    NI 5412 Specifications NI PXI/PCI-5412 14-Bit 100 MS/s Arbitrary Waveform Generator Unless otherwise noted, the followi ng conditions were used for each specification: • Interpola tion set to maximum allo wed factor f or a gi v en sampl e rate. • Signals terminated with 50 Ω . • Lo w-Gain Amplif ie r Path set to 2 V pk-pk , and High-Gain Am ...

  • National Instruments NI 5412 - page 2

    NI 5412 Specifications 2 ni.com CH 0 (Channel 0 Analog Output , Front Panel Connector) Table 1. Specif ication Va l u e Comments Number of Channels 1 — Connector SMB (jack) — Output Voltage Characteristics Output Paths The software-selectable Main Ou tput Path setting provides full-scale vo ltages from 12.00 V pk-pk to 5.64 mV pk-pk into a 50 ? ...

  • National Instruments NI 5412 - page 3

    © National Instruments Corpor ation 3 NI 5412 Specifications Maximum Output Voltage Maximum Output Vo l t a g e Pat h Load Maximum Output V oltage (V pk -pk ) The Maxi mum Output V oltage of the NI 5412 is determined by the Amplitude Range and the Of fset Range. Low- Gain Amplif ier 50 Ω ±1.000 1 k Ω ±1.905 Open ±2.000 High- Gain Amplif ier ...

  • National Instruments NI 5412 - page 4

    NI 5412 Specifications 4 ni.com Frequency and Transient Response Bandwidth 20 MHz –3 dB Digital Interpolation Filter Software-selectable Finite Im pulse Response (FIR) filter . A vailable interpolation f actors are 2, 4, or 8. — Passband Flatness Low-Gain and High-Gain Amplifiers Path — ±1.0 dB from DC to 6 MHz Pulse Response Pat h All value ...

  • National Instruments NI 5412 - page 5

    © National Instruments Corpor ation 5 NI 5412 Specifications Figure 1. Pulse Response, Low-Gai n Amplifier Path 50 Ω Load –2.0 –1.6 –0.4 –1.2 0.0 –0.8 0.4 0.8 1.2 1.6 2.0 0 0.2 0.4 0.6 0.8 1.0 Time ( µ s ) Amplitude (V) ...

  • National Instruments NI 5412 - page 6

    NI 5412 Specifications 6 ni.com Table 1. (Continued) Specification Va l u e Comments Suggested Maximum Frequencies for Common F unctions Function Path — Lo w-Gain Amplif ier High-Gain Amplifier Sine 20 MHz 20 MHz Square 5 MHz 5 MHz Ramp 1 MHz 1 MHz T r iangle 1 MHz 1 MHz Spectral Characteristics Spurious-Free Dynamic Range (SFDR) without Harmonic ...

  • National Instruments NI 5412 - page 7

    © National Instruments Corpor ation 7 NI 5412 Specifications Sample Clock Spectral Characteristics (Continued) A verage Noise Density Pat h Amplitude Range A verage Noise Density A verage Noise Density at small amplitudes is limited by a –148 dBm/Hz noise floor . V pk-pk dBm dBm/Hz dBFS/ Hz Lo w-Gain 2 10 45 –134 –144 High- Gain 12 25.6 251 ...

  • National Instruments NI 5412 - page 8

    NI 5412 Specifications 8 ni.com Sample Rate Rang e and Resolution Sample Clock Source Sample Rate Range Sample Rate Resolution — Divide-b y- N 23.8 4 S/s to 100 MS/s Settable to ( 100 MS/s) / N (1 ≤ N ≤ 4,194,304) High- Resolution 10 S/s to 100 MS/s 1.06 µHz CLK IN 200 kS/s to 1 05 MS/s Resolution determined by external clock source. Externa ...

  • National Instruments NI 5412 - page 9

    © National Instruments Corpor ation 9 NI 5412 Specifications System Phase Noise and Jitter (10 MHz Carrier) Sample Clock Source System Phase Noise Density (dBc/Hz) Offset System Output Jitter (Integrated from 100 Hz to 100 kHz) 1. High- Resolution specifications va r y wi t h Sample Rate. 2. All values are typical. 100 Hz 1 kHz 10 kHz NI PXI-5412 ...

  • National Instruments NI 5412 - page 10

    NI 5412 Specification s 10 ni.com Onboard Clock (Internal VCXO) Phase-Locked Loop (PLL) Reference Clock Table 3. Specif ication Va l u e Comments Clock Source Internal sample clocks can either be locked to a Reference Clock using a phase-locked loop or be deriv ed from th e onboard VCXO frequency reference. — Frequenc y Accuracy ±25 ppm — Tabl ...

  • National Instruments NI 5412 - page 11

    © National Instruments Corpor ation 11 NI 5412 Specifications CLK IN (Sample Clock and Reference Cloc k Input, Front Panel Connector) Table 5. Specif ication Va l u e Comments Connector SMB (jack) — Direction Input — Destinations 1. Sample Clock 2. PLL Reference Clock — Frequenc y Range 1 MHz to 105 MHz (Sample Clock destination and sine wa ...

  • National Instruments NI 5412 - page 12

    NI 5412 Specification s 12 ni.com PFI 0 and PFI 1 (Programmable Function Interfac e, Front Panel Connectors) Table 6. Specif ication Va l u e Comments Connectors T wo SMB (jack) — Direction Bi-directional — Frequenc y Range DC to 105 MHz — As an Input (Trigger) Destinations Start T rigger — Maximum Input Overload –2 V to +7 V — V IH 2.0 ...

  • National Instruments NI 5412 - page 13

    © National Instruments Corpor ation 13 NI 5412 Specifications Start T rigger V OH Minimum: 2.9 V (open load), 1.4 V (50 Ω load) Output dri vers are +3.3 V TTL compatible. Measured with a 1 m cable. V OL Maximum: 0.2 V (open load), 0.2 V (50 Ω load) Rise/F all T ime (20% to 80%) ≤ 2.0 ns Load of 10 pF . Table 7. Specif ication Va l u e Commen ...

  • National Instruments NI 5412 - page 14

    NI 5412 Specification s 14 ni.com Delay from Start T rigger to CH 0 Analog Output Interpolation Factor T ypical Delay Refer to t s2 at NI Signal Generators Help»Devices» NI 5412» NI <bus>-5412» Triggering» Trigger Timing . Digital Inter polation F ilter disabled. 43 Sample Clock Periods + 110 ns 2 57 Sample Clock Periods + 110 ns 4 63 Sa ...

  • National Instruments NI 5412 - page 15

    © National Instruments Corpor ation 15 NI 5412 Specifications Markers Table 8. Specif ication Va l u e Comments Destinations 1. PFI<0..1> (SMB front panel connectors) 2. NI PXI-5412 —PXI_T rig<0..6> (backplane connector) NI PCI-5412 —R TSI<0..6> — Quantity One Marker per Se gment. — Quantum Marker position must be placed ...

  • National Instruments NI 5412 - page 16

    NI 5412 Specification s 16 ni.com W aveform and Instruction Memor y Utilization Table 9. Specif ication Va l u e Comments Memory Usage The NI 5412 uses the Synchronization and Memory Core (SMC) technology in which wav eforms and instructions share onboard memory . Parameters, such as number of segments in sequence list, maximum number of waveforms ...

  • National Instruments NI 5412 - page 17

    © National Instruments Corpor ation 17 NI 5412 Specifications Loop Count 1 to 16,777,215. Burst trigger: Unlimited — Quantum W aveform size must be an integer multiple of four samp les. — Memory Limits 8 MB Standard 32 MB Option 256 MB Option All trigger modes except where noted. Arbitrary W av eform Mode, Maximum W av eform Memory 4,194,176 S ...

  • National Instruments NI 5412 - page 18

    NI 5412 Specification s 18 ni.com Calibration Power Table 10. Specif ication Va l u e Comments Self-Calibrati on An onboard, 24-bit ADC and precision voltage reference are used to calibrate the DC gain and offset. The self-calibration is initiated by the user through the software and takes approximately 75 seconds to complete. — External Calibrat ...

  • National Instruments NI 5412 - page 19

    © National Instruments Corpor ation 19 NI 5412 Specifications Software Table 12. Specif ication Va l u e Comments Dri ver Software NI-FGEN 2.3 or later ve rsion. NI-FGEN is an IVI-compliant driv er that all ows you to conf igure, control , and calibrate the NI 5412. NI -FGEN prov ides application programming interfaces for man y de velopment en vi ...

  • National Instruments NI 5412 - page 20

    NI 5412 Specification s 20 ni.com Environment NI PXI-5412 Environment Note T o ensure that the NI PXI-5412 cools ef fectively , follo w the guidelines in the Maintain Forced-Air Cooling Note to Users incl uded in the NI 5412 kit. The NI PXI-5412 is intended for indoor use only . Table 13. Specifications Va l u e Comments Operating T emperature 0 º ...

  • National Instruments NI 5412 - page 21

    © National Instruments Corpor ation 21 NI 5412 Specifications NI PCI-5412 Environment Note T o ensure that the NI PCI-5412 cools ef fectively , follo w the guidelines in the Maintain Forced -Air Cooling Note to Users inclu ded in the NI 5412 kit. Also, to maxim ize airflo w and extend the life of the device, lea ve any adjacent PCI slots empty . T ...

  • National Instruments NI 5412 - page 22

    NI 5412 Specification s 22 ni.com Safety , Electromagnetic Compatibility , and CE Compliance Table 15. Specification Va l u e Comments Safety The NI 5412 meets the requirements of the following standards for safety and electrical equipment for measurement, control, and laboratory use: • IEC 61010 -1, EN 61010-1 • UL 61010-1 • CAN/CSA C22.2 No ...

  • National Instruments NI 5412 - page 23

    © National Instruments Corpor ation 23 NI 5412 Specifications EMC/EMI CE, C-T ick, and FCC Part 15 (Class A) Compliant Notes : 1. This device is not intended for , and is restricted from, use in residential areas. 2. For EMC compliance, op erate this device with shielded cabling. 3. When connected to other test objects, thi s product may cause rad ...

  • National Instruments NI 5412 - page 24

    NI 5412 Specification s 24 ni.com Physical Table 16. Specif ication Va l u e Comments Dimensions NI PXI-5412 NI PCI-5412 — Single 3U PXI slot. CompactPCI compatib le. 2.0 × 13.0 × 21.6 cm (0.8 × 5.1 × 8.5 inches) 34.07 × 10.67 × 2.03 cm (13.4 × 4.20 × 0.8 inches) We i g h t 340 g (11 oz) 480 g (17 oz) — Front Panel Connectors Label Func ...

  • National Instruments NI 5412 - page 25

    © National Instruments Corpor ation 25 NI 5412 Specifications Where to Go for Support The National Instrum ents W eb site is y our complete resource for technical support. At ni.com/support you hav e access to ev erything from troubleshooting and ap plication dev el opment self-help resources to email and phone assistance from NI Application Engin ...

  • National Instruments NI 5412 - page 26

    National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more information a bout National Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s . For patents covering Natio ...

Fabbricante National Instruments Categoria Portable Generator

I documenti che otteniamo dal fabbricante del dispositivoNational Instruments NI 5412 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici National Instruments
- manuali per l’uso NI 5412
- schede prodotto National Instruments
- opuscoli
- o etichette energetiche National Instruments NI 5412
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso National Instruments NI 5412.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione National Instruments NI 5412, manuali di manutenzione, brevi manuali o manuali utente National Instruments NI 5412. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto National Instruments NI 5412.

Manuale completo del dispositivo National Instruments NI 5412, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare National Instruments NI 5412 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti National Instruments NI 5412.

Il manuale per l’uso completo National Instruments, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale National Instruments NI 5412 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice National Instruments NI 5412, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per National Instruments NI 5412 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo National Instruments NI 5412 - che ci dovrebbe aiutare nei primi passi durante l'uso National Instruments NI 5412
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del National Instruments NI 5412
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio National Instruments NI 5412 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su National Instruments NI 5412?

Utilizza il modulo sottostante

Se non hai risolto il problema con National Instruments NI 5412 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con National Instruments NI 5412 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)