    www .omega.com e-mail: info@omega.com U ser ’ s Guide Shop online at CDTX-90 Series Conductivity/Resistivity T ransmitters Conductivity/Resistivity 62.50 uS/cm 25.0 ° C ENTER ...

  • Omega CDTX-90 - page 2

    page 2 It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct, ...

  • Omega CDTX-90 - page 3

    page 3 CAUTION! • Remove power to unit before wiring input and output connections. • Follow instructions carefully to avoid personal injury. Omega CD TX -90 Series Conductivity/Resistivity T ransmitter Contents 1. Installation 2. Specifications 3. Electrical Connections 4. Menu Functions 2. Specifications General Compatible electrodes: Omega CD ...

  • Omega CDTX-90 - page 4

    page 4 Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument. Wiring Procedure 1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end. 2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws. 3. Insert exposed (non-insulated) wire end in terminal hole until i ...

  • Omega CDTX-90 - page 5

    page 5 CDTX-90-1 System Power/Loop Connections Stand-alone application, no current loop used T ransmitter 4 3 2 1 Sys. Pwr . Loop - Sys. Pwr . Loop + AUX Power - AUX Power + Power Supply DC 12 - 24 V Power Supply Power Supply Connection to a PLC with built-in power supply PLC Power Supply Ground Power Supply Loop Input 4-20 mA Loop Input 4-20 mA In ...

  • Omega CDTX-90 - page 6

    page 6 Stand-alone application, no current loop used T ransmitter T erminals 4 3 2 1 Sys. Pwr . Loop - Sys. Pwr . Loop + AUX Power - AUX Power + Power Supply DC 12 - 24 V Power Supply Power Supply Connection to a PLC with built-in power supply PLC Te r m inals Power Supply Ground Power Supply Loop Input 4-20 mA Loop Input 4-20 mA Internal PLC Conne ...

  • Omega CDTX-90 - page 7

    page 7 Stand-alone application, no current loop used T ransmitter T erminals 6 5 4 3 2 1 Loop2- Loop2+ Sys. Pwr . Loop - Sys. Pwr . Loop + AUX Power - AUX Power + Power Supply DC 12 - 24 V Power Supply Power Supply Connection to a PLC with built-in power supply PLC T erminals Power Supply Ground Power Supply Channel 2 4-20 mA in Channel 2 4-20 mA i ...

  • Omega CDTX-90 - page 8

    page 8 VIEW menu • During normal operation, the transmitter displays the VIEW menu. • When editing the CALIBRATE or OPTIONS menus, the transmitter will return to the VIEW menu if no activity occurs for 10 minutes. • To select a VIEW display, press the UP or DOWN arrow keys. The selections will scroll in a continuous loop. • Changing the VIE ...

  • Omega CDTX-90 - page 9

    page 9 Notes on Steps 5 and 6: • All output functions remain active during editing. • Only the flashing element can be edited. • RIGHT ARROW key advances the flashing element in a continuous loop. • Edited value is effective immediately after pressing ENTER key. • If no key is pressed for 10 minutes unit will restore the last saved value ...

  • Omega CDTX-90 - page 10

    page 10 Select CUSTOM only if you are connecting a certified conductivity sensor. Select STANDARD for all other systems. For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0. For CUSTOM sensors: Enter the precise cell constant from the certificate provided with your sensor, or from the information label on the sensor. Selec ...

  • Omega CDTX-90 - page 11

    page 11 Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In general, select lower contrast if the display is in warmer ambient surroundings. Set the decimal to the best resolution for your application. The display will auto- matically scale down to this restriction. Select *****., ****.*, ***.** **.*** or *.* ...

  • Omega CDTX-90 - page 12

    page 12 Cell Constant: Standard > Cell: Standard 1> Cell: Custom 1.0000 > Cond Units: µ S> PPM Factor: 2.00 > Set: Temperature > Set: Conductivity > Loop Source: Cond > Loop Range: µ S 0.0000 → → → → → 100.000 > Relay1 Source: Cond > Relay1 Mode: Low > Relay1 Setpnt: 10.0000 µ S> Relay1 Hys: µ S 0.5 ...

  • Omega CDTX-90 - page 13

    page 13 Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In general, select lower contrast if the display is in warmer ambient surroundings. Set the decimal to the best resolution for your application. The display will automatically scale down to this restriction. Select *****., ****.*, ***.** **.*** or *.*** ...

  • Omega CDTX-90 - page 14

    page 14 Select CUSTOM only if you are connecting a certified conductivity sensor. Select STANDARD for all other systems. For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0. For CUSTOM sensors: Enter the precise cell constant from the certificate provided with your sensor, or from the information label on the sensor. Selec ...

  • Omega CDTX-90 - page 15

    page 15 Contrast: 3> Cond Decimal: ****.* > Averaging Off > Loop1 Adjust: 4.00 mA > Loop1 Adjust: 20.00 mA > Temp Display: º C> Temperature Comp %: 2.00 > Output1 Active: Low > Test Loop 1: > Test Output 1: > Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In general, select lo ...

  • Omega CDTX-90 - page 16

    page 16 Calibration Procedure 1. Requirements The CDTX-90 Transmitter is factory calibrated using simulated input signals. System calibration will reduce errors caused by sensor wire lengths longer than the standard 15 ft. length. Wire lengths of 100 feet are acceptable; cable shield must be maintained through cable splice. Calibration may be done ...

  • Omega CDTX-90 - page 17

    page 17 Display Condition Suggested Solutions Possible Causes "- - - -" Value Must be less than 3 Value must be greater than 0 "Value must be 400 or less" Display erratic, shows "0" or "------" intermittently "Too much Error Check Sensor" Display is over range. This may be a normal condition if your ...

  • Omega CDTX-90 - page 18

    page 18 Notes ...

  • Omega CDTX-90 - page 19

    page 19 W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OME GA ’ s W ARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA ? ...

  • Omega CDTX-90 - page 20

    6-8850.090-OM/(C-8/02) M3632/0802 Where Do I Find Ever ything I Need for Process Measurement and Control? OMEGA…Of Course! Shop online at www.omega.com TEMPERA TURE   Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies   Wire: Thermocouple, RTD & Thermistor   Calibrators & Ice Point References ...

メーカー Omega カテゴリー Stud Sensor

Omega CDTX-90のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- CDTX-90の取扱説明書
- Omegaの製品カード
- パンフレット
- またはOmega CDTX-90の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Omega CDTX-90の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Omega CDTX-90の取り付け説明書、サービスマニュアル、簡易説明書、またはOmega CDTX-90のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Omega CDTX-90の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Omega CDTX-90デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがOmega CDTX-90を使用するのを助ける技術的文書のことです。説明書は通常、全てのOmega CDTX-90ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Omegaの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Omega CDTX-90の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははOmega CDTX-90の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Omega CDTX-90に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Omega CDTX-90デバイスの基本機能を使うにあたってのヒント。 Omega CDTX-90のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Omega CDTX-90に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるOmega CDTX-90のメーカー/サービスへの連絡先に関する情報。

Omega CDTX-90についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもOmega CDTX-90の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがOmega CDTX-90で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)