    MODEL 6300 VISIBLE RANGE SPECTROPHOTOMETER OPERATING MANUAL 630 026 ...

  • IBM 6300 - page 2

    6300/REV B/12-96 SAFETY Please read this information carefully prior to installing or using this equipment. 1. The unit described in this manual is designed to be operated only by trained personnel. Any adjustments, maintenance and repair must be carried out as defined in this manual, by a person qualified to be aware of the hazards involved. 2. It ...

  • IBM 6300 - page 3

    6300/REV B/12-96 6300 SPECTROPHOTOMETER OPERATING MANUAL CONTENTS SECTION 1 INTRODUCTION Instrument Description 1.1 Instrument Specification 1.2 SECTION 2 INSTALLATION Unpacking 2.1 Installation 2.2 Displays 2.3 Controls 2.4 Outputs 2.5 SECTION 3 OPERATION Initial Set-Up 3.1 Sample Measurement 3.2 Good Practice Guidelines 3.3 SECTION 4 MAINTENANCE ...

  • IBM 6300 - page 4

    SECTION 1 INTRODUCTION 1.1 INSTRUMENT DESCRIPTION The Model 6300 is a microprocessor controlled Visible Range Spectrophotometer covering the wave- length range of 320 to 950nm with a 10nm bandwidth. The monochromator is of a modified Czerny Turner design, incorporating a stepper motor driven 1200 lines/mm holographic diffraction grating and featuri ...

  • IBM 6300 - page 5

    SECTION 2 INSTALLATION 2.1 UNPACKING Remove the Model 6300 from the packaging and ensure the following items are present: 1. Model 6300 Spectrophotometer 2. Mains Cable 3. Pack 100 disposable Cuvettes (060 084) 4. Optional Accessories (as ordered) Any shortages or damage should be reported immediately to the Manufacturer or your local Distributor. ...

  • IBM 6300 - page 6

    2.3 DISPLAYS 1. Primary display area - Transmission, Absorbance, Concentration 2. Primary display adjust annunciator 3. Secondary display area - Wavelength, Factor 4. Primary display units 5. Secondary display adjust annunciator 6. Operation with PC 7. Menu options - %T ABS CONC FACTOR UNITS 8. Menu pointers (for 7) 3 6300/REV B/12-96 ...

  • IBM 6300 - page 7

    2.4 CONTROLS 1. used to adjust values on the selected display 2. used to move horizontally between menu options 3. used to select the displayed menu option 4. initiates a calibration routine 5. Print key. Provides a printout of the current reading with an incremental sample number. When pressed for the first time after a calibration the print out w ...

  • IBM 6300 - page 8

    2.5 INPUTS/OUTPUTS Fig. 2.5.1 Rear Panel Layout 1. LAMP ACCESS PANEL This panel allows the user to gain access to the tungsten halogen lamp when replacement is necessary. NOTE: The Lamp Access Panel and all ventilation slots must not be covered or obstructed at any time. 2. ROCKER SWITCH On/Off switch for the unit. 3. POWER IN SOCKET IEC type conne ...

  • IBM 6300 - page 9

    SECTION 3 OPERATION 3.1 INITIAL SET-UP NOTE: If the unit is in continuous use it is recommended that it should be left switched on to obtain maximum lamp life. Do not open the sample chamber door during initialisation. Connect the unit to the correct mains supply and switch on. After switch on, the Model 6300 automatically re-aligns the monochromat ...

  • IBM 6300 - page 10

    7 The complimentary colour method of selecting wavelength may not be applicable in all situations, perhaps because the solution has no distinct colour, or the solution is of a complex nature and the absorbing species of interest is not the predominant colour. In these circumstances it will be necessary to scan across the spectrum to determine point ...

  • IBM 6300 - page 11

    8 Concentration Mode This mode is selected by moving the cursor to the CONC menu option using the LEFT or RIGHT arrow keys. The primary display will show the concentration, with ppm, mgl -1 , gl -1 , M, % or blank units. The wavelength can be adjusted using the UP and DOWN arrow keys. Pressing the CAL key places the 6300 in a concentration cal mode ...

  • IBM 6300 - page 12

    3.3 GOOD PRACTICE GUIDELINES 1. For optimum performance a calibration routine should be carried out at the beginning and end of every sample batch. 2. To ensure accurate results are obtained the sample area lid should be kept in the closed position during measurement. 3. The styrene cuvettes supplied with the unit are disposable (i.e; ideally they ...

  • IBM 6300 - page 13

    SECTION 4 MAINTENANCE 4.1 GENERAL The 6300 has been designed to give optimum performance with minimum maintenance. It is only necessary to keep the external surfaces clean and free from dust. The sample area should always be kept clean and any accidental spillage should be wiped away immediately. To give added protection when not in use, the unit s ...

  • IBM 6300 - page 14

    Fig. 4.2.1 Lamp Fitting 6. Close the lamp access panel and re-tighten the thumbscrew. NOTE: It is essential that only the specified replacement lamp should be used. Accuracy of optical alignment and performance cannot be guaranteed using alternative manufactured lamps. SECTION 5 OPTIONAL ACCESSORIES 5.1 OPTIONAL ACCESSORIES The following list of it ...

  • IBM 6300 - page 15

    SECTION 6 INTERFACING Serial Interface The Model 6300 has a bi-directional RS232 interface set to: 1200 baud 7 data bits odd parity 1 stop bit The 25 way D connector allows a standard one-to-one interconnection lead to be used, as supplied with the 40 column printer. A printout is initiated by pressing the PRINT key. If the sample munber is unity, ...

  • IBM 6300 - page 16

    ASCII Fxxxx.x<CR> Sets the concentration factor to xxxx.x. For example: F1000<CR> will set the factor to 1000. Note <CR> is an ASCII carriage return character. The last three commands provide an output which can readily be incorporated into most spreadsheet software packages. 6.2 RS232 Output The bi-directional RS232 interface is ...

  • IBM 6300 - page 17

    EC Declaration of Conformity Jenway Model 6300 Spectrophotometer complies with the following European Standards: EN 50081-1:1992 Electromagnetic compatibility - Generic emission standard EN 50082-1:1992 Electromagnetic compatibility - Generic immunity standard (Performance criterion B) EN 61010-1:1993 Safety requirements for electrical equipment fo ...

Fabbricante IBM Categoria Power Supply

I documenti che otteniamo dal fabbricante del dispositivoIBM 6300 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici IBM
- manuali per l’uso 6300
- schede prodotto IBM
- opuscoli
- o etichette energetiche IBM 6300
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso IBM 6300.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione IBM 6300, manuali di manutenzione, brevi manuali o manuali utente IBM 6300. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto IBM 6300.

Manuali per l’uso simili

Manuale completo del dispositivo IBM 6300, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare IBM 6300 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti IBM 6300.

Il manuale per l’uso completo IBM, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale IBM 6300 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice IBM 6300, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per IBM 6300 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo IBM 6300 - che ci dovrebbe aiutare nei primi passi durante l'uso IBM 6300
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del IBM 6300
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio IBM 6300 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su IBM 6300?

Utilizza il modulo sottostante

Se non hai risolto il problema con IBM 6300 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con IBM 6300 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)