen- sions and material of each calibration standards or from its actual measured ...

  • Agilent Technologies 8510-58 - page 6

    6 T able 1. Standard definitions table T able 2. Standard class assignments ...

  • Agilent Technologies 8510-58 - page 7

    7 Modification procedure Calibration kit modif ication pro vides the capability to adapt t o measurement calibrations in other con- nector types or to generate more precise error models fr om existing kits. Pro vided the appropri- ate standards are available, cal kit modif ication can be used to establish a reference plane in t he same transmission ...

  • Agilent Technologies 8510-58 - page 8

    8 Define standards A glossar y of standard definition parameters used with the Agilent 8510 is included in this section. Each parameter is described and appropriate con- versions are listed f or im plement ation with t he 8510. T o illustrate, a calibration kit f or WR-62 rec- tangular wav eguide (operating frequency range 12.4 to 18 GHz) will be d ...

  • Agilent Technologies 8510-58 - page 9

    9 Each standard is described using the Standard Definition Table in accordance wit h t he 1- or 2- port model. The St andard Definition t able for a wav eguide calibration kit is sho wn in Table 1. Each standard type (shor t, open, load, thru, and arbi- trar y impedance) ma y be def ined by the parame- ters as specif ied below . • Standard number ...

  • Agilent Technologies 8510-58 - page 10

    10 It is not possible t o remov e fringing capacit ance, but the result ant phase shif t can be modeled as a function of frequency using C 0 t hr ough C 3 (C 0 +C l  f + C 2  f 2 + C 3  f 3 ,with units of F(Hz), C 0 (fF), C 1 (10 -27 F/Hz), C 2 (10 -36 F/Hz 2 ) and C 3 (10 -45 F/Hz 3 ), which are the coeff icients for a cubic polynomial th ...

  • Agilent Technologies 8510-58 - page 11

    11 Note In some cases (when t he phase response is linear with respect t o frequency) the response of an open can be modeled as an equivalent “increment al” length. This method will serve as a f irst order appro xima- tion onl y , but can be useful when data or st an- dards for t he abov e modeling techniques are not available. F or the wa vegu ...

  • Agilent Technologies 8510-58 - page 12

    The inductance as a function of frequency can be modeled b y specifying t he coef ficients of a third- order polynomial (L 0 + L 1  f + L 2  f 2 + L 3  f 3 ), with units of L 0 (nH), L 1 (10 -24 H/Hz), L 2 (10 -33 H/Hz 2 ) and L 3 (10 -42 H/Hz 3 ). F or the wa veguide exam ple, the inductance of t he offset short circuits is negligible. L ...

  • Agilent Technologies 8510-58 - page 13

    13 The conv ention f or definition of offset dela y in wav eguide requires entr y of t he delay assuming no dispersion. F or wa veguide transmission line, the Agilent 8510 calculat es t he ef f ects of dispersion as a function of frequency as f ollows: f co = lo wer cutof f frequency f = measurement frequency Note T o assure accurat e def inition o ...

  • Agilent Technologies 8510-58 - page 14

    14 µ r = relativ e permeability constant of t he medium (equal to 1.0 in air) ε r = relativ e permittivity constant of t he medium (equal to 1.000649 in air) D = inside diameter of outer conductor d = outside diameter of inner conductor The 8510 requires that the charact eristic imped- ance of wav eguide transmission line is assigned t o be equal ...

  • Agilent Technologies 8510-58 - page 15

    15 Therefore, for the WR -62 wav eguide standard defi- nition table, offset loss of zero ohm/sec is entered for all f our standards. Lower/minimum frequency Lo wer frequency defines the minimum frequency at which the st andard is to be used f or the purposes of calibration. Note When defining coaxial of fset st andar ds, it ma y be necessar y t o u ...

  • Agilent Technologies 8510-58 - page 16

    16 Upper/maximum frequency This specifies the maximum frequency at which the standard is valid. In broadband applications, a set of banded standards ma y be necessar y to pr o- vide constant response. For example, coaxial offset standards (i.e., 1 / 4 λ offset short) are generally spec- ified o ver bandwidths of an octav e or less. Bandwidth speci ...

  • Agilent Technologies 8510-58 - page 17

    17 Note Mathematical operations on measurements (and display ed data) after calibration are not corrected for dispersion. Enter W A VEGUIDE int o t he st andar d definition table for all f our standards. Standard labels Labels are entered through the title menu and may contain up to 10 characters. Standard Labels are entered to facilitate menu driv ...

  • Agilent Technologies 8510-58 - page 18

    18 S 11 A,B,C and S 22 A,B,C S 11 A, B,C and S 22 A,B,C cor respond to the S 11 and S 22 ref lection calibrations for por t 1 and port 2 respectivel y. These three classes are used by the Agilent 8510 t o solve for t he systematic errors; directivity , source mat ch, and ref lection tracking. The three classes used b y the 7-mm cal kit are labeled ...

  • Agilent Technologies 8510-58 - page 19

    19 TRL Thru TRL Thru corresponds t o t he measurement of t he S-parameters of a zero-length or shor t t hru connec- tion betw een port 1 and por t 2. The Thr u, Ref lect and Line classes are used ex clusively for t he t hree steps of t he TRL 2-PORT calibration. T ypically , a “delay/thru” with zer o (or the smallest) Of fset Delay is specified ...

  • Agilent Technologies 8510-58 - page 20

    20 Each adapter is specified as a sing le delay/thru standard and up to sev en standards numbers can be specified into t he adapt er class. Standard Class labels Standard Class labels are entered t o facilit ate menu-driven calibration. A label can be any user- selected term which best describes the de vice or class of devices t hat t he operator s ...

  • Agilent Technologies 8510-58 - page 21

    21 Again, cal kit labels should be chosen t o best describe the calibration devices. The “B.1” default suff ix cor responds t o the kit’s mechanical revision (B) and mathematical revision (1). Note T o prev ent confusion, if any st andard definitions in a calibration kit are modified but a new kit label is not entered, the default label will ...

  • Agilent Technologies 8510-58 - page 22

    22 User modified cal kits and Agilent 8510 specifications As noted previously , the resultant accuracy of the 8510 when used with an y calibration kit is depend- ent on how well its standards are def ined and is verified through measurement of a device wit h traceable frequency response. The published Measurement Specifications f or the 8510 N etwo ...

  • Agilent Technologies 8510-58 - page 23

    23 Appendix A Calibration kit entry procedure Calibration kit specif ications can be ent ered into the Agilent 8510 using the 8510 disk drive, a disk drive connected to the system bus, by front panel entr y , or t hr ough program contr ol by an external controller . Disk procedure This is an im por tant feature since the 8510 can internally store o ...

  • Agilent Technologies 8510-58 - page 24

    24 Front panel procedure: (P -band waveguide example) 1. Prior to modifying or generating a cal kit, store one or both of the cal kits in t he 8510’s non- volatile memor y t o a disk. 2. Select CAL menu, MORE. 3. Prepare to modify cal kit 2: press MODIFY 2. 4. T o define a standard: press DEFINE ST AND ARD. 5. Enable st andar d no. 1 to be modifi ...

  • Agilent Technologies 8510-58 - page 25

    25 4. Change the class label for S 11 A: L ABEL CL ASS, S 11 A, ERASE TITLE. 5. Enter t he label of PSHORT 1 b y using the knob, the SELECT sof t key and the SP A CE soft ke y. 6. Com plete the label entry procedure: TITLE DONE, LABEL DONE. F ollow a similar procedure to enter t he remaining standard classes and labels as shown in the table belo w: ...

  • Agilent Technologies 8510-58 - page 26

    26 Appendix B Dimensional considerations in coaxial connectors This appendix describes dimensional considera- tions and required conv entions used in deter min- ing the physical offset lengt h of calibration standards in sexed coaxial connector families. Precise measurement of the ph ysical offset length is required to deter mine the OFFSET DELAY o ...

  • Agilent Technologies 8510-58 - page 27

    27 Female type-N Male type-N 7 mm Coaxial connector T ype-N coaxial connector interface The location of the “calibration plane” in T ype-N standards is the outer conductor mating surfaces as shown below . Note: 1.0mm, 1.85mm and 2.4mm connectors not shown, but similar to 3.5mm calibration planes. ...

  • Agilent Technologies 8510-58 - page 28

    28 ...

  • Agilent Technologies 8510-58 - page 29

    29 Appendix C Cal coefficients model Offset devices like of fset shor ts and offset opens can be modeled by t he follo wing signal f low g raph : Figure 1 Signal flow graph model of offset devices The offset portion of t he open or shor t, is modeled as a per f ectly unif orm lossy air dielectric transmis- sion line. The expected coefficient of ref ...

  • Agilent Technologies 8510-58 - page 30

    30 Their first or der appro ximations, R is small and G=0, are: Equation 3 Since Equation 4 F or coaxial devices ...

  • Agilent Technologies 8510-58 - page 31

    31 then: Equation 5 Equation 6 If the Of fset delay=0, then t he coef ficient of ref lection, Γ = Γ L . ...

  • Agilent Technologies 8510-58 - page 32

    Agilent T echnologies’ T est and Measurement Support, Services, and Assistance Agilent T echnologies aims to maximize t he value y ou receive, while minimizing your risk and problems. W e striv e to ensure that y ou get t he test and measurement capabilities you paid f or and obt ain t he support you need. Our extensive suppor t resources and ser ...

Fabbricante Agilent Technologies Categoria Switch

I documenti che otteniamo dal fabbricante del dispositivoAgilent Technologies 8510-58 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Agilent Technologies
- manuali per l’uso 8510-58
- schede prodotto Agilent Technologies
- opuscoli
- o etichette energetiche Agilent Technologies 8510-58
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Agilent Technologies 8510-58.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Agilent Technologies 8510-58, manuali di manutenzione, brevi manuali o manuali utente Agilent Technologies 8510-58. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Agilent Technologies 8510-58.

Manuale completo del dispositivo Agilent Technologies 8510-58, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Agilent Technologies 8510-58 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Agilent Technologies 8510-58.

Il manuale per l’uso completo Agilent Technologies, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Agilent Technologies 8510-58 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Agilent Technologies 8510-58, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Agilent Technologies 8510-58 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Agilent Technologies 8510-58 - che ci dovrebbe aiutare nei primi passi durante l'uso Agilent Technologies 8510-58
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Agilent Technologies 8510-58
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Agilent Technologies 8510-58 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Agilent Technologies 8510-58?

Utilizza il modulo sottostante

Se non hai risolto il problema con Agilent Technologies 8510-58 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Agilent Technologies 8510-58 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)