    Teledyne Analytical Instruments QUICKS T AR T GUIDE M O D E L 3 0 0 0 T A T ra ce Oxygen Analyzer ...

  • Teledyne 3000TA - page 2

    Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments ...

  • Teledyne 3000TA - page 3

    Model 3000TA Quickstart Guide Teledyne Analytical Instruments 3 GETTING S T ARTED This Quickstart Guide is designed to get you s et up and ope rating your Teledyne An a lytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss. This Quicksta rt Guide should be used in conjunct ...

  • Teledyne 3000TA - page 4

    Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 4 • Left & Right Arrows Select between functi ons currently displayed on the VF D screen. • Up & Down Arrows Increment or decrement values of functions currently displayed. • Enter Advances VFD display to the next screen in a series or re turns to the Analyze scree ...

  • Teledyne 3000TA - page 5

    Model 3000TA Quickstart Guide Teledyne Analytical Instruments 5 • Analog Output C onnecti on: Terminals for current and volt age a n alog output signals. • Digital Input Terminals: Termi n als for connecti ng d igital input signals. • Alarm Connections: 2 concentration alarms and 1 system alarm. • Calibration Contact : To notify e xternal e ...

  • Teledyne 3000TA - page 6

    Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 6 SETUP AND INS T ALL Mount the Ana lyzer The Model 3000T A is for indoor use in a general purpose are a. It is NOT for hazardous environme nts of any type. It is design ed fo r f lush p anel m ount ing. Us e t he fou r m ount ing h ole s— on e i n each corn er of t he r igid ...

  • Teledyne 3000TA - page 7

    Model 3000TA Quickstart Guide Teledyne Analytical Instruments 7 4. If using remote calibra tion feature: • Connect zero gas to t he Z ERO IN port. • Connect span ga s to the SP AN IN port. 5. Otherwise, c a li b ra t io n ga se s m u st b e T e e' d i nt o t h e s a m p l e i n le t wi t h a p p r o p ri a t e v a l ve s . 6. Regulate the ...

  • Teledyne 3000TA - page 8

    Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 8 • Connect primary input power: Insert the power cord into the power cord rece ptacle. • Install the fuse: Fuses are not installed at the fac tory. Be sure to install the proper fuse as par t of installati on. Place small scre wdr iver in notch, and pry cover off, as shown. ...

  • Teledyne 3000TA - page 9

    Model 3000TA Quickstart Guide Teledyne Analytical Instruments 9 • Connect Range ID Contacts: There a re four ded icated Range ID relay contacts on the rear panel. The first three ranges are assigned to relays in ascending order. Range 1 ID— Low range Range 2 ID— Medium ra nge Range 3 ID— High range Range 4 ID—Air Cal Ra nge (25%). • Net ...

  • Teledyne 3000TA - page 10

    Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 10 9. With t h e O-ring in place, align the guide pin with the hole on the cell holder. Then, with the hol d er, lift cell into the cell block. 10. Push the gate on the cell block down so that the slots on the side of the gate engage the locati ng screws on the side of the block ...

  • Teledyne 3000TA - page 11

    Model 3000TA Quickstart Guide Teledyne Analytical Instruments 11 BASIC OPER A TION Once the a naly zer has been installed, it can be configured for your application. Prior to using the instrument you m ust : • Calibrate t he instrument • Define the analysis r anges and choose whether to use autoranging or a fixed range. • Set ala rm setpoints ...

  • Teledyne 3000TA - page 12

    Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 12 A vailab le Functions The six function butt ons are used to control the analyzer. Some functi ons have submenus that allow additional choices as s hown below. Any function ca n be sel ected at any time by pressing the appropriate button (unless password restrictions appl y). ...

  • Teledyne 3000TA - page 13

    Model 3000TA Quickstart Guide Teledyne Analytical Instruments 13 Password Protection If a password is assigned, then setting the following system parameters can be done only after t he passw ord is entered: span and zero settings, alarm setpoints, analysis range definitions, switching bet ween autoranging and manual overri de, setting up an auto-ca ...

  • Teledyne 3000TA - page 14

    Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 14 TRAK/HLD Auto — Cal PSWD Logou t M ore • Use the ◄► arrow keys to scroll the blin king over to PSWD , • Press Enter to select the password function. Eithe r the default TE TAI password or AAAAA place holders for an existing password will appear on screen depending o ...

  • Teledyne 3000TA - page 15

    Model 3000TA Quickstart Guide Teledyne Analytical Instruments 15 Auto Mode Zero Calibration • Press Zero to enter the zero function mode. • Use the ▲▼ arrow keys to toggle between AUT O and MAN zero settling. Stop when AUT O appears blinking. Zero : Settling: AUTO <ENT> To B e gin • Press Enter to begin zeroing. #### PP M Zero Slop ...

  • Teledyne 3000TA - page 16

    Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 16 Span Calibration Make sure an appropriate span gas is connecte d to the sample port. Auto Mode Spanning: • Press Span to enter the span function. • Use the arrow keys to toggle between AUTO and MAN span settling. Stop when AUT O appears, bli nking on the display. Span: Se ...

  • Teledyne 3000TA - page 17

    Model 3000TA Quickstart Guide Teledyne Analytical Instruments 17 Calib. Holding time Cal hold: 5 min • Press Enter key to continue. Span Val : 000008.00 <ENT>Span <UP>Mod # • Press ▲ (<UP>) to permit modification ( Mod # ) of span value. • Use the arrow keys to enter the oxygen c o ncentration of t he span gas you ar e usi ...

  • Teledyne 3000TA - page 18

    Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 18 • Latc hing? Ltch–Y/N (Yes/No). • To change any of th e alarm parameters, use the ◄► arrow keys to move the blinkin g over to the specific par ameter. • Use the ▲▼ arrow keys to change t he va lue of the parameter. Holding down the key speeds up the inc rem en ...

  • Teledyne 3000TA - page 19

    Model 3000TA Quickstart Guide Teledyne Analytical Instruments 19 • Use the ▲▼ arrow keys to switch from AUTO t o FX/L O , FX/MED , or F X /HI to set the instrument on the desired fixe d range (low, medium , or h igh). L—100 M—1000 H—10000 Mode—FX/LO —or— L—1 M—10 H—100 Mode—FX/MED —or— L—1 M—10 H—100 Mode—FX/HI ...

Fabricant Teledyne Catégorie Respiratory Product

Les documents que nous recevons du fabricant de l'appareilTeledyne 3000TA peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Teledyne
- manuels d’utilisations 3000TA
- fiches produit Teledyne
- dépliants
- ou étiquettes-énergie Teledyne 3000TA
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Teledyne 3000TA.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Teledyne 3000TA, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Teledyne 3000TA. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Teledyne 3000TA.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Teledyne 3000TA, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Teledyne 3000TA par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Teledyne 3000TA.

Le manuel d’utilisation complet Teledyne, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Teledyne 3000TA - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Teledyne 3000TA, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lTeledyne 3000TA qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Teledyne 3000TA - qui devraient nous aider dans les premières étapes lors de l'utilisation Teledyne 3000TA
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Teledyne 3000TA
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Teledyne 3000TA dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Teledyne 3000TA?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Teledyne 3000TA, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Teledyne 3000TA il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)