Manuale per l’uso Omega CDTX-90

20 pagine 0.25 mb
Scarica

Vai alla pagina of 20

Summary
  • Omega CDTX-90 - page 1

    www .omega.com e-mail: info@omega.com U ser ’ s Guide Shop online at CDTX-90 Series Conductivity/Resistivity T ransmitters Conductivity/Resistivity 62.50 uS/cm 25.0 ° C ENTER ...

  • Omega CDTX-90 - page 2

    page 2 It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct, ...

  • Omega CDTX-90 - page 3

    page 3 CAUTION! • Remove power to unit before wiring input and output connections. • Follow instructions carefully to avoid personal injury. Omega CD TX -90 Series Conductivity/Resistivity T ransmitter Contents 1. Installation 2. Specifications 3. Electrical Connections 4. Menu Functions 2. Specifications General Compatible electrodes: Omega CD ...

  • Omega CDTX-90 - page 4

    page 4 Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument. Wiring Procedure 1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end. 2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws. 3. Insert exposed (non-insulated) wire end in terminal hole until i ...

  • Omega CDTX-90 - page 5

    page 5 CDTX-90-1 System Power/Loop Connections Stand-alone application, no current loop used T ransmitter 4 3 2 1 Sys. Pwr . Loop - Sys. Pwr . Loop + AUX Power - AUX Power + Power Supply DC 12 - 24 V Power Supply Power Supply Connection to a PLC with built-in power supply PLC Power Supply Ground Power Supply Loop Input 4-20 mA Loop Input 4-20 mA In ...

  • Omega CDTX-90 - page 6

    page 6 Stand-alone application, no current loop used T ransmitter T erminals 4 3 2 1 Sys. Pwr . Loop - Sys. Pwr . Loop + AUX Power - AUX Power + Power Supply DC 12 - 24 V Power Supply Power Supply Connection to a PLC with built-in power supply PLC Te r m inals Power Supply Ground Power Supply Loop Input 4-20 mA Loop Input 4-20 mA Internal PLC Conne ...

  • Omega CDTX-90 - page 7

    page 7 Stand-alone application, no current loop used T ransmitter T erminals 6 5 4 3 2 1 Loop2- Loop2+ Sys. Pwr . Loop - Sys. Pwr . Loop + AUX Power - AUX Power + Power Supply DC 12 - 24 V Power Supply Power Supply Connection to a PLC with built-in power supply PLC T erminals Power Supply Ground Power Supply Channel 2 4-20 mA in Channel 2 4-20 mA i ...

  • Omega CDTX-90 - page 8

    page 8 VIEW menu • During normal operation, the transmitter displays the VIEW menu. • When editing the CALIBRATE or OPTIONS menus, the transmitter will return to the VIEW menu if no activity occurs for 10 minutes. • To select a VIEW display, press the UP or DOWN arrow keys. The selections will scroll in a continuous loop. • Changing the VIE ...

  • Omega CDTX-90 - page 9

    page 9 Notes on Steps 5 and 6: • All output functions remain active during editing. • Only the flashing element can be edited. • RIGHT ARROW key advances the flashing element in a continuous loop. • Edited value is effective immediately after pressing ENTER key. • If no key is pressed for 10 minutes unit will restore the last saved value ...

  • Omega CDTX-90 - page 10

    page 10 Select CUSTOM only if you are connecting a certified conductivity sensor. Select STANDARD for all other systems. For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0. For CUSTOM sensors: Enter the precise cell constant from the certificate provided with your sensor, or from the information label on the sensor. Selec ...

  • Omega CDTX-90 - page 11

    page 11 Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In general, select lower contrast if the display is in warmer ambient surroundings. Set the decimal to the best resolution for your application. The display will auto- matically scale down to this restriction. Select *****., ****.*, ***.** **.*** or *.* ...

  • Omega CDTX-90 - page 12

    page 12 Cell Constant: Standard > Cell: Standard 1> Cell: Custom 1.0000 > Cond Units: µ S> PPM Factor: 2.00 > Set: Temperature > Set: Conductivity > Loop Source: Cond > Loop Range: µ S 0.0000 → → → → → 100.000 > Relay1 Source: Cond > Relay1 Mode: Low > Relay1 Setpnt: 10.0000 µ S> Relay1 Hys: µ S 0.5 ...

  • Omega CDTX-90 - page 13

    page 13 Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In general, select lower contrast if the display is in warmer ambient surroundings. Set the decimal to the best resolution for your application. The display will automatically scale down to this restriction. Select *****., ****.*, ***.** **.*** or *.*** ...

  • Omega CDTX-90 - page 14

    page 14 Select CUSTOM only if you are connecting a certified conductivity sensor. Select STANDARD for all other systems. For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0. For CUSTOM sensors: Enter the precise cell constant from the certificate provided with your sensor, or from the information label on the sensor. Selec ...

  • Omega CDTX-90 - page 15

    page 15 Contrast: 3> Cond Decimal: ****.* > Averaging Off > Loop1 Adjust: 4.00 mA > Loop1 Adjust: 20.00 mA > Temp Display: º C> Temperature Comp %: 2.00 > Output1 Active: Low > Test Loop 1: > Test Output 1: > Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In general, select lo ...

  • Omega CDTX-90 - page 16

    page 16 Calibration Procedure 1. Requirements The CDTX-90 Transmitter is factory calibrated using simulated input signals. System calibration will reduce errors caused by sensor wire lengths longer than the standard 15 ft. length. Wire lengths of 100 feet are acceptable; cable shield must be maintained through cable splice. Calibration may be done ...

  • Omega CDTX-90 - page 17

    page 17 Display Condition Suggested Solutions Possible Causes "- - - -" Value Must be less than 3 Value must be greater than 0 "Value must be 400 or less" Display erratic, shows "0" or "------" intermittently "Too much Error Check Sensor" Display is over range. This may be a normal condition if your ...

  • Omega CDTX-90 - page 18

    page 18 Notes ...

  • Omega CDTX-90 - page 19

    page 19 W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OME GA ’ s W ARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA ? ...

  • Omega CDTX-90 - page 20

    6-8850.090-OM/(C-8/02) M3632/0802 Where Do I Find Ever ything I Need for Process Measurement and Control? OMEGA…Of Course! Shop online at www.omega.com TEMPERA TURE   Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies   Wire: Thermocouple, RTD & Thermistor   Calibrators & Ice Point References ...

Fabbricante Omega Categoria Stud Sensor

I documenti che otteniamo dal fabbricante del dispositivoOmega CDTX-90 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Omega
- manuali per l’uso CDTX-90
- schede prodotto Omega
- opuscoli
- o etichette energetiche Omega CDTX-90
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Omega CDTX-90.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Omega CDTX-90, manuali di manutenzione, brevi manuali o manuali utente Omega CDTX-90. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Omega CDTX-90.

Manuali per l’uso simili

Manuale completo del dispositivo Omega CDTX-90, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Omega CDTX-90 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Omega CDTX-90.

Il manuale per l’uso completo Omega, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Omega CDTX-90 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Omega CDTX-90, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Omega CDTX-90 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Omega CDTX-90 - che ci dovrebbe aiutare nei primi passi durante l'uso Omega CDTX-90
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Omega CDTX-90
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Omega CDTX-90 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Omega CDTX-90?

Utilizza il modulo sottostante

Se non hai risolto il problema con Omega CDTX-90 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Omega CDTX-90 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)